K151896 · Cruzar Medsystems, Inc. · DQY · Dec 7, 2015 · Cardiovascular
Device Facts
Record ID
K151896
Device Name
Houdini Catheter
Applicant
Cruzar Medsystems, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Dec 7, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Houdini Catheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the peripheral vasculature and for guidewire exchange. The Houdini Catheter is indicated for use in the illac, femoral, ilio-femoral and popliteal arteries.
Device Story
Single-use, dual-lumen intravascular catheter; used with 0.014"-0.035" steerable guidewire. Physician-operated in peripheral vasculature (iliac, femoral, popliteal arteries). Distal tip contains inflatable balloon; inflated to 6atm with radiopaque contrast media via manual inflator. Balloon anchors catheter within vessel; stabilizes guidewire. Physician manually advances guidewire through proximal Y-connector. Radiopaque marker aids fluoroscopic visualization. Provides distal anchoring and support for guidewire advancement; facilitates access to remote cardiovascular locations. Benefits patient by enabling navigation and stabilization in complex peripheral anatomy.
Clinical Evidence
No clinical data. Bench testing only. Testing included biocompatibility (ISO 10993), simulated use (insertion, deployment, removal), radio-detectability (ASTM F640-12), peak tensile force, pressurization performance (leakage, burst pressure per ISO 10555-1), dimensional verification, torque load, crossing profile, and kink resistance.
Technological Characteristics
Dual-lumen catheter; 5Fr shaft; 100cm effective length. Distal inflatable balloon (4-8mm diameter). Materials biocompatible per ISO 10993. Sterilization via Ethylene Oxide. Radiopaque distal marker. Manual inflation to 6atm (max working 12atm; burst 16atm). Standards: ISO 10555-1 (catheter performance), ASTM F640-12 (radio-detectability).
Indications for Use
Indicated for patients requiring access to discrete regions of the peripheral vasculature, specifically the iliac, femoral, ilio-femoral, and popliteal arteries, for guidewire exchange and support.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of three abstract human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 7, 2015
Cruzar Medsystems, Inc. Ms. Gina To Regulatory Consultant 50 Braintree Hill Office Park, Suite 301 Braintree, MA 02184
Re: K151896
Trade/Device Name: Houdini Catheter Regulation Number: 21 CFR 870.1250 Regulatory Class: Class II Product Code: DQY Dated: October 19, 2015 Received: October 22, 2015
Dear Ms. To:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K151896
Device Name
Houdini Catheter
Indications for Use (Describe)
The Cruzar Medsystems Houdini Catheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the peripheral vasculature and for guidewire exchange.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cruzar Medsystems. The word "CRUZAR" is in gray, with a curved orange line above the "Z". Below "CRUZAR" is the word "MEDSYSTEMS" in orange. The logo is simple and modern.
# 510(k) Summary
# Cruzar Medsystems Houdini™ Catheter
### 1. MANUFACTURER AND SUBMITTER
| Company Name: | Cruzar Medsystems, Inc. |
|------------------|--------------------------------------------------------------------------------------------|
| Company Address: | 50 Braintree Hill Office Park, Suite 301<br>Braintree, MA 02184<br>Telephone: 781-223-0508 |
| Contact Person: | Gina To, Regulatory Consultant |
| Date Prepared: | November 23, 2015 |
#### 2. DEVICE NAME AND CLASSIFICATION
| Trade Name: | Houdini™ Catheter |
|----------------------|---------------------------------------------|
| Model Numbers: | CM-3400, CM-3500, CM-3600, CM-3700, CM-3800 |
| Classification Name: | Percutaneous catheter |
| Regulation Number: | 21 CFR 870.1250 |
| Regulatory Class: | II |
| Product Code: | DQY |
# 3. PREDICATE DEVICE
| Device Name: | ENDOCROSS, ENABLER-P CATHETER, 1000 PART<br>NUMBER: 1160 |
|-------------------|----------------------------------------------------------|
| 510(k) Applicant: | ENDOCROSS, LTD |
| 510(k): | K083833 |
There has been no recall on the Endocross Enabler-P device.
#### 4. DEVICE DESCRIPTION
The Houdini Catheter is a single use, dual-lumen intravascular catheter intended for percutaneous use. It is designed for use in conjunction with a 0.014" - 0.035" quide wire to gain access to locations within the cardiovascular system that are remote from the site of insertion. Once accessed, quide wires may be exchanged within the catheter. In addition, the Houdini Catheter can provide distal anchoring and supports the advancement of the guidewire.
The effective length of the Houdini Catheter is a nominal 100cm. The distal tip incorporates a radiopaque marker to aid in visualization under fluoroscopy. The inner
{4}------------------------------------------------
diameter of the Houdini Catheter shaft will accommodate a standard commercially available 0.014" - 0.035" guidewire.
The balloon catheter is inflated to 6atm with radiopaque contrast media with a standard manual inflator to anchor the catheter. Internal forces are placed on the inner lumen to the point that it grips and stabilizes the guidewire. Once pressurized to 6 atm, the guidewire is advanced by manual control of the proximal Y-connector by the physician.
The Houdini Catheter requires the use of traditional manual inflation devices (not included).
### 5. INTENDED USE/INDICATIONS FOR USE
The Houdini Catheter is intended to be used in conjunction with a steerable guidewire to access discrete regions of the peripheral vasculature and for quidewire exchange.
The Houdini Catheter is indicated for use in the illac, femoral, ilio-femoral and popliteal arteries.
The intended use and indications for use for the subject device are the same as the predicate device.
#### 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Houdini Catheter and the predicate Endocross Enabler-P Catheter (K083833) are both sterile, single-use devices designed to access discrete regions of the peripheral vasculature. The mechanism of operation of the non-therapeutic inflatable balloon elements at the distal tip of the Houdini Catheter is similar to the Enabler and facilitates stability, visual orientation and placement of the central quidewire. The Houdini Catheter has the same intended use and indication for use as the Enabler.
| Characteristic | Houdini Catheter – Subject Device | Predicate Device Endocross, Enabler-P Catheter (K083833) |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Intended to be used in conjunction with a steerable guidewire to access discrete regions of the peripheral vasculature and for guidewire exchange. | Intended to be used in conjunction with a steerable guidewire to access discrete regions of the peripheral vasculature and for guidewire exchange |
| Nominal Balloon Diameter | Model CM-3400: 4mm<br>Model CM-3500: 5mm<br>Model CM-3600: 6mm | 6mm |
A comparison the technological characteristics of the Houdini Catheter and the predicate Endocross Enabler-P Catheter (K083833) is tabulated below.
{5}------------------------------------------------
| Characteristic | Houdini Catheter – Subject<br>Device | Predicate Device Endocross,<br>Enabler-P Catheter (K083833) |
|-----------------------------------|------------------------------------------|-------------------------------------------------------------|
| | Model CM-3700: 7mm<br>Model CM-3800: 8mm | |
| Catheter Effective<br>Length | 100cm | 103cm |
| Maximum Balloon<br>Working Length | 2cm | 2cm |
| Anchoring Inflation<br>Pressure | 6atm | 1.5atm |
| Maximum Working<br>Pressure | 12atm | 6atm |
| Rated Burst<br>Pressure | 16atm | 10atm |
| Catheter Shaft<br>Diameter | 5Fr | 5Fr |
| Guidewire Size | 0.014 - 0.035 inches | 0.035 inches |
| Introducer Sheath<br>Size | 6Fr | 7Fr |
| Peak Force Delivery | 0.35kgf to 0.71kgf | 0.026Kgf to 0.185 |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide |
# 7. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility Testing
The biocompatibility evaluation for the Houdini Catheter was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests: cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, hemocompatibility, and thrombogenicity. Biocompatibility test results indicate that the device materials are biocompatible.
# Simulated Use Testing
The Houdini Catheter was subjected to simulated use testing to assess device insertion, deployment and device removal.
{6}------------------------------------------------
# Performance Bench Testing
The Houdini Catheter has completed the performance bench tests as listed below. Test results indicate that the device satisfies functional performance requirements when used as indicated.
- Simulated use validation ●
- Device insertion, deployment, removal
- o Surface, Distal Tip, ISO 10555-1
- o Introducer sheath sizing
- Radio-detectability, ASTM F640-12 ●
- Peak tensile force, ISO 10555-1 Annex B ●
- . Pressurization Performance
- Freedom from fluid leakage, ISO 10555-1 Annex C
- o Freedom from Air leakage, Hubs, ISO 10555-1 Annex D
- Burst pressure, ISO 10555-1 Annex F o
- o Inflation pressure vs. diameter
- Peak delivery force versus inflation pressure
- Dimensional verification ●
- Tensile strength
- Torque load ●
- . Crossing Profile
- Kink Resistance
# 8. CONCLUSIONS
A comparison of the technological characteristics between the Houdini Catheter and the predicate Endocross, Enabler-P Catheter shows that they are largely similar. While there are some differences in technological characteristics between the subject device and the predicate device, these differences do not raise new questions of safety or effectiveness. The performance data for the Houdini Catheter demonstrate that it should perform as intended for the indicated use, the same indicated use as the predicate device. Therefore, the subject device is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.