MagnetOs Flex Matrix is intended to fill bony voids or gaps of the skeletal system, i.e., posterolateral spine. In the posterolateral spine, MagnetOs Flex Matrix must be hydrated with bone marrow aspirate and used as an extender to autograft bone. The osseous defects may be surgically created or the result of traumatic injury to the bone that are not intrinsic to the stability of the bony structure. MagnetOs Flex Matrix resorbs and is replaced with bone during the healing process.
Device Story
MagnetOs Flex Matrix is a resorbable, osteoconductive bone void filler; composed of medical-grade bovine type I collagen matrix containing hydroxyapatite and β-tricalcium phosphate ceramic granules. Device is supplied as a ready-to-use, moldable strip. Used in clinical settings by surgeons to fill skeletal voids or gaps; specifically as an autograft extender in posterolateral spine procedures. Upon hydration with bone marrow aspirate, the matrix is shaped to conform to defect contours. The collagen component resorbs within 6 weeks; the ceramic granules provide a porous, trabecular structure with submicron needle-shaped surface features that guide 3D bone regeneration. The graft is gradually replaced by host bone during natural remodeling. Benefits include providing a scaffold for bone growth in non-load-bearing defects.
Clinical Evidence
No human clinical data provided. Substantial equivalence supported by bench-top physicochemical/crystallographic characterization and in vivo animal performance testing. Rabbit posterolateral spine fusion model compared subject device, primary predicate (MASTERGRAFT® Strip), and autograft (positive control) over 12 weeks. Endpoints included manual palpation, range of motion/flexibility testing, radiography, micro-CT, and histomorphometric analysis. Results confirmed performance equivalence to predicate devices.
Technological Characteristics
Resorbable bone void filler; materials: bovine type I collagen, hydroxyapatite, and β-tricalcium phosphate. Form factor: moldable strip. Sensing/actuation: passive osteoconductive scaffold. Sterilization: gamma-irradiated. Connectivity: none. Software: none.
Indications for Use
Indicated for filling bony voids or gaps of the skeletal system, specifically the posterolateral spine, in patients requiring bone graft extenders for surgically created or traumatic osseous defects not intrinsic to bony stability. Must be hydrated with bone marrow aspirate and used with autograft.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
April 14, 2022
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Kuros Biosciences BV Sonja van der Meer Regulatory Affairs Expert Professor Bronkhorstlaan 10, Building 48 Bilthoven, 3723 MB, The Netherlands
Re: K213959
Trade/Device Name: MagnetOs Flex Matrix Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: March 9, 2022 Received: March 9, 2022
Dear Sonja van der Meer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213959
Device Name MagnetOs Flex Matrix
#### Indications for Use (Describe)
MagnetOs Flex Matrix is intended to fill bony voids or gaps of the skeletal system, i.e., posterolateral spine. In the posterolateral spine, MagnetOs Flex Matrix must be hydrated with bone marrow aspirate and used as an extender to autograft bone. The osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure.
MagnetOs Flex Matrix resorbs and is replaced with bone during the healing process.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> <path d="M2,2 L14,14 M14,2 L2,14" stroke="black" stroke-width="2"></path> </svg> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> </svg> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Kuros Biosciences Traditional 510(k) For MagnetOs Flex Matrix
## 510(k) Summary
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92.
#### Submitter's Name:
Kuros Biosciences B.V.
#### Submitter's Address:
Kuros Biosciences B.V. Prof. Bronkhorstlaan 10, building 48 3723 MB Bilthoven The Netherlands
### Establishment Registration Number:
3008147766
#### Contact Person:
Sonja van der Meer
Telephone +31 (0)30 229 7280
#### Date Prepared:
14th April 2022
{4}------------------------------------------------
## Kuros Biosciences Traditional 510(k) For MaqnetOs Flex Matrix
The Device Classification Information regarding MagnetOs Flex Matrix is summarized below.
### Primary Product Code
| Regulation<br>Number | Device | Device Class | Product Code | Classification<br>Panel |
|----------------------|---------------------------------------------|--------------|--------------|-------------------------|
| 888.3045 | Resorbable calcium<br>salt bone void filler | 2 | MQV | Orthopedic |
#### Device Trade Name
MaqnetOs Flex Matrix
### Device Common Name
Resorbable calcium salt bone void filler
#### Device Description
MagnetOs Flex Matrix is a resorbable, osteoconductive bone void filler, intended to fill bony voids or gaps of the skeletal system, i.e., posterolateral spine, In the posterolateral spine, MagnetOs Flex Matrix must be hydrated with bone marrow aspirate and used as an extender to autograft bone. The product is biocompatible.
MaqnetQs Flex Matrix is a mixture of medical grade collagen and hydroxyapatite and ß-tricalcium phosphate ceramic granules. The collagen is produced from highly purified bioresorbable bovine split skin and consists mainly of collagen type I. The ceramic portion of MagnetOs Flex Matrix consists of β-tricalcium phosphate and hydroxyapatite.
MagnetOs Granules has a porous trabecular structure that resembles the structure and interconnected porosity of human cancellous bone. The surface of MagnetOs Granules is covered with needle-shaped features that are submicron in size.
The collagen matrix has a fibrillar porous structure which allows for the exposure of MagnetOs Granules' surface structure without interfering with its mode of action. While the collagen sponge matrix is readly resorbed in the first 6 weeks after implantation, MagnetOs Granules guides the three-dimensional regeneration of bone in the defect into which it is implanted. New bone will be deposited on the surface of the graft when it is placed next to viable host bone. The graft will be resorbed and replaced by bone during the natural process of bone remodelling.
MagnetOs Flex Matrix is a ready-to-use product. Upon hydration, the material is moldable and allows users to shape MagnetOs Flex Matrix to conform to the contours of bony defects. MagnetOs Flex Matrix is gamma-sterilized, sterile packaged and ready for single patient use only.
#### Intended Use
MagnetOs Flex Matrix is intended to fill bony voids or gaps of the skeletal system, i.e., posterolateral spine. In the posterolateral spine, MagnetOs Flex Matrix must be hydrated with bone marrow aspirate and used as an extender to autograft bone. The osseous defects may be surgically created or the result of traumatic injury to the bone that are not intrinsic to the stability of the bony structure.
MagnetOs Flex Matrix resorbs and is replaced with bone during the healing process.
{5}------------------------------------------------
## Kuros Biosciences Traditional 510(k) For MaqnetOs Flex Matrix
#### Summary of Substantial Equivalence
MagnetOs Flex Matrix is substantially equivalent in indications and design principles to the following cleared and legally marketed predicate devices:
| Table 1.1: Subject device and Predicate devices general information | | | | |
|---------------------------------------------------------------------|-----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Property | 510(k) Subject Device | Primary Predicate<br>Device | Equivalent Device | Equivalent Device |
| Device Name | MagnetOs Flex Matrix | MASTERGRAFT® Strip | Vitoss Scaffold Foam | MagnetOs Granules |
| Device<br>Manufacturer | Kuros Biosciences B.V. | Medtronic Sofamor<br>Danek USA | Othovia, Inc | Kuros Biosciences B.V. |
| 510(k) | No | K082166 | K032288 | K161859 |
| Device<br>Classification | II | II | II | II |
| Product Code | MQV | MQV | MQV | MQV |
| Use<br>Environment | Medical / Clinical | Medical / Clinical | Medical / Clinical | Medical / Clinical |
| Anatomic<br>location | Bony voids or gaps of<br>the skeletal system, i.e.,<br>posterolateral spine | Bony voids or gaps of<br>the skeletal system, i.e.,<br>the extremities, spine<br>and pelvis | Bony voids or gaps of<br>the skeletal system, i.e.,<br>the extremities, spine<br>and pelvis | Bony voids or gaps of<br>the skeletal system, i.e.,<br>posterolateral spine |
| Materials | hydroxyapatite and β-<br>tricalcium phosphate<br>mixture | hydroxyapatite and β-<br>tricalcium phosphate<br>mixture | β-tricalcium phosphate<br>mixture | hydroxyapatite and β-<br>tricalcium phosphate<br>mixture |
| | Bovine collagen type I | Bovine collagen type I | Bovine collagen type I | - |
| Dosage form | Strips | Strips | Strips | Granules |
Table 1 1: Subjoct dovice and Prodicato dovices gonoral information
MaqnetQs Flex Matrix is substantially equivalent to the predicate device MASTERGRAFT® Strip and to equivalent devices Vitoss Foam and MagnetOs Granules with respect to design, structure, materials, and mechanism of action, and has similar Indications for Use.
For the two MagnetOs devices, the MagnetOs Granules component is identical.
MASTERGRAFT® Strip is the primary predicate device used for animal performance testing studies. MagnetOs Granules is the reference standard comparators used for comparison animal studies.
## Non-clinical Testing (Performance/Physical Data):
Characterization of MagnetOs Flex Matrix materials has been conducted in compliance with the Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Resorbable Calcium Salt Bone Void Filler Device (Document issued on: June 2, 2003) and the requirements of other applicable standards.
Bench-top testing was performed to evaluate the physicochemical and crystallographic characteristics of MaqnetOs Flex Matrix, which demonstrated that the MagnetOs Granules component in MagnetOs Flex Matrix is identical to MagnetOs Granules device. With SEM and XRD it is confirmed that the collagen matrix does not change the MagnetOs Granules specifications. This ensures that the mode of action is not compromised.
## Biocompatibility testing:
The biocompatibility of the MaqnetOs Flex Matrix is demonstrated by ISO 10993 testing and the long history of clinical use of the collagen and calcium phosphate materials for the same intended use.
Bacterial endotoxin testing was performed using the limulus amebocyte lysate (LAL) method and showed that the device meets the endotoxin limits of established guidelines.
{6}------------------------------------------------
## Kuros Biosciences Traditional 510(k) For MaqnetOs Flex Matrix
#### Animal studies:
Substantial equivalence in performance of MagnetOs Flex Matrix to Mastergraft Strip was provided by in vivo animal data in the intended use.
Mastergraft Strip is the primary predicate device used for animal performance testing studies. The performance of the MagnetOs Flex Matrix was established in a posterolateral spine fusion animal model that justifies that MagnetOs Flex Matrix can be used as an autograft extender (1:1 volume ratio) in posterolateral spine.
Animal testing was performed in a rabbit posterolateral spine fusion model to demonstrate substantial equivalence to the primary predicate device. Animals were evaluated after implantation with the subject device, the primary predicate device, and autograft (positive control) up to 12 weeks. Evaluation endpoints included manual palpation, range of motionflexibility testing, plain and high-resolution radiography, microcomputed tomography (micro-CT) imaging, undecalcified histologic evaluation, and histomorphometric analysis. Decalcified paraffin histology sections also were graded according to ISO 10993-6 (Annex E).
The exact same animal model and study design were used for pre-clinical performance testing of MagnetQs Granules (K161859, K21311). By comparing the outcomes of both studies, substantial equivalence in performance of MagnetOs Flex Matrix and MagnetOs Granules is confirmed.
From all available pre-clinical data, it is concluded that the performance of MagnetOs Flex Matrix is substantially equivalent to both the primary predicate device, Mastergraft Strip, and MagnetOs Granules.
#### Safety and Effectiveness/Conclusion:
Based on the information presented in these 510(k) premarket notifications, MagnetOs Flex Matrix is considered substantially equivalent to predicate devices. It is our determination that MagnetOs Flex Matrix is as safe and effective as currently marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.