K220230 · Amsino International, Inc. · LZH · Nov 10, 2022 · General Hospital
Device Facts
Record ID
K220230
Device Name
AMSure Enteral Feeding Pump
Applicant
Amsino International, Inc.
Product Code
LZH · General Hospital
Decision Date
Nov 10, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AMSure® Enteral Feeding Pump and Feeding Set are intended to deliver nutritional formula to the gastrointestinal system of adult patients who are physically unable to eat and swallow or who are unable to get sufficient nutrition through eating and swallowing. Only intended for adult patients, not for pediatric use. The feeding pump and feeding sets are intended to be used in healthcare facilities by licensed healthcare professional users only. The AMSure® Enteral feeding pump is for use only with Amsino AMSure® feeding sets.
Device Story
Rotary peristaltic pump; delivers nutritional formula to gastrointestinal system via enteral feeding sets. Inputs: user-defined feed rates/volumes via 4.3" color LCD touch screen. Operation: stepper motor drives tubing in tension against rotor. Sensors: infrared (upstream occlusion, empty bag, free flow), ultrasonic (air-in-line), pressure (downstream occlusion). Output: controlled liquid nutrition delivery. Used in healthcare facilities by licensed professionals. Benefits: maintains patient nutrition; includes KTO (keep-to-open) feature to maintain tube patency. Features: auto-prime, intermittent/continuous modes, RFID capability, lithium-ion battery. Alarms: occlusion, air-in-line, free flow, battery/motor errors, power failure.
Clinical Evidence
No clinical data. Bench testing only: software verification/validation (Major Level of Concern), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), performance testing (flow rate accuracy, alarm detection, reliability), human factors/usability engineering, and biocompatibility (ISO 10993).
Indicated for adult patients physically unable to eat/swallow or requiring supplemental nutrition. Not for pediatric use. For use in healthcare facilities by licensed professionals only.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below.
November 10, 2022
Amsino International Inc. Jane Gao VP of R&D and RA 708 Corporate Center Drive Pomona, California 91768
Re: K220230
Trade/Device Name: AMSure Enteral Feeding Pump and Feeding Set Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: LZH Dated: October 12, 2022 Received: October 12, 2022
Dear Jane Gao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pm.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the
{1}------------------------------------------------
Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Courtney H. Lias -S
Courtney H. Lias, Ph.D. Director OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220230
Device Name AMSure® Enteral Feeding Pump
#### Indications for Use (Describe)
The AMSure® Enteral Feeding Pump is intended to work with the disposable enteral feeding sets to deliver nutritional formula to the gastrointestinal system of adult patients who are physically unable to eat and swallow or who are unable to get sufficient nutrition through eating and swallowing.
Only intended for adult patients, not for pediatric use.
The AMSure® Enteral feeding pump is intended to be used in healthcare facilities by licensed healthcare professional users only.
The AMSure® Enteral feeding pump is for use only with Amsino AMSure® feeding sets.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) SUMMARY
| Preparation Date: | 11/8/2022 |
|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer's Name: | Amsino International Inc. |
| Correspondions official : | 708 Corporate Center Drive, Pomona, CA91768<br>Jane Gao<br>VP of R&D and RA of Amsino<br>Mobile: +86 139 1614 7664 |
| Email Address: | Jane_gao@amsino.com |
| Trade Name:<br>Common or Usual Name:<br>Regulation Name:<br>Regulation Number:<br>Product Code:<br>Device Class: | AMSure® Enteral Feeding Pump and Feeding Set<br>Enteral Feeding Pump and Set<br>Infusion Pump<br>21 CFR880.5725<br>LZH<br>Class II |
| Predicate Device(s): | SENTINEL Enteral Feeding pump (K011587) |
#### Device Description:
The AMSure® enteral feeding pump is intended to work with the disposable enteral feeding sets that deliver formula via rotary peristaltic pumping system to provide nutrition for those who do not have the ability to orally ingest food
#### Indication for Use:
The AMSure® Enteral Feeding Pump is intended to work with the disposable enteral feeding sets to deliver nutritional formula to the gastrointestinal system of adult patients who are physically unable to eat and swallow or who are unable to get sufficient nutrition through eating and swallowing.
Only intended for adult patients, not for pediatric use.
The AMSure® Enteral feeding pump is intended to be used in healthcare facilities by licensed healthcare professional users only.
The AMSure® Enteral feeding pump is for use only with Amsino AMSure® feeding sets.
#### Substantial Equivalence Discussion
Intended use comparison
The table below includes a comparison of the intended use between the new device and the predicate device:
| Characteristic | Subject<br>Device: AMSure<br>Enteral feeding pump<br>K220230 | Predicate: SENTINEL Enteral<br>feeding pump K011587 | Discussion between subject<br>device and predicate |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Indications for use | The AMSure® Enteral Feeding<br>Pump and Feeding Set are<br>intended to deliver nutritional<br>formula to the gastrointestinal<br>system of adult patients who<br>are physically unable to eat and<br>swallow or who are unable to<br>get sufficient nutrition through<br>eating and swallowing.<br>Only intended for adult patients,<br>not for pediatric use.<br>The feeding pump and feeding<br>sets are intended to be used in<br>healthcare facilities by licensed<br>healthcare professional users<br>only.<br>The AMSure® Enteral feeding<br>pump is for use only with<br>Amsino AMSure® feeding sets. | The enteral feeding pump<br>accurately controls the flow of<br>liquid feeding solution to<br>patients who are unable or<br>unwilling to consume adequate<br>nutrients. Liquid feedings may<br>consist of commercially<br>prepared formulas or<br>blenderized foods and are most<br>often delivered by means of<br>nasogastric, nasoduodenal of<br>nasojenjunal feeding tubes. In<br>some cases, a surgically placed<br>esophagostomy, gastrostomy or<br>enterostomy tube may be used.<br>The pump is not intended for<br>use with blood or blood<br>products. The pump is not for<br>intravenous delivery. | No differences |
{4}------------------------------------------------
| Prescription only or over the<br>counter | Prescription only | Prescription only | No differences |
|------------------------------------------|-----------------------|-----------------------|-------------------------------------------------------------------------|
| Intended population | Adult | Adult, pediatric | Different<br>Subject device for adults uses only, not for pediatric use |
| Environment of use | Healthcare facilities | Healthcare facilities | No differences |
Discussions of differences in indications for use statement:
The subject device has same indications for use with the predicate, but not for pediatric use. This difference does not raise different questions of safety and effectiveness than the predicate device.
#### Technological characteristics
The table below includes a comparison of the technological characteristics between the new device and the predicate device:
| Characteristic | Subject Device:<br>AMSure Enteral feeding pump<br>and feeding set | Predicate:<br>SENTINEL Enteral feeding<br>pump<br>K011587 | Discussion between subject<br>device and predicate |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design (pump) | The pump incorporates a menu-<br>controlled operating system which<br>contains on board custom<br>software designed to allow the<br>user to set feed rates and<br>volumes as well as other feeding<br>options. | The pump incorporates a menu-<br>controlled operating system which<br>contains on board custom<br>software designed to allow the<br>user to set feed rates and<br>volumes as well as other feeding<br>options. | Different.<br>The hardware and software are<br>different. The subject device<br>contains 2 separate MCUs, and<br>each MCU embeds a specific<br>software in order to secure the<br>pump's running. These<br>differences do not raise different<br>questions of safety and<br>effectiveness than the predicate<br>device. |
| Mode of Action | Tubing in tension against rotor | Tubing in tension against rotor | |
| Motor type | Stepper motor | Stepper motor | |
| Sensors and flow monitoring<br>alarms | The device incorporates 3<br>sensors.<br>Infrared for<br>• Upstream occlusion/empty bag<br>• Free flow<br>Ultrasonic for<br>• Air in line<br>Pressure sensors for<br>• Downstream occlusion | The device incorporates 1 sensor.<br>Infrared sensors for<br>• Upstream occlusion/empty bag<br>• Free flow | The subject device has added an<br>ultrasonic sensor to detect air in<br>line alarm based on bubble<br>accumulation calculation; also<br>added a pressure sensor to detect<br>downstream occlusion and<br>distinguish upstream or<br>downstream occlusion.<br>The predicate device has one<br>infrared sensor to monitor all the<br>flow related events including<br>occlusion/empty bag/free flow.<br>The alarms of subject device have<br>been verified and the results<br>passed the criteria.<br>These differences do not raise<br>different questions of safety and<br>effectiveness than the predicate<br>device. |
| Pump control alarms | • Upstream Tube Alarm<br>• Downstream Occlusion<br>• Free Flow<br>• Air in Line<br>• Feeding Complete<br>• Out of Battery<br>• Battery Error<br>• Motor Error<br>• Communication Fault<br>• Total power failure alarm<br>• Single Dose Complete<br>• Low Battery<br>• Battery Degradation<br>• No Operation | • Feeding complete<br>• Single dose complete<br>• Low battery<br>• Out of battery<br>• Battery error<br>• Motor error<br>• Communication fault<br>• No operation | Different<br>The subject device has more<br>control alarms than the predicate. |
| Display | Color 4.3" LCD touch screen | 4-digit LED | Different<br>The subject device has a larger |
| Characteristic | Subject Device:<br>AMSure Enteral feeding pump<br>and feeding set | Predicate:<br>SENTINEL Enteral feeding<br>pump<br>K011587 | Discussion between subject<br>device and predicate |
| | | | differences do not raise different<br>questions of safety and<br>effectiveness than the predicate<br>device. |
| Self-test | Automated self-tests at system<br>start-up | No | Different.<br>The self-test at system start-up to<br>check the functioning of major<br>components, motor, sensors etc.<br>In case of component failure, the<br>pump pop-up warning information<br>and the pump cannot be used.<br>These differences do not raise<br>different questions of safety and<br>effectiveness than the predicate<br>device. |
| Maximum occlusion pressure | 11 psi | 15 psi | Different.<br>The occlusion pressure of subject<br>device is lower than the predicate.<br>These differences do not raise<br>different questions of safety and<br>effectiveness than the predicate<br>device. |
| Electrical Safety<br>Standards | IEC 60601-1<br>IEC 60601-1-1<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 60601-1-8<br>IEC 62304<br>IEC 62366 | UL544 | Different<br>The subject device is designed to<br>follow the currently recognized<br>standards. These differences do<br>not raise different questions of<br>safety and effectiveness than the<br>predicate device. |
| EMC/ EMI and ESD standards | IEC 60601-1-2 | IEC601-1-1-2 | Different |
| Degree of Protection against<br>Electrical Shock | Class 2 Type BF per IEC<br>Standards<br>Pass | Class 2 Type BF<br>Pass | Different |
| Degree of Protection for liquid<br>ingress | IPX4<br>Pass | IPX1<br>Pass | Different<br>The subject device has higher<br>level of liquid ingress protection<br>than the predicate. These<br>differences do not raise different<br>questions of safety and<br>effectiveness than the predicate<br>device. |
| Enclosure | Flame retarded plastic | Flame retarded plastic | Different |
| Power adapter | Yes | Yes | Different |
| Power requirement | 100-240VAC, 50-60Hz | 120VAC,60Hz | Different |
| Rechargeable battery | Lithium-ion | Sealed lead acid | Different<br>The subject device is using<br>Lithium-Ion battery. This<br>difference does not raise safety<br>and effectiveness than the<br>predicate device. |
| Battery life | 7 hours at 120 mL/h | 12 hours at 125 mL/h | Different<br>The subject device has shorter<br>battery life than predicates.<br>These differences do not raise<br>different questions of safety and<br>effectiveness than the predicate<br>device. |
| Characteristic | Subject Device:<br>AMSure Enteral feeding pump<br>and feeding set | Predicate:<br>SENTINEL Enteral feeding<br>pump<br>K011587 | Discussion between subject<br>device and predicate |
| Battery recharge time | 5hr | 10hr | Different<br>The recharge time of subject<br>device is much shorter than the<br>predicate's |
| Battery operation indicator | Yes | Yes | Different |
| Feeding mode | Both continuous mode and<br>intermittent mode. | No | Different<br>The subject device has<br>intermittent feed mode, but the<br>predicate doesn't have. These<br>differences do not raise different<br>questions of safety and<br>effectiveness than the predicate<br>device. |
| Flow range | 1-400 mL/hr in 1 mL increments | 5-295 mL/hr in 1mL increments | Different.<br>The subject device has a wider<br>flow rate range to meet the update<br>market needs.<br>These differences do not raise<br>different questions of safety and<br>effectiveness than the predicate<br>device. |
| Volume range | 1-9999 mL 1 mL increments | Adjustable 1-2000 mL 1 mL<br>increments, and 2000-9999ml in<br>5ml increments | Different.<br>The subject device has a smaller<br>increment. These differences do<br>not raise different questions of<br>safety and effectiveness than the<br>predicate device. |
| Flow rate accuracy | $\pm$ 10% of selected rate under<br>normal or worst-case condition | $\pm$ 10% of selected rate under<br>normal or worst-case condition | Different |
| Auto prime | Yes - flow rate 1200 mL/h | No | Different<br>The subject device has auto prime<br>function, but the predicate doesn't<br>have this function . These<br>differences do not raise different<br>questions of safety and<br>effectiveness than the predicate<br>device. |
| Pump weight | 2.38pounds | 5.2pounds | Different<br>The pump weight of subject<br>device is less than the predicate's.<br>These differences do not raise<br>different questions of safety and<br>effectiveness than the predicate<br>device. |
| Pump size | 5.2"x4.3"x6.7" | 7.5"x9"x5.2" | Different<br>The pump size of subject device is<br>less than the predicate's. These<br>differences do not raise different<br>questions of safety and<br>effectiveness than the predicate<br>device. |
| Pole clamp mountable | Yes | Yes | Different |
| RFID | Yes | No | Different<br>As an optional feature, the subject<br>device incorporates RFID<br>technology. These differences do<br>not raise different questions of<br>safety and effectiveness than the<br>predicate device. |
| KTO | Yes | No | Different<br>As an optional feature, KTO<br>feature may help maintain tube<br>patency during the pump stop<br>interval. The KTO feature has |
| Characteristic | Subject Device:<br>AMSure Enteral feeding pump<br>and feeding set | Predicate:<br>SENTINEL Enteral feeding<br>pump<br>K011587 | Discussion between subject<br>device and predicate |
| | | | been verified to meet the specifications, this difference does not raise different questions of safety and effectiveness than the predicate device. |
| Operating Temperature Range | 16°-40°C (60.8°-104°F) | 16°-40°C (60.8°-104°F) | |
| Storage Temperature Range | -20°- 60°C (-4°-140°F) | -10°- 50°C (14°-122°F) | Different<br>The subject device has wider storage temperature range. These differences do not raise different questions of safety and effectiveness than the predicate device. |
| Anti-free flow | Yes | Yes | No differences |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
## Performance Tests:
A safety assurance case as recommended by the FDA guidance document, "Infusion pumps total product life cycle" was provided for the AMSure® Enteral Feeding Pump and Feeding Set.
The stated top-level goal of the safety assurance case is the pump and set are adequately safe for its intended use.
The following specific evidence were included within the assurance case to demonstrate that the subject device is verified and validated for its intended use and to demonstrate substantial equivalence to the predicate devices.
| Software | Software verification and validation per FDA guidance for the "Content of premarket<br>submissions for software contained in medical devices" for a Major Level of Concern<br>and FDA guidance document "Infusion pump total product life cycle" |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical safety | The electrical safety evaluation of the medical electrical equipment was performed<br>per standards IEC60601-1 medical electrical equipment part1: General requirements<br>for basic safety and essential performance, IEC60601-1 Canadian and US national<br>differences test. |
| EMC | The AMSure® pump was evaluated to IEC 60601-1-2: Medical electrical equipment-<br>Part 2: General requirement for basic safety and essential performance - Collateral<br>standard: electromagnetic compatibility - Requirements and tests. |
| Device performance | The essential performance requirements of the device (including feeding sets) were<br>verified through performance testing in accordance with the intended use of the<br>device and in accordance with the FDA Guidance "Infusion Pumps Total Product Life<br>Cycle" including:<br>• Performance testing of essential performance attributes<br>• Reliability testing<br>• Flow rate accuracy testing across all operating conditions<br>• Alarm detection |
| Human factors | Follow FDA Guidance Applying Human Factors and Usability Engineering to Medical<br>Devices (February 3, 2016), the human factors studies were conducted with the<br>intended user population, use environment, and use scenarios to simulate clinical<br>conditions. Results of the human factors testing demonstrate validation of the device<br>per the intended use. |
| Biocompatibility | The biocompatibility test reports provided were conducted per ISO10993 series<br>standard following Good Laboratory Practices and the representative product tested<br>passed all acceptance criteria. |
{8}------------------------------------------------
# Clinical Tests
Not Applicable
## Conclusions
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The AMSure® Enteral Feeding Pump and Feeding Set is substantially equivalent to the SENTINEL Enteral Feeding pump (K011587) with respect to the indications for use, target populations, treatment method, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.