K200589 · Medacta International S.A. · LLZ · Feb 19, 2021 · Radiology
Device Facts
Record ID
K200589
Device Name
MyPAO Planning Report
Applicant
Medacta International S.A.
Product Code
LLZ · Radiology
Decision Date
Feb 19, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Medacta MyPAO surgical planning report is intended to provide users with patient's pre-operative information and an estimation of the post-operative parameters after the realignment of the acetabulum following the osteotomy. It is intended to be used for the pre-operative planning phase of Peri-Acetabular Osteotomy surgeries and it consists of a report in .pdf format elaborated on the basis of Patient's Anatomical Data derived from the CT scan of the patient. The planning report is generated through a validated internal Medacta Manufacturing software and with a highly specialized engineering manual landmark selection. The Medical Device, in accordance with a written prescription of the Surgeon, is intended for the specific patient for which the case is created. Medacta MyPAO surgical planning is intended to be used on a skeletally mature population for which a PAO surgery is required.
Device Story
Device processes patient CT scans to generate PDF surgical planning reports for Peri-Acetabular Osteotomy (PAO). Workflow: surgeon uploads CT images; Medacta operators perform manual image quality control, bone segmentation, and landmark selection; software calculates anatomical parameters (e.g., LCE/MCE angles, Tonnis angle, femoral head coverage, femoral neck anteversion); report provides pre-operative analysis and proposed acetabular repositioning. Surgeons validate the plan. Used in clinical settings to assist pre-operative planning; output informs surgical decision-making for hip dysplasia correction. Benefits include standardized, patient-specific anatomical assessment and post-operative parameter estimation.
Clinical Evidence
No clinical studies were conducted. Bench testing included software verification and validation, and measurement repeatability tests (intra- and inter-observer) on PAO cases. Intraclass correlation coefficients (ICC) showed excellent agreement (95% CI > 0.9) for key hip dysplasia parameters; paired measurement differences were < ±5°.
Technological Characteristics
Image processing software; inputs CT scan data; outputs PDF report. Workflow involves manual segmentation and landmark selection by trained operators. Connectivity: cloud/networked upload of images. Software-based analysis; no specific hardware materials or energy sources described.
Indications for Use
Indicated for skeletally mature patients requiring Peri-Acetabular Osteotomy (PAO) surgery.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Move Forward 3D Motion Simulation Service (K162559)
Submission Summary (Full Text)
{0}------------------------------------------------
February 19, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG' and 'ADMINISTRATION' in blue text.
Medacta International SA % Mr. Chris Lussier Director, Quality and Regulatory Medacta USA 3973 Delp Street MEMPHIS TN 38118
Re: K200589
Trade/Device Name: MyPAO Planning Report Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: January 11, 2021 Received: January 12, 2021
Dear Mr. Lussier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name MyPAO Planning Report
### Indications for Use (Describe)
Medacta MyPAO surgical planning report is intended to provide users with patient's pre-operative information and an estimation of the post-operative parameters after the realignment of the acetabulum following the osteotomy. It is intended to be used for the pre-operative planning phase of Peri-Acetabular Osteotomy surgeries and it consists of a report in .pdf format elaborated on the basis of Patient's Anatomical Data derived from the CT scan of the patient. The planning report is generated through a validated internal Medacta Manufacturing software and with a highly specialized engineering manual landmark selection.
The Medical Device, in accordance with a written prescription of the Surgeon, is intended for the specific patient for which the case is created. Medacta MyPAO surgical planning is intended to be used on a skeletally mature population for which a PAO surgery is required.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size: 100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <span style="font-size: 100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary 2.0
### K200589
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory Affairs Manager, Medacta International SA Applicant Correspondent: Christopher Lussier, Director of Quality and Regulatory, Medacta USA Date Prepared: March 4, 2020 Date Revised: January 11, 2021
#### II. Device
| Device Proprietary Name: | MyPAO Planning Report |
|--------------------------|----------------------------------------------------|
| Common or Usual Name: | Picture archiving and communications system (PACS) |
| Classification Name: | System, Image Processing, Radiological |
| Primary Product Code: | LLZ |
| Regulation Number: | 21 CFR 892.2050 |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following device:
- Move Forward 3D Motion Simulation Service, K162559, Biomet Inc. ●
#### Device Description IV.
MyPAO uses the patients' CT scan to segment the patient's bones in order to provide the surgeons with a report (MyPAO Surgical Planning Report) with some parameters of interest that can be used to properly plan the peri-acetabular osteotomy (PAO) and the manual repositioning of the acetabulum during standard PAO procedures. Specifically, the MyPAO Planning Report includes all the relevant preoperative anatomic data of the patient's hip, a proposal of the repositioning of the acetabulum, following the manual osteotomy, and a calculation of the expected postoperative results in relation to the anatomical reference data.
{4}------------------------------------------------
#### V. Indications for Use
Medacta MyPAO surgical planning report is intended to provide users with patient's pre-operative information and an estimation of the post-operative parameters after the realignment of the acetabulum following the osteotomy.
It is intended to be used for the pre-operative planning phase of Peri-Acetabular Osteotomy surgeries and it consists of a report in .pdf format elaborated on the basis of Patient's Anatomical Data derived from the CT scan of the patient. The planning report is generated through a validated internal Medacta Manufacturing software and with a highly specialized engineering manual landmark selection.
The Medical Device, in accordance with a written prescription of the Surgeon, is intended for the sole use of the specific patient for which the case is created. Medacta MyPAO surgical planning is intended to be used on a skeletally mature population for which a PAO surgery is required.
#### Comparison of Technological Characteristics VI.
The MyPAO Planning Report and the predicate device, Move Forward 3D Motion Simulation Service (K162559), share the following characteristics:
- design concept/principle of operation; ●
- workflow;
- image processing method; ●
- user interactions. ●
The subject MyPAO Planning Report and the predicate device, Move Forward 3D Motion Simulation Service (K162559), are technologically different with respect to the specific contents of the report and the provided parameters.
A comparison of key technological features between the subject and predicate device is provided in the table below.
| Parameters | MyPAO Planning Report<br>[Subject device] | Move Forward 3D Motion Simulation<br>Service (K162559)<br>[Predicate device] |
|--------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Input images | CT images | CT or MRI images |
| Design | Image processing software providing a<br>report with parameters of interest<br>calculated on the pre-operative images<br>of the patient | Same |
| Report<br>generation<br>workflow | 1) Surgeon uploading of the patient's<br>CT images<br>2) Images quality control check | Same |
| Parameters | MyPAO Planning Report<br>[Subject device] | Move Forward 3D Motion Simulation<br>Service (K162559)<br>[Predicate device] |
| | 3) Segmentation and 3D bone model<br>reconstruction<br>4) Anatomic landmark selection<br>5) Surgery plan upload<br>6) Surgery plan validation by the surgeon<br>7) Report provision | |
| Image quality<br>control,<br>segmentation<br>and landmark<br>acquisition | Performed manually by Medacta<br>operators who have been specifically<br>trained for this purpose | Performed by Zimmer Biomet operators<br>who have been specifically trained for<br>this purpose |
| User<br>interactions | Clinicians do not interact with the image<br>analysis software directly. The image<br>analysis software is only operated by<br>Medacta operators who have been<br>specifically trained for this purpose.<br>The user is only required to upload the<br>patients' images and validate the plan | Same |
| Report content | Parameters of interest calculated on<br>• the pre-operative anatomical<br>situation of the patient's pelvis<br>• Proposed correction of the<br>pathological situation (Hip<br>dysplasia).<br>• Calculation of the same parameters<br>referred to the proposed post-op<br>position. | Parameters of interest calculated on<br>• the pre-operative anatomical situation<br>of the patient's pelvis<br>• Proposed correction of the<br>pathological FAI situation. |
| Parameters and<br>evaluation<br>provided in the<br>report | • Center Edges Angle (LCE-MCE)<br>• Tonnis angle<br>• Femoral Head coverage<br>• Range of Motion analysis<br>• Femoral neck anteversion<br>• PAO correction proposal | • Femoral Head and neck diameters<br>• Alpha angles<br>• Center Edge Angle (LCE)<br>• Femoral Head coverage<br>• Acetabular orientation analysis<br>(inclination and version)<br>• Range of Motion analysis along<br>defined path<br>• FAI treatment proposal |
{5}------------------------------------------------
### Discussion
The subject and predicate devices are substantially equivalent with reference to the intended use, design concept, workflow, image processing method and user interactions. The differences in the report contents and provided parameters do not raise new questions of safety or effectiveness as
{6}------------------------------------------------
demonstrated by design validation testing. The comparison of technological characteristics and performance data provided within this submission, shows that there are no new risks associated with the subject device design, and supports the substantial equivalence of the MyPAO Planning Report to the identified predicate device.
#### VII. Performance Data
Based on the risk analysis, verification and validation activities were conducted according to written protocols with defined acceptance criteria. The following tests are being provided in support of a substantial equivalence determination:
## Non-Clinical Studies
- software verification and validation;
- measurements repeatability test based on an analysis intra and inter-observer of two different parameters, chosen for their relevance to detect Hip Dysplasia, on different PAO cases planned with MyPAO software. Intraclass correlation coefficients (ICC) analysis revealed excellent intra-observer and intra-observer agreement for both the relevant parameters measurements, with the whole 95% confidence interval lying above 0.9. All differences between paired measurements were less than ±5°. Therefore, test results have been met the established acceptance criteria.
## Clinical Studies
- No clinical studies were conducted.
## VIII. Conclusion
The information provided above supports that the MyPAO Planning Report is as safe and effective as the identified predicate devices.
Substantial equivalence has been demonstrated through a comparison of intended use, design and technological characteristics, as well as performance evaluations provided within this submission. Therefore, it is concluded that the MyPAO Planning Report can be considered substantially equivalent to the identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.