K032288 · Orthovita, Inc. · MQV · Dec 19, 2003 · Orthopedic
Device Facts
Record ID
K032288
Device Name
VITOSS SCAFFOLD FOAM BONE GRAFT MATERIAL
Applicant
Orthovita, Inc.
Product Code
MQV · Orthopedic
Decision Date
Dec 19, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Vitoss Scaffold Foam Bone Graft Material is intended for use as a bone void filler for voids or gaps that are not intrinsic to the stability of the bony structure. Vitoss Scaffold Foam is indicated for use in the treatment of surgically created osseous defects or osseous defects created from traumatic injury to the bone. Vitoss Scaffold Foam should not be used to treat large defects that in the surgeon's opinion would fail to heal spontaneously. Vitoss Scaffold Foam Bone Graft Material is intended to be used for filling bony voids or gaps of the skelctal system (i.e., the extremities, spine and pelvis). Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process.
Device Story
Porous calcium phosphate resorbable material admixed with Type I bovine collagen; trabecular structure mimics human cancellous bone; pore diameters 1-1000 um. Implanted by surgeons in OR/clinic to fill bony voids/gaps in extremities, spine, and pelvis. May be mixed with autogenous blood or bone marrow prior to placement. Osteoconductive scaffold guides 3D bone regeneration; resorbs concurrently with new bone ingrowth and remodeling. Does not provide mechanical stability. Benefits patient by facilitating natural bone repair in non-load-bearing defects.
Clinical Evidence
No clinical data. Pre-clinical animal data demonstrated scaffold supports bone growth into metaphyseal defects and resorbs concurrently with bone remodeling. In-vitro data supports safety and effectiveness equivalence to predicates.
Technological Characteristics
Material: Beta-tricalcium phosphate (Ca3(PO4)2) and Type I bovine collagen. Structure: Trabecular, porous (1-1000 um). Form: Sterile cylinders or strips. Energy: None. Connectivity: None. Sterilization: Gamma irradiation. Mechanical: Non-load bearing.
Indications for Use
Indicated for patients with surgically created or traumatic osseous defects in the skeletal system (extremities, spine, pelvis). Contraindicated for large defects that would not heal spontaneously or defects intrinsic to bony stability.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Predicate Devices
Vitoss® Scaffold Synthetic Cancellous Bone Void Filler
Healos® Bone Graft Material
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary Vitoss® Scaffold Foam Bone Graft Material
| Submitted by | Address | Telephone | Contact Person |
|---------------------|----------------------------------------------|----------------|-------------------------------------------------------------------------------------------|
| Orthovita, Inc. | 45 Great Valley Parkway<br>Malvern, PA 19355 | (610) 640-1775 | Gina M. Nagvajara, Ph.D.<br>Director of Research and<br>Intellectual Property |
| | Subject Device | | Predicate Devices |
| Trade Name | Vitoss® Scaffold Foam<br>Bone Graft Material | | Vitoss® Scaffold Synthetic<br>Cancellous Bone Void Filler;<br>Healos® Bone Graft Material |
| Common Name | Bone Void Filler | | Bone Void Filler |
| Classification Name | Filler, Calcium Sulfate<br>Preformed Pellets | | Filler, Calcium Sulfate<br>Preformed Pellets |
#### Device Description:
DEC 1 9 2003
Vitoss Scaffold Foam is a porous calcium phosphate resorbable material admixed with Type I bovine collagen for the repair of bony defects. It is an osteoconductive porous implant with a trabecular structure that resembles the multidirectional interconnected porosity of human cancellous bone. Pore diameters in the scaffold range from 1 um to 1000 um (1 mm). The implant is provided sterile in both cylinder and strip forms. Vitoss Scaffold Foam may be mixed with autogenous blood or bone marrow.
Vitoss Scaffold Foam guides the three-dimensional regeneration of bone in the defect site into which it is implanted. When Vitoss Scaffold Foam is placed in direct contact with viable host bone, new bone grows in apposition to the surfaces of the implant. As the implant resorbs, bone and other connective tissues grow into the space previously occupied by the scaffold.
## Intended Use:
Vitoss Scaffold Foam Bone Graft Material is intended for use as a bone void filler for voids or gaps that are not intrinsic to the stability of the bony structure. Vitoss Scaffold Foam is indicated for use in the treatment of surgically created osseous defects or osseous defects created from traumatic injury to the bone. Vitoss Scaffold Foam should not be used to treat large defects that in the surgeon's opinion would fail to heal spontaneously.
Vitoss Scaffold Foam Bone Graft Material is intended to be used for filling bony voids or gaps of the skelctal system (i.e., the extremities, spine and pelvis). Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows handwritten text that appears to be a combination of letters and numbers. The text includes 'FO32288' on the top line, followed by 'aofa' on the second line. The handwriting is somewhat stylized, with some characters slightly distorted or overlapping, giving it a unique and informal appearance.
510(k) Summary Vitoss Foam Orthovita, Inc.
## Comparison to Predicate:
| | NEW DEVICE:<br>VITOSS SCAFFOLD<br>FOAM | PREDICATE:<br>VITOSS SCAFFOLD<br>SYNTHETIC | PREDICATE:<br>HEALOS BONE<br>GRAFT MATERIAL |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Intended Use | Bone Void Filler | Bone Void Filler | Bone Void Filler |
| Target Population | Individuals with bony<br>defects resulting from<br>surgery or trauma | Individuals with bony<br>defects resulting from<br>surgery or trauma | Individuals with bony<br>defects resulting from<br>surgery or trauma |
| Anatomical Locations | Bony voids or gaps of<br>the skeletal system, i.e.,<br>the extremities, spine<br>and pelvis | Bony voids or gaps of<br>the skeletal system, i.e.,<br>the extremities, spine<br>and pelvis | Bony voids or gaps of<br>the skeletal system, i.e.,<br>the extremities, spine<br>and pelvis |
| Labeling | Vitoss Scaffold Foam's product label contains the same intended use,<br>contraindications, warnings, precautions, and adverse events as for the Vitoss<br>Scaffold Synthetic Cancellous Bone Void Filler with the addition of collagen<br>related contraindications and adverse events from the Healos Bone Graft<br>Material product label; wording is not identical but is substantially equivalent | | |
| Materials | | | |
| Chemical<br>Composition | Calcium salt with Type<br>I bovine collagen | Calcium salt | Calcium salt with Type I<br>bovine collagen |
| Mineral Phase(s) | $β$ -Tricalcium phosphate<br>$Ca_3(PO_4)_2$ | $β$ -Tricalcium phosphate<br>$Ca_3(PO_4)_2$ | Hydroxyapatite<br>$Ca_{10}(PO_4)_6(OH)_2$ |
| Design | | | |
| Physical<br>Structure | Trabecular structure<br>similar to cancellous<br>bone | Trabecular structure<br>similar to cancellous<br>bone | Three dimensional,<br>open-cell matrix |
| Pore Size (range) | 1-1000 μm | 1-1000 μm | 4-200 μm |
| Performance | | | |
| Osteoconductivity | Osteoconductive | Osteoconductive | Osteoconductive |
| Mechanical<br>Strength | Does not impart<br>mechanical strength to<br>surgical site | Does not impart<br>mechanical strength to<br>surgical site | Does not impart<br>mechanical strength to<br>surgical site |
| Sterility | Sterilized by gamma<br>irradiation, single use<br>only | Sterilized by gamma<br>irradiation, single use<br>only | Sterilized by e-beam,<br>single use only |
| Biocompatibility | Established | Established | Established |
| Dosage Form(s) | Cylinders (Blocks),<br>Strips | Blocks, Morsels | Strips, Pads |
#### Non-clinical Performance Data:
Pre-clinical animal data demonstrate that Vitoss Scaffold Foam supports bone growth into a metaphyseal defect. These data show that Vitoss Scaffold Foam is resorbed concurrently with bone ingrowth and remodeling. These results, in conjunction with in-vitro data, demonstrate that Vitoss Scaffold Foam is as safe and as effective as the predicate devices, Vitoss Scaffold Synthetic Cancellous Bone Void Filler and Healos Bonc Graft Material.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle with three wavy lines representing its wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 9 2003
Gina M. Nagvajara, Ph.D. Director of Research and Intellectual Property Orthovita, Inc. 45 Great Valley Parkway Malvern, Pennsylvania 19355
Re: K032288
K032206 Regulation Number: 21 CFR 888.3045 Regulation Name: Bone void filler Regulatory Class: II Product Code: MQV Dated: December 1, 2003 Received: December 2, 2003
Dear Dr. Nagvajara:
We have reviewed your Section 510(k) premarket notification of intent to market the device wt nave reviewed your bection of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosure) to regary the Medical Device Amendments, or to commerce province to may 20, 1978, the excordance with the provisions of the Federal Food. Ding, devices that have tocer rechassined in avere approval of a premarket approval application (PMA). and Cosmette Act (11ct) market the device, subject to the general controls provisions of the Act. The 1 the may, therefore, thans of the Act include requirements for annual registration, listing of general controls provisions practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is enabilional controls. Existing major regulations affecting your device can may be sunyeet to suen additions, Title 21, Parts 800 to 898. In addition, FDA may be found in the councements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination docs not mean I has be advised that I Dr o waters at your device complics with other requirements of the Act that 1177 hab intatutes and regulations administered by other Federal agencies. You must of any i-cutcal statutes and regaraments, including, but not limited to: registration and listing (21 Comply with an the Hot 3 requirements)); good manufacturing practice requirements as set Of It fart 6077, abounting (21 CFR Part 820); and if applicable, the electronic Routh m the quants (Sections (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
#### Page 2 - Mr. Lewis Ward
This letter will allow you to begin marketing your device as described in your Section 510(k) rms lotter with and wyourse FDA finding of substantial equivalence of your device to a legally premance notification in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please in you attire office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark A. Millican
Celia M. Witten, PhD, MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE
## STATEMENT
#### 510(k) Number (If Known): K032288
Device Name: Vitoss® Scaffold Foam Bone Graft Material
#### Indications For Use:
Vitoss Scaffold Foam Bone Graft Material is intended for use as a bone void filler for voids or gaps that are not intrinsic to the stability of the bony structure. Vitoss Scaffold Foam is indicated for use in the treatment of surgically created osseous defects or osseous defects created from traumatic injury to the bone. Vitoss Scaffold Foam should not be used to treat large defects that in the surgeon's opinion would fail to heal spontaneously.
Vitoss Scaffold Foam Bone Graft Material is intended to be used for filling bony voids or gaps of the skeletal system (i.e., the extremities, spine and pelvis). Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process.
Prescription Use_ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for Mark N Milker
Division of General, Restorative
and Neurological Devices
K032288
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.