HI-SLIP AND HI-SLIP PLUS SINGLE USE HYDROPHILIC URINARY CATHETERS AND KIT
K062444 · Oasis Medikal Urunler Kimya Sanayi VE Tic A.S. · EZD · Feb 22, 2007 · Gastroenterology, Urology
Device Facts
Record ID
K062444
Device Name
HI-SLIP AND HI-SLIP PLUS SINGLE USE HYDROPHILIC URINARY CATHETERS AND KIT
Applicant
Oasis Medikal Urunler Kimya Sanayi VE Tic A.S.
Product Code
EZD · Gastroenterology, Urology
Decision Date
Feb 22, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Hi-Slip, Hi-Slip Plus and Hi-Slip Kit are launched for Clean Intermittent Catheterization-CIC treatment and is indicated for use by patients with chronic urine retention. The catheter is inserted into the bladder through the urethra for emptying the bladder.
Device Story
Hi-Slip family consists of single-use, hydrophilic-coated PVC urinary catheters for clean intermittent catheterization (CIC). Input: manual insertion into urethra. Transformation: hydrophilic coating (polyvinyl pyrrolidone) binds water to surface to reduce friction during insertion. Output: bladder drainage. Variants: Hi-Slip (standard), Hi-Slip Plus (includes integrated sterile water sachet for activation), Hi-Slip Kit (includes catheter, water sachet, urine collection bag, povidone iodine swab). Used in clinical or home settings by patients or caregivers. Benefits: reduced urethral friction, simplified catheterization process, sterile drainage. Healthcare providers use output (urine drainage) to manage chronic retention.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993 standards, including acute systemic toxicity (ISO 10993-11), sensitization (ISO 10993-10), genotoxicity (ISO 10993-3), cytotoxicity (ISO 10993-5), and skin irritation (ISO 10993-10). All tests met requirements.
Technological Characteristics
Materials: Medical grade polyvinyl chloride (PVC) catheter coated with hydrophilic polymer (polyvinyl pyrrolidone/PVP). Dimensions: Various lengths (20-40 cm) and French sizes (CH 06-24). Energy source: None (mechanical). Sterilization: Sterile. Connectivity: None. Software: None.
Indications for Use
Indicated for patients with chronic urine retention requiring clean intermittent catheterization to empty the bladder via urethral insertion.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
K062444
pg 1 of 9
# 510(k) SUMMARY
FEB 2 2 2007
#### 3.1. Submitter
Contact Person: Phone number:
Fax number:
Oasis Medikal Ürünler Kimya Sanayi ve Tic.A.S. Yıdızevler Mah. 4. Cad. 82. Sok. Cankaya 06550, ANKARA TURKEY Ayşe Yerlikaya +90312 4388353 +90312 4411220
Date of Preparation of the Summary: August 16th, 2006
### 3.2. Device Name
#### A-Device Name-CATHETER
Classification Name: Urological Catheter Urinary Catheter for Intermittent Use Common/Usual Name: Proprietary Name: Hi-Slip
#### B-Device Name-CATHETER
| Classification Name: | Urological Catheter |
|----------------------|---------------------------------------|
| Common/Usual Name: | Urinary Catheter for Intermittent Use |
| Proprietary Name: | Hi-Slip Plus |
# C-Device Name-CATHETER
Classification Name: Common/Usual Name: Proprietary Name:
Urological Catheter Kit Urinary Catheter Kit for Intermittent Use Hi-Slip Kit
# 3.3. Device Classification
Hi-Slip, Hi-Slip Plus and Hi-Slip Kit all have been classified by the FDA under the heading of Urological Catheters and Accessories as a Class II device in accordance with the regulations under 21 CFR 876.5130. Thev are all Prescription Devices meaning they can be sold by or on the offer of a physician.
The appropriate panel is Gastroenterology or Urology and the Product Code is GBM/Catheter,Urethral.
{1}------------------------------------------------
K 062444
J72gg
# 3.4. Device Description
The Hi-Slip Hydrophilic Urinary Catheter is a single use urinary catheter designed for Clean Intermittent Catheterization, drainage of bladder. The device consists of disposable polyvinyl chloride catheter (medical grade PVC) coated with hydrophilic polymer. When the catheter is immersed in water for 30 seconds, it becomes slippery and ready to use. The catheter is provided in a variety of sizes and lengths.
The Hi-Slip Plus Hydrophilic Urinary Catheter is a single use urinary catheter designed for Clean Intermittent Catheterization, drainage of bladder. The device consists of disposable PVC (medical grade) catheter coated with hydrophilic polymer and water sachet (sterile water). Coating is activated by the water of integrated water sachet. Water is let down by squeezing the water sachet. When the catheter is immersed in water for 30 seconds, it becomes slippery and ready to use. The PLUS is provided in a variety of sizes and lengths.
Hi-Slip Kit includes a single use coated PVC (medical arade) catheter, urine bag (medical grade PVC urine bag), water sachet (sterile water) and povidone iodine swab for cleaning urethra outlet prior to catheterization. The catheter is coated with a hydrophilic polymer (polyvynil pyrolidone, PVP) which makes the catheter surface slipperv when dipped into the water. Patient clean the urethra outlet by using povidone iodine swab and lets the water fill the catheter place in the urine bag by squeezing the sachet. When the catheter is immersed in water for 30 seconds, it becomes slippery and ready to use. Patient disposes the urine directly into urine bag. The KIT is provided in a variety of sizes and lenaths.
#### 3.5. Intended Use
Hi-Slip, Hi-Slip Plus and Hi-Slip Kit are all launched for Clean Intermittent Catheterization-CIC treatment and indicated for use by patients with chronic urine refention. The catheter is inserted into the bladder through the urethra for emptying the bladder.
# 3.6. Summary of Technological Characteristics/Statement of Substantial Equivalence:
All predicate catheters are single use and intended for single intermittent catheterization.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains handwritten text. The top line reads "Kob24444". The second line reads "Pg 3 of 9". The text appears to be written in black ink on a white background.
The catheter is pre-lubricated with a coating containing polyvinyl pyrrolidone- a kind of hydrophilic polymer (PVP), which binds the water molecules to the surface of the catheter to reduce the risk of friction while inserting the catheter into the urethra.
Hi-Slip is substantially equivalent to the following predicate devices:
- · LoFric®Single Use Urinary Catheter K896750, Astra Tech Inc.
- LoFric®Plus Single Use Urinary Catheter K012374, Astra Tech Inc. .
- SpeediCath K023254, Coloplast Corp. .
- FloaCath Catheter K000070, Rüsch International. .
Hi-Slip Plus (with water sachet) is substantially equivalent to the following predicate devices:
- · LoFric® Primo Single Use Urinary Catheter K050874, Astra Tech Inc.
- SpeediCath K023254, Coloplast Corp. .
Hi-Slip Kit (with urine bag, water sachet and povidone iodine swab) is substantially equivalent to the following predicate devices:
- · LoFric®Plus Hydro-Kit II Single Use Urinary Catheter K043241, Astra Tech Inc.
- EasiCath Set K973070, Coloplast Corp. Branded as the SureCath Set . in the US.
{3}------------------------------------------------
K062444
Ry489
# 3.7. Substantial Equivalence Comparison:
Boys/Pediatric
30 cm, CH06-10
Girls/Pediatric
20 cm, CH06-10
Tiemann 40 cm,
CH10-18
| Table 3.1.<br>Properties | Hi-Slip-Substantial Equivalence Comparison - General | | | |
|--------------------------|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Hi-Slip<br>Hydrophilic<br>Urinary<br>Catheter | LoFric® Single<br>Use Urinary<br>Catheter, Astra<br>Tech Inc. | LoFric® Plus<br>Single Use<br>Urinary Catheter,<br>Astra Tech Inc. | SpeediCath<br>Coloplast<br>Corp. |
| 510 (k)<br>number | | K896750 | K012374 | K023254 |
| Device<br>composition | Polyvinyl<br>chloride (PVC)<br>catheter<br>coated<br>with<br>polyvinylpyr-<br>rolidone. | Polyvinylchloride<br>(PVC) catheter<br>coated with<br>polyvinylpyr-<br>rolidone and<br>salt. | Polyether block<br>amide catheter<br>coated with<br>polyvinyl-<br>pyrrolidone and<br>salt. | Polyurethane<br>catheter<br>coated<br>with<br>polyvinylpyr-<br>rolidone,<br>placed in<br>a saline<br>solution<br>containing<br>polivinyl-<br>pyrrolidone. |
| Sizes | Male 40 cm,<br>CH08-24<br><br>Female 20 cm,<br>CH08-18 | Male 40 cm<br>CH 08-24<br><br>Female 15 cm<br>CH 08-14<br>Female 20 cm | Male 40 cm<br>CH 08-24<br><br>Female 20 cm<br>CH 08-18 | Male 40 cm<br>CH08-18<br><br>Female 20 cm<br>CH 06-16 |
CH 08-18
Pediatric 20 cm
CH06-10
Pediatric 30 cm
CH06-10
Tiemann 40 cm CH 10-18
Pediatric 20 cm
CHOP-10
Pediatric 30 cm
CHO6-10
Tiemann 40 cm
CH 10-18
Pediatric 20
cm
CH 06-10
Tiemann 40
cm
CH 10-14
{4}------------------------------------------------
K062444
pg 5089
| | Hi-Slip<br>Hydrophilic<br>Urinary Catheter | LoFric® Single<br>Use Urinary<br>Catheter, Astra<br>Tech Inc. | LoFric® Plus<br>Single Use Uri-<br>nary Catheter,<br>Astra Tech Inc. | SpeediCath<br>Coloplast<br>Corp. |
|------------------------------|--------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Function of<br>the<br>device | Inserted into the<br>bladder through<br>the urethra for<br>emptying the<br>bladder | Inserted into the<br>bladder through<br>the urethra for<br>emptying the<br>bladder | Inserted into<br>the bladder<br>through the<br>urethra for<br>emptying the<br>bladder | Inserted into<br>the bladder<br>through the<br>urethra for<br>emptying the<br>bladder |
| Indication<br>for use | Intermittent<br>catheterization<br>(Because of<br>chronic urine<br>retention and<br>other voiding<br>dysfunctions. | Intermittent<br>catheterization | Intermittent<br>catheterization | Chronic urine<br>re-<br>tention,<br>post-void<br>residual<br>volume (PVR)<br>and voiding<br>dysfunctions. |
| Features<br>of the<br>device | Hydrophilic<br>coated. Slippery<br>surface. Low<br>friction between<br>catheter and<br>urethral mucosa. | Hydrophilic<br>coated. Low<br>friction between<br>catheter<br>and urethral<br>mucosa. | Hydrophilic<br>coated. Low<br>friction<br>between<br>catheter<br>and urethral<br>mucosa. | Hydrophilic<br>coated.<br>Low friction<br>between<br>catheter and<br>urethral<br>mucosa.<br>Ready to use. |
| Sterility | Sterile | Sterile | Sterile | Sterile |
| Packaging | Peel Pack | Peel Pack | Peel Pack | Peel Pack |
| | | Table 3.2. Hi-Slip-Substantial Equivalence Comparison: Indications for Use |
|--|--|----------------------------------------------------------------------------|
|--|--|----------------------------------------------------------------------------|
. 11
{5}------------------------------------------------
Kob2444
Pg6qgg
| Properties | Hi-Slip Plus<br>Hydrophilic Urinary<br>Catheter | LoFric® Primo Single<br>Use Urinary<br>Catheter, Astra<br>Tech Inc. | SpeediCath<br>Coloplast<br>Corp. |
|-----------------------|-------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| 510 (k)<br>number | | K050874 | K023254 |
| Device<br>composition | Polyvinyl chloride<br>(PVC)<br>catheter coated<br>with polyvinylpyr<br>rolidone and sterile<br>water sachet | Polyvinylchloride<br>(PVC) catheter<br>coated with<br>polyvinylpyr-<br>rolidone and salt and<br>water sachet | Polyurethane<br>catheter coated<br>with polyvinylpyr-<br>rolidone, placed in<br>a saline solution<br>containing<br>polivinyl-<br>pyrrolidone. |
| Sizes | Male 40 cm,<br>CH08-24<br>Female 20 cm,<br>CH08-18 | Male 40 cm<br>CH 10-18<br>Female 20 cm<br>CH 08-18 | Male 40 cm<br>CH08-18<br>Female 20 cm<br>CH 06-16 |
| | Boys/Pediatric<br>30 cm, CH06-10<br>Girls/Pediatric<br>20 cm, CH06-10 | Pediatric 20 cm<br>CH08-10 | Pediatric 20 cm<br>CH 06-10<br>Tiemann 40 cm<br>CH 10-14 |
| | | Table 3.3. Hi-Slip Plus-Substantial Equivalence Comparison: General | | |
|------------|--|---------------------------------------------------------------------|--|--|
| Properties | | | | |
:
{6}------------------------------------------------
K062444
pg 7 of 9
| Use | Hi-Slip Plus Hydrophilic<br>Urinary Catheter | LoFric® Primo Single<br>Use Urinary<br>Catheter, Astra<br>Tech Inc. | SpeediCath<br>Coloplast<br>Corp. |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Function of<br>the<br>device | Inserted into the<br>bladder through the<br>urethra for emptying<br>the bladder | Inserted into the<br>bladder through the<br>urethra for emptying<br>the bladder | Inserted into the<br>bladder through<br>the urethra for<br>emptying the<br>bladder |
| Indication for<br>use | Intermittent<br>catheterization<br>(Because of chronic<br>urine retention and<br>other voiding<br>dysfunctions. | Intermittent<br>catheterization | Chronic urine re-<br>tention,<br>post-void residual<br>volume (PVR)<br>and voiding<br>dysfunctions. |
| Features<br>of the<br>device | Hydrophilic coated.<br>Slippery surface. Low<br>friction between<br>catheter and<br>urethral mucosa.<br>Ready to use. | Hydrophilic coated.<br>Low friction between<br>catheter<br>and urethral mucosa.<br>Ready to use. | Hydrophilic<br>coated.<br>Low friction<br>between catheter<br>and urethral<br>mucosa. Ready to<br>use. |
| Sterility | Sterile | Sterile | Sterile |
| Packaging | Peel Pack | Closed PVC | Peel Pack |
Table 3.4. Hi-Slip Plus-Substantial Equivalence Comparison: Indications for lise
:
{7}------------------------------------------------
K062444
Pg 8 of 9
| Properties: | Hi-Slip Kit-Substantial Equivalence Comparison: | | General |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Hi-Slip Hydrophilic<br>Urinary Catheter Kit | LoFric® Plus<br>Hydro-Kit II Single<br>Use Urinary<br>Catheter Astratech<br>Inc. | EasiCath Set<br>(identical to<br>the SureCath<br>Set) Coloplast<br>Corp. |
| 510 (k)<br>number | | K043241 | K973070 |
| Device<br>composition | PVC catheter coated<br>with PVP. Then sterile<br>water sachet povidone<br>iodine swab and the<br>catheter are placed<br>into the urine bag and<br>the bag is sealed. | Polyvinylchloride<br>catheter coated<br>with polyvinylpyr<br>rolidone and salt.<br>Includes water<br>sachet and urine<br>bag. | Polyvinylchloride<br>catheter coated with<br>polyvinylpyrrolidone,<br>packed<br>with an ampoule with<br>sterile saline solution and<br>sealed in a urine<br>collection bag. |
| Sizes | Male 08-18 CH<br>Female 08-16 CH<br>Pediatric Girls 06-10 CH<br>Pediatric Boys 06-10<br>CH<br>Tiemann 10-18 CH | Male CH 08-18<br>Female CH 08-18<br>Pediatric CH 06-10<br>Tiemann CH 10-18 | Male CH 08-18<br>Female CH 08-14<br>Pediatric CH 06-10 |
. . . .
### Table 3.6. Hi-Slip Kit-Substantial Equivalence Comparison: Indications for Use
| | Hi-Slip Hydrophilic<br>Urinary Catheter Kit | Lofric Plus<br>Hydro-Kit II Single<br>Use Urinary Catheter<br>Astratech Inc. | EasiCath Set<br>(identical to<br>the SureCath<br>Set) ColoplastCorp |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Function of<br>the device | Inserted into the<br>urethra till catheter<br>reaches bladder and<br>allows urine to drain<br>into urine collection<br>bag. | Inserted into the<br>urethra till catheter<br>reaches bladder and<br>allows urine to drain<br>into urine collection<br>bag. | Inserted into the urethra<br>till catheter reaches<br>bladder and allows urine<br>to drain into urine<br>collection bag. |
| Indication<br>for use | Intermittent<br>catheterization<br>because of chronic<br>urine retention and<br>other voiding<br>dysfunctions. | Chronic urine<br>retention and<br>voiding dysfunctions. | Chronic urine retention<br>and voiding<br>dysfunctions. |
| Features<br>of the<br>device | Hydrophilic coated.<br>Slippery surface. Low<br>friction between<br>catheter and<br>urethral mucosa.<br>Ready to use. | Hydrophilic coated.<br>Low friction between<br>catheter and<br>urethral mucosa.<br>Ready to use. | Hydrophilic coated. Low<br>friction between<br>catheter and urethral<br>mucosa. Ready to use. |
{8}------------------------------------------------
K062444
Jy 9 of 9
| · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·<br>Sterily | and second address of there a a supportunes of<br>Sterle | a sual challed and different were any are any and<br>when and fully for the province of<br>Sterle | A FOR A MONE AND ENGINEER AND AND<br>Sterile |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Packagina | A | of the property of the country of the first of the first of the first of the first of the first of the first of the first of the first of the first of the first of the first | 200000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 |
| | Papl | Peel Pack | 000 |
| | ปากม | assessfully and the research as a may as any assessment | a si i i i i i i a |
| | LA MA LE MINNER FOR LEASE LE / CONSULE CONSULTION CARREED IN AND AND AND AND IN AND WINNER OF | results and relation of the laboration and compress of the compress of the compress of the compress of the compress of the compress of the compress of the compressional progr | control can be and the program and the line of the line of the line of the line in the line in the line in the line in the line in the line in the line in the line in the lin |
# 3.8. Summary of Safety Testing
. . .
A summary of the safety testing performed on the coated catheter is listed below.
# Table: 3.7 Summary of Safety Tests
| Test | Reference | Results |
|--------------------------------------------------------|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Investigaton of<br>Acute Systemic<br>Toxicity Test | Bioserv Analytik<br>Und<br>Medizinprodukte | Did not cause any acute toxicity,<br>meeting the requirements of ISO<br>10933-11 |
| Closed Patch<br>Sensitization Test | | Did not cause any sensitization,<br>meeting the requirements of DIN<br>ISO 10933-10 |
| Salmonella<br>typhimurium<br>Reverse<br>Mutation Assay | | Did not cause any genotoxic<br>activity, meeting the requirements<br>of DIN ISO 10933-3 |
| Cytotoxicity<br>Assay | | Caused no<br>toxicological/biological critical<br>cell damages and growth<br>inhibition. Under these conditions<br>the test material is considered<br>non-cytotoxic and meets the<br>requirements of DIN ISO 10993-5<br>(EN 30993-5) |
| Skin Irritation<br>Test | | Did not cause any irritation under<br>the selected test conditions<br>P.I.O.:0,0., meeting the<br>requirements of DIN ISO 10993-10 |
Conclusion: 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 -
{9}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/9/Picture/16 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, facing left. The eagle is composed of thick, curved lines. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OASIS Medikal Ürünler c/o Mr. Charles H. Kyper Kyper & Associates 208 Barrington Overlook DURHAM NC 27703
FEB 2 2 2007
Re: K062444
Trade/Device Name: Hi-Slip Single Use Hydrophilic Urinary Catheter Hi-Slip Plus Single Use Hydrophilic Urinary Catheter with water sachet Hi-Slip Single Use Hydrophilic Urinary Catheter Kit with water sachet, urine bag and povidone iodine swab
Regulation Number: 21 CFR §876.5130
Regulation Name: Urological catheter and accessories Regulatory Class: II
Product Code: EZD Dated: January 24, 2007
Received: January 26, 2007
Dear Mr. Kyper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all
{10}------------------------------------------------
Page 2 - Mr. Charles Kyper
the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
In addition, we have determined that your device kit contains a povidone iodine swab which is subject to regulation as a drug.
Our substantially equivalent determination does not apply to the drug component of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug component. For information on applicable Agency requirements for marketing this drug, we suggest you contact:
Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation, please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification'' (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (240) 276-3150, or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
David A. Segerson
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number (if known): K062444
Device Name:
Hi-Slip Single Use Hydrophilic Urinary Catheter
Hi-Slip Plus Single Use Hydrophilic Urinary Catheter with water sachet
Hi-Slip Single Use Hydrophilic Urinary Catheter Kit with water sachet, urine bag and povidone iodine swab
Indications for Use:
Hi-Slip, Hi-Slip Plus and Hi-Slip Kit are launched for Clean Intermittent Catheterization-CIC treatment and is indicated for use by patients with chronic urine retention. The catheter is inserted into the bladder through the urethra for emptying the bladder.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use -(21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices 510(k) Number.
Section-2, Indication for Use Statement, Page 1/1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.