The Polso™ Watch is a wrist-worn device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (%SpO2), pulse rate, and respiration rate. The Polso™ Watch is a prescription-use device intended for spot-checking measurements of patients aged 18 years and older in the home use environment in non-motion conditions. The device is not intended to be used for diagnostic applications nor is it intended to independently guide treatment decisions. The device is not an apnea monitor. Physiological parameters will later be provided for remote review by a clinician.
Device Story
Wrist-worn device; uses PPG sensors facing radial artery and dorsal wrist to measure SpO2, pulse rate, and respiration rate. Operates in non-motion conditions; user-initiated spot-checking. Data displayed on device screen; transmitted via Bluetooth Low Energy (BLE) to companion Android mobile app. App stores up to 3 months of data for remote clinician review. Benefits include non-invasive physiological monitoring in home settings. Physician reviews data during clinic visits or via remote communication. Powered by rechargeable lithium-polymer battery.
Clinical Evidence
Two clinical studies conducted. SpO2/Pulse Rate study: 10 healthy adults (plus 2 additional subjects with Fitzpatrick scores IV/VI) compared against arterial blood CO-Oximetry gold standard. Respiration Rate study: 31 adult subjects with varied skin types/health conditions compared against clinician-scored End Tidal CO2 monitoring. Testing performed per ISO 80601-2-61 and FDA pulse oximeter guidance. Results demonstrate accuracy in non-motion conditions.
Technological Characteristics
Wrist-worn form factor; PPG sensing principle. AMOLED display. Rechargeable lithium-polymer battery. Connectivity: Bluetooth Low Energy (BLE) to Android mobile app. Biocompatibility per ISO 10993-1/5/10. Electrical safety/EMC per IEC 60601-1, 60601-1-2, 60601-1-11, 62133, ISO 80601-2-61. Software V&V per IEC 62304.
Indications for Use
Indicated for patients aged 18+ for spot-checking SpO2, pulse rate, and respiration rate in home environments during non-motion conditions. Not for diagnostic use, independent treatment guidance, or apnea monitoring.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 18, 2022
Chronisense Medical, Ltd. % Allison Komiyama President, AcKRS Ram+ 2251 San Diego Ave. Ste B-257 San Diego, California 92110
Re: K220351
Trade/Device Name: Polso Watch Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: October 14, 2022 Received: October 31, 2022
Dear Allison Komiyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James J. Lee -S
James J. Lee, Ph.D. Divisiion Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220351
Device Name Polso™ Watch
Indications for Use (Describe)
The Polso™ Watch is a wrist-worn device indicated for use in measuring and displaying function of arterial hemoglobin (%SpO2), pulse rate, and respiration rate. The Polso™ Watch is a prescription-use device intended for spot-checking measurements of patients aged 18 years and older in the home use environment in non-motion conditions.
The device is not intended to be used for diagnostic applications nor is it intended to independently guide treatment decisions. The device is not an apnea monitor. Physiological parameters will later be provided for remote review by a clinician.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Polso" in a sans-serif font. The letters are a muted purple color. The "o" in the word is replaced with a circle of dots, some of which are a lighter blue color.
## 510(k) Summary K220351
#### DATE PREPARED
November 10, 2022
#### MANUFACTURER AND 510(k) OWNER
ChroniSense Medical, Ltd. 2 Hacarmel St., P.O.B. 399, Yokneam Illit, Israel 2066724 Telephone: Official Contact: Bridget A. Ross, CEO
#### REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., RAC Pierre Bounaud, Ph.D. RQM+ 2251 San Diego Ave., Ste B-257, San Diego, CA 92110 Telephone: +1 (412) 816-8253 Email: akomiyama@rqmplus.com, pbounaud@rqmplus.com
#### DEVICE INFORMATION
| Proprietary Name/Trade Name: | Polso™ Watch |
|------------------------------|----------------------------------------------------------------------|
| Common Name: | Oximeter, breathing frequency monitor |
| Regulation Number: | 21 CFR 870.2700, 21 CFR 868.2375 |
| Class: | II |
| Product Code: | DQA, BZQ |
| Premarket Review: | Anesthesiology |
| Review Panel: | OHT1/ Division of Anesthesia, Respiratory, and Sleep Devices (DHT1C) |
#### PREDICATE DEVICE IDENTIFICATION
The Polso™ Watch is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary<br>Predicate | Reference<br>Device |
|---------------|----------------------------------------------------------------------|----------------------|---------------------|
| K181352 | Loop System / Spry Health, Inc. | ✓ | |
| K181956 | Masimo MightySat Rx Fingertip Pulse<br>Oximeter / Masimo Corporation | | ✓ |
The predicate devices have not been subject to a design related recall.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Polso" in a sans-serif font. The letters are a dark purple color. In the middle of the word, where the "l" should be, there is a circle of dots. Most of the dots are the same dark purple color as the letters, but one of the dots is a lighter blue color.
#### DEVICE DESCRIPTION
The Polso™ Watch is a wearable device intended for non-invasive measurements and display of pulse rate (PR), blood oxygen saturation (SpO2), and respiration rate (RR). It is a prescriptionuse device intended for spot-checking measurements of patients aged 18 years and older in the home use environment in non-motion conditions. The device is not intended to be used for diagnostic applications nor it is intended to independently guide treatment decisions.
The Polso™ Watch is worn on the wrist and features sensors that are located facing the radial artery and at the dorsal wrist, for measurement of physiological information. The Polso™ Watch screen displays measurements, battery status, Bluetooth connection status, and time of day. The Control Button allows navigation between different display screens. The Polso™ Watch is available in sizes S, M and L to cover wrist sizes from 5.5″ to 8.4″ circumference. The Polso™ Watch is powered by a rechargeable lithium-polymer battery.
The Polso™ Watch is paired with a companion mobile application (Polso™ App) on an Android device. The App is used to set up the Polso™ Watch and to review up to 3 months of measurement data.
#### INDICATIONS FOR USE
The Polso™ Watch is a wrist-worn device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (%SpO2), pulse rate, and respiration rate. The Polso™ Watch is a prescription-use device intended for spot-checking measurements of patients aged 18 years and older in the home use environment in non-motion conditions. The device is not intended to be used for diagnostic applications nor is it intended to independently guide treatment decisions. The device is not an apnea monitor. Physiological parameters will later be provided for remote review by a clinician.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
ChroniSense Medical, Ltd. believes that the Polso™ Watch is substantially equivalent to the predicate devices based on the information summarized here:
The subject device has the same intended use as the device cleared in K181352. Both devices are intended to measure pulse rate (PR), blood oxygen saturation (SpO2), and respiration rate (RR) in non-moving adult patients in home environments. The intended population for the subject device includes transitional adolescents, aged 18 to 21. For the intended use of the device, the inclusion of this pediatric population does not raise new questions of safety or effectiveness and is validated by two clinical studies.
The subject device has a similar design as the device cleared in K181352. Both devices have a 'watch' form factor that is worn on the wrist, with measurements displayed on the wearable
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Polso" in a sans-serif font. The letters are a light purple color. The "l" in "Polso" is replaced by a circle of dots, with some of the dots being a lighter blue color.
device and transmitted to another location (mobile app, web interface) for additional monitoring. Both devices have internal rechargeable batteries.
The subject device uses similar technology for measurements as the device cleared in K181352. Both devices have PPG sensors to gather data from peripheral arteries.
The main technological differences of the subject device, as compared to the predicate device cleared in K181352, are:
- . Sampling frequency. The subject device is intended for spot-checking under non-motion conditions, meaning the user determines when to check for measurements. By contrast, the predicate device is designed for nearly continuous measuring and recording of physiological data when the patient is not in motion. The difference in sampling frequency does not raise new questions of safety or effectiveness.
- . SpO2 accuracy. The subject device display range is 70% to 100% and accuracy is 2% Arms while the predicate device has the same display range, but accuracy is 3% Arms. This difference is within the values of the predicate device and is therefore considered equivalent.
- RR range. The subject device has a RR range of 5-45 BPM while the predicate device has a RR range of 4-40 RPM. This technical difference has undergone testing to ensure the subject device is as safe and effective as the predicate device.
- Data communication. Both devices transmit measurements wirelessly to either a mobile device (subject device) or a web server (predicate). The subject device uses Bluetooth low energy (BLE) methodology while the predicate uses a cellular connection. This technical difference has undergone testing to ensure the subject device is as safe and effective as the predicate device.
- Data communication to physician. The predicate device automatically transmits collected data to a remote server for review by a physician, while the subject device stores data on the Watch and the mobile application. Review of the data by the physician is done during clinic visits or data is communicated remotely to the physician by the patient. This technical difference has undergone human factor testing to ensure the subject device is as safe and effective as the predicate device. In addition, the reference device cleared in K181956 uses similar spot-checking technology in a home use environment. Collected data is stored on the reference device. Bluetooth communication to a mobile app is an optional feature for data transfer and review by a physician.
The Polso™ Watch is substantially equivalent to the predicate devices based on the information summarized in the following table:
{6}------------------------------------------------
# P:::lso
| | Subject Device | Predicate Device | Reference Device |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | ChroniSense Medical Ltd | Spry Health, Inc. | Masimo Corporation |
| | Polso™ Watch | Loop System | Masimo MightySat Rx<br>Fingertip Pulse Oximeter |
| | K220351 | K181352 | K181956 |
| Indications for<br>Use | The Polso™ Watch is a wrist-<br>worn device indicated for use<br>in measuring and displaying<br>functional oxygen saturation<br>of arterial hemoglobin<br>(%SpO2), pulse rate, and<br>respiration rate. The Polso™<br>Watch is a prescription-use<br>device intended for spot-<br>checking measurements of<br>patients aged 18 years and<br>older in the home use<br>environment in non-motion<br>conditions.<br>The device is not intended to<br>be used for diagnostic<br>applications nor is it intended<br>to independently guide<br>treatment decisions. The<br>device is not an apnea<br>monitor. Physiological<br>parameters will later be<br>provided for remote review<br>by a clinician. | The Loop System is intended<br>for adult patients in the home<br>environment for passive, non-<br>invasive, intermittent data<br>collection of physiological<br>parameters that will later be<br>transmitted to a web server<br>for remote review by a<br>clinician.<br>The Loop System measures<br>and records:<br>• arterial oxygen<br>saturation (SpO2)<br>• heart rate (HR)<br>• respiration rate (RR)<br>All of these measurements<br>are made when no motion is<br>detected by the System. The<br>Loop System device does not<br>provide physiological alarms | The Masimo MightySat Rx<br>Fingertip Pulse Oximeter is<br>intended for hospitals,<br>hospital-type facilities, home<br>environments, and transport.<br>The Masimo MightySat Rx<br>Fingertip Pulse Oximeter is<br>indicated for the noninvasive<br>spot checking of functional<br>oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse<br>rate (PR) for adult and<br>pediatric patients during both<br>no motion and motion<br>conditions, and for patients<br>who are well or poorly<br>perfused.<br>The Masimo MightySat Rx<br>Fingertip Pulse Oximeter is<br>indicated for the noninvasive<br>spot checking of respiration<br>rate (RRp) for adult patients. |
| Product Codes<br>/ Regulation<br>Number | DQA / 21 CFR 870.2700<br>BZQ / 21 CFR 868.2375 | DQA / 21 CFR 870.2700<br>BZQ / 21 CFR 868.2375 | DQA / 21 CFR 870.2700<br>BZQ / 21 CFR 868.2375 |
| Rx/OTC | Rx | Rx | Rx |
| Intended users | Adults (aged 18 and older) | Adults | Adults/pediatric (SpO2, pulse<br>rate)<br>Adults (respiration rate) |
| Intended use<br>environments | Home environment | Home environment | Hospitals, hospital-type<br>facilities,<br>home environments, and<br>transport |
| Technological Characteristics | | | |
| Device | • Polso™ Watch | • Loop band | • Fingertip device |
| Components | • Polso™ App | • Spry server | • Battery |
| | • Polso™ Watch Charger<br>Stand with micro USB cable<br>and power adapter | • Charging station | • Carrying case |
| | | | |
| Anatomical site<br>of<br>measurements | Wrist | Wrist | Finger |
| Sampling<br>frequency | Spot-checking in non-motion<br>conditions | Intermittent data collection<br>when patient is not in motion | Spot-checking in non-motion<br>and motion conditions |
| SpO2<br>measurement<br>method | Photoplethysmography | Photoplethysmography | Photoplethysmography |
| SpO2 display<br>range | 70% to 100% | 70% to 100% | 70% to 100% (no motion) |
| SpO2 Accuracy | 2% Arms | 3% Arms | 2% Arms (no motion) |
| Pulse rate /<br>heart rate<br>measurement<br>method | Photoplethysmography | Photoplethysmography | Photoplethysmography |
| Pulse rate /<br>heart rate<br>display range | 25-250 beats per minute<br>(BPM) | 25-250 BPM | 25-240 BPM |
| Pulse rate /<br>heart rate<br>accuracy | 3 BPM Arms | 3 BPM Arms | 3 BPM Arms |
| RR<br>measurement<br>method | Photoplethysmography | Photoplethysmography | Photoplethysmography |
| RR display<br>range | 5-45 breaths per minute<br>(BPM) | 4-40 respirations per minute<br>(RPM) | 4-70 respirations per minute<br>(RPM) |
| RR accuracy | 3 BPM Arms | 3 RPM Arms | 3 RPM Arms |
| Display | AMOLED color display | LED color display | OLED color display |
| Power source | Internal rechargeable battery | Internal rechargeable<br>batteries | Two 1.5V AAA batteries |
| Data<br>communication | Wireless (Bluetooth BLE<br>pairing) with Android mobile<br>device | Wireless (cellular connection)<br>via charging station to Spry<br>Server | Optional wireless (Bluetooth<br>LE) |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Polso" in a sans-serif font. The letters are a dark purple color. The "o" in "Polso" is replaced with a circle of dots, with some of the dots being a lighter blue color.
### SUMMARY OF NON-CLINICAL TESTING
The following tests were performed to demonstrate safety and performance based on current industry standards:
- . Biocompatibility testing per ISO 10993-1, 10993-5, 10993-10
- . Software V&V per IEC 62304
- Electrical Safety and Electromagnetic Compatibility testing per IEC 60601-1, IEC 60601-1-. 2, IEC 60601-1-11, IEC 62133, ISO 80601-2-61
- . Compliance to CFR 47 Part 15 Subpart C for BLE transmission
- . Bench performance testing for RR range and accuracy
The results of these tests indicate that the Polso™ Watch is substantially equivalent to the predicate device.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Polso. The logo is purple and features the word "Polso" in a sans-serif font. The "o" in "Polso" is replaced by a circle of dots, with some of the dots being a lighter blue color. The logo is simple and modern.
#### SUMMARY OF CLINICAL TESTING
Two clinical studies were conducted to demonstrate the performance of the Polso™ Watch. Evaluation included participants with a range of demographics including a mix of gender, skin tone (Fitzpatrick scale I-VI), age, and BMI (based on height and weight).
- SpO2 and Pulse Rate Validation Study. The objective of this study was to validate the SpO2 accuracy and the pulse rate accuracy of the Polso™ Watch pulse oximeter during non-motion conditions in 10 healthy adult subjects with a range of skin types (2 having a Fitzpatrick score of V or VI). The accuracy of the Polso™ Watch was compared to the gold standard, arterial blood samples assessed by SaO2 CO-Oximetry. Per FDA's request, data on two additional subjects with Fitzpatrick scores of IV and VI were provided to demonstrate the device works on the intended population.
- Respiration Rate Validation Study. The objective of this study was to validate the . respiration rate accuracy of the Polso™ Watch pulse oximeter during non-motion conditions in 31 adult subjects with a range of skin types and health conditions. The accuracy of the Polso™ Watch was compared to manual, clinician-scored End Tidal Carbon Dioxide monitoring as the gold standard.
This testing was conducted in accordance with ISO 80601-2-61 Medical electrical equipment — Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment, and FDA's Guidance Pulse Oximeters - Premarket Notifications Submissions.
#### CONCLUSION
Based on the testing performed, including biocompatibility, software V&V, EMC and electrical safety, wireless transmission testing, bench performance testing, and clinical testing, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The similar indications for use, technological characteristics, and performance characteristics for the proposed Polso™ Watch are assessed to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.