K161425 · Motion Composites · IOR · Dec 5, 2016 · Physical Medicine
Device Facts
Record ID
K161425
Device Name
APEX Manual Wheelchair
Applicant
Motion Composites
Product Code
IOR · Physical Medicine
Decision Date
Dec 5, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Indications for Use
The indications for use of the APEX Manual Wheelchair are to provide mobility to persons limited to a sitting position.
Device Story
APEX Manual Wheelchair is a manually operated, user-propelled, rigid mechanical wheelchair. Device consists of aluminum alloy or carbon fiber frame, backrest, seat/back upholstery, rear wheels with hand rims, and front swivel casters. User controls motion, speed, and direction by manual propulsion of hand rims. Device features adjustable axle plates for center of gravity and height adjustment, and a patented caster housing system allowing angle adjustment for user comfort. Intended for use by individuals with mobility limitations. No complex scientific concepts or automated functions; device function is entirely dependent on user input.
Clinical Evidence
Bench testing only. Device tested to ISO 7176-1 (static stability), ISO 7176-3 (brakes), ISO 7176-5 (dimensions/mass/maneuvering), ISO 7176-7 (seating/wheel dimensions), ISO 7176-8 (static impact/fatigue strength), and ISO 7176-15 (labeling). No clinical data provided.
Technological Characteristics
Rigid frame construction using aluminum alloy or carbon fiber. Components include aluminum wheels, steel fasteners, polyurethane tires, and fabric-covered foam upholstery. Features adjustable axle plates and patented caster housing. Manual operation; no energy source. Complies with ISO 7176 series standards.
Indications for Use
Indicated for persons limited to a sitting position requiring a manual wheelchair for mobility.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Predicate Devices
Sunrise Medical Quickie Q7 Manual Wheelchair (K123975)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 5, 2016
Motion Composites Stephane Le Beau Marketing and International sales coordinator 519 J. Oswald Forest, suite 101 Saint-Roch-de-l'Achigan, Quebec City, J0K 3H0 Canada
Re: K161425
Trade/Device Name: APEX Manual Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: October 28, 2018 Received: November 7, 2016
Dear Stephane Le Beau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161425
Device Name APEX Manual Wheelchair
Indications for Use (Describe)
The indications for use of the APEX Manual Wheelchair are to provide mobility to persons limited to a sitting position.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
APEX Manual Wheelchair
Image /page/3/Picture/1 description: The image shows the logo for Motion Composites. The logo consists of a stylized "M" inside of a circle on the left, and the words "MOTION" and "COMPOSITES" in a bold, sans-serif font on the right. Below the logo is the text "510(k) Premarket Notification".
# Motion Composites APEX Manual Wheelchair 510(k) Summary
#### l. SUBMITTER
Motion Composites 519 J - Oswald Forest Suite 101 Saint-Roch-de-l'Achigan Quebec, Canada JOK 3H0 Phone: (450) 588-6555 Fax: (450) 588-0200 Email: info@motioncomposites.com Web Site: www.motioncomposites.com
#### II. DEVICE
Name of Device: APEX Manual Wheelchair Common or Usual Name: Manual Wheelchair Classification Name: Wheelchair, Mechanical (21 CFR 890.3850) Regulatory Class: I Product Code: IOR
#### lll. PREDICATE DEVICE
Sunrise Medical Quickie Q7 Manual Wheelchair K123975 Helio C2 Manual Wheelchair K120628
#### IV. DEVICE DESCRIPTION
The Motion Composites APEX Manual Wheelchair is manually operated, user propelled, manual, mechanical wheelchairs. Their intended function and use is to provide mobility to persons limited to a sitting position.
The Apex wheelchairs are traditional rigid wheelchairs. They are made of aluminum alloy or carbon fiber. The frame which utilizes a standard geometry that creates a cantilever like rigid assembly. Upon the outside of this framework, and to the rear, are assembled two aluminum axle plates. Wheels of varying size and type are connected to the carbon fiber camber tube via stainless steel axles.
On the front end of the frame are assembled two aluminum caster mount. Caster forks are mounted to these mounts via steel axles. A variety of caster wheels and tires are then connected to the fork based on user preference.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Motion Composites. The logo consists of a stylized circle with a black swoosh inside, followed by the words "MOTION" and "COMPOSITES" stacked on top of each other in a bold, sans-serif font. Below the logo, the text "510(k) Premarket Notification" is displayed.
## Device Function
Device function is dependent solely upon the wheelchair user. It does not function on its' own in any manner. The wheelchair user controls motion, speed and direction by propelling themselves using the hand rims located on the rear wheels.
## Scientific Concepts
There are no complex scientific concepts related to the APEX manual wheelchair. It's a simple, basic, manually operated mobility device.
## Significant Physical and Performance Characteristics:
## Design:
- The APEX utilizes a 2 sided aluminum or carbon fiber frame tubes that are linked by rigid ● cross-members improving greatly rigidity over regular interconnected side frames.
- . The 2 rigid links called «Rigidizers» by Motion Composites are set at the front and rear of the frame camber tube. The 3 components of the front rigidizer tube, rear rigidizer tube and the camber tube serves to increase the rigidity of the frame. The camber tube also serves to receive the rear wheel on each side.
- A set of adjustable axle plates are attached vertically to each side frame and provide the . wheelchair with ability to move the rear wheel position forward and backward in order to give the user an adjustment as to the center of gravity of the wheelchair. The vertical position can be also adjusted to set the height of the user on the wheelchair. These rear axle plates are linked by the camber tube to limits torsion and flexion for improved propulsion.
## Materials:
Materials used are:
- Aluminum or Carbon fiber frame, support members ●
- Aluminum or Carbon fiber Components ●
- Aluminum wheels
- Steel fasteners and components
- Polyurethane tires
- Fabric covered foam upholstery ●
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Motion Composites. The logo consists of a stylized circle with a curved line inside, followed by the words "MOTION" and "COMPOSITES" stacked on top of each other in a bold, sans-serif font. Below the logo, the words "510(k) Premarket Notification" are written in a smaller font.
# Physical Properties:
The APEX consist primarily of an aluminum or carbon fiber frame assembly, a back rest frame, seat and back rest upholstery, large rear wheels with hand rims for self-propelling the chair and front swivel pivoting casters for turning.
#### V. INDICATIONS FOR USE
The indications for use of the APEX Manual Wheelchair are to provide mobility to persons limited to a sitting position. This is identical to the predicate device.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The technology and principle of operation for the APEX, Helio C2 and the Quickie Q7 are identical. They all consist of an aluminum frame or carbon fiber frame with a seat and large rear wheels with hand rims for propelling the device. Smaller, pivoting type casters are mounted on the front of the chairs for steering and turning.
Device function is dependent solely upon the wheelchair user. They do not function on their own in any manner. The wheelchair user controls motion, speed and direction by propelling themselves using the hand rims located on the rear wheels.
If both rear wheels are propelled at the same time, the chair will move forward in a straight direction. If only the left rear wheel is propelled the chair will turn to the right. If only the right rear wheel is propelled the chair will turn to the left.
## Discussion of Similarities and Differences:
The APEX wheelchairs are substantially equivalent to the predicate Quickie Q7 (K123975) or Helio C2 wheelchairs (K120628) in technology, function, performance and materials. They have the same indications for use which is to provide mobility for persons restricted in a sitting position.
The frame width, depth and the back cane heights vary slightly between the devices, but since these are only used to better fit the device to the user's needs, these slight differences do not cause any concerns for the safety and effectiveness of the weight limit is similar within 15 pounds for all devices. The Quickie Q7 is at 265 lbs. and the Helio C2 and APEX is at 250 lbs. and is clearly stated in the proposed device labeling.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Motion Composites. The logo consists of a stylized circle with a black design inside, along with the words "MOTION" and "COMPOSITES" in a bold, sans-serif font. Below the logo, the text "510(k) Premarket Notification" is displayed.
With regard to accessories and add-ons, all devices offer the same types of armrests, backrests, hangers and footblates. These accessories are made of the exact same materials for all devices and thus do not raise any questions for the safety and effectiveness of the APEX wheelchairs. The same can be said of back types and extension tubes.
Axle plates for all devices are made out of the same material; however, the actual mounting systems differ slightly. The Quickie Q7 and APEX axle plates are mounted vertically on a single tube on each side of the frame and the Helio C2 are mounted vertically on 2 tubes on each side of the frame.
The mechanical difference found on the Quickie Q7 and Apex permits a larger range of adjustment of the Center of Gravity for the user. It does not raise any concerns for the safety and effectiveness of the device and actually improves the precision of the adjustment.
The same can be said for the caster housing that holds the front wheel into place. While the caster housing of the predicate device and those of the APEX are made all made of aluminum, the mounting system of the APEX casters housing is patented and thus slightly different from the one found on the Quickie Q7 or Helio C2 wheelchairs.
This mounting system embeds the caster housing directly into the fork stem support, and allows adjusting the angle of the castes for an efficient and more comfortable position for the user. The caster housing also imbeds a level to make sure the fork and casters are perpendicular to the ground, thus providing a better overall experience to the wheelchair user when turning and moving straight forward.
This small difference as well does not raise any concerns as far as the safety of the overall product against the predicate device, since all casters offer the same effectiveness as per the APEX successfully obtaining its ISO certification.
The seat and back upholstery offering of the APEX is equivalent to the one found on the predicate devices. It is made of high-quality durable sailcloth, which consists of 1000 denier fill and 250 denier wrap.
The rear wheels and hand rims of the predicate device and the APEX wheelchairs are offered in a wide range of different options. They are made of the same materials and are widely used in the wheelchair industry. The specific type, material and size of the wheels are chosen depending on the user's needs. These differences do not raise any concerns for the safety and effectiveness of the APEX wheelchairs. We also note the wheel locks are the same on all devices. The predicated devices were tested under the older ISO 7176-8 (1st Edition, 1998-07-15) standard for parking brakes, it's uncertain that these brakes could meet the new ISO 7176-8 (2nd Edition 2014-12-15) standard. The APEX push-to-lock and Graid-Aid brakes pass the new standard but when testing the pull-to-lock parking brakes we found that they failed. For this reason, the APEX will not offer an option with the pull to lock parking brake. This does not influence the ability of the user to park the
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Motion Composites. The logo consists of a stylized circle with a wing-like shape inside, followed by the words "MOTION" and "COMPOSITES" stacked on top of each other. Below the logo, the text "510(k) Premarket Notification" is displayed.
wheelchair, it only limits the number of options of parking brakes available. For this reason, Motion Composites feel that the APEX is offering a safer parking brake option.
Based on the above discussion, Motion Composites believes that the APEX wheelchairs are substantially equivalent to the Quickie Q7 wheelchair (K123975) and the Helio C2 (K120628). While there are some differences between the APEX and its predicate, these differences are minor and do not alter the intended function and use of the device, nor do they raise any new questions pertaining to safety or effectiveness.
#### VII. PERFORMANCE DATA
The APEX Manual Wheelchairs have been tested to the following standards;
- ISO 7176-1:1999 Determination of Static Stability
- ISO 7176-3:2012 Determination of Effectiveness of Brakes
- ISO 7176-5:2008 Determination of Overall Dimensions, Mass and Maneuvering
- Space
- ISO 7176-7:1998 Determination of Seating and Wheel Dimensions
- ISO-7176-8:1998 Requirements and Test Method for Static Impact and Fatigue Strength
- ISO 7176-15 Requirements for Information Disclosure, Documentation and Labeling
The performance data for the Quickie Q7 is not published. However, on page 13 of the Quickie Q7 Owner's Manual, the following statement is made;
This wheelchair has been dynamically tested in a forward-facing mode with the specified crash test dummy, restrained by both pelvic and upper-torso belts in accordance with ANSI/RESNA WC Vol 4 Section
This statement indicates that the Sunrise Quickie Q7 has been tested to at least some of the ANS//RESNA standards. These standards are comparable to the ISO 7176 standards which were used to test the Motion Composites APEX wheelchairs. (Note that a copy of the Quickie Q7 and Helio C2 Owner's Manual are provided in Volume 10 of this submission).
Further, manual wheelchairs are subject to the 1995 FDA guidance document entitled: "Guidance Document for the Preparation of Premarket Notification [(510(k)] Applications Mechanical and Powered Wheelchairs, and Motorized Three-Wheeled Vehicles." This guidance includes various safety and performance requirements that the Quickie 7 would have to have met or addressed prior to being cleared by FDA. Meeting these requirements also serves as a basis for substantial equivalence of the APEX to its' predicate.
#### VIII. CONCLUSIONS:
Performance testing and compliance with FDA guidance for manual wheelchairs supports the safety of the APEX manual wheelchair. Since the predicate device was cleared based in part on these data, conformance with the same requirements demonstrates that the APEX will perform as intended in the specified use conditions.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Motion Composites. The logo consists of a stylized circle with a wing-like shape inside, followed by the words "MOTION COMPOSITES" in a bold, sans-serif font. Below the logo, the text "510(k) Premarket Notification" is displayed.
Further, technology, principle of operation and indications for use between APEX, the Quickie Q7 and Helio C2 are identical. This demonstrates that they perform comparably to the predicate device that is currently marketed for the same intended use. Therefore, the APEX manual wheelchair is substantially equivalent to the Quickie Q7 and Helio C2 devices.
{9}------------------------------------------------
# APEX Predicate Device Comparison
| Feature/Specification | Sunrise Medical LLC Quickie Q7<br>(K123975) | Motion Composites Helio<br>(K120628) | APEX |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Quickie Q7 Wheelchair is a<br>manually operated device intended to<br>be used as a means of mobility for<br>persons restrict to a sitting position. | The Helio wheelchair is a manually<br>operated device intended to be used<br>as a means of mobility for persons<br>restrict to a sitting position. | The APEX wheelchair is a manually<br>operated device intended to be used<br>as a means of mobility for persons<br>restrict to a sitting position. |
| Primary Materials | Aluminum | Carbon Fiber | Carbon Fiber and Aluminum |
| Folding Method | Rigid wheelchair | Collapsible Cross-brace | Rigid wheelchair |
| Frame Width | 12″ to 20″ | 14″ to 22″ | 12″ to 20″ |
| Overall width | 19″ to 27″ | 21 1/2" to 29 ½" | 18 ½" to 33 ¼" |
| Seat Depth | 12″ to 20″ | 15″ to 21″ | 12" to 20" |
| Back Heights | 8″ to 20″ | 9″ to 21″ | 9″ to 21″ |
| Weight Limit : | 265 lbs. | 250 lbs. | 250 lbs. |
| Front Seat height : | 16″ to 21″ | 13½" to 21 ¼" | 14" to 20" |
| Chair Weight<br>Standard configuration | 14. lbs.<br>Without wheels | 16.5 lbs.<br>Without footrests and wheels | 10.7 lbs. without wheels Aluminum<br>model<br>9.2 lbs. without wheels carbon fiber<br>model |
| Warranty | Lifetime on frame | Lifetime on frame | Lifetime on frame |
| Armrests | -Padded Swing Away<br>-Single Post Height Adjustable<br>-T-Shaped Armrests - Height adjustable<br>Tubular Swing-away | -Flip back armrest, removable, height<br>adjustable (by user, no tool required)<br>-T-Shaped, Single Post removable,<br>Height adjustable (by user, no tool<br>required)<br>-Tubular Swing-away | -T-Shaped armrest, removable, height<br>adjustable (by user, no tool required)<br>-Tubular Swing-away |
| Front end types | Welded to frame | Mounted to frame Swing-in Swing-out | Factory clamped to the frame |
| Back Type | -Folding Lock-Down Angle Adjustable<br>-Non-Folding Angle-Adjustable<br>-Depth Adjustable | -Straight<br>-8 degree bend back<br>-Adjustable angle- straight or 8 degree<br>bend | -Folding Lock-Down Angle Adjustable<br>straight with or with-out push handle<br>- Ultralight Folding without push<br>Handle |
| Footrest | Integrated to the frame | Flip in, flip out; 60°, 70° , 90°,<br>elevating legrest | Integrated to the frame |
| Footplates | -Aluminum Tubular<br>-Aluminum Tubular w/ Snap-on Cover<br>-Angle Adjustable Flip Back<br>-Platform<br>-Ultra Lite Platform | -Standard non adjustable<br>-Angle Adjustable Oversized | -Aluminum Tubular with or without<br>platform<br>-Platform angle adjustable<br>- Flip-back platform |
| Back Upholstery | Standard, Tension Adjustable | Standard, Tension Adjustable | Standard, Tension Adjustable |
| Axle Plates | -Adjustable COG (1"-4")<br>-Adjustable Extended COG (0"-3") | Standard adjustable<br>Extended adjustable for Amputee | Standard post type Adjustable<br>COG (0"-5") |
| Wheel sizes | 20,22,24, 25,26 " | 20 ,22,24, 25, 26 " | 20 ,22,24, 25, 26 " |
| Wheel types | -Spoke<br>-Composite Mag<br>-Spinergy<br>-Mountain | -Spoke<br>-Composite Mag<br>-Spinergy | -Spoke<br>-Composite Mag<br>-Spinergy |
| Tire types | -Pneumatic regular and high pressure<br>-Pneumatic w/ airless insert<br>-Full and low profile polyurethane | -Pneumatic regular and high pressure<br>-High and low profile polyurethane<br>-Hard Urethane | -Pneumatic regular and high pressure<br>-High and low profile polyurethane<br>-Hard Urethane |
| Handrims | -Aluminum Anodized<br>-Plastic Coated<br>-Natural Fit<br>-Spinergy Flex Rim | -Aluminum Anodized<br>-Plastic Coated<br>-Natural Fit<br>-Spinergy Flex Rim<br>-Surge Handrim | -Aluminum Anodized<br>-Plastic Coated<br>-Natural Fit<br>-Spinergy Flex Rim<br>-Surge Handrim |
| Caster Sizes | 3", 4",5", 6" | 3″, 4″,5″, 6″, 8″ | 3", 4",5", 6" |
| Feature/Specification | Sunrise Medical LLC Quickie Q7<br>(K123975) | Motion Composites Helio (K120628) | APEX Wheelchair |
| Caster types | Polyurethane, semi-pneumatic, Soft<br>Roll, Pneumatic | Composites wheels, Pneumatic, Soft<br>Roll | Composites wheels, Pneumatic, Soft<br>Roll |
| Fork Sizes | 3",4",5",6",7" | 3",4",5",7" | 4 ¾" |
| Fork Stem Sizes | Std,+3/4",+1 ½" | Std, +1, +2 | Std, +1, +2 |
| Caster Options | Multi-position fork caster, pin locks | Multi-position fork caster, pin locks | Multi-position fork caster |
| Wheel Locks | Push to lock, pull to lock, scissor lock | Push to lock, pull to lock, scissor lock | Push to lock, scissor lock |
| Anti-tip tubes | Yes | Yes | Yes |
| Target population | Restricted to a sitting position | Restricted to a sitting position | Restricted to a sitting position |
| Standards | Unknown | ISO 7176 - (1, 3, 5, 7, 8 & 15) | ISO 7176 - (1, 3, 5, 7, 8 & 15) |
{10}------------------------------------------------
# APEX Predicate Device Comparison
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.