SAGE 1-STEP WITH HUMAN SERUM ALBUMIN, SAGE 1-STEP WITH SERUM PROTEIN SUPPLEMENTS
Applicant
Origio A/S
Product Code
MQL · Obstetrics/Gynecology
Decision Date
Apr 16, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6180
Device Class
Class 2
Indications for Use
This product is intended for the in vitro fertilization and culture of human gametes and embryos from fertilization until Day 5/6 of development. The medium can also be used for embryo transfer.
Device Story
SAGE 1-Step™ is a ready-to-use, aseptically filtered, non-viscous culture medium for assisted reproductive technology (ART). It supports human gametes and embryos from fertilization through the blastocyst stage (Day 5/6) and is used for embryo transfer. The device is used in clinical laboratory settings by trained professionals. It provides a controlled environment containing physiological salts, amino acids, energy sources (glucose, calcium-L-lactate, sodium pyruvate), and protein supplements (Human Serum Albumin or Serum Protein Supplement). One version includes sodium hyaluronate. The medium is supplied in PETG bottles. By providing a stable, nutrient-rich environment, the device facilitates successful embryo development in vitro, ultimately supporting clinical decision-making regarding embryo selection for transfer to the patient's uterus.
Clinical Evidence
Bench testing only. Evidence includes product specification comparisons (pH, osmolality, endotoxin, sterility), Mouse Embryo Assay (MEA) results (≥80% for SAGE 1-Step™), and stability studies validating a 14-week shelf life. Biological safety evaluation conducted per ISO 10993-1.
Technological Characteristics
Ready-to-use, aseptically filtered liquid medium. Components: physiological salts, amino acids, energy sources (glucose, calcium-L-lactate, sodium pyruvate), sodium bicarbonate buffer, phenol red, gentamicin sulphate, and protein (HSA or Serum Protein Supplement). One version contains sodium hyaluronate. Packaged in 10ml/60ml PETG bottles with HDPE caps. pH 7.2-7.4; osmolality 257-273 mOsm/kg; endotoxin <0.15 EU/mL.
Indications for Use
Indicated for in vitro fertilization and culture of human gametes and embryos from fertilization through Day 5/6 of development, and for embryo transfer.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
# a CooperSurgical Company
K133707 Page 1 of 4
APR 1 6 2014
K-133707 -- SAGE 1-Step™
# 510(K) SUMMARY
# Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination
| Submitted by: | ORIGIO a/s<br>Knardrupvej 2,<br>2760 Måløv,<br>Denmark |
|--------------------------------------|---------------------------------------------------------------------------|
| Contact person: | Tove Kjær<br>Director Corporate Regulatory Affairs<br>ORIGIO a/s |
| | Phone: +45 4679 0220<br>Fax: +45 4679 0300 |
| | Date Submitted: April 10, 2014 |
| Device Identification<br>Trade name: | SAGE 1-Step™ supplemented with Human Serum Albumin |
| | SAGE 1-Step™ supplemented with Serum Protein Supplement |
| Common name: | SAGE 1-Step™ |
| Classification name: | Reproductive media and supplements (21 CFR 884.6180, Product<br>Code MQL) |
| Predicate device:<br>LifeGlobal: | LifeGlobal Global® Total® (K112083). |
#### Description
The SAGE 1-Step 11 medium is intended for in vitro fertilization and culture of human gametes and embryos from fertilization until Day 5/6 of development. The medium can also be used for transfer.
Two versions of the SAGE 1-Step™ medium are available:
- Catalogue no. 6701: SAGE 1-Step™ supplemented with Human Serum Albumin .
- Catalogue no. 6702: SAGE 1-Step™ supplemented with Serum Protein Supplement .
Both versions of SAGE 1-Step™ are aseptically filtered, light pink, non viscous solutions, which are ready to use by professionals within assisted reproduction.
The SAGE 1-Step™ media products are contained in 10 ml or 60 ml transparent Polyethylene Terephthalate Glvcol (PETG) bottles with high density polyethylene (HDPE) cap, available in a card boxes of 1 x 10 ml and 1 x 60 ml bottles. The bottles are individually labeled. The boxes also contain instruction for use provided as package insert.
{1}------------------------------------------------
# Compositions:
The different types of components included in SAGE 1-Step™ supplemented with Human Serum Albumin and SAGE™ 1 step supplemented with Serum Protein Supplement are provided in table 1.
# Indication for use
This product is intended for in vitro fertilization and culture of human gametes and embryos from fertilization until Day 5/6 of development. The medium can also be used for embryo transfer.
### Technological Characteristics
The design of the SAGE 1-Step™ as well as the predicate listed in this submission, are based on the concept to offer the embryos various components at any time, from fertilization until blastocyst stage. Table 1 compares the technological characteristics of SAGE 1-Step™ with the predicate Global® Total®. Both similarities and differences are illustrated.
SAGE 1-Step™ is for fertilization: culture and embryo transfer. Global® Total @ is intended for culture of human embryos from zygote to blastocyst and embryo transfer. In vivo the gametes and zyqote are exposed to the same environment as the spermatozoa fertilizes the oocyte and forms the zyqote in the part of the fallopian tube called the ampulla. Thus, the intended use for the SAGE 1-StepTM is considered comparable to the predicate and the differences are not considered to represent a new intended use nor does it pose any safety or effectiveness issues.
| Product | SAGE 1-Step™<br>(REF 6701) | SAGE 1-Step™<br>(REF 6702) | Global® Total® |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Intended use | This product is<br>intended for the <i>in vitro</i><br>fertilization and culture<br>of human gametes<br>and embryos from<br>fertilization until Day<br>5/6 of development.<br>The medium can also<br>be used for embryo<br>transfer. | This product is<br>intended for the <i>in vitro</i><br>fertilization and culture<br>of human gametes<br>and embryos from<br>fertilization until Day<br>5/6 of development.<br>The medium can also<br>be used for embryo<br>transfer. | Culture of human<br>embryos from zygote<br>to blastocyst, and<br>embryo transfer. |
| <b>Product specification</b> | | | |
| pH | 7.2-7.4 | 7.2-7.4 | 7.2-7.4 |
| Osmolality (mOsm/kg) | 257-273 | 257-273 | 260-270 |
| Endotoxin (EU/mL) | <0.15 | <0.15 | <0.50 |
| Sterility | No growth | No growth | negative |
| 1-cell MEA | ≥80% | ≥80% | >80% |
| <b>Formulation</b> | | | |
| Physiological salts | Magnesium Sulphate<br>Potassium Chloride<br>Potassium Phosphate<br>Sodium Chloride | Magnesium Sulphate<br>Potassium Chloride<br>Potassium Phosphate<br>Sodium Chloride | Calcium Chloride<br>Magnesium Sulphate<br>Potassium Chloride<br>Potassium Phosphate<br>Sodium Chloride |
| Amino Acids | Yes | Yes | Yes |
# Table 1 Comparison of SAGE 1-Step™ with the predicate Global® Total®
{2}------------------------------------------------
| Product | SAGE 1-Step™<br>(REF 6701) | SAGE 1-Step™<br>(REF 6702) | Global® Total® |
|----------------------------------------------|-------------------------------------------------|-------------------------------------------------|-------------------------------------------------|
| Stable form of L-Glutamine | Alanyl-Glutamine | Alanyl-Glutamine | Glycyl-Glutamine |
| Energy sources | Glucose<br>Calcium-L-Lactate<br>Sodium Pyruvate | Glucose<br>Calcium-L-Lactate<br>Sodium Pyruvate | Glucose<br>Sodium-DL-Lactate<br>Sodium Pyruvate |
| EDTA | Yes | Yes | Yes |
| Buffer | Sodium Bicarbonate | Sodium Bicarbonate | Sodium Bicarbonate |
| Phenol Red | Yes | Yes | Yes |
| Sodium hyaluronate | 0.1 mg/mL | - | - |
| Protein source | | | |
| HSA (proteins: albumin) | 5 mg/mL | | - |
| PPF (proteins: albumin, α- and β-globulins) | - | 5 mg/mL | - |
| LGPS (proteins: albumin, α- and β-globulins) | - | - | 5 mg/mL |
| Drugs | | | |
| Gentamicin sulphate | 0.01 mg/mL | 0.01 mg/mL | 0.01 mg/mL |
The base medium of SAGE 1-Step™, as well as the predicate, is similar to the first single-step medium developed by Lawitts and Biggers (1991) and later supplemented with amino acids.
The technological characteristics of SAGE 1-Step™ are comparable to those of the predicate device. The main differences are:
- Protein supplementation: SAGE 1-Step™ is available in two ready-to-use formulations with . either Serum Protein Supplementation (REF 6702) or Human Serum Albumin (REF 6701) as protein source. The protein source of the predicate is LGPS which is similar to Serum Protein Supplementation as they both contain albumin and globulins. Human Serum Albumin is different from the protein source in the predicate, it is however a standard protein supplement used in other cleared devices for continuous culture such as Irvine Scientific Continuous Single Culture Complete (K121572).
- Sodium Hyaluronate: SAGE 1-Step™ supplemented with Human Serum Albumin is further . added Sodium Hyaluronate as a supplement to albumin. Sodium Hyaluronate is a well known component in other ART media (VitroLifes G-series (K081114, K081117, K031015)) and hyaluronan is naturally found in the female reproductive tract. Thus the addition of Sodium Hyaluronate in SAGE 1-Step™ supplemented with Human Serum Albumin (REF 6701) is not considered to represent a new technology when compared to the predicate.
- Lactate: SAGE 1-Step™ contains Ca-L-Lactate whereas the predicate contains Na-DL-. Lactate.. The concentration of L-lactate in SAGE 1-Step™ is comparable with the concentration of L-lactate in the predicate.
- Magnesium Sulphate is added in a higher concentration compared to the predicate. . However other culture media e.g Quinn 's Advantage media (K002836) also contain higher concentrations of Magnesium Sulphate and it is judged that it will not influence the performance or safety compared to the predicate device .
- Calcium: Calcium is added in the form of Ca-Lactate whereas the predicate contains Na-. Lactate. The Calcium concentration in SAGE 1-Step™ is similar to the predicate and therefore it has no impact.
{3}------------------------------------------------
- Stable L-Glutamine: Both SAGE 1-Step™ and the predicate contains a stable form of glutamine. Both Alanyl-Glutamine used in SAGE 1-STEP™ and Glycyl-Glutamine used in the predicate are widely used in ART media and have a history of safe use.
The differences in composition do not impact the substantial equivalence and do not raise any new types of safety or effectiveness concern.
#### Performance data
The product specifications for SAGE 1-Step™ and the predicate are similar regarding sterility, pH, and Mouse Embryo Assay (MEA) test. However, the endotoxin levels varies. Endotoxin is <0.50 EU/mL for Global® Total®. For SAGE 1-Step™ the endotoxin limit is <0.15 EU/mL.
Regarding the osmolality, the specification limit is a bit wider for SAGE 1-Step™ (257-273 mOsm/kg) than for the predicate (260-270 mOsm/kg).
The shelf life of SAGE 1-Step™ has been validated in stability studies to 14 weeks. The parameters which have been tested in the stability studies through shelf life includes pH, osmolality, endotoxin. HSA concentration, MEA, and sterility.
In general, SAGE 1-Step" medium is subject to the same control methods and, to a significant dearee, contain the same components as the predicate device. SAGE 1-Step™ has similar handling procedures and storage conditions. Therefore, SAGE 1-Step™ is considered substantially equivalent to the predicate device Global® Total®.
#### Biocompatibility
SAGE 1-Step™ is categorized as a medium in direct contact with gametes and embryos from OHOE Y-otep " is categorized as a median in direct contact with gameter and childryou franser, it is also in direct contact with the uterus (patient). The biological safety evaluation (ISO 10993-1) demonstrates that SAGE 1-Step™ medium, consist of well tested components and is non-toxic in use. SAGE 1-Step™ is therefore considered safe for fertilization and culture of human gametes and embryos as well as transfer of embryos back into the patient (uterus).
#### Conclusion
The conclusion from the performance and safety data, intended use comparison, product formulation comparison, and test specification comparison, demonstrates that the SAGE 1-Step™ supplemented with Human Serum Albumin and SAGE 1-Step™ supplemented with Serum Protein Supplement are suitable for its intended use, and meets the criteria in the comparison to the predicate device LifeGlobal Global® Total® (K112083).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a person embracing another person, or a stylized caduceus.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 16, 2014
ORIGIO a/s Tove Kjaer Director Corporate Regulatory Affairs Knardrupvej 2 Måløv 2760 Denmark
Re: K133707
> Trade/Device Name: SAGE 1-Step™ supplemented with Human Serum Albumin and SAGE 1-Step™ supplemented with Serum Protein Supplement Regulation Number: 21 CFR§ 884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: II Product Code: MQL Dated: March 17, 2014 Received: March 19, 2014
Dear Tove Kjaer,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see aboye) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{5}------------------------------------------------
Page 2 - Tove Kjaer
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/5/Picture/7 description: The image shows the word "Benjamin" followed by the letters "FDA" in a stylized font. The letters "FDA" are enclosed in a box-like shape. The word "sher-S" is written after the letters "FDA".
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
### 510(k) Number (if known) K 133707
#### Device Name
SAGE I-Step™ with Human Serum Albumin and SAGE 1-Step™ with Serum Protein Supplement
#### Indications for Use (Describe)
These products are intended for the in vitro fertilization and culture of human gametes and embryos from fertilization uni! Day 5/6 of development. The media can also be used for transfer.
Type of Use (Select one or both, as applicable)
D Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# Please do not write below this line -- continue on a separate page if needed.
BARTER FOR FOR USE ONLY TO A WARRELLER FOR SELECT FOR A BEA 2017 11 ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------(· Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Concurrence of Center for Devices and Radiological Health (CDRH) [signature]
Image /page/6/Picture/14 description: The image shows the name "Benjamin R. Fisher" and the date "2014.04.16" followed by the time "08:03:02" and some illegible text. The text is in a simple, sans-serif font and is printed in black ink. The image appears to be a scan or photocopy of a document.
FORM FDA 3881 (9/13)
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.