ARTSMedia In Vitro Culture Medium (AM-IVC Medium) is a culture medium intended for in vitro fertilization and in vitro culture of human gametes and embryos from fertilization until the blastocyst stage of development (day 5). The medium can also be used for embryo transfer.
Device Story
AM-IVC Medium is a sterile, single-use culture medium for human gametes and embryos; used in clinical embryology laboratories for fertilization, culture to blastocyst stage, and embryo transfer. The medium is provided in 10 mL glass bottles with Fluorotec-coated stoppers. It functions as a nutrient-rich environment supporting embryo development. Healthcare providers (embryologists/clinicians) use the medium to maintain optimal physiological conditions (pH, osmolality) for gametes and embryos outside the body. The device benefits patients by providing a controlled environment for successful fertilization and embryo development prior to transfer.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing: biocompatibility (cytotoxicity, sensitization, vaginal irritation per ISO 10993), aseptic processing validation, shelf-life stability (real-time aging to 10 months), transportation testing (ASTM D4169-22), and 1-cell Mouse Embryo Assay (MEA) showing ≥80% development to expanded blastocysts at 96 hours.
Technological Characteristics
Formulation includes physiological salts, amino acids, vitamins, calcium lactate, EDTA, gentamicin sulphate, glucose, L-glutamine, human serum albumin, sodium bicarbonate, sodium hyaluronate, sodium pyruvate, phenol red, selenite, ethanolamine, sodium citrate, and human insulin. Sterilized via aseptic filtration. pH 7.2-7.45; osmolality 257-275 mOsm/kg; endotoxin <0.05 EU/mL. Packaging: 10 mL glass bottles with Fluorotec-coated stoppers and aluminum caps. Shelf-life: 10 months at 2-8°C.
Indications for Use
Indicated for in vitro fertilization and culture of human gametes and embryos from fertilization through the blastocyst stage (day 5) and for embryo transfer procedures.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 10, 2023
ARTSMedia Denmark ApS Lotte Stroebech CEO Kongevejen 149 Virum. 2830 Denmark
Re: K220715
> Trade/Device Name: ARTSMedia In Vitro Culture Medium (AM-IVC Medium) Regulation Number: 21 CFR§ 884.6180 Regulation Name: Reproductive Media and Supplements Regulatory Class: II Product Code: MQL Dated: January 9, 2023 Received: January 11, 2023
Dear Lotte Stroebech:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Monica D. Garcia -S
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K220715
Device Name ARTSMedia In Vitro Culture Medium (AM-IVC Medium)
#### Indications for Use (Describe)
ARTSMedia In Vitro Culture Medium) is a culture medium intended for in vitro fertilization and in vitro culture of human gametes and embryos from fertilization until the blastocyst stage of development (day 5). The medium can also be used for embryo transfer.
| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 5px;"> <div style="display:inline-block; vertical-align:middle; width:15px; height:15px; border: 1px solid black;"> <div style="text-align: center;">X</div> </div> </span>Prescription Use (Part 21 CFR 801 Subpart D) </span> <span style="padding-left: 20px;"> <span style="padding-right: 5px;"> <div style="display:inline-block; vertical-align:middle; width:15px; height:15px; border: 1px solid black;"></div> </span>Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
==============================================================================================================================================================================
{3}------------------------------------------------
## 510(k) Summary K220715
510(k) Owner
ARTSMedia Denmark ApS Kongevejen 149 2830 Virum Denmark Lotte Stroebech Phone: +45 53504530
Date Prepared
Trade Name Common Name Regulation Name Regulation Number Class Product Code
Predicate Device
February 8, 2023
ARTSMedia In Vitro Culture Medium (AM-IVC Medium) Assisted Reproduction Media Reproductive Media and Supplements 21 CFR 884.6180 Class II MQL (Media, Reproductive)
K133707 ORIGIO als SAGE 1-StepTM supplemented with Human Serum Albumin
The predicate device has not been subject to a designrelated recall.
#### Device Description
ARTSMedia In Vitro Culture Medium (AM-IVC Medium) is a medium for in vitro fertilization and culture of human gametes and embryos from fertilization until the blastocyst stage of development (day 5). The medium can also be used for embryo transfer procedures.
The medium is aseptically filtered and provided in in a volume of 10 mL in pre-sterilized 10 mL glass bottles closed with Fluorotec-coated stoppers and aluminum caps. The medium has a shelflife of 10 months when stored at 2-8℃ and is for single-use only. Additional information on the formulation and specifications of AM-IVC Medium are provided in the Comparison of the Subject and Predicate Device Intended Use and Technological Characteristics section of this summary.
#### Indications for Use
ARTSMedia In Vitro Culture Medium (AM-IVC Medium) is a culture medium intended for in vitro fertilization and in vitro culture of human gametes and embryos from fertilization until the blastocyst stage of development (day 5). The medium can also be used for embryo transfer.
{4}------------------------------------------------
### Comparison of the Subject and Predicate Device Intended Use and Technological Characteristics
A comparison of the intended use and technological characteristics of the subject and predicate devices are shown in the table below:
| Parameters | K220715 | K133707 | Comparison |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | AM-IVC Medium | SAGE 1-StepTM | |
| | Subject Device | Predicate Device | |
| Indications<br>for Use | ARTSMedia In Vitro<br>Culture Medium (AM-IVC<br>Medium) is a culture<br>medium intended for in<br>vitro fertilization and in<br>vitro culture of human<br>gametes and embryos<br>from fertilization until the<br>blastocyst stage of<br>development (day 5). The<br>medium can also be used<br>for embryo transfer. | This product is intended<br>for the in vitro<br>fertilization and culture<br>of human gametes and<br>embryos from<br>fertilization until Day 5/6<br>of development. The<br>media can also be used<br>for embryo transfer. | The subject device<br>indications for use<br>statement is not<br>identical to the predicate<br>device, as it is only for<br>use up to day 5 of<br>development. However,<br>the overall intended use<br>of the subject and<br>predicate device is the<br>same (i.e., for<br>fertilization and culture<br>of embryos to the<br>blastocyst stage of<br>development and for<br>use in embryo transfer<br>procedures). |
| Conditions<br>for Use | Prescription Use Only | Prescription Use Only | Same |
| Formulation | Physiological Salts<br>Amino Acids<br>Vitamins<br>Calcium Lactate<br>EDTA<br>Gentamicin Sulphate<br>Glucose<br>L-glutamine<br>Human Serum<br>Albumin (HSA)<br>Sodium Bicarbonate<br>Sodium Hyaluronate<br>Sodium Pyruvate<br>Phenol Red<br>Selenite<br>Ethanolamine<br>Sodium citrate<br>Human Insulin (from<br>yeast) | Physiological Salts<br>Amino Acids<br>Calcium -L-Lactate<br>EDTA<br>Gentamicin Sulphate<br>Glucose<br>L-glutamine<br>Human Serum Albumin<br>(HSA)<br>Sodium Bicarbonate<br>Sodium Hyaluronate<br>Sodium Pyruvate<br>Phenol Red | Different: The<br>formulations of the<br>subject and predicate<br>devices are not the<br>same. Differences in<br>device formulations do<br>not raise different<br>questions of safety and<br>effectiveness (S&E). |
| Sterilization | Aseptic filtration | Aseptic filtration | Same |
| Sterility | No growth | No growth | Same |
| pH | 7.2-7.45 | 7.2-7.4 | Similar |
| Osmolality<br>(mOsm/kg) | 257-275 | 257-273 | Similar |
| Mouse<br>Embryo<br>Assay (MEA) | 1-cell Mouse Embryo<br>Assay (MEA) ≥ 80%<br>embryos developed to<br>expanded blastocysts at<br>96 hours. | 1-cell MEA ≥ 80% | Similar |
| Endotoxin<br>(EU/ml) | <0.05 | <0.15 | Different: The subject<br>device has a lower<br>endotoxin level than the<br>predicate device. This<br>difference in endotoxin<br>specification does not<br>raise different questions<br>of S&E. |
| Shelf life | 10 months | 14 weeks | Different: The subject<br>device has a longer<br>shelf-life than the<br>predicate device.<br>Differences in shelf-life<br>do not raise different<br>questions of S&E. |
{5}------------------------------------------------
As shown in the table above, there are differences in the indications for use statements and technological characteristics of the subject and predicate devices. However, as stated in the table, the differences in indications for use do not represent a new intended use and the differences in technological characteristics do not raise different questions of safety and effectiveness.
#### Summary of Non-Clinical Performance Testing
The following studies have been performed to support substantial equivalence to the predicate device:
- Biocompatibility testing was conducted in support of the subject device that will have direct . contact with the patient during embryo transfer procedures. Testing was conducted in accordance with the 2020 FDA guidance Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process. Testing included:
- Cytotoxicity per ISO 10993-5:2009 -
- Sensitization per ISO 10993-10:2010 -
- -Vaginal Irritation per ISO 10993-10:2010
The testing demonstrated the device formulation to be non-cytotoxic, non-sensitizing, and non-irritating.
- . Aseptic filtration and aseptic filling validation, per ISO 13408-1:2008 & A1:2013 and ISO 13408- 2:2018.
{6}------------------------------------------------
- Shelf-life testing was conducted to support a 10-month shelf-life for the subject device . through demonstration that the product specifications (shown below) were met at time 0 and after 10 months of real-time aging:
- Appearance: Clear and particulate free -
- pH, per USP <791>: 7.2-7.45
- Osmolality, per USP <785>: 257-275 mOsm/kg -
- Endotoxin, per USP <85>: < 0.05 EU/mL ।
- MEA testing, in accordance with the 2021 FDA guidance Mouse Embryo Assay for । Assisted Reproduction Technology Devices: 1-Cell Mouse Embryo Assay (MEA) ≥80% embryos developed to expanded blastocyst at 96 hours.
- Sterility, per USP <71>: No growth -
- . Transportation testing per ASTM D4169-22 and cap/seal leak testing using a method equivalent to USP <1207.2> on transportation-conditioned devices.
#### Conclusions
The results of the performance testing described above demonstrate that AM-IVC Medium is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.