K220722 · Well Lead Medical Co., Ltd. · EZD · May 22, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K220722
Device Name
PVC Hydrophilic Urethral Catheter
Applicant
Well Lead Medical Co., Ltd.
Product Code
EZD · Gastroenterology, Urology
Decision Date
May 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Well Lead PVC Hydrophilic Urethral Catheter is launched for clean intermittent catheterization-CIC treatment and indicated for use by patients with chronic urine retention. The catheter is inserted into the bladder through the urethra for emptying the bladder.
Device Story
Flexible, single-use polyvinyl chloride (PVC) urethral catheter; features polyacrylamide hydrophilic coating to facilitate insertion; includes optional water sachet for coating activation. Used in professional medical facilities and home environments by patients or clinicians. Inserted through urethra into bladder to drain urine; connects to urine collection apparatus or drains directly into toilet. Benefits patient by enabling bladder emptying in chronic retention cases. Intended for use <24 hours.
Clinical Evidence
No clinical data. Safety and performance established via bench testing, including biocompatibility (ISO 10993 series), seal integrity (ASTM F1929, F88, F1886), accelerated aging (ASTM F1980-16), and mechanical performance (flow rate, kink stability, tensile force, coefficient of friction).
Indicated for clean intermittent catheterization (CIC) in patients with chronic urine retention. Suitable for pediatric and adult patients.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Predicate Devices
Well Lead PVC Hydrophilic Urethral Catheter (K133615)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 22, 2023
Well Lead Medical Co., Ltd. Jenny Zhu RA Specialist No. 47 Guomao Avenue South, Hualong, Panyu Guangzhou, Guangdong 511434 China
Re: K220722
> Trade/Device Name: Wellead® PVC Hydrophilic Urethral Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: EZD Dated: April 21, 2022 Received: April 24, 2022
Dear Jenny Zhu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica K. Nguyen -S
Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220722
Device Name Wellead® PVC Hydrophilic Urethral Catheter
Indications for Use (Describe)
The Well Lead PVC Hydrophilic Urethral Catheter is launched for clean intermittent catheterization-CIC treatment and indicated for use by patients with chronic urine retention. The catheter is inserted into the urethra for emptying the bladder.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <div> <span>☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <span>☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summarv
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
# 1. Submitter's Information
- 510(k) Owner's Name: Well Lead Medical Co., Ltd. �
- Address: No.47, Guomao Avenue South, Hualong, Panyu, Guangzhou, China � 511434
- Tel: 86 20 8475 8878-8628 �
- Contact Person (including title): Miss. Zhiqin Liang (RA Specialist); Ms. Jenny Zhu � (RA Supervisor)
- � E-mail: liangzhigin@welllead.com.cn ; jenny zhu@welllead.com.cn
# 2. Subject Device Information
- Type of 510(k) submission: Traditional �
- � Common Name: Urethral Catheter
- � Trade Name: Wellead® PVC Hydrophilic Urethral Catheter
- � Device: Catheter, Urological
- Requlation Name: Urological catheter and accessories �
- � Regulation Medical Specialty: Gastroenterology/Urology
- Review Panel: Gastroenterology/Urology �
- Product Code: EZD �
- Regulation Number: 876.5130 �
- Regulation Class: 2 �
{4}------------------------------------------------
| | Predicate Device | Reference Device |
|----------------------|------------------------------------------------|---------------------------------------------------|
| Sponsor | WELL LEAD MEDICAL CO.<br>LTD | OASIS MEDIKAL URUNLER<br>KIMYA SANAYI VE TIC A.S. |
| Device<br>Name | Well Lead PVC Hydrophilic<br>Urethral Catheter | Urinary Catheter for<br>Intermittent Use |
| 510(k)<br>Number | K133615 | K062444 |
| Product<br>Code | EZD | EZD |
| Regulation<br>Number | 876.5130 | 876.5130 |
| Regulation<br>Class | 2 | 2 |
#### 3. Predicate Device and Reference Device Information
The predicate device has not been subject to a design related recall.
#### 4. Device Description
The device is a urethral catheter used for intermittent catheterization. The device is a flexible single use urinary catheter inserted through the urethra to drain the bladder. The device is a disposable polyvinyl chloride (PVC) catheter coated with a hydrophilic coating (polyacrylamide) and an optional water sachet is included, which can activate the lubricant coating on the PVC. The device is generally for use in professional medical facilities and home environments.
The catheters come in sizes from 6Fr – 22Fr for the Male Model, 6Fr – 22Fr for the Female Model and 10Fr – 22Fr for the Tiemann Model. The catheter can connect to two different shapes of connector (called connector I and connector II) provided with the device. The connector enables the catheter to be connected to a urine bag or other collection apparatus. Urine can also be drained directly into the toilet.
The catheter is intended to be used for less than 24 hours based on the biocompatibility
{5}------------------------------------------------
information provided. It is expected that patients will use a new intermittent catheter multiple times a day (each time urine needs to be drained).
# 5. Indications for Use
The Well Lead PVC Hydrophilic Urethral Catheter is launched for clean intermittent catheterization-CIC treatment and indicated for use by patients with chronic urine retention. The catheter is inserted into the bladder through the urethra for emptying the bladder.
# 6. Test Summary
PVC Hydrophilic Urethral Catheter has been evaluated for safety and performance by lab bench testing as follows:
- ASTM F1929 Standard Test Method for Detecting Seal Leaks in Porous Medical � Packaging by Dye Penetration
- � ASTM F88 Standard Test Method for Seal Strength of Flexible Barrier Materials
- � ASTM F1886 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems � for Medical Devices
- ISO 10993-1:2018 Biological Evaluation of Medical Devices Part 1: Evaluation and � Testing Within A Risk Management Process
- � ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021 Biological evaluation of medical devices Part 10: Tests for � irritation and skin sensitization
- ISO 10993-23:2021 Biological evaluation of medical devices Part 23: Tests for � irritation;
- ♦ ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity;
{6}------------------------------------------------
- ISO 10993-18: 2020, "Biological Evaluation of Medical Devices -Part 18: Chemical � characterization of medical device materials within a risk management process"
- � Surface finish
- Outer diameter �
- � Catheter length
- � Strength
- Connector security �
- Flow rate �
- Kink stability �
- Peak tensile force �
- � Coefficient of Friction
| Elements of<br>Comparison | Subject Device | Predicate Device | Reference Device | Remark |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Device Name<br>and Model | Well Lead PVC<br>Hydrophilic<br>Urethral<br>Catheter | Well Lead PVC<br>Hydrophilic Urethral<br>Catheter | Urinary Catheter for<br>Intermittent Use | -- |
| 510 (K)<br>Number | K220722 | K133615 | K062444 | -- |
| Regulation<br>number | 21 CFR<br>876.5130 | 21 CFR 876.5130 | 21 CFR 876.5130 | SE |
| Regulation<br>description | Urological<br>Catheter and<br>Accessories | Urological Catheter<br>and Accessories | Urological Catheter<br>and Accessories | SE |
| Product code | EZD | EZD | EZD | SE |
| Class | II | II | II | SE |
| Indication for<br>Use | The Well Lead<br>PVC Hydrophilic<br>Urethral<br>Catheter is<br>launched for<br>clean<br>intermittent | The Well Lead PVC<br>Hydrophilic Urethral<br>Catheter is<br>launched for clean<br>intermittent<br>catheterization-CIC<br>treatment and | Hi-Slip, Hi-Slip Plus<br>and Hi-Slip Kit are<br>all launched for<br>Clean Intermittent<br>Catheterization-CIC<br>treatment and<br>indicated for use by | SE |
| Elements of<br>Comparison | Subject Device | Predicate Device | Reference Device | Remark |
| | catheterization-<br>CIC treatment<br>and indicated for<br>use by patients<br>with chronic<br>urine retention.<br>The catheter is<br>inserted into the<br>bladder through<br>the urethra for<br>emptying the<br>bladder. | indicated for use by<br>patients with<br>chronic urine<br>retention. The<br>catheter is inserted<br>into the bladder<br>through the urethra<br>for emptying the<br>bladder. | patients with<br>chronic urine<br>retention. The<br>catheter is inserted<br>into the bladder<br>through the urethra<br>for emptying the<br>bladder | |
| Prescription<br>use | Yes | Yes | Yes | SE |
| Sterility | Sterile | Sterile | Sterile | SE |
| Sterilization<br>Method | Water sachet is<br>irradiation<br>sterilized, and<br>the final product<br>is EO sterilized. | EO sterilized | EO sterilized | SE<br>Note 1 |
| Use | Single Use | Single Use | Single Use | SE |
| Type and Size | Standard<br>Pediatric (6Fr,<br>8Fr, 10Fr)<br>Adult (12Fr,<br>14Fr, 16Fr,<br>18Fr, 20Fr,<br>22Fr) | Male (12Fr, 14Fr,<br>16Fr, 18Fr, 20Fr,<br>22Fr, 24Fr) | Hi-Slip (urinary<br>catheter)<br>(Male 40 cm,<br>CH08-24;<br>Female 20 cm,<br>CH08-18;<br>Boys/Pediatric 30<br>cm, CH06-10;<br>Girls/Pediatric 20<br>cm, CH06-10,<br>Tiemann 40 cm,<br>CH10-18) | SE<br>Note 2 |
| | Female<br>Pediatric (6Fr,<br>8Fr, 10Fr)<br>Adult (12 Fr, 14<br>Fr, 16Fr, 18Fr,<br>20 Fr, 22Fr) | Female (12Fr,<br>14Fr, 16Fr, 18Fr,<br>20Fr, 22Fr, 24Fr) | Hi-Slip Plus<br>(urinary catheter<br>and water sachet)<br>(Male 40 cm,<br>CH08-24;<br>Female 20 cm, | |
| Elements of<br>Comparison | Subject Device | Predicate Device | Reference Device | Remark |
| | | | CH08-18;<br>Boys/Pediatric 30<br>cm, H06-10;<br>Girls/Pediatric 20<br>cm, CH06-10) | |
| | | Pediatric (6Fr, 8Fr,<br>10Fr) | Hi-Slip Kit (urinary<br>catheter, water<br>sachet, Urine bag<br>and Iodine swab) | |
| | Tiemann (10Fr,<br>12 Fr, 14Fr,<br>16Fr, 18Fr,<br>20Fr, 22Fr) | Tiemann (6Fr, 8Fr,<br>10Fr, 12Fr, 14Fr,<br>16Fr, 18Fr, 20Fr,<br>22Fr, 24Fr) | (Male 08-18CH;<br>Female CH08-<br>16CH;<br>Pediatric Girls 06-<br>10 CH;<br>Pediatric Boys 06-<br>10 CH;<br>Tiemann 10-18 CH) | |
| Catheter<br>Materials | Polyvinylchloride<br>catheter coated<br>with<br>polyacrylamide | Polyvinylchloride<br>catheter coated<br>with<br>polyvinylpyrrolidone | Polyvinyl chloride<br>catheter coated<br>with<br>polyvinylpyrrolidone | SE<br>Note 3 |
| Components | Water sachet | None | Water sachet<br>Urine bag<br>Iodine swab | SE<br>Note 4 |
# 7. Substantial Equivalence Discussion
{7}------------------------------------------------
{8}------------------------------------------------
# Comparison in Detail(s):
# Note 1:
The terminal sterilization method of the subject device is EO Sterilization, just like the predicate device according to standard ISO 11135:2014 and has a sterility assurance level (SAL) of 10^-6. The water sachet is sterilized through radiation prior to final assembly.
# Note 2:
{9}------------------------------------------------
The subject device adds - 6, 8, and 10 Fr- for the Male and Female catheter. The predicate device also has 6, 8, and 10 Fr. The reference device covers the 8-10 Fr Male and Female catheters, as well as the Tiemann tips. FDA has cleared 6Fr catheters in previous submissions, so it does not present different questions of safety and effectiveness. The sizes 6-10 Fr are indicated for pediatric patients, just like the predicate device.
#### Note 3:
The hydrophilic coating of subject device is polyacrylamide (PAM), while the predicate device is polyvinylpyrrolidone (PVP). Performance and biocompatibility testing was conducted to show that the device is substantially equivalent to the predicate device.
#### Note 4:
The subject device has an optional water sachet accessory. The water can lubricate the PVC catheter which has hydrophilic-coating, so it is easier for the catheter to pass through the urethra during urinary catheterization and into the bladder to drain urine. The water sachet of subject device are same as the Hi-Slip Plus Type (the reference device). The water sachet are sterilized via radiation. There are no different questions of safety or effectiveness raised.
# Final Conclusion:
The technological characteristics and indication for use of the Wellead PVC Hydrophilic Urethral Catheter is substantially equivalent to the predicate device. The differences between the subject device and predicate devices do not raise different questions of safety or effectiveness.
# 8. Date of the summary prepared: May 16, 2023
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.