K220698 · Panthera AB · IOR · Feb 16, 2023 · Physical Medicine
Device Facts
Record ID
K220698
Device Name
Bambino 3, S3 and S3 Swing, U3 and U3 Light, X
Applicant
Panthera AB
Product Code
IOR · Physical Medicine
Decision Date
Feb 16, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
Panthera mechanical wheelchairs are manually operated multifunctional wheelchairs designed for indoor / outdoor use and intended to provide mobility to persons that have the capability of operating a mechanical wheelchair.
Device Story
Manually operated, multifunctional mechanical wheelchairs; intended for indoor/outdoor mobility for users capable of self-propulsion. Device consists of a spoke and hub chassis frame, rubber tires, semi-solid polymer casters, and flame-retardant polyurethane-coated polyester fabric seat/back cushions. Models vary by dimensions, weight capacity, and adjustability. The 'X' model utilizes a carbon fiber reinforced composite (CFRC) chassis, while others use chrome-molly steel. Operated by the patient; no external power source or electronic interface. Provides mobility assistance to individuals with physical limitations requiring a wheelchair.
Clinical Evidence
No clinical data performed. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993-1, -5, -17, -18), static stability, brake performance, and fatigue strength (ISO 7176 series, EN 12183).
Technological Characteristics
Chassis: Chrome-molly steel or carbon fiber reinforced composite (CFRC). Tires: Rubber. Casters: Semi-solid polymer. Upholstery: Flame-retardant polyurethane-coated polyester fabric over foam. Standards: ISO 10993 (biocompatibility), ISO 7176 (stability, strength, crash, ignition resistance), EN 12183 (manual wheelchair requirements). Manual operation; no energy source.
Indications for Use
Indicated for persons restricted to a sitting position who have the capability of operating a manual, multifunctional mechanical wheelchair for indoor/outdoor mobility.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 16, 2023
Panthera AB % Stuart Goldman Sr. Consultant RA/QA Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, Texas 78746
Re: K220698
Trade/Device Name: Bambino 3, S3 and S3 Swing, U3 and U3 Light, X Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I, reserved Product Code: IOR Dated: July 7, 2022 Received: July 11, 2022
Dear Stuart Goldman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Julia E. Slocomb -S Julia E. 2023.02.16 15:08:16 Slocomb -S -05'00'
for Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220698
Device Name Bambino 3, S3, S3 Swing, U3, U3 Light and X
Indications for Use (Describe)
Panthera mechanical wheelchairs are manually operated multifunctional wheelchairs designed for indoor / outdoor use and intended to provide mobility to persons that have the capability of operating a mechanical wheelchair.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Panthera Mechanical Wheelchairs (Bambino 3, S3, S3 Swing, U3, U3 Light and X)
### 1. Submission Sponsor
Panthera AB Gunnebogatan 26 SE-163 53 Spanga Stockholm, Sweden +46-8-7615040-73869 Fredrik Skantz, VP/General Manager
#### 2. Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746, USA +1.512.327.9997 Stuart R. Goldman, Sr. Consultant, RA/QA
#### 3. Date Prepared
February 15, 2023
#### 4. Device Identification
| Trade/Proprietary Name: | Bambino 3, S3, S3 Swing, U3, U3 Light and X |
|-------------------------|---------------------------------------------|
| Common/Usual Name: | Wheelchair |
| Classification Name: | Mechanical wheelchair |
| Regulation Number: | 890.3850 |
| Product Code: | IOR |
| Class: | 1 |
| Classification Panel: | Physical Medicine |
### 5. Legally Marketed Predicate Device
| Device name: | Bambino, S2, S2 Swing, U2, U2 Light |
|----------------|-------------------------------------|
| 510(k) number: | K083231 |
| Manufacturer: | Panthera AB |
The subject devices are modified versions of the predicate devices have not been subject to a design related recall. The following reference device is also included in this submission:
{4}------------------------------------------------
| Trade Name: | APEX Manual Wheelchair |
|---------------|------------------------|
| 510(k) No.: | K161425 |
| Manufacturer: | Motion Composites |
### 6. Indication for Use
Panthera mechanical wheelchairs are manually operated multifunctional wheelchairs designed for indoor / outdoor use and intended to provide mobility to persons that have the capability of operating a mechanical wheelchair.
#### Device Description 7.
Panthera mechanical wheelchairs are of a spoke and hub design and made of chrome-molly steel to construct the chassis (frame and back rest), except for the X model which uses carbon fiber reinforced composite (CFRC) instead of chrome-molly steel. All Panthera mechanical wheelchairs have rubber tires, while the casters are of a semi-solid polymer-based design. The seat cushions and backs are made from flame retardant polyurethane coated polyester fabric, over a foam core. The primary differences in the models of wheelchairs are in their dimensions, gross weight carrying capacities, adjustability features, wheels, accessories and hardware options.
#### 8. Device Modifications
The minor modifications made to the predicate Bambino, S2, S2 Swing, U2 and U2 Light model wheelchairs to create the subject Bambino 3, S3, S3 Swing, U3 and U3 Light model wheelchairs are primarily related to dimensional changes, as well as minor modifications to some of the components and accessories used on the wheelchairs. These modifications were undertaken in order to make these devices more modular in their design, construction and features than the predicate devices. In addition, the X model wheelchair was added as a product line extension of the U2 Light model, as the X model uses CFRC instead of chrome-molly steel to construct the frame chassis.
#### g. Substantial Equivalence Discussion
The following tables compare the modified subject devices to the unmodified predicate devices with respect to their intended uses, indications for use and technological characteristics and forms the basis for the determination of substantial equivalence. The subject devices do not raise any new questions of safety or effectiveness compared to the predicate devices.
| Attributes | Predicate Devices<br>(K083231) | Subject Devices | Comparison |
|--------------|----------------------------------------|------------------------------------------------|---------------------------------------------------------------------------------|
| Manufacturer | Panthera AB | Panthera AB | Same |
| Models | Bambino, S2, S2 Swing, U2, U2<br>Light | Bambino 3, S3, S3 Swing, U3, U3<br>Light and X | Equivalent. Model X is<br>a product line<br>extension of the U2<br>light model. |
| | Table 5-1 - Comparison of Characteristics of Panthera Subject and Predicate Devices |
|--|-------------------------------------------------------------------------------------|
| | |
{5}------------------------------------------------
| Attributes | Predicate Devices<br>(K083231) | Subject Devices | Comparison |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Provide mobility to persons<br>restricted to a sitting position. | Provide mobility to persons<br>restricted to a sitting position. | Same |
| Indications for<br>Use | Panthera mechanical wheelchairs<br>are manually operated<br>multifunctional wheelchairs<br>designed for indoor / outdoor use<br>and intended to provide mobility<br>to persons that have the capability<br>of operating a mechanical<br>wheelchair. | Panthera mechanical wheelchairs<br>are manually operated<br>multifunctional wheelchairs<br>designed for indoor / outdoor use<br>and intended to provide mobility<br>to persons that have the capability<br>of operating a mechanical<br>wheelchair. | Same |
| Mode of<br>Operation | User propelled, manual,<br>mechanical wheelchair. | User propelled, manual,<br>mechanical wheelchair. | Same |
| Models<br>Bambino (G356) vs. Bambino 3 (G517) | | | |
| Frame Width<br>(cm) | 52-61 | 55-64 | Similar |
| Seat Depth<br>(cm) | 23-30 | 25-30 | Similar |
| Back Rest<br>Angle (°) | 12-(-7.7) | 11.5-(-7.5) | Similar |
| Back Rest<br>Height (cm) | 17-25 (standard)<br>27-35 (optional) | 20-28 (standard)<br>27-35 (optional) | Similar<br>Same |
| Footrest Angle<br>(°) | 103 | 100 | Similar |
| Footrest Width<br>(cm) | 29.5-38.5 | 32.5-38.5 | Similar |
| Total Weight<br>(kg) | 8.1-8.8 | 9.8-10.4 | Similar |
| Castors<br>(inches) | 5 | 4.73 | Similar |
| Models<br>S2 (G348) vs. S3 (G548) | | | |
| Back Rest<br>Angle (°) | 11.5-(-7.3) | 17.3-(-5) | Similar |
| Footrest Width<br>(cm) | 46-64 | 46-63 | Similar |
| Total Weight<br>(kg) | 8.3-8.8 | 8.2-8.8 | Similar |
| Attributes | Predicate Devices<br>(K083231) | Subject Devices | Comparison |
| Castors<br>(inches) | 5 | 4.73 | Similar |
| Models<br>S2 Swing (G347) vs. S3 Swing (G547) | | | |
| Total Height<br>(cm) | 64-84 | 93 | Similar |
| Back Rest<br>Angle (°) | 7-(8.5) | 13.3-(-12) | Similar |
| Frame Width<br>(cm) | 57/60/63/66 | 57/60/63/66/71 | Similar |
| Seat Width<br>(cm) | 36/39/42/45 | 36/39/42/45/50 | Similar |
| Footrest Width<br>(cm) | 49-59 | 49-63 | Similar |
| Total Weight<br>(kg) | 9.6-10.0 | 9.4-10.3 | Similar |
| Castors<br>(inches) | 5 | 4.73 | Similar |
| Models<br>U2 (G351) vs. U3 (G551) | | | |
| Total Length<br>(cm) | 84.5 | 84 | Similar |
| Back Rest<br>Angle (°) | 11.5-(-7.3) | 17.3-(-5) | Similar |
| Total Weight<br>(kg) | 8.2-8.7 | 8.0-8.5 | Similar |
| Castors<br>(inches) | 3 | 3.54 | Similar |
| Models<br>U2 Light (G346) vs. U3 Light (G557) | | | |
| Back Rest<br>Height (cm) | 22-35 | 25-35 | Similar |
| Footrest Angle<br>(°) | 100 | 100 | Same |
| Total Weight<br>(kg) | 6.8-7.2 | 6.0-6.2 | Similar |
| Castors<br>(inches) | 3 | 2.95 | Similar |
| Attributes | Predicate Devices<br>(K083231) | Subject Devices | Comparison |
| Models<br>U2 Light (G346) vs. X (G350) | | | |
| Frame<br>Material | Chrome-molly chassis | CFRC | Different. CFRC is also<br>used to construct the<br>frame of the reference<br>device (K161425). |
| Total Weight<br>(kg) | 6.8-7.2 | 4.4-4.6 | Different. The<br>reference device<br>(K161425) has a similar<br>weight (4.9 kg with<br>wheels and 4.2 kg<br>without wheels) to the<br>X model subject device<br>as a result of its CFRC<br>chassis. |
| Safety and<br>Performance<br>Testing | Biocompatibility per ISO 10993-1. Requirements and Test Methods of Manual Wheelchairs per EN 12183. Resistance to Ignition of Postural Support Devices per ISO 7176-16. | Biocompatibility per ISO 10993-1. Requirements and Test Methods of Manual Wheelchairs per EN 12183. Resistance to Ignition of Postural Support Devices per ISO 7176-16. | Same |
{6}------------------------------------------------
{7}------------------------------------------------
# 10. Non-Clinical Performance Data
To demonstrate substantial equivalence of the subject Bambino 3, S3, S3 Swing, U3, U3 Light and X model wheelchairs to the predicate Bambino, S2, S2 Swing, U2 and U2 wheelchairs, Panthera submitted their devices for the following non-clinical tests. Results confirm that the design inputs and performance specifications for the modified subject devices have been met.
- Biocompatibility testing per: .
o ISO 10993-1
- . Cytotoxicity testing per:
- o ISO 10993-5
- . Chemical Characterization testing per:
- o ISO 10993-18
- . Toxicological Risk Assessment per:
- o ISO 10993-17
- . Wheelchair Performance Testing per:
{8}------------------------------------------------
- ISO 7176-16, Resistance to Ignition of Postural Support Devices O
- EN 12183, Manual Wheelchairs. Requirements and Test Methods (equivalent to): O
- l ISO 7176-1 (static stability)
- ISO 7176-3 (brakes)
- ISO 7176-8 (static, impact and fatigue strengths)
- ISO 7176-11 (test dummies)
- ISO 7176-15 (information, documentation and labeling)
- ISO 7176-19 (crash)
- l ISO 7176-22 (Set-up procedures)
Panthera also performed the following internal quality control measures during the design, development and manufacturing of their Bambino 3, S3, S3 Swing, U3, U3 Light and X wheelchairs:
- . Compliance with the following FDA guidance document:
- Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for O Mechanical and Powered Wheelchairs, and Motorized Three-Wheeled Vehicles
- Compliance with 21 CFR Part 820 Quality System Regulation
- (including Design Controls) O
- Compliance with 21 CFR Part 801 Labeling
- Compliance with 21 CFR Part 830 Unique Device Identification
- . Dimensional Verification per:
- Panther internal specifications O
- . Risk Analysis (FMEA) per:
- o ISO 14971, Medical Devices Application of Risk Management to Medical Devices
In conclusion, the performance of the subject devices met the non-clinical bench testing and demonstrate that the device is as safe, as effective, and performs as well as or better than the predicate devices, and therefore support substantial equivalence to them.
#### 11. Clinical Performance Data
Clinical testing on the subject devices have not been performed.
#### 12. Substantial Equivalence Conclusion
The Bambino 3, S3, S3 Swing, U3, U3 Light and X wheelchairs have the same intended use and indications for use as the Bambino, S2, S2 Swing, U2 and U2 wheelchairs, and the same or similar technological characteristics. The differences in technological characteristics between the subject and predicate devices do not raise new or different questions of safety and effectiveness. Performance testing has demonstrated that the subject devices are as safe and effective as the predicate devices and substantially equivalent to them.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.