K163174 · Boston Scientific Corporation · LOX · Dec 14, 2016 · Cardiovascular
Device Facts
Record ID
K163174
Device Name
Emerge PTCA Dilatation Catheter
Applicant
Boston Scientific Corporation
Product Code
LOX · Cardiovascular
Decision Date
Dec 14, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Emerge Over-The-Wire and Emerge Monorail PTCA Dilatation Catheters (1.20 mm balloon models) are indicated as pre-dilatation catheters in the stenotic portion of a coronary artery or bypass graft stenosis (≥70% stenosis). The Emerge Over-The-Wire and Emerge Monorail PTCA Dilatation Catheters (balloon models 1.50-4.00 mm) are indicated for the balloon catheter dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The Emerge Over-The-Wire and Emerge Monorail PTCA Dilatation Catheters (balloon models 2.00-4.00 mm) are also indicated for the post-delivery expansion of balloon expandable stents (bare metal and drug-eluting).
Device Story
Emerge PTCA Dilatation Catheter is a sterile, single-use intravascular device for coronary angioplasty. It features a dual-lumen polymeric shaft with a distal inflatable balloon and an inner lumen for guidewire advancement. Radiopaque marker bands under the balloon facilitate positioning; proximal shaft marks indicate exit from the guide catheter. Hydrophilic and hydrophobic coatings enhance insertion/withdrawal. Used in clinical settings by physicians to dilate stenotic coronary arteries or bypass grafts and expand stents. The device provides mechanical dilatation of vessels, improving myocardial perfusion and ensuring proper stent apposition. It is a mechanical tool; no electronic or software components are involved.
Clinical Evidence
Bench testing only. Performance testing included corrosion resistance, proximal mark abrasion resistance, proximal shaft marks, repeat inflation, midshaft bond tensile, and shaft/bond burst pressure. Biocompatibility and chemical characterization included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic injection, materials-mediated pyrogenicity, mutagenicity, hemolysis, PTT, in vitro hemocompatibility, complement activation, USP physicochemical, latex assay, nonvolatile residue, and SEM analysis.
Technological Characteristics
Dual-lumen polymeric shaft catheter with distal balloon; Monorail (MR) and Over-The-Wire (OTW) configurations. Materials include polymers, radiopaque marker bands, and hydrophilic/hydrophobic coatings. Sterilization via Ethylene Oxide (EO). Mechanical sensing/actuation via manual inflation/deflation. No software or electronic components.
Indications for Use
Indicated for patients requiring pre-dilatation of stenotic coronary arteries or bypass grafts (≥70% stenosis), balloon dilatation of stenotic coronary arteries or bypass grafts to improve myocardial perfusion, or post-delivery expansion of bare metal or drug-eluting stents.
Regulatory Classification
Identification
Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.
Special Controls
*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 14, 2016
Boston Scientific Corporation Jane Horan Senior Regulatory Affairs Specialist Three Scimed Place Maple Grove, Minnesota 55311
Re: K163174
Trade/Device Name: Emerge PTCA Dilatation Catheter Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter Regulatory Class: Class II Product Code: LOX Dated: November 10, 2016 Received: November 14, 2016
Dear Jane Horan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Brian D. Pullin -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K163174
Device Name Emerge PTCA Dilatation Catheter
#### Indications for Use (Describe)
The Emerge Over-The-Wire and Emerge Monorail PTCA Dilatation Catheters (1.20 mm balloon models) are indicated as pre-dilatation catheters in the stenotic portion of a coronary artery or bypass graft stenosis (≥70% stenosis).
The Emerge Over-The-Wire and Emerge Monorail PTCA Dilatation Catheters (balloon models 1.50-4.00 mm) are indicated for the balloon catheter dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion.
The Emerge Over-The-Wire and Emerge Monorail PTCA Dilatation Catheters (balloon models 2.00-4.00 mm) are also indicated for the post-delivery expansion of balloon expandable stents (bare metal and drug-eluting).
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510k Summary
# Per 21 CFR §807.92
| Common or<br>Usual Name | PTCA Dilatation Catheter |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name(s) | Emerge™ Monorail PTCA Dilatation Catheter<br>Emerge™ Push Monorail PTCA Dilatation Catheter<br>Emerge™ Over-The-Wire Monorail PTCA Dilatation Catheter<br>Emerge™ Push Over-The-Wire Monorail PTCA Dilatation Catheter |
| Product Code | LOX - Percutaneous Transluminal Coronary Angioplasty<br>(PTCA) Catheter |
| Classification of<br>Device | Class II, 21 CFR 870.5100 |
| Submitter's Name<br>and Address | Boston Scientific Corporation<br>Three Scimed Place<br>Maple Grove, MN 55311-1566 |
| Contact Name<br>and Information | Jane Horan<br>Senior Regulatory Affairs Specialist<br>Phone: 763-494-2572<br>Fax: 763-494-2222<br>Email: Jane.Horan@bsci.com |
| Date Prepared | 7 Dec. 2016 |
| Section 514 of the<br>Act Performance<br>Standards | No performance standards have been established under<br>Section 514 of the Food, Drug and Cosmetic Act for<br>percutaneous catheters. |
{4}------------------------------------------------
| Establishment<br>Registration<br>Numbers | Owner<br>/Operator: | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, MA 01752<br>ERN: 9912058 |
|------------------------------------------|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| | Manufacturing<br>Facility: | Boston Scientific Corporation<br>Two Scimed Place<br>Maple Grove, MN 55311<br>ERN: 2134265 |
| | Sterilization<br>Facilities: | Boston Scientific Corporation –<br>Coventry<br>8 Industrial Drive<br>Coventry, RI 02816<br>ERN: 1000121056 |
| | | STERIS Isomedix Services<br>3459 South Clinton Avenue<br>South Plainfield, NJ 07080<br>ERN: 2246552 |
| | | Synergy Health Ireland Ltd.<br>(Tullamore)<br>IDA Business & Technology Park<br>Tullamore, County Offaly<br>Ireland<br>ERN: 3002807314 |
| | | Synergy Health AST, SRL<br>B13. 1 Street 4, Avenue 1<br>El Coyol Free Zone<br>El Coyol, Alajeula 20102<br>Costa Rica<br>ERN: 3010273872 |
| | | Synergy Health Venlo<br>Faunalaan 38<br>Venlo Limburg, Netherlands<br>5928 RZ<br>ERN: 3009337401 |
{5}------------------------------------------------
| Predicate Devices | Emerge™ PTCA Dilatation Catheter, K113220, cleared 22<br>March 2012. |
|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Emerge™ PTCA Dilatation Catheter, K121196, cleared 31<br>August 2012. |
| | Emerge™ PTCA Dilatation Catheter, K130391, cleared 10<br>July 2013. |
| Reference<br>Device | NC Emerge™ PTCA Dilatation Catheter, K141236,<br>cleared 7 August 2014 |
| Device<br>Description | The Boston Scientific Emerge™ PTCA Dilatation Catheter is a sterile (EO), single-use, intravascular medical device for use in coronary angioplasty procedures. The catheter consists of a dual-lumen polymeric shaft with a balloon near the distal tip. The outer lumen is used for inflation of the balloon, and the inner lumen permits the use of a guidewire to facilitate advancement of the catheter to the appropriate location. The balloon is designed to provide an inflatable segment of known diameter and length at recommended pressures. The Emerge™ PTCA Dilatation Catheter is offered in both Monorail (MR) and Over-The- Wire (OTW) platforms. Marks on the proximal portion of the catheter shaft (at 90 cm and 100 cm) indicate the exit of the balloon catheter tip out of the guide catheter. Radiopaque marker bands located under the balloon aid in positioning the system during the procedure. Hydrophilic and hydrophobic coatings are applied to the balloon and catheter to enhance insertion and withdrawal performance. The Emerge™ PTCA Dilatation Catheters (1.20 mm and 1.50 mm balloon models) are available in balloon lengths from 8 mm to 20 mm. The Emerge™ PTCA Dilatation Catheters (2.00 - 4.00 mm balloon models) are available in balloon lengths from 8 mm to 30 mm. |
| Intended Use/<br>Indications for<br>Use | The Emerge Over-The-Wire and Emerge Monorail PTCA<br>Dilatation Catheters (1.20 mm balloon models) are<br>indicated as pre-dilatation catheters in the stenotic portion<br>of a coronary artery or bypass graft stenosis (≥ 70%<br>stenosis). |
| | The Emerge Over-The-Wire and Emerge Monorail PTCA<br>Dilatation Catheters (balloon models 1.50-4.00 mm) are<br>indicated for the balloon catheter dilatation of the stenotic<br>portion of a coronary artery or bypass graft stenosis for<br>the purpose of improving myocardial perfusion. |
| | The Emerge Over-The-Wire and Emerge Monorail PTCA<br>Dilatation Catheters (balloon models 2.00 - 4.00 mm) are<br>also indicated for the post-delivery expansion of balloon<br>expandable stents (bare metal and drug-eluting). |
| Comparison of<br>Required<br>Technological<br>Characteristics | The proposed Emerge™ PTCA Dilatation Catheter<br>incorporates substantially equivalent device materials and<br>design, packaging materials and design, fundamental<br>technology, manufacturing processes, sterilization<br>process and intended use as those featured in the Boston<br>Scientific predicate devices, Emerge™ PTCA Dilatation<br>Catheter K113220 (cleared March 22, 2012); Emerge™<br>PTCA Dilatation Catheter K121196 (cleared August 31,<br>2012); and Emerge™ PTCA Dilatation Catheter K130391<br>(cleared July 10, 2013). |
| | The proposed and predicate devices share the following<br>technological characteristics: |
| | Fundamental catheter design and intended use Shaft, balloon, marker band, and coating materials Packaging materials and design Sterilization method Fundamental manufacturing processes |
| | The proposed and predicate devices differ in the following<br>technological characteristic: |
| | Method of making proximal marks on the shaft |
{6}------------------------------------------------
{7}------------------------------------------------
| Summary of Non-Clinical Test Summary | Bench testing was performed to support a determination of substantial equivalence. The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the device testing. | | | | | | | | | | | | | | |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|-----------------------------------|----------------------|----------------------------|---------------------------|-------------------------------|--------------------------|---------------------|---------------------------------|-------------|--------------|---------------------|-----------|-----|
| | The following performance tests were completed on the Emerge™ PTCA Dilatation Catheter: | | | | | | | | | | | | | | |
| | <table><tr><td>Corrosion Resistance</td><td>Proximal Mark Abrasion Resistance</td></tr><tr><td>Proximal Shaft Marks</td><td>Repeat Inflation</td></tr><tr><td>Midshaft Bond Tensile</td><td>Shaft and Bond Burst Pressure</td></tr></table> | Corrosion Resistance | Proximal Mark Abrasion Resistance | Proximal Shaft Marks | Repeat Inflation | Midshaft Bond Tensile | Shaft and Bond Burst Pressure | | | | | | | | |
| Corrosion Resistance | Proximal Mark Abrasion Resistance | | | | | | | | | | | | | | |
| Proximal Shaft Marks | Repeat Inflation | | | | | | | | | | | | | | |
| Midshaft Bond Tensile | Shaft and Bond Burst Pressure | | | | | | | | | | | | | | |
| | The following biocompatibility and chemical characterization tests were completed on the Emerge™ PTCA Dilatation Catheter: | | | | | | | | | | | | | | |
| | <table><tr><td>Cytotoxicity</td><td>Partial Thromboplastin Time</td></tr><tr><td>Sensitization</td><td>In Vitro Hemocompatibility</td></tr><tr><td>Intracutaneous Reactivity</td><td>Complement Activation</td></tr><tr><td>Acute Systemic Injection</td><td>USP Physicochemical</td></tr><tr><td>Materials Mediated Pyrogenicity</td><td>Latex Assay</td></tr><tr><td>Mutagenicity</td><td>Nonvolatile Residue</td></tr><tr><td>Hemolysis</td><td>SEM</td></tr></table> | Cytotoxicity | Partial Thromboplastin Time | Sensitization | In Vitro Hemocompatibility | Intracutaneous Reactivity | Complement Activation | Acute Systemic Injection | USP Physicochemical | Materials Mediated Pyrogenicity | Latex Assay | Mutagenicity | Nonvolatile Residue | Hemolysis | SEM |
| Cytotoxicity | Partial Thromboplastin Time | | | | | | | | | | | | | | |
| Sensitization | In Vitro Hemocompatibility | | | | | | | | | | | | | | |
| Intracutaneous Reactivity | Complement Activation | | | | | | | | | | | | | | |
| Acute Systemic Injection | USP Physicochemical | | | | | | | | | | | | | | |
| Materials Mediated Pyrogenicity | Latex Assay | | | | | | | | | | | | | | |
| Mutagenicity | Nonvolatile Residue | | | | | | | | | | | | | | |
| Hemolysis | SEM | | | | | | | | | | | | | | |
| Conclusion | Based on the indications for use, technological characteristics, and safety and performance testing, the proposed Emerge™ PTCA Dilatation Catheter has been shown to be appropriate for its intended use and is considered to be substantially equivalent to Emerge™ PTCA Dilatation Catheter (K113220), (K121196) and (K130391). | | | | | | | | | | | | | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.