K161165 · Next Science, LLC · FQH · Sep 7, 2016 · General Hospital
Device Facts
Record ID
K161165
Device Name
Next Science Irrigation Solution
Applicant
Next Science, LLC
Product Code
FQH · General Hospital
Decision Date
Sep 7, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5475
Device Class
Class 2
Indications for Use
Next Science™ Irrigation Solution is to be used with a jet lavage system and is indicated for use in cleansing and removal of debris, including microorganisms, from wounds.
Device Story
Next Science Irrigation Solution is an aqueous solution for wound irrigation and debridement; used to remove debris and microorganisms from wounds. Device functions via mechanical action of fluid moving across wound surface (hydrodynamic shear). Provided in 1000 mL polypropylene containers; labeled 'not for IV use'. Used in clinical settings; applied via jet lavage system. Healthcare providers use the solution to irrigate wounds; mechanical action aids in cleaning and removal of foreign material. Benefits include effective wound cleansing and debris removal, supporting wound healing.
Clinical Evidence
Bench testing and animal study provided. Biocompatibility testing included ISO Intracutaneous Reactivity Test and ISO Guinea Pig Maximization Sensitization Test. Animal study: Porcine Dermal Full-Thickness Wound Wash Study using 8 Yorkshire pigs (64 total wounds). Results showed normal wound healing comparable to control and untreated wounds over 24 days.
Technological Characteristics
Aqueous solution containing ethanol (100g/L), acetic acid (50g/L), sodium acetate (30g/L), benzalkonium chloride (1.3g/L), and purified water. Packaged in 1000 mL polypropylene containers. Classified as a surface device with limited contact per ISO 10993-1:2007. Mechanical sensing/actuation via hydrodynamic shear.
Indications for Use
Indicated for use with a jet lavage system for cleansing and removing debris, including microorganisms, from wounds.
Regulatory Classification
Identification
A jet lavage is a device used to clean a wound by a pulsatile jet of sterile fluid. The device consists of the pulsing head, tubing to connect to a container of sterile fluid, and a means of propelling the fluid through the tubing, such as an electric roller pump.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, layered on top of each other to create a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 7, 2016
Next Science, LLC Matthew Myntti, Ph.D. President 8130 Baymeadows Way West, Suite 200 Jacksonville, FL 33256
Re: K161165
Trade/Device Name: Next Science Irrigation Solution Regulation Number: 21 CFR 880.5475 Regulation Name: Jet lavage Regulatory Class: Class II Product Code: FQH Dated: August 9, 2016 Received: August 9, 2016
Dear Dr. Myntti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices
> Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| ̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣̣<br>리를 | of the looking to the first to the first to the forme the forme the forme the forme to the forme of the forme of the forme of the formation of the formation of the formation | 00 000 000 000 000 000 000 000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | the production in a stamilies nebud zich guident commont to notismoni to noticello and weiver bris<br>a mil<br>ent gribulari şenodser ned strong et belemires si noticemoni to noireline sint not emit nebud edit<br>(inisi in introlution, in surposition, in resultant that in is in isti<br>o uoticello e 'o' puodsar of painted tou si uossed e pue 'sosuods 10 louploo you yew your of the prove a<br>NOJE SEBROCA JIAMS FRATS ARA JHT OT MAO3 COTTLISMOS AUCY COMBE TOM OD<br>ວຍອາ ໄດ້ ເວີລ ການປ່ານປອກ ສາວຫາອຸດສາ ອຸດັງ ໄດ້ ໄດ້ ຢາກອາຫອງແນງອາ ດຳ ປຸ່ມາ ຂອງແຜນລາວລາວໄທ<br>้าอย่านมา BMO bilev ปุ่นตอนนว ธ รับธุรกิจ วัน รรอบม ของเธนนองก่<br>Department of Health and Human Services<br>apperwork Reduction Art (PRA) Stat<br>vog.zdyd.leb]@MelSAAAq<br>Office of Chief Information Office<br>For a ministration Drug Asministration | |
| | ค้าที่ 1 คือ ค.ศ. 1994 ค.ศ. 1994 ค.ศ. 1995 น. 1995 บทเป็นที่เกิด 2 | |
| O | (D heading 108 973 12 heat 21 CFR 800 Support D | Type of Use (Select one or both, as applicable) |
| | | |
| | ໄດ ໂຮບດຕອງ ປ່າ ຂອງຕ່າງປີ ອອນ ແລະ ອອນ ອອກສາ ລາວ ຂອງຣາຍໄລວ໌ ຄ.ສ. ຄ.ສ. 2012 ເດີຍນາ ປະຊາທິນາປຸລິຍາມາລາວ | spunow mont , sustime , sustims , from wound<br>Next<br>Indications for Use (Describe) |
| | | Nort Secience Irrigation Solution<br>Device Name |
| | | 510(k) Number (if known<br>K1616165 |
| sing the statement below<br>iration Date: January 31, 2017 | Food animbA Quog Aministration<br>dications for Us | |
| Form Approved: OMB No. 0910-0120 | гээгдэг иамин ома нттаан то тиэмтяачээ | |
{3}------------------------------------------------
## 510(k) Summary: K161165
## Next Science™ Irrigation Solution
| Submitter: | Next Science TM, LLC<br>8130 Baymeadows Way West<br>Suite 200<br>Jacksonville, Florida 32256 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Matthew Myntti, Ph.D.<br>President, Next Science TM, LLC<br>855-564-2762 ext. 1004 |
| Date Prepared: | September 2, 2016 |
| Device Trade Name: | Next Science TM Irrigation Solution |
| Device Common Name: | Irrigation Solution |
| Classification Name: | Dressing, Wound, Drug |
| Product Code: | FRO and FQH |
| Classification: | Unclassified |
| Predicate Device: | Prontosan Wound Irrigation Solution (K072876) |
| Reference Device: | Atteris Antimicrobial Skin & Wound Cleanser<br>(K160192). This device is referenced because, like<br>the proposed device, it is intended for cleansing and<br>removal of debris, including microorganisms, from<br>wounds. |
| Device Description: | Next Science TM Irrigation Solution is an aqueous<br>solution for irrigation and debridement of wounds.<br>The solution is a clear, colorless, low-odor aqueous<br>solution that is used to remove debris, including<br>microorganisms, from wounds.<br>The mechanical action of fluid moving across the<br>wound provides the mechanism of action and aids in<br>the removal of debris, including microorganisms, from<br>wounds.<br>Next Science TM Irrigation Solution will be provided in<br>a 1000 mL polypropylene plastic container with a |
{4}------------------------------------------------
single port and will be labeled "not for IV use". The container will be provided in a polyethylene overwrap and packaged with six (6) containers per case. The formulation for Next Science™ Irrigation Solution is composed of ethanol, acetic acid, sodium acetate, benzalkonium chloride and water.
- Indications For Use: Next Science™ Irrigation Solution is to be used with a jet lavage system and is indicated for use in cleansing and removal of debris, including microorganisms, from wounds.
- The proposed device and the predicate device are Substantial Equivalence: intended for use in the irrigation of wounds. Although the proposed device and the predicate device consist of different ingredients, these differences do not raise different questions of safety or effectiveness. The safety and effectiveness of the Next Science™ Irrigation Solution is supported by biocompatibility testing, functional performance testing, animal testing, and shelf-life testing conducted on the proposed device.
| Product | Next Science™ Irrigation<br>Solution | Prontosan Wound Irrigation<br>Solution (K072876) |
|----------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Company | Next Science, LLC | B. Braun |
| 510(k) Number | K161165 | K072876 |
| Classification | FQH, Jet Lavage<br>FRO, Dressing, Wound,<br>Drug | FRO, Dressing, Wound, Drug |
| Composition | Ethanol, 100g/L<br>Acetic acid, 50g/L<br>Sodium acetate, 30g/L<br>Benzalkonium chloride,<br>1.3g/L<br>Purified water | 0.1% Undecylenamidopropyl<br>betaine<br>0.1% Polyaminopropyl<br>biguanide<br>Sodium hydroxide<br>Purified water |
| Description | Next Science™ Irrigation<br>Solution is a clear, colorless<br>liquid containing ethanol,<br>acetic acid, sodium acetate, | Prontosan Wound Irrigation<br>Solution is a clear, colorless<br>liquid containing<br>undecylenamidopropyl betaine, |
{5}------------------------------------------------
| Product | Next Science™ Irrigation<br>Solution | Prontosan Wound Irrigation<br>Solution (K072876) |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | and benzalkonium chloride. | polyaminopropyl biguanide,<br>sodium hydroxide and purified<br>water. |
| Indications | Next Science Irrigation<br>Solution is to be used with a<br>jet lavage system and is<br>indicated for use in<br>cleansing and removal of<br>debris, including<br>microorganisms, from<br>wounds. | Prontosan Wound Irrigation<br>Solution is intended for cleaning<br>wounds and for moistening and<br>lubricating absorbent wound<br>dressings for ulcers, burns,<br>post-surgical wounds and<br>abrasions. |
| Mechanism(s)<br>of Action | Mechanical removal of<br>debris via hydrodynamic<br>shear.<br>The mechanical action of<br>moving across the wound<br>aids in the removal of<br>foreign material such as dirt<br>and debris. | Mechanical removal of debris<br>via hydrodynamic shear.<br>The mechanical action of<br>moving across the wound aids<br>in the removal of foreign<br>material such as dirt and debris. |
{6}------------------------------------------------
## Performance testing:
Next Science™ Irrigation Solution is categorized as a surface device with limited contact per ISO10993-1:2007, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" and FDA's Draft Guidance for Industry and Food and Drug Administration Staff "Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" (April 23, 2013).
Biocompatibility testing was conducted in accordance with International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA, and included the following:
- > ISO Intracutaneous Reactivity Test to determine if the solution was capable of causing local dermal irritation.
- > ISO Guinea Pig Maximization Sensitization Test (GLP Method for Liquid Test Article) to evaluate the solution's allergenic potential or sensitizing capacity.
Animal testing:
A Porcine Dermal Full-Thickness Wound Wash Study was conducted to evaluate the effect of the Irrigation Solution components on the wound healing process.
In the animal study conducted, eight Yorkshire pigs were used each receiving eight wounds, four on either side of the dorsal midline. The four wounds were treated with the same control or test solution, or an empty control (no treatment). The empty control article was included to serve as a baseline.
Under the conditions of the porcine wound healing study, Next Science Irrigation Solution resulted in normal wound healing and was comparable to wounds that received the control article and wounds without treatment over 24 days in this porcine full-thickness dermal wound model.
Substantial Equivalence Conclusions:
The Indications for Use statement for Next Science™ Irrigation Solution is similar to that for the predicate device, and the differences in ingredients do not raise different questions of safety or effectiveness. The performance testing and animal testing demonstrate that Next Science™ Irrigation Solution is at least as safe and effective as the predicate device. Therefore, the information in this submission demonstrates that Next Science™ Irrigation Solution is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.