extriCARE 1000 Negative Pressure Wound Therapy System
K213906 · Alleva Medical, Ltd. · OMP · Mar 8, 2023 · General, Plastic Surgery
Device Facts
Record ID
K213906
Device Name
extriCARE 1000 Negative Pressure Wound Therapy System
Applicant
Alleva Medical, Ltd.
Product Code
OMP · General, Plastic Surgery
Decision Date
Mar 8, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The extriCARE® 1000 NPWT System is intended to generate negative pressure to remove wound exudates, infectious material, and tissue debris from the wound bed. The extriCARE® 1000 NPWT System is intended for use in wounds with low to moderate levels of exudate. Appropriate wound types include: Chronic wounds Acute wounds Traumatic wounds Subacute and dehisced wounds Partial-thickness burns Ulcer wounds (such as diabetic or pressure) Flaps and grafts Closed surgical incisions The extriCARE 1000 NPWT System is intended for use in healthcare facilities.
Device Story
Portable, battery-powered NPWT pump; generates negative pressure to remove wound exudates, infectious material, and tissue debris. Operates via air pump, solenoid valve, and pressure sensor; user selects 80, 100, or 125mmHg. Used in healthcare facilities by clinicians. System monitors wound site pressure via proprietary algorithm; maintains target vacuum by activating pump for leaks or using valves to vent excess pressure. Output is vacuum therapy; healthcare providers use to promote wound healing. Benefits include exudate management and wound bed preparation.
Clinical Evidence
Bench testing only. Testing included electromagnetic compatibility (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11) and functional performance testing (attribute, measurement, pneumatic, exudate removal, destructive, degradation, temperature, and transportation simulation). All requirements met.
Technological Characteristics
Portable battery-powered pump (2x AA 1.5V alkaline). Components: pump unit and 50cc collection canister. Sensing: pressure sensor. Actuation: air pump, solenoid valve, check valve. Vacuum settings: 80, 100, 125mmHg. Flow rate: ~0.5LPM. Connectivity: standalone. Software: moderate level of concern.
Indications for Use
Indicated for patients with chronic, acute, traumatic, subacute, dehisced wounds, partial-thickness burns, diabetic/pressure ulcers, flaps, grafts, and closed surgical incisions with low to moderate exudate. Contraindicated for exposed vessels, organs, or nerves; anastomotic sites; fistulas (unexplored/non-enteric); untreated osteomyelitis; malignancy in the wound; necrotic tissue with eschar; wounds too large/deep for dressing; inability to follow medical professional instructions; and allergy to silicone dressings/adhesives.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
PICO 14 Single Use Negative Pressure Wound Therapy System (K191760)
extriCARE 2400 Negative Pressure Wound Therapy System (K140634)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 8, 2023
Alleva Medical Ltd Max Choi CEO Suite M-Q, 12th Floor, Kings Wing Plaza 2, 1 On Kwan Street Shek Mun, New Territories Hong Kong
Re: K213906
Trade/Device Name: extriCARE® 1000 Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: November 30, 2021 Received: December 14, 2021
Dear Max Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Julie A. Morabito -S
Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K213906
#### Device Name
extriCARE® 1000 Negative Pressure Wound Therapy System
#### Indications for Use (Describe)
The extriCARE® 1000 NPWT System is intended to generate negative pressure to remove wound exudates, infectious material, and tissue debris from the wound bed.
The extriCARE® 1000 NPWT System is intended for use in wounds with low to moderate levels of exudate. Appropriate wound types include:
- Chronic wounds
- Acute wounds
- · Traumatic wounds
- · Subacute and dehisced wounds
- · Partial-thickness burns
- · Ulcer wounds (such as diabetic or pressure)
- · Flaps and grafts
- · Closed surgical incisions
The extriCARE 1000 NPWT System is intended for use in healthcare facilities.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" width="12" x="0" y="0"></rect> <path d="M2,2 L10,10 M2,10 L10,2" stroke="black"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <svg height="12" width="12"> <rect fill="none" height="12" stroke="black" width="12" x="0" y="0"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Alleva Medical. The logo consists of the word "ALLEVA" in a bold, dark blue font, with a stylized "A" symbol to the left. Below the word "ALLEVA" is the word "MEDICAL" in a smaller, dark blue font.
## 510(k) Summary
[as required by 21 CFR 807.92(c)]
## extriCARE 1000 Negative Pressure Wound Therapy System
#### 510(k) K213906
| DATE PREPARED: | March 8, 2023 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT | Alleva Medical Ltd<br>Suite M-Q, 12th Floor, Kings Wing Plaza 2<br>1 On Kwan Street, Shek Mun, Shatin<br>New Territories, Hong Kong |
| CONTACT | Mr. Max Choi<br>Chief Executive Officer<br>Phone: +852-3166-7239<br>Fax: +852-2355-7663<br>Email: max@allevamedical.com |
| TRADE NAME: | extriCARE 1000 NPWT System |
| DEVICE CLASSIFICATION: | Class II per 21 CFR 878.4780 |
| CLASSIFICATION NAME: | Negative Pressure Wound Therapy Pump |
| PRODUCT CODE | OMP |
| PREDICATE DEVICE: | PICO 14 Single Use Negative Pressure Wound Therapy System<br>(K191760) |
| REFERENCE DEVICES: | SVED Wound Treatment System (K142916)<br>extriCARE 2400 Negative Pressure Wound Therapy System (K140634) |
{4}------------------------------------------------
# MEDICAL
#### Indications for Use:
The extriCARE® 1000 NPWT System is intended to generate negative pressure to remove wound exudates, infectious material, and tissue debris from the wound bed.
The extriCARE 1000 NPWT System is intended for use in wounds with low to moderate levels of exudate. Appropriate wound types include:
- Chronic wounds
- Acute wounds
- Traumatic wounds
- Subacute and dehisced wounds
- Partial-thickness burns
- · Ulcer wounds (such as diabetic or pressure)
- Flaps and grafts
- Closed surgical incisions
The extriCARE 1000 NPWT System is intended for use in healthcare facilities.
#### Device Description:
The extriCARE 1000 Negative Pressure Wound Therapy (NPWT) system consists of a portable, battery-powered pump unit intended to generate vacuum pressure to remove low to moderate level of wound exudates, infectious material, and tissue debris from the wound bed. Three preset vacuum pressures can be selected by the user - 80, 100, and 125mmHg. The extriCARE 1000 is packaged and provided with the following components:
- 1. extriCARE 1000 Negative Pressure Wound Therapy Pump
- 2. extriCARE 1000 50cc Collection Canister
The extriCARE 1000 Negative Pressure Wound Therapy (NPWT) Pump accessory that is included in this 510(k) applications and will be commercialized separately is the extriCARE 1000 50cc Collection Canister (additional cannister replacements).
{5}------------------------------------------------
#### Summary of Technological Characteristics
The extriCARE 1000 pump, similar to its predicate and reference device, operates by having an air pump, a solenoid valve, and a pressure sensor working in conjunction to deliver a specific target vacuum pressure on the wound bed. When the therapy begins, air and wound exudate are removed from the dressing's underside (wound bed) via the connection tubing into the collection canister until the target vacuum is reached. A pressure sensor inside the pump, along with a proprietary algorithm, approximates and monitors the pressure on the wound site. If a leakage on the wound dressing occurs, the device activates its air pump to remove air until the target vacuum is reestablished. In the event that actual vacuum pressure exceeds target vacuum pressure, a solenoid valve and a check valve will act to allow ambient air into the system and lower the vacuum pressure to the desired level.
#### Summary of Non-Clinical Testing:
The following non-clinical testing has been completed to demonstrate that the extriCARE 1000 NPWT pump is performing as intended, has performance characteristic that are substantially equivalent to the listed predicate devices, and conforms to internal standards and regulations.
#### Electromagnetic Compatibility and Electrical Safety Testing
| Standards and Regulations Applied | Results |
|--------------------------------------------------------------------------|---------|
| IEC 60601-1:2005 (Third Edition) + A1:2012(or IEC 60601-1: 2012 reprint) | Passed |
| IEC 60601-1-2:2014 | Passed |
| IEC 60601-1-6:2010 Edition 3.1, AMD1:2013 | Passed |
| IEC 60601-1-11:2015 Edition 2.0 | Passed |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Alleva Medical. The logo consists of a stylized letter A with a blue triangle in the upper left corner, followed by the word "ALLEVA" in dark blue, sans-serif font. Below "ALLEVA" is the word "MEDICAL" in a smaller, dark blue, sans-serif font.
#### Functional Performance Testing
| Test Items | Results |
|--------------------------------|----------------------|
| Attribute Test | All requirements met |
| Measurement Test | All requirements met |
| System and Pneumatic Test | All requirements met |
| Exudate Removal Test | All requirements met |
| General Destructive Test | All requirements met |
| Performance Degradation Test | All requirements met |
| Hot and Cold Operation Test | All requirements met |
| Hot and Cold Temperature Test | All requirements met |
| Transportation Simulation Test | All requirements met |
### Software Verification and Validation
Software verification and validation activities were completed and there were no unresolved anomalies. The software was determined to be of a "moderate" level of concern, since a failure or latent flaw in the software could directly result in minor injury to the user (and delayed therapy).
{7}------------------------------------------------
| Selected<br>Features | extriCARE 1000<br>NPWT System<br>(Subject Device) | Smith & Nephew<br>PICO 14 NPWT<br>System<br>(Predicate)<br>K191760 | SVED Wound<br>Treatment System<br>(Reference)<br>K142916 | extriCARE 2400 NPWT<br>System<br>(Reference)<br>K140634 |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use/<br>Indications<br>for Use | The extriCARE®<br>1000 NPWT<br>System is intended<br>to generate<br>negative pressure<br>to remove wound<br>exudates,<br>infectious material,<br>and tissue debris<br>from the wound<br>bed.<br><br>The extriCARE<br>1000 NPWT<br>System is intended<br>for use in wounds<br>with low to<br>moderate levels of<br>exudate.<br>Appropriate wound<br>types include:<br>• Chronic wounds<br>• Acute wounds<br>• Traumatic<br>wounds<br>• Subacute and<br>dehisced wounds<br>• Partial-thickness<br>burns<br>• Ulcer wounds<br>(such as diabetic or<br>pressure)<br>• Flaps and grafts<br>• Closed surgical | PICO is indicated<br>patients who<br>would benefit from<br>a suction device<br>(NPWT) as it may<br>promote would<br>healing via<br>removal of low to<br>moderate levels of<br>exudate and<br>infectious<br>materials.<br><br>Appropriate<br>wound types<br>include<br>• chronic,<br>• acute,<br>• traumatic,<br>• subacute and<br>dehisced<br>wounds,<br>• partial-<br>thickness<br>burns<br>• flaps and<br>grafts<br>• closed surgical<br>incisions<br>• ulcers (such<br>as diabetic or<br>pressure) | The Cardinal Health<br>NPWT SVED system<br>is an integrated wound<br>management system,<br>indicated for the<br>application of<br>continual or<br>intermittent negative<br>pressure wound<br>therapy to the wound<br>as the device may<br>promote wound<br>healing by the removal<br>of fluids, including<br>wound exudates,<br>irrigation fluids, body<br>fluids and infectious<br>materials. The system<br>is intended for patients<br>with chronic, acute,<br>traumatic, subacute<br>and dehisced wounds,<br>partial-thickness<br>burns, ulcers (such as<br>diabetic or pressure),<br>flaps and grafts. The<br>system is intended for<br>use in acute, extended<br>and home care<br>settings. | The extriCARE® NPWT<br>foam kit is intended to be<br>used in conjunction with<br>the extriCARE® 2400<br>NPWT pump.<br><br>The extriCARE® 2400<br>Negative Pressure Wound<br>Therapy System is<br>indicated for wound<br>management via the<br>application of negative<br>pressure to the wound by<br>the removal of wound<br>exudate,<br>infectious<br>materials, and tissue debris<br>from the wound bed.<br><br>The extriCARE Negative<br>Pressure Wound Therapy<br>System is indicated for the<br>following wound types:<br>chronic, acute, traumatic,<br>subacute and dehisced<br>wounds, partial-thickness<br>burns, ulcers (such as<br>diabetic or pressure), flaps<br>and grafts. |
| Selected<br>Features | extriCARE 1000<br>NPWT System<br>(Subject Device) | Smith & Nephew<br>PICO 14 NPWT<br>System<br>(Predicate) | SVED Wound<br>Treatment System<br>(Reference) | extriCARE 2400 NPWT<br>System<br>(Reference) |
| | | K191760 | K142916 | K140634 |
| | incisions<br>The extriCARE<br>1000 NPWT<br>System is intended<br>for use in<br>healthcare<br>facilities. | PICO 14 Single<br>Use Negative<br>Pressure Wound<br>Therapy System is<br>suitable for use<br>both in a hospital<br>and homecare<br>setting. | | |
| Contra-<br>indications<br>for Use | Exposed vessels, organs, or nerves. Anastomotic sites. Exposed arteries or veins in a wound. Fistulas, unexplored or non-enteric. Untreated osteomyelitis. Malignancy in the wound. Excess amount of necrotic tissue with eschar. Wounds which are too large or too deep to be accommodated by the dressing. Inability to be followed by a medical professional or to keep scheduled appointments. | Patients with malignancy in the wound bed or margins of the wound (except in palliative care to enhance quality of life) Previously confirmed and untreated osteomyelitis Non-enteric and unexplored fistulas Necrotic tissue with eschar present Exposed arteries, veins, nerves or organs Exposed anastomotic sites The PICO 14 | The SVED is contraindicated for patients with malignancy in the wound, untreated osteomyelitis, non-enteric and unexplored fistulas, or necrotic tissue with eschar present. Do not place the NPWT Dressing over exposed blood vessels or organs. The NPWT Dressings are also contraindicated for hydrogen peroxide and solutions which are alcohol based or contain alcohol. It is not recommended to deliver fluids to the thoracic cavity. | Exposed vessels, organs, or nerves. Anastomotic sites. Exposed arteries or veins in a wound. Fistulas, unexplored or non-enteric. Untreated osteomyelitis. Malignancy in the wound. Excess amount of necrotic tissue with eschar. Wounds which are too large or too deep to be accommodated by the dressing. Inability to be followed by a medical professional or to keep scheduled appointments. Allergy to urethane dressings and adhesives. Use of topical products which must |
| Selected<br>Features | extriCARE 1000<br>NPWT System<br>(Subject Device) | Smith & Nephew<br>PICO 14 NPWT<br>System<br>(Predicate)<br>K191760 | SVED Wound<br>Treatment System<br>(Reference)<br>K142916 | extriCARE 2400 NPWT<br>System<br>(Reference)<br>K140634 |
| | Allergy to silicone dressings and adhesives. Use of topical products which must be applied more frequently than the dressing change schedule allows | System should not be used for the purposes of: Emergency airway aspiration Pleural, mediastinal or chest tube drainage Surgical suction | | be applied more frequently than the dressing change schedule allows |
| Use | Multi-patient | Single-patient | Multi-patient | Single-patient |
| Prescription<br>Use | Yes | Yes | Yes | Yes |
| Vacuum<br>mode | Continuous and<br>Intermittent | Continuous only | Continuous and<br>Intermittent | Continuous and<br>Intermittent |
| Pressure<br>settings | 3, selectable;<br>80/100/125mmHg | Vary from 60 to<br>100 mm Hg | 70, 120 and 150mmHg | 40 - 140mmHg |
| Flow Rate | ~0.5LPM | ~0.5LPM | Unknown | About 1.0LPM @<br>100mmHg |
| Canister | 50cc | No | 300cc and 500cc | 100cc and 400cc |
| Battery | External alkaline,<br>2 x AA 1.5V | External alkaline,<br>2 x AA 1.5V | Internal Rechargeable<br>Battery | Internal Rechargeable<br>Lithium battery, 3.7V,<br>1500mAh |
| Use<br>barometric<br>environment | 800hPa – 1060hPa | 800hPa – 1060hPa | 50kPa-110kPa | 800hPa – 1060hPa |
## Summary of Substantial Equivalence
{8}------------------------------------------------
# �ALLEVA MEDICAL
{9}------------------------------------------------
# �ALLEVA MEDICAL
{10}------------------------------------------------
# ALLEVA MEDICAL
Based on its operating principles, specifications, indications for use, the extriCARE 1000 is substantially equivalent to the predicate devices that were previously cleared. The system does not raise any new risks when compared to the predicate devices.
#### Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this pre-market notification, Alleva Medical Ltd, believes that the extriCARE 1000 NPWT system performs as well as the predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.