K212495 · Renovia, Inc. · HIR · Sep 8, 2021 · Obstetrics/Gynecology
Device Facts
Record ID
K212495
Device Name
Leva Pelvic Health System
Applicant
Renovia, Inc.
Product Code
HIR · Obstetrics/Gynecology
Decision Date
Sep 8, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.1425
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The leva Pelvic Health System is intended for: 1) Strengthening of the pelvic floor muscles; 2) Rehabilitation and training of weak pelvic floor muscles for the treatment of stress, mixed and mild to moderate urgency urinary incontinence (including overactive bladder) in women. This device interacts with the user via smart phone technology.
Device Story
Prescription intra-vaginal device for pelvic floor muscle (PFM) rehabilitation and training. Probe contains six accelerometers to measure relative intensity of PFM contractions. Device transmits real-time performance data wirelessly to a dedicated mobile application. Used by women at home for repeated sessions. App provides graphical and numeric feedback based on applied bending to guide user training; helps monitor progress. Enables patient-directed exercise; facilitates PFM strengthening to treat urinary incontinence.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, irritation). Software/firmware design verification and manufacturing testing protocols were conducted to validate modifications.
Technological Characteristics
Intra-vaginal probe with six accelerometers; Thermoplastic Elastomer (TPE) probe material; Acrylonitrile Butadiene Styrene (ABS) battery pack. Wireless data transmission to mobile app. Clean, non-sterile. Single-patient, reusable. No muscle stimulation. Biocompatibility per ISO 10993-5 and ISO 10993-10.
Indications for Use
Indicated for women with weak pelvic floor muscles requiring strengthening or rehabilitation for stress, mixed, or mild to moderate urgency urinary incontinence, including overactive bladder.
Regulatory Classification
Identification
A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 8, 2021
Renovia, Inc. % Jacqueline Schmainda Director, QA/RA Bold Type 2100 N. Alafaya Trail Orlando, FL 32826
Re: K212495
Trade/Device Name: leva Pelvic Health System Regulation Number: 21 CFR §884.1425 Regulation Name: Perineometer Regulatory Class: II Product Code: HIR Dated: August 6, 2021 Received: August 9, 2021
Dear Jacqueline Schmainda:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
K212495
Device Name leva Pelvic Health System
Indications for Use (Describe)
The leva Pelvic Health System is intended for:
1) Strengthening of the pelvic floor muscles;
2) Rehabilitation and training of weak pelvic floor muscles for the treatment of stress, mixed and mild to moderate urgency urinary incontinence (including overactive bladder) in women.
This device interacts with the user via smart phone technology.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
This 510(k) Summary has been prepared in accordance with 21 CFR 807.92.
### 1. Submitter
| Name and Address: | Renovia Inc.<br>263 Summer Street<br>Boston, MA 02210 |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact: | Jacqueline Schmainda<br>Regulatory Consultant on behalf of Renovia<br>Renovia Inc.<br>263 Summer Street<br>Boston, MA 02210<br>Phone: 763-269-2069<br>Email: <a href="mailto:jackie@boldtype.com">jackie@boldtype.com</a> |
Date Prepared: September 8, 2021
### 2. Device Information
| Trade Name: | leva Pelvic Health System |
|------------------------|---------------------------|
| Model: | leva-02 |
| Common Name: | Perineometer |
| Product Code: | HIR |
| Regulatory Class: | Class II |
| Classification Number: | 21 CFR 884.1425 |
| Classification Name: | Perineometer |
| Review Panel: | Gastroenterology/Urology |
### 3. Predicate Information
| 510(k) Number | Trade Name | Model | Submitter |
|---------------|-----------------------------------|---------|--------------|
| K192270 | leva Pelvic Digital Health System | Leva-02 | Renovia Inc. |
### 4. Device Description
The leva Pelvic Health System ("leva System") is a prescription intra-vaginal device designed to allow the user (or woman) to rehabilitate and strengthen their pelvic floor muscles (PFM) as well as allow them to monitor their progress during pelvic floor muscle training (PFMT). The leva system is designed to wirelessly facilitate PFMT in women and to transmit real-time performance data through a dedicated mobile application that has been downloaded onto the patient's mobile device. The leva system is designed to be used vaginally and is intended to be used repeatedly by a single patient.
The leva PHS consists of a probe, storage case, associated batteries, and the Renovia Digital Health App (App). Thermoplastic elastomer (TPE) was used as the material overlay for the electronics and six accelerometers are contained within the probe. Additional electronics are contained in the storage case to transmit data wirelessly between the device and the App.
{4}------------------------------------------------
## 5. Indications for Use
The leva Pelvic Health System is intended for:
- 1) Strengthening of the pelvic floor muscles;
- 2) Rehabilitation and training of weak pelvic floor muscles for the treatment of stress, mixed and mild to moderate urgency urinary incontinence (including overactive bladder) in women.
This device interacts with the user via smart phone technology.
### 6. Comparison of Technological Characteristics
The following table provides a comparison of the leva Pelvic Health System to the predicate device.
| Characteristic | Predicate Device | Subject Device |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Information | | |
| Device Name | leva Pelvic Digital Health System | leva Pelvic Health System |
| Manufacturer | Renovia Inc. | Renovia Inc. |
| Technological Characteristics | | |
| Principle of<br>Operation | Provides indication of relative<br>intensity of pelvic floor muscle<br>contraction using accelerometers | Provides indication of relative<br>intensity of pelvic floor muscle<br>contraction using accelerometers |
| Muscle Stimulation | No | No |
| Intended Anatomical<br>Location | Vagina | Vagina |
| Use Model | Single patient, reusable | Single patient, reusable |
| Sterility | Clean, non-sterile | Clean, non-sterile |
| Display Information | Graphical and numeric based on<br>applied bending, anatomical overlay | Graphical and numeric based on<br>applied bending, anatomical overlay |
| Device Materials | • Probe: Thermoplastic Elastomer<br>(TPE)<br>• Probe Battery Pack: Acrylonitrile<br>Butadiene Styrene (ABS) | • Probe: Thermoplastic Elastomer<br>(TPE)<br>• Probe Battery Pack: Acrylonitrile<br>Butadiene Styrene (ABS) |
| Patient Contact<br>(Contact Duration) | • Probe: Direct (<24 hours)<br>• Probe Battery Pack: Direct (<24<br>hours)<br>• Vaginal Probe Hub: Indirect<br>(Incidental)<br>• Battery Pack Cap: Indirect<br>(Incidental) | • Probe: Direct (<24 hours)<br>• Probe Battery Pack: Direct (<24<br>hours)<br>• Vaginal Probe Hub: Indirect<br>(Incidental)<br>• Battery Pack Cap: Indirect<br>(Incidental) |
See Section 7 for discussion of modifications made to the leva Pelvic Health System.
{5}------------------------------------------------
## 7. Summary of Nonclinical Testing and Risk Analysis
The following non-clinical performance testing and risk analysis was performed to support modifications to the leva Pelvic Health System:
### A. Biocompatibility
Biocompatibility assessment was done, as needed, based on the changes made to the device. To ensure patientsafety, all the patient-contacting material was subjected to biocompatibility testing in compliance with ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, for mucosal surface for limited, less than 24 hours duration including:
- Cytotoxicity (ISO 10993-5)
- . Sensitization (ISO 10993-10)
- Irritation (ISO 10993-10)
Biocompatibility testing demonstrated that the subject device, leva Pelvic Health System, is as safe and effective, and performs as well as the predicate device, leva Pelvic Digital Health System.
### B. The sponsor reported other changes in this special 510 (k) which are stated below:
- Hardware (H): Change of the adhesive on the battery case, Printed Circuit Board (PCB)/Printed Circuit . Board Assemblies (PCBA), battery vendor change, and product labeling modifications.
- Embedded software (firmware) (F): Software changes related to performance improvement, communications, diagnostic and bug fix modifications.
- . Mobile Application Software (RA): Updates and enhancements for bug fixes and addition/modification of general content (e.g., educational videos added) to provide additional information and clarity to improve the user experience.
- Manufacturing (M) and Risk Analysis (D): Manufacturing quality control related changes and updates to the risk analysis modifications.
The following testing was done on the subject device to ensure that the device continues to meet the requirements for substantial equivalence:
- Software and firmware design verification testing
- Manufacturing testing protocol verification
All the changes reported above are minor and did not alter the substantial equivalence of the leva Pelvic Health System.
### 8. Conclusion
Based on the comparison and analysis above, Renovia has demonstrated that the leva Pelvic Health System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.