K213466 · Cruzar Medsystems, Inc. · DQY · Oct 20, 2022 · Cardiovascular
Device Facts
Record ID
K213466
Device Name
Houdini Cross Support Catheter
Applicant
Cruzar Medsystems, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Oct 20, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Houdini Cross Support Catheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the peripheral vasculature and for guidewire exchange.
Device Story
Bi-lumen intravascular catheter; used in peripheral vasculature (iliac, femoral, ilio-femoral, popliteal arteries). Operates with standard 0.014"-0.018" guidewire. Features distal anchoring balloon (inflated to 6 atm via manual inflation device with contrast media) and physician-controlled inner shaft extension (up to 2 cm) to support guidewire advancement. Radiopaque stainless-steel design allows fluoroscopic visualization. Used by clinicians in interventional settings to access remote vascular locations and facilitate guidewire exchange. Provides distal stability and support for guidewire navigation; benefits patient by enabling access to difficult peripheral lesions.
Clinical Evidence
No clinical data. Evidence consists of bench testing, biocompatibility evaluation, and simulated use testing. Bench testing per ISO 10555-1 included dimensional verification, crossing profile, radiopacity, corrosion resistance, tensile strength, pressurization, leakage, inflation fatigue, torque load, kink resistance, and particulate testing. Biocompatibility testing followed ISO 10993-1, covering cytotoxicity, sensitization, irritation, pyrogenicity, acute systemic toxicity, and hemocompatibility (hemolysis, PTT, platelet/leukocyte count, complement activation).
Technological Characteristics
Bi-lumen catheter; 5 Fr outer diameter; 100 cm effective length. Materials include radiopaque stainless steel. Distal anchoring balloon (5mm/6mm diameter, 2 cm length). Energy source: manual inflation device (hydraulic). Connectivity: none. Sterilization: Ethylene Oxide. Standards: ISO 10555-1 (performance), ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for use in conjunction with a steerable guidewire to access discrete regions of the peripheral vasculature (iliac, femoral, ilio-femoral, and popliteal arteries) and for guidewire exchange in patients requiring peripheral vascular access.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
October 20, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" written in a clear, sans-serif font. The logo is simple and professional, reflecting the organization's role in regulating food and drugs.
Cruzar Medsystems Inc Adrienne Foller Consultant Strategic Quality Solutions LLC 1594 E Monaco Ave Salt Lake City, Utah 84121
Re: K213466
Trade/Device Name: Houdini Cross Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: January 13, 2022 Received: January 18, 2022
Dear Adrienne Foller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 5 10(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213466
Device Name Houdini Cross Support Catheter
Indications for Use (Describe)
The Houdini Cross Support Catheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the peripheral vasculature and for guidewire exchange.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Cruzar Medsystems, Inc. Houdini® Cross Support Catheter
#### 1. MANUFACTURER AND SUBMITTER
- Company Name: Cruzar Medsystems, Inc. Company Address: 50 Braintree Hill Office Park, Suite 301 Braintree, MA 02184
- Contact Person: Adrienne von Foller, Strategic Quality Solutions, LLC Regulatory Consultant Telephone: 801-916-8188 Email: Adrienne@device2market.com
Telephone: 781-223-0508
26 October 2021 Date Prepared:
#### 2. DEVICE NAME AND CLASSIFICATION
| Trade Name: | Houdini® Cross Support Catheter |
|----------------------|---------------------------------|
| Common Name: | Percutaneous catheter |
| Model Numbers: | CM-3502, CM-3602 |
| Classification Name: | Percutaneous catheter |
| Regulation Number: | 21 CFR 870.1250 |
| Regulatory Class: | II |
| Product Code: | DQY |
#### 3. PREDICATE DEVICE
Device Name: Houdini® Catheter 510(k) Applicant: Cruzar Medsystems, Inc. 510(k): K151896 There has been no recall on the Cruzar Medsystems Houdini® Catheter device.
#### 4. REFERENCE DEVICE
| Device Name: | XO Cross Support Catheter |
|-------------------|-------------------------------------------------------------------------------|
| 510(k) Applicant: | Transit Scientific |
| 510(k): | K193420 |
| : | There has been no recall on the Transit Scientific XO Cross Support Catheter. |
{4}------------------------------------------------
## 5. DEVICE DESCRIPTION
The Houdini® Cross Support Catheter is a single use, bi-lumen intravascular catheter intended for percutaneous use. The Houdini® Cross Support Catheter is intended for use in the iliac. femoral, ilio-femoral and popliteal arteries. It is designed for use in conjunction with a standard commercially available off the shelf (OTS) 0.014" - 0.018" quidewire (not included) to gain access to locations within the cardiovascular system that are remote from the site of insertion. Once accessed, guidewires may be exchanged within the catheter. In addition, the Houdini® Cross Support Catheter provides the option of distal anchoring as well as physician-controlled extension of the inner shaft (up to 2 cm) to support the OTS guidewire during advancement.
The effective length of the Houdini® Cross Support Catheter is a nominal 100 cm. The inner lumen accommodates an OTS 0.014" - 0.018" spring-tipped guidewire. The inner lumen shaft allows for visualization under fluoroscopy due to the radiopague stainlesssteel design and additionally allows for continuous clinician control of the OTS guidewire for appropriate positioning.
If distal anchoring is desired, the Houdini® Cross Support Catheter requires the use of a commercially available OTS manual inflation device (not included). Once positioned within 2 cm of the user's target location, the catheter is inflated to 6 atm using diluted radiopaque contrast media to anchor the distal end of the catheter in the blood vessel. For additional support of the quidewire, the clinician may extend the location of the inner lumen shaft up to 2 cm distally without the need to reposition the anchoring balloon.
## 6. INDICATIONS FOR USE
The Houdini® Cross Support Catheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the peripheral vasculature and for quidewire exchange.
The intended use and indications for use of the modified device are the same as the cleared, leqally marketed predicate device (K151896) and reference device (K193420).
### 7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Houdini® Cross Support Catheter and the predicate Houdini® Catheter (K151896) are both sterile, single-use devices designed to access discrete regions of the peripheral vasculature. A comparison of the technological characteristics of the subject device, the predicate Houdini® Catheter (K151896), and the reference XO Cross Support Catheter (K193420) is described below in Table 1.
{5}------------------------------------------------
| | Subject Device | Predicate Device | Reference Device |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Houdini Cross<br>Support Catheter | Houdini Catheter<br>(K151896) | XO Cross Support<br>Catheter (K193420) |
| Indications for Use | Intended to be used in<br>conjunction with a<br>steerable guidewire to<br>access discrete<br>regions of the<br>peripheral vasculature<br>and for guidewire<br>exchange. | Intended to be used in<br>conjunction with a<br>steerable guidewire to<br>access discrete<br>regions of the<br>peripheral vasculature<br>and for guidewire<br>exchange. | Intended to guide and<br>support a guidewire<br>during access of the<br>peripheral vasculature,<br>allow for wire<br>exchanges and provide<br>a conduit for the<br>delivery of saline<br>solutions or diagnostic<br>contrast agents. |
| Guidewire<br>Compatibility | 0.014" - 0.018" | 0.014" - 0.035" | 0.014" - 0.035" |
| Sheath Compatibility | 6 Fr | 6 Fr | 4 Fr |
| Catheter Outer<br>Diameter | 5 Fr | 5 Fr | 2 Fr - 4 Fr |
| Catheter Effective<br>Length | 100 cm | 100 cm | 90, 135, 150, 175 cm |
| Labeled Balloon<br>Diameter | 5mm<br>6mm | 4mm<br>5mm<br>6mm<br>7mm<br>8mm | No balloon component |
| Balloon Length | 2 cm | 2 cm | No balloon component |
| Anchoring Inflation<br>Pressure | 6 atm | 6 atm | |
| Maximum Working<br>Pressure | 9 atm | 12 atm | |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
## 8. PERFORMANCE DATA
The following performance testing has been completed in support of the substantial equivalence determination, including bench testing, biocompatibility, and simulated use testing.
{6}------------------------------------------------
## Biocompatibility Testing
The biocompatibility evaluation for the Houdini® Cross Support Catheter was conducted in accordance with the FDA Guidance "Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" dated September 04, 2020. The biological evaluation included the following:
- Material characterization .
- Cytotoxicity ●
- Sensitization ●
- Irritation ●
- Material-mediated Pyrogenicity ●
- Acute Systemic Toxicity
- Hemocompatibility: Hemolysis, Direct and Extract Methods ●
- Hemocompatibility: Partial Thromboplastin Time
- Hemocompatibility: Platelet and Leukocyte Count ●
- Hemocompatibility: Complement Activation .
#### Simulated Use Testing
The Houdini® Cross Support Catheter was subjected to simulated use testing to verify ability to access discrete regions of the peripheral vasculature and for guidewire exchange. Test results indicate that the device satisfies simulated use testing requirements when used as indicated.
### Performance Bench Testing
The Houdini® Cross Support Catheter has completed the following performance bench tests per ISO 10555-1. Test results indicate that the device satisfies functional performance requirements when used as indicated.
- Dimensional verification ●
- Crossing profile ●
- Radiopacity ●
- Corrosion resistance ●
- Tensile strength
- Pressurization performance
- Freedom from leakage
- Inflation fatigue
- Torque load ●
- Kink resistance ●
- Particulate testing ●
{7}------------------------------------------------
### 9. CONCLUSIONS
A comparison of the design and technological characteristics between the Houdini® Cross Support Catheter and the predicate Houdini® Catheter demonstrates substantial equivalence. The performance data for the subject device, including biological evaluation, bench testing, and simulated use, further supports the substantial equivalence of the Houdini® Cross Support Catheter.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.