K220759 · Armis Biopharma, Inc. · FQH · May 25, 2023 · General Hospital
Device Facts
Record ID
K220759
Device Name
Armis VeriCyn Wound Wash
Applicant
Armis Biopharma, Inc.
Product Code
FQH · General Hospital
Decision Date
May 25, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5475
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Armis VeriCyn® Wound Wash is to be used with a lavage system to create mechanical movement at the wound surface by delivery of a solution and is indicated for use in cleansing and removal of foreign material including microorganisms and debris from wounds (such as stage I-IV pressure ulcers, diabetic foot ulcers, post surgical wounds, first degree and partial thickness burns, grafted and donor sites).
Device Story
VeriCyn® Wound Wash is an aqueous irrigation solution containing hydrogen peroxide, acetic acid, disodium EDTA, and purified water. Used with a lavage system to provide mechanical debridement of wounds; solution facilitates removal of debris and microorganisms from wound surfaces. Intended for clinical use by healthcare professionals on various wound types including pressure ulcers, diabetic ulcers, and burns. Device acts via mechanical flushing; solution composition supports preservative effectiveness. Healthcare providers apply the solution via lavage to cleanse the wound bed, aiding in the removal of foreign material to support wound healing.
Clinical Evidence
Bench testing only. No clinical data. Testing included visual/chemical analysis (clarity, color, pH, viscosity), preservative effectiveness (USP <51>), bioburden (USP <61>, <62>), and biocompatibility (cytotoxicity ISO 10993-5, sensitization ISO 10993-10, irritation ISO 10993-10, acute systemic toxicity ISO 10993-11, material-mediated pyrogenicity ISO 10993-11). All tests passed.
Technological Characteristics
Aqueous solution containing hydrogen peroxide, acetic acid, disodium EDTA, and purified water. pH 3.0-3.8. Non-sterile. Single-patient use. Biocompatible per ISO 10993. Mechanical action via lavage system delivery.
Indications for Use
Indicated for cleansing and removal of foreign material, microorganisms, and debris from wounds including stage I-IV pressure ulcers, diabetic foot ulcers, post-surgical wounds, first-degree and partial-thickness burns, and grafted/donor sites. Prescription use only.
Regulatory Classification
Identification
A jet lavage is a device used to clean a wound by a pulsatile jet of sterile fluid. The device consists of the pulsing head, tubing to connect to a container of sterile fluid, and a means of propelling the fluid through the tubing, such as an electric roller pump.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
May 25, 2023
Armis Biopharma, Inc. Angela Mallery Principle Product Development Specialist 2950 E. Harmony Road Ste. 252 Fort Collins, Colorado 80528
Re: K220759
Trade/Device Name: Armis VeriCyn Wound Wash Regulation Number: 21 CFR 880.5475 Regulation Name: Jet Lavage Regulatory Class: Class II Product Code: FQH, FRO Dated: April 18, 2023 Received: April 18, 2023
Dear Angela Mallery:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Tek N. I Tek N. Lamichhane -S Lamichhane -S Date: 2023.05.25 13:09:44 -04'00' For Julie A. Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220759
Device Name VeriCyn® Wound Wash
#### Indications for Use (Describe)
The Armis VeriCyn® Wound Wash is to be used with a lavage system to create mechanical movement at the wound surface by delivery of a solution and is indicated for use in cleansing and removal of foreign material including microorganisms and debris from wounds (such as stage I-IV pressure ulcers, diabetic foot ulcers, post surgical wounds, first degree and partial thickness burns, grafted and donor sites).
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span> <span style="font-size: 10pt;"> <b> <span style="font-family: Symbol;">☑</span> </b> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| | <span> <span style="font-size: 10pt;"> <b> <span style="font-family: Symbol;">☐</span> </b> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY K220759
| Date Prepared | May 25, 2023 |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | Armis Biopharma<br>2950 E. Harmony Road Ste. 252<br>Fort Collins, CO, 80528 |
| Contact Person | Franklin Okumu, Ph.D.<br>Vice President of Product Development<br>908-635-4172 |
| Device Trade Name | VeriCyn® Wound Wash |
| Device Classification | Class 2<br>21 CFR 880.5475<br>FQH / FRO / General Hospital<br>Lavage, Jet<br>VeriCyn® Wound Wash |
| Predicate Device | K161165 Next Science™ Irrigation Solution |
| Reference Device | K151186 Stay Fresh Hydrocolloid Dressing<br>K160192 Atteris Antimicrobial Skin & Wound Cleanser Rochal Industries |
| Indications for Use | The Armis VeriCyn® Wound Wash is to be used with a lavage system to create mechanical movement at<br>the wound surface by delivery of a solution and is indicated for use in cleansing and removal of foreign<br>material including micro-organisms and debris from wounds (such as stage I-IV pressure ulcers, diabetic<br>foot ulcers, post surgical wounds, first degree and partial thickness burns, grafted and donor sites). |
| Device Description | The ARMIS VeriCyn® Wound Wash is an aqueous solution for irrigation and debridement of wounds.<br>The solution is a clear, colorless, no-odor aqueous solution that is used to remove debris, including<br>microorganisms from wounds though the use of a lavage system.<br>ARMIS VeriCyn® Wound Wash contains: Hydrogen Peroxide, Acetic Acid, Disodium EDTA and<br>Purified water.<br>VeriCyn® Wound Wash has been tested for compliance with ISO 10993. |
| Comparison of<br>Technological<br>Characteristics | Minor differences exist between VeriCyn® and Next Science™ Irrigation Solution, these differences do<br>not raise different questions of safety and effectiveness when compared to the predicate device.<br>Therefore, VeriCyn® is considered substantially equivalent to Next Science™ Irrigation Solution. |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The difference in materials and composition do not raise any new questions of safety and effectiveness. The materials of construction of VeriCyn® are non-novel to wound wash products cleared by FDA. Each ingredient has a stated purpose, literature data to support its function, and testing demonstrated there are no new questions raised related to safety or effectiveness compared to the predicate. The difference in composition do not raise any new questions of safety and effectiveness. The composition of the materials of the lavage solution are similar between the subject device and the predicate device. VeriCyn® uses a two-preservative system designed to provide preservative effectiveness for the shelf life of the product. |
| | The difference in pH do not raise any new questions of safety and effectiveness. During intended use of the product the pH of VeriCyn® increase to a higher, physiological, pH upon contact with the skin or wound; this is based on the low concentration of acetic acid. The predicate contains a higher concentration of acetic acid and sodium acetate as a pH buffer system that is used to control changes in pH and would maintain lower pH upon contact with skin or wounds than the non-buffered VeriCyn® product. Therefore, no additional risks have been identified related to pH. |
| | The difference in sterility does not raise any new questions of safety and effectiveness VeriCyn® has been tested against USP<51>. equivalent to a non-sterile reference device. |
{4}------------------------------------------------
| Equivalence Comparison | | | | |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device<br>VeriCyn® Wound Wash<br>(K220759) | Predicate Device<br>Next Science™<br>Irrigation Solution<br>(K161165) | Reference Device<br>Stay Fresh Hydrocolloid Dressing<br>(K151186) | Reference Device<br>Atteris Antimicrobial Skin &<br>Wound Cleanser Rochal<br>Industries<br>(K160192) |
| Rx / OTC | Rx | Rx | Rx and OTC | Rx and OTC |
| Product Code | FQH / FRO | FQH / FRO | FRO | FRO |
| Intended Use | Wound lavage | Wound lavage | Wound dressing | Wound Dressing |
| Indications for<br>Use | Prescription (Rx) use: The Armis<br>VeriCyn® Wound Wash is to be used<br>with a lavage system to create mechanical<br>movement at the wound surface by<br>delivery of a solution and is indicated for<br>use in cleansing and removal of foreign<br>material including micro-organisms and<br>debris from wounds (such as stage I-IV<br>pressure ulcers, diabetic foot ulcers, post<br>surgical wounds, first degree and second<br>degree partial thickness burns, grafted<br>and donor sites). | To be used with a jet<br>lavage system and is<br>indicated for use in<br>cleansing and removal<br>of debris, including<br>microorganisms, from<br>wounds. | For Over-the-Counter Use: The<br>Stay Fresh Hydrocolloid dressing<br>acts as a barrier to bacterial<br>penetration and is indicated for first<br>aid to cover minor cuts, minor<br>abrasions, and minor lacerations<br>Rx Use: Under the supervision of a<br>healthcare professional, the Stay<br>Fresh Hydrocolloid dressing is<br>intended for use as a primary<br>dressing for exuding wounds that<br>acts as a barrier to bacterial<br>penetration, for use on first and<br>second degree burns, surgical<br>wounds, pressure ulcers, dermal<br>ulcers, as well as minor cuts,<br>abrasions, lacerations. | For Over-the-Counter Use: Atteris<br>Antimicrobial Skin & Wound<br>Cleanser is intended for physical<br>cleaning and removal of dirt and<br>debris, from skin scrapes, cuts,<br>lacerations, minor irritations, exit<br>sites and unbroken skin.<br>Rx Use: Atteris Antimicrobial Skin<br>& Wound Cleanser is intended for<br>mechanical cleansing and removal<br>of debris, dirt and foreign materials,<br>including microorganisms from<br>wounds such as stage I-IV pressure<br>ulcers, diabetic foot ulcers, post-<br>surgical wounds, first and second<br>degree burns, grafted and donor<br>sites. |
| Composition | Purified Water<br>Acetic Acid<br>Hydrogen Peroxide<br>Disodium EDTA | Purified water<br>Ethanol<br>Acetic acid<br>Sodium acetate<br>Benzalkonium chloride | Hydrogen Peroxide<br>Inert polymer matrix<br>Superabsorbent particles<br>Polyurethane film | Purified Water<br>Poloxamer 407<br>Sodium Chloride<br>EDTA<br>Hypromellose<br>Sensivas<br>Polyaminopropyl<br>Biguanide |
| Color | Clear | Clear | N/A | Clear |
| Clarity | No visible particles | No visible particles | N/A | No visible particles |
| pH | 3.0-3.8 | unknown | N/A | unknown |
| Mechanism of<br>Action | Mechanical removal of debris | Mechanical removal of debris | Wound dressing | Mechanical removal of debris |
| Packaging | Single use package | Single use package | Pouch | Single use package |
| Biocompatibility | Biocompatible per ISO 10993 | Biocompatible per ISO<br>10993 | Biocompatible per ISO 10993 | Biocompatible per ISO 10993 |
| How supplied | Non-sterile | Sterile | Sterile | Non-sterile |
| Single patient<br>Use | Single patient Use | Single patient Use | Single patient Use | Single patient Use |
| Non-Clinical Tests<br>Performed | Non-Clinical testing was conducted, and results were substantially equivalent the prior non-clinical testing<br>including visual and chemical tests, biocompatibility, and a wound-wash study. |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Testing Clarity, Color, pH, and Viscosity were tested. Device met internal specifications Preservative Effectiveness USP <51> (pass)<br>Bioburden USP <61> and <62> (pass)<br>Cytotoxicity per ISO 10993-5 (pass)<br>Sensitization per ISO 10993-10 (pass)<br>Irritation ISO 10993-10 (pass)<br>Acute Systemic Toxicity per ISO 10993-11 (pass)<br>Material Mediated Pyrogenicity ISO 10993-11 (pass) |
| Conclusion | Conclusion(s) drawn from the nonclinical tests demonstrate the device is as safe, as effective, and performs<br>as well as the identified legally marketed predicate device (K161165 Next Science™ Irrigation Solution). |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.