The Caretaker Advanced Hemodynamic Parameters provides calibrated cardiac output/stroke volume (CO/SV), left ventricular ejection time (LVET), and heart rate variability (HRV) in adult patients to the existing Caretaker Remote Display App And Caretaker Software Library (K181196) via Pulse Decomposition Analysis ("PDA")(K211588, K163255, K151499). To provide CO/SV measurements. the Caretaker platform is to be calibrated with a thermodilution measurement, or other accurate reference determination of cardiac output, to ensure accuracy. The device is intended for use by physicians or other properly trained medical personnel in a hospital or other appropriate clinical setting.
Device Story
Firmware upgrade for Caretaker Physiological Monitor platform; utilizes Pulse Decomposition Analysis (PDA) to process arterial pulse signals from finger cuff. Transforms pulse contour into hemodynamic parameters: cardiac output (CO), stroke volume (SV), left ventricular ejection time (LVET), and heart rate variability (HRV). CO/SV requires calibration via thermodilution or other reference method. Used in hospitals/clinical settings by physicians or trained personnel. Output displayed on Caretaker Remote Display App/Software Library. Assists clinicians in hemodynamic assessment and management of patients with complex cardiac conditions. Benefits include non-invasive, continuous monitoring of advanced hemodynamic metrics.
Clinical Evidence
Clinical validation performed across multiple cohorts. CO/SV validation involved 47 subjects (mean age 66.8, 57.5% male) undergoing cardiac catheterization; compared against reference methods. LVET validation compared against central aortic BP tracings. HRV validation involved 60 subjects (mean age 50.9) undergoing physiological interventions (ice-bath, Valsalva, etc.). Results demonstrate equivalent performance to predicate devices.
Technological Characteristics
Firmware-based hemodynamic parameter extraction; utilizes finger cuff arterial pulse sensor. Employs Pulse Decomposition Analysis (PDA) and modified Wesseling/Windkessel modeling. Connectivity via Caretaker platform to Remote Display App. Complies with IEC 60601 safety standards.
Indications for Use
Indicated for adult patients requiring hemodynamic monitoring, including those with challenged cardiac physiologies, hypertension, diabetes, or undergoing cardiac catheterization. No specific contraindications listed.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
March 30, 2022
Caretaker Medical, LLC Jeff Pompeo President & CEO 941 Glenwood Station Ln, Suite 301 Charlottesville, Virginia 22901
Re: K213699
Trade/Device Name: Caretaker Advanced Hemodynamic Parameters Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN, DRG Dated: March 1, 2022 Received: March 2, 2022
Dear Jeff Pompeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K213699
Device Name
# Caretaker Advanced Hemodynamic Parameters
Indications for Use (Describe)
The Caretaker Advanced Hemodynamic Parameters provides calibrated cardiac output/stroke volume (CO/SV), left ventricular ejection time (LVET), and heart rate variability (HRV) in adult patients to the existing Caretaker Remote Display App And Caretaker Software Library (K181196) via Pulse Decomposition Analysis ("PDA")(K211588, K163255, K151499). To provide CO/SV measurements. the Caretaker platform is to be calibrated with a thermodilution measurement, or other accurate reference determination of cardiac output, to ensure accuracy. The device is intended for use by physicians or other properly trained medical personnel in a hospital or other appropriate clinical setting.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# 510(k) Summary of Safety and Effectiveness
- 1) Preparation Date: 29 Mar 2022
## 2) Submitted by:
Caretaker Medical, LLC 941 Glenwood Station Ln, Suite 301 Charlottesville, Virginia 2290 User Fee Organization Number 397095 Owner/Operator #: 10054848
3) Contact Person/Prepared by: Jeff Pompeo President & CEO CareTaker Medical 941 Glenwood Station Ln, Suite 301 Charlottesville, Virginia 22901-1455 Phone: 434-409-1945 Email: Jeff@caretakermedical.net
#### 4) Device Identification:
| Trade Name: | Caretaker Advanced Hemodynamic Parameters |
|-----------------------|------------------------------------------------------------------------|
| Common Name: | Caretaker Advanced Hemodynamic Parameters |
| Product Code(s): | DXN -System, Measurement, Blood-Pressure, Non-Invasive |
| | DRG - Transmitters and Receivers, Physiological signal, Radiofrequency |
| Regulation Number(s): | 21 CFR 870.1130 |
| | 21 CFR 870.2910 |
| Device Class: | II |
Device Class: Il
- ર) Nexfin_HD (for cardiac output/stroke volume, K072049), Finapres Predicate Devices: Portapres model 2 (for left ventricular ejection time, K023338), and ANSWatch (for heart rate variability measures, K123130).
- 6) Device Description: The Caretaker Advanced Hemodynamic Parameters is a firmware upgrade that runs on the Caretaker platform to provide additional hemodynamic measures to the Caretaker Remote Display App And Caretaker Software Library, K181196, and CareTaker Physiological Monitor, (K211588, K163255, K151499). These parameters are not intended to predict or detect cardiovascular mortality or any other condition, disease, and/or patient outcome.
- 7) Intended Use: The Caretaker Advanced Hemodynamic Parameters provides calibrated cardiac output/stroke volume (CO/SV), left ventricular ejection time (LVET), and heart rate variability (HRV) in adult patients to the existing Caretaker Remote Display App And Caretaker Software Library (K181196) via Pulse Decomposition Analysis ("PDA")(K211588, K163255, K151499). To provide CO/SV measurements, the Caretaker platform is to be calibrated with a thermodilution measurement, or other accurate reference determination of cardiac output, to ensure
{4}------------------------------------------------
accuracy. The device is intended for use by physicians or other properly trained medical personnel in a hospital or other appropriate clinical setting.
Page 2 of 3
### 8) Comparison to Predicates:
### ● Cardiac Output / Stroke Volume
The patient groups represent challenged cardiac physiologies that necessitated the monitoring of cardiac output among other vital signs. Common to all groups is a high incidence of hypertension, 76%, as well as diabetes, 26%. Less than 16%, across all cohorts, have a normal BMI<25. The CTM PDA model currently does not use patient demographic data because internally obtained correlations between such data and hemodynamic parameters such as blood pressure, CO etc. are weak.
CTM's approach to calculating CO incorporates refined approaches to modeling efforts that others have pursued. The Windkessel, Area-Under-the-Curve, approach that was originally proposed by Warner 1 is based on two concepts: A) mass is conserved, meaning the volume of blood entering a vessel is equal to the volume leaving, B) compliance is part of the model since during systole, as pressure inside the vessel is increasing, the vessel expands and absorbs some of the blood that would otherwise pass through it. During diastole, as the pressure inside the vessel drops, the vessel contracts and the additional blood that was stored during systole is expelled. After dividing the pulse contour into systolic and diastolic regions, flow equations for both regions can be established that relate stroke volume to the ratio of systolic to diastolic pulse areas and the end-systolic pressure while aortic resistance is incorporated via a proportionality constant.
The Caretaker Physiological Monitor System is substantially equivalent to the NexFin HD in terms of providing the hemodynamic parameters cardiac output and stroke volume based on arterial pulse signals obtained from a finger cuff, preferably with calibration via thermodilution, although Caretaker uses a slightly different methodology (PDA/modified Wesseling) to obtain these hemodynamic parameters. Caretaker has clinically validated evidence attesting equivalent performance to that of the predicate device and followed quality management system rigor to comply with IEC 60601 safety standards.
## Left Ventricular Ejection Time (LVET) ●
Baseline demographic information is presented below
| Characteristic | Mean (SD) or N (%) |
|-----------------------------------------------------------|--------------------|
| Age - mean (std. dev.) – yr. | 66.8 (9.8) |
| Body Mass Index - mean (std. dev.) – kg/m² | 31.4 (7.3) |
| Male Sex - no. (%) | 27 (57.5) |
| Tobacco Use - no. (%) | 31 (66.0) |
| Hypertension – no. (%) | 40 (85.1) |
| Diabetes Mellitus - no. (%) | 22 (46.8) |
| Indication for Cardiac Catheterization - no. (%) | |
| Acute Coronary Syndrome, Stable Angina, or known Coronary | 28 (59.6) |
{5}------------------------------------------------
K213699
| Artery Disease | |
|--------------------------------|-----------|
| Valvular Disease | 12 (25.5) |
| Other Indications | 7 (14.9) |
| Vascular Access Site - no. (%) | |
| Radial Artery | 26 (55.3) |
| Femoral Artery | 21 (44.7) |
Primary indications for cardiac catheterization procedures mostly consisted of coronary evaluation (37/47, 78%). Other primary indications included hemodynamic assessment for a suspected severe aortic stenosis. The radial artery was more commonly used for a vascular access (27/47, 57%) compared to the femoral artery. Of the study participants, 21% underwent percutaneous coronary intervention.
LVET was obtained from central aortic BP tracings by measuring the time interval between the initiating point of upstroke of the aortic pressure wave form and the dichrotic notch according to established guidelines.
The Caretaker Physiological Monitor System is substantially equivalent to the Portapres model 2 in terms of providing the LVET hemodynamic parameter based on arterial pulse signals obtained from a finger cuff. Both the predicate device and the Caretaker use pulse analysis of the arterial pressure pulse to determine LVET. Caretaker has clinically validated evidence attesting equivalent performance to that of the predicate device and followed quality management system rigor to comply with IEC 60601 safety standards.
## ● Heart Rate Variabilitv
Overall subject demographics were as follows: age (mean: 50.9 y (standard deviation: 16.8)), sex (34 m/26 f), height (67.1 inch (4.43)), weight (179 lbs. (36.5 BMI (28.2 (5.41)).
The protocol of the interventions utilized hand in ice-bath immersions, ice bag on forehead, light valsalva maneuvers, ball gripping and leg raises. In addition there were extended period of supine relaxation.
The Caretaker Physiological Monitor System is substantially equivalent to the ANSWatch in terms of providing heart rate variability measures derived from a peripheral mechanical arterial pulse sensor. The Caretaker has clinically validated evidence attesting performance equivalent to or better than the predicate device.
## Conclusions: 9)
- . Cardiac Output / Stroke Volume The CareTaker Monitor is substantially equivalent to the NexFin HD.
- Left Ventricular Ejection Time (LVET) The CareTaker Monitor is substantially equivalent to the ● Portapres model 2.
- Heart Rate Variability The CareTaker Monitor is substantially equivalent to the ANSWatch.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.