K213913 · Renovia, Inc. · HIR · Jun 30, 2022 · Obstetrics/Gynecology
Device Facts
Record ID
K213913
Device Name
leva Pelvic Health System
Applicant
Renovia, Inc.
Product Code
HIR · Obstetrics/Gynecology
Decision Date
Jun 30, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1425
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The leva Pelvic Health System is intended for: 1) Strengthening of the pelvic floor muscles; 2) Rehabilitation and training of weak pelvic floor muscles for the treatment of stress, mixed and mild to moderate urgency urinary incontinence (including overactive bladder) in women; 3) Rehabilitation and training of weak pelvic floor muscles for the first-line treatment of chronic fecal incontinence (>3-month uncontrolled passage of feces) in women. This device interacts with the user via smart phone technology.
Device Story
Intravaginal device for pelvic floor muscle (PFM) rehabilitation and training. Device uses internal accelerometers to detect relative position changes during PFM contraction (lift and squeeze). Data transmitted via Bluetooth Low Energy (BLE) to a dedicated smartphone application. App provides real-time biofeedback to the user, allowing monitoring of training progress. Used by patients at home; prescription-only. Output consists of graphical and numeric biofeedback displayed on the mobile app. Helps patients perform correct PFM exercises, potentially improving symptoms of urinary and fecal incontinence.
Clinical Evidence
Evidence includes a 10-week single-arm study (REN-17, n=31) and a subset analysis of a randomized controlled trial (REN-19, n=92). REN-17 showed improvement in St. Mark's Score (14.6 to 11.6, p=0.0047). REN-19 subset analysis compared leva-02 to Kegel exercises; leva-02 arm showed statistically superior improvement in CRADI-8 and CRAIQ-7 scores at 6 months compared to control. No serious device-related adverse events reported.
Technological Characteristics
Intravaginal probe made of thermoplastic elastomer and polypropylene. Uses accelerometers for sensing relative position. Powered by non-rechargeable 1.5V silver oxide coin cell. Connectivity via Bluetooth Low Energy (BLE). Non-sterile, reusable, single-patient device. Software provides biofeedback via mobile app.
Indications for Use
Indicated for women with weak pelvic floor muscles requiring strengthening or rehabilitation for stress, mixed, or mild-to-moderate urgency urinary incontinence (including overactive bladder), and for first-line treatment of chronic fecal incontinence (>3 months).
Regulatory Classification
Identification
A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 30, 2022
Renovia Inc. % Jacqueline Schmainda Regulatory Consultant Bold Type 2100 N Alafaya Trail Orlando, FL 32826
Re: K213913
> Trade/Device Name: leva Pelvic Health System Regulation Number: 21 CFR 884.1425 Regulation Name: Perineometer Regulatory Class: Class II Product Code: HIR Dated: June 1, 2022 Received: June 2, 2022
Dear Jacqueline Schmainda:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Je Hi An, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213913
Device Name leva Pelvic Health System
Indications for Use (Describe)
The leva Pelvic Health System is intended for:
1) Strengthening of the pelvic floor muscles;
2) Rehabilitation and training of weak pelvic floor muscles for the treatment of stress, mixed and mild to moderate urgency urinary incontinence (including overactive bladder) in women;
3) Rehabilitation and training of weak pelvic for the first-line treatment of chronic fecal incontinence (>3month uncontrolled passage of feces) in women.
This device interacts with the user via smart phone technology.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the word "Renovia" in a stylized font. The word is in a purple color. Above the right side of the word are three curved lines.
This 510(k) Summary has been prepared in accordance with 21 CFR 807.92.
#### 1. Submitter
| Name and Address: | Renovia Inc.<br>263 Summer Street<br>Boston, MA 02210 |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact: | Jacqueline Schmainda<br>Regulatory Consultant on behalf of Renovia<br>Renovia Inc.<br>263 Summer Street<br>Boston, MA 02210<br>Phone: 763-269-2069<br>Email: jackie@boldtype.com |
| Date Prepared: | June 27, 2022 |
#### 2. Device Information
| Trade Name: | leva® Pelvic Health System |
|--------------------|----------------------------|
| Model: | leva-02 |
| Common Name: | Perineometer |
| Product Code: | HIR |
| Regulatory Class | Class II |
| Regulation Number: | 21 CFR 884.1425 |
| FDA Panel: | Obstetrics / Gynecology |
#### 3. Predicate Information
| 510(k) Number | Trade Name | Applicant |
|---------------|---------------------------|--------------|
| K212495 | leva Pelvic Health System | Renovia Inc. |
#### 4. Device Description
The leva Pelvic Health System ("leva PHS" or "leva System") is a prescription intra-vaginal device designed to allow the user to rehabilitate and strengthen their pelvic floor muscles ("PFM") as well as allow them to monitor their progress during pelvic floor muscle training ("PFMT"). The leva system is designed to wirelessly facilitate PFMT in women and to transmit real-time performance data through a dedicated mobile application that has been downloaded onto the patient's mobile device. The leva system is designed to be used vaginally and is intended to be used repeatedly by a single patient.
{4}------------------------------------------------
#### 5. Indications for Use
The /eva Pelvic Health System is intended for:
- 1) Strengthening of the pelvic floor muscles;
- Rehabilitation and training of weak pelvic floor muscles for the treatment of stress, mixed and mild 2) to moderate urgency urinary incontinence (including overactive bladder) in women;
- 3) Rehabilitation and training of weak pelvic floor muscles for the first-line treatment of chronic fecal incontinence (≥3-month uncontrolled passage of feces) in women.
This device interacts with the user via smart phone technology.
#### 6. Comparison of Technological Characteristics
The following table provides a comparison of the leva Pelvic Health System to the predicate device.
| Characteristic | Subject Device | Predicate Device |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | leva PHS | leva PHS |
| | This Submission | K212495 |
| Regulatory Information | | |
| Brand Name | leva Pelvic Health System | leva Pelvic Health System |
| Manufacturer | Renovia Inc. | Renovia Inc. |
| Model Number | leva-02 | leva-02 |
| Common or Usual<br>Name | Perineometer | Perineometer |
| Regulation | 21 CFR 884.1425 | 21 CFR 884.1425 |
| Class | Class II | Class II |
| Product Code | HIR | HIR |
| Intended Use / Indications for Use | | |
| Intended Use | Strengthen pelvic floor muscles for the<br>treatment of urinary incontinence and fecal<br>incontinence | Strengthen pelvic floor muscles for the<br>treatment of urinary incontinence |
| Indications for Use<br>Statement | The leva Pelvic Health System is intended for:<br>1) Strengthening of the pelvic floor muscles;<br>2) Rehabilitation and training of weak pelvic<br>floor muscles for the treatment of stress,<br>mixed and mild to moderate urgency<br>urinary incontinence (including overactive<br>bladder) in women;<br>3) Rehabilitation and training of weak pelvic<br>floor muscles for the first-line treatment of<br>chronic fecal incontinence (>3-month<br>uncontrolled passage of feces) in women.<br>This device interacts with the user via smart<br>phone technology. | The leva Pelvic Health System is intended for:<br>1) Strengthening of the pelvic floor muscles;<br>2) Rehabilitation and training of weak pelvic<br>floor muscles for the treatment of stress,<br>mixed and mild to moderate urgency<br>urinary incontinence (including overactive<br>bladder) in women.<br>This device interacts with the user via smart<br>phone technology. |
| | Subject Device | Predicate Device |
| Characteristic | leva PHS | leva PHS |
| | This Submission | K212495 |
| Technological Characteristics | | |
| Principle of Operation | Provides indication of relative position of<br>pelvic floor muscle contraction using<br>accelerometers | Provides indication of relative position of<br>pelvic floor muscle contraction using<br>accelerometers |
| Mechanism of Action | Lift and Squeeze for pelvic floor muscle<br>strengthening. The leva sensor hardware<br>(probe) transmits biofeedback to leva App<br>which is displayed to user. | Lift and Squeeze for pelvic floor muscle<br>strengthening. The leva sensor hardware<br>(probe) transmits biofeedback to leva App<br>which is displayed to user. |
| Electrical Stimulation | No | No |
| Parameter | Relative position of device | Relative position of device |
| Anatomical Use | Vaginal only | Vaginal only |
| Single Patient Device | Yes | Yes |
| Reusable | Yes | Yes |
| Sterility | Non-sterile, clean | Non-sterile, clean |
| Information Display | Graphical and numeric based on implied<br>bending, anatomical overlay | Graphical and numeric based on implied<br>bending, anatomical overlay |
| Patient Contact<br>(Contact Duration) | • Probe: Direct (<24 hours)<br>• Probe Battery Pack: Direct (<24 hours)<br>• Vaginal Probe Hub: Indirect (Incidental)<br>• Battery Pack Cap: Indirect (Incidental) | • Probe: Direct (<24 hours)<br>• Probe Battery Pack: Direct (<24 hours)<br>• Vaginal Probe Hub: Indirect (Incidental)<br>• Battery Pack Cap: Indirect (Incidental) |
| Patient Contact<br>Materials | • Probe: Thermoplastic Elastomer<br>• Probe Hub: Polypropylene<br>• Battery Pack: Hard Plastic | • Probe: Thermoplastic Elastomer<br>• Probe Hub: Polypropylene<br>• Battery Pack: Hard Plastic |
| Probe Shape | • Oval | • Oval |
| Storage | • Storage case provided with device | • Storage case provided with device |
| Primary System<br>Elements | • Wireless Vaginal Probe<br>• User Interface Device (i.e., smartphone,<br>tablet)<br>• Mobile Application | • Wireless Vaginal Probe<br>• User Interface Device (i.e., smartphone,<br>tablet)<br>• Mobile Application |
| Power Source | • Device: Non-rechargeable silver oxide 1.5V<br>Coin Cell<br>• Storage Case: Non-rechargeable CR2032 3V<br>Coin Cell | • Device: Non-rechargeable silver oxide 1.5V<br>Coin Cell<br>• Storage Case: Non-rechargeable CR2032 3V<br>Coin Cell |
| Wireless Technology | • Bluetooth Low Energy (BLE) | • Bluetooth Low Energy (BLE) |
{5}------------------------------------------------
Renovía
{6}------------------------------------------------
#### Summary of Nonclinical Testing and Risk Analysis 7.
Nonclinical testing of the leva Pelvic Health System was not required for the indications for use statement. Minor modifications made to the device since last clearance were summarized in this submission and included reference to design verification and/or design validation activities and risk analyses, where applicable.
#### A. Hardware Changes
There were no hardware changes made to the leva Pelvic Health System to support the expanded indications for use statement.
#### B. Software Changes
No software changes were made to the medical device functionality of the leva Pelvic Health System. More specifically, there were no changes to the mobile application which modified how the user interacts with the device. All software modifications were reviewed and evaluated following the software life cycle processes for medical device software as defined by IEC 62304. The only changes to the software of the leva Pelvic Health System relate to non-medical device functionality (aka "other functionality") under FDA's guidance on "Multiple Function Device Products: Policy and Considerations" (2020). Software modifications related to "other functionality" included the addition of educational information regarding fecal incontinence to the mobile application. Regression analyses were conducted to determine the scope of software testing to be reexecuted related to these modifications and the test results supported the implementation of updated software version.
#### 8. Clinical Testing
The mechanism of action of the leva Pelvic Health System has been the subject of multiple published, peerreviewed clinical trials studying its efficacy, including the two studies outlined below. The data from the referenced clinical studies supports the leva Pelvic Health System can serve as a safe and to provide feedback when performing pelvic floor training in the treatment of fecal incontinence. In all studies, the leva device was used vaginally to perform the same type of pelvic floor muscle training.
### A. REN-17 Study (NCT04027335):
The REN-17 Study was a ten (10) week, single-arm, open-label study involving thirty-one (31) participants who engaged in pelvic floor muscle training (PFMT) using the leva Pelvic Health System (leva-01). Subject selection was based on meeting pre-defined screening criteria, including fecal incontinence ("FI") defined as any uncontrolled loss of liquid or solid fecal material that occurs at least monthly over the last three (3) months and is bothersome enough to desire treatment.
{7}------------------------------------------------
| REN-17 – Summary | |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Level of Evidence: | Single-arm study with Performance Goals |
| Location of Study: | United States only |
| Primary Effectiveness Endpoint: | Change in St. Mark's Score |
| Was the study primary endpoint met? | Yes, participants reported a mean St. Mark's Score of 14.6 at baseline<br>which improved to 11.6 at ten weeks (p=0.0047). The sample size<br>needed to avoid a type β error was met. |
| Adverse events and complications: | There were three adverse events reported as possibly device related<br>– two urinary tract infections and one yeast infection. None of these<br>adverse events were considered serious and the participants<br>continued to participate in the study through the ten-week endpoint. |
| REN-17 - Patient Accountability | | | | | |
|-----------------------------------------|------------------------------|--|--|--|--|
| Stage | Treatment Group (Single-Arm) | | | | |
| Enrollment | 31 participants | | | | |
| Treatment | 27 participants | | | | |
| Primary Effectiveness Endpoint Analysis | 26 participants | | | | |
| Primary Safety Endpoint Analysis | 27 participants | | | | |
| REN-17 – Study Results | | | | |
|---------------------------|----|-----------------|-----------------|---------------------------|
| Endpoint | n | Baseline | Week 10 | Wilcoxon signed rank test |
| St. Mark's Score (Vaizey) | 26 | $14.6 \pm 4.37$ | $11.6 \pm 5.12$ | 0.008 |
| FIQoL - Lifestyle | 26 | $2.7 \pm 0.85$ | $3.1 \pm 0.84$ | <0.001 |
| FIQoL - Coping | 26 | $2.2 \pm 0.73$ | $2.5 \pm 0.81$ | <0.001 |
| FIQoL - Depression | 26 | $3.0 \pm 0.68$ | $3.2 \pm 0.70$ | <0.001 |
| FIQoL - Embarrassment | 26 | $2.1 \pm 0.84$ | $2.3 \pm 0.88$ | 0.06 |
| Bowel Diary | 21 | $8.4 \pm 8.73$ | $4.8 \pm 3.79$ | 0.052 |
*Bold text identifies statistical significance <0.05.
{8}------------------------------------------------
### B. REN-19 (NCT04508153):
The REN-19 Study was an eight (8) week, randomized controlled trial ("RCT") that included two hundred ninety-nine (299) women with stress and stress-dominant mixed urinary incontinence ("UI") that included completion of pelvic floor muscle training ("PFMT") for all participants (control and treatment groups) and evaluated the participants as follows:
- o Control Group: Performed PFMT after receiving a standardized written and verbal instructions but without the aid of any device, effectively performing what are commonly called Kegel exercises.
- Treatment Group: Performed PFMT using the leva Pelvic Health System (leva-02). O
While the primary endpoints of the Ren-19 RCT involved PFMT for the treatment of stressdominant mixed UI, the trial also included several secondary endpoints including the Colorectal-Anal Distress Inventory, Short Form (the "CRADI-8"). The CRADI-8 is a validated patient-reported symptom scale, validated, in part, based on the outcomes of bowel diaries.
Renovia completed a subset analysis of participants from the Ren-19 RCT who indicated they had fecal incontinence ("FI") symptoms. Participants included in the subset analysis met all of the inclusion criteria for the larger study and, in addition, indicated they "usually lose stool beyond their control" for well-formed or loose stool, with a bother of at least "somewhat" in their responses to the CRADI-8. Based on the same modified intent to treat plan executed for the larger study, a subset of 92 participants was identified as meeting this additional inclusion criteria.
The purpose of the subset analysis was to evaluate symptom improvement of the 92 participants with Fl over the study period within their assigned treatment group and to compare the results between the 48 participants in the control group (at-home PFMT without leva-02) and the 44 participants in the intervention group (at-home PFMT with leva-02). By focusing on participants with FI, it is possible to better assess the impact of the control treatment vs. the intervention treatment on FI symptoms.
| REN-19 Subset Analysis – Summary | |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Level of Evidence: | Randomized, multi-arm, unblinded study with an active control |
| Location of Study: | United States only |
| Effectiveness Endpoints: | 1) Change in participants' symptoms using the Colorectal-Anal Distress<br>Inventory, Short Form (the “CRADI-8”)<br>2) Change in participants' condition-specific quality of life assessment using<br>the Colorectal-Anal Impact Questionnaire Short Form (the “CRAIQ-7”) |
| REN-19 Subset Analysis - Patient Accountability | | |
|-------------------------------------------------|------------------------------------|------------------------|
| Stage | Treatment Group ( <i>leva</i> -02) | Control Group (Active) |
| Treatment | 44 participants | 48 participants |
| Primary Effectiveness Endpoint Analysis | 44 participants | 48 participants |
| Primary Safety Endpoint Analysis | 44 participants | 48 participants |
{9}------------------------------------------------
# Safety:
There were no treatment-related severe adverse events in either arm of the Ren-19 Study. Looking at only those 92 participants within the Fl subset analysis, there were two (2) urinary tract infections in the control (Kegel) group, neither of which were deemed to be intervention related, and none in the intervention (leva-02) group. There was one (1) adverse event, vaginal spotting, in one subject which was reported as "possibly" intervention related occurring in the intervention (leva-02) arm. This resolved before the end of the study and the participant resumed use of the device without complication through the study endpoint at 8 weeks.
## Endpoints:
The Ren-19 subset analysis compared participants' overall CRADI-8 scores and the two individual questions within the CRADI-8 specifically addressing Fl, and participants' condition-specific quality of life assessment using the Colorectal-Anal Impact Questionnaire Short Form (the "CRAIQ-7") – each at baseline, 8-weeks, and 6-months.
The results of the FI subset analysis demonstrate that FI symptom improvement of those in the leva-02 arm was statistically superior to that of participants in the Kegel arm, as summarized in the Results table below.
### CRADI-8 Results:
On the CRADI-8, both the Kegel arm and the leva-02 arm demonstrated statistically significant improvement that met the minimum clinical important difference ("MCID") (-4.68) for the CRADI-8 at 8 weeks. Jelovsek, Chen et. al. found in their 2014 publication, Minimum Important Differences for Scales Assessing Symptom Severity and Quality of Life in Patients with Fecal Incontinence (available at doi: 10.1097/SPV.00000000000000078), that the MCID for the CRADI-8 is a reduction of 4.68.
While the improvement in the leva-02 arm was greater, there was not a statistically significant difference between the groups at 8 weeks (P=.54). However, the CRADI-8 symptom improvement of the leva-02 group was statistically significantly greater at 6 months. This difference at 6 months was driven by the fact that the CRADI-8 results of participants in the Kegel arm did not continue to improve between 8 weeks and 6 months. In the leva-02 arm, however, the CRADI-8 results continued to improve between 8 weeks and 6 months.
This result of stronger continued improvement in Fl symptoms through 6 months within the leva-02 arm is also evident in the results of the questions 3 and 4 of CRADI-8, which are the two questions within the survey that specifically address loss of stool.2 As shown in Table 4 below, the symptom improvement specific to the control of loose stool was statistically superior within the leva-02 arm at 6 months.
## CRAIQ-7 Results:
In the CRAIQ-7 assessing participants' condition-specific quality of life, both the Kegel arm and the leva-02 arm demonstrated statistically significant improvement at 8 weeks. However, only participants in the leva-02 arm reported results that met the MCID (-8.01) for the CRAIQ-7 at 8 weeks (see Jelovsek, Chen et. dl. for MCID on CRAIQ-7).
While the CRAIQ-7 improvement in the leva-02 arm was greater, there was not a statistically significant difference between the groups at 8 weeks (P=.39). However, the CRAIQ-7 improvement of the leva-02 group demonstrated improvement that was statistically significantly greater at 6 months (p=0.02). This difference at 6 months was driven by the fact that the CRAIQ-7 results of participants in the Kegel arm did not continue
<sup>1</sup> Question 3 of the CRADI-8 asks: Do you usually lose stool beyond your control if your stool is well formed? Question 4 of the CRADI-8 asks: Do you usually lose stool beyond your control if your stool is loose?
{10}------------------------------------------------
to improve between 8 weeks and 6 months. In the leva-02 arm, however, the CRAIQ-7 results continued to improve between 8 weeks and 6 months.
| REN-19 Subset Analysis – Results | | | | | | | | | | | |
|----------------------------------|-------------------------------------------------|-----------------|------------------------|------------------|--------------------|-----------------|-----------------|------------------------|------------------|--------------------|--|
| Baseline to 8 weeks | | | Baseline to 6 months | | | | | | | | |
| | Baseline | 8<br>Weeks | Abs Mean<br>Difference | Paired<br>t-test | Students<br>t-test | Baseline | 6<br>Months | Abs Mean<br>Difference | Paired<br>t-test | Students<br>t-test | |
| | CRADI-8 (full survey) | | | | | | | | | | |
| Kegel arm | 40.9<br>(±19.9) | 29.2<br>(±22.4) | -5.3<br>(±14.3) | <0.001 | | 40.9<br>(±19.9) | 29.6<br>(±20.2) | -6.8<br>(±15.0) | <0.001 | 0.01 | |
| leva-02 arm | 37.7<br>(±20.7) | 27.8<br>(±20.6) | -3.0<br>(±15.0) | <0.001 | 0.54 | 37.7<br>(±20.7) | 17.2<br>(±19.4) | -8.6<br>(±13.8) | <0.001 | | |
| | CRADI-8: Question 3 (Loss of Well-Formed Stool) | | | | | | | | | | |
| Kegel arm | 0.9<br>(±1.6) | 0.7<br>(±1.4) | -0.1<br>(±1.0) | 0.3523 | 0.53 | 0.9<br>(±1.6) | 0.5<br>(±1.3) | -0.1<br>(±1.1) | 0.1914 | 0.64 | |
| leva-02 arm | 0.6<br>(±1.3) | 0.5<br>(±1.3) | 0.1<br>(±0.7) | 0.7656 | | 0.6<br>(±1.3) | 0.2<br>(±0.8) | -0.1<br>(±0.7) | | | |
| | CRADI-8: Question 4 (Loss of Loose Stool) | | | | | | | | | | |
| Kegel arm | 3.2<br>(±1.0) | 2.0<br>(±1.7) | -0.2<br>(±1.4) | <0.001 | 0.45 | 3.2<br>(±1.0) | 2.0<br>(±1.9) | -0.2<br>(±1.4) | <0.001 | 0.04 | |
| leva-02 arm | 3.09<br>(±1.1) | 1.6<br>(±1.6) | 0.2<br>(±1.5) | <0.001 | | 3.09<br>(±1.1) | 1.0<br>(±1.6) | -0.4<br>(±1.5) | <0.001 | | |
| | CRAIQ-7 | | | | | | | | | | |
| Kegel arm | 23.5<br>(±27.4) | 16.7<br>(±21.7) | -4.3<br>(±14.6) | 0.02 | | 23.5<br>(±27.4) | 17.5<br>(±24.7) | -3.8<br>(±13.1) | 0.1175 | | |
| leva-02 arm | 23.5<br>(±25.0) | 12.3<br>(±17.0) | -5.0<br>(±13.7) | <0.001 | 0.39 | 23.5<br>(±25.0) | 10.4<br>(±18.3) | -4.2<br>(±15.5) | <0.001 | 0.02 | |
*Bold text identifies statistical significance <0.05.
#### 9. Risk Review
In support of the expanded indications for use statement, a risk review was conducted for the use of the leva Pelvic Health System as a first-line treatment of chronic fecal incontinence in women. The risk review confirmed that there were no new or increased risks associated with the device which would raise new questions of safety or effectiveness.
#### 10. Conclusion
Renovia has demonstrated that the leva Pelvic Health System is as safe and effective as the predicate device for its intended use and substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.