Midas Rex MR8 Depth Stop Attachment, Midas Rex MR8 Depth Stop Tool 2.4mm, Midas Rex MR8 Depth Stop Tool 3.2mm, Midas Rex MR8 Depth Stop Tool 3.2mm Long, Midas Rex MR8 Depth Stop Tool 7.5mm
Applicant
Medtronic Powered Surgical Solutions
Product Code
HBE · Neurology
Decision Date
Nov 23, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Midas Rex MR8 Depth Stop attachment and tools are indicated for the incision, cutting, removing, and drilling of soft and hard tissue during cranial surgical procedures with the intent to create a hole through the cranium to allow surgeons access to desired surgical locations and/or to facilitate insertion, placement of other surgical devices during such procedures.
Device Story
System comprises depth stop attachments and surgical dissecting tools used with MR8 high-speed drill systems (pneumatic/electric) and Stealth Autoguide system; creates cranial access holes of controlled diameter and depth. Used in cranial surgery by surgeons. Device provides mechanical depth control during drilling; output is physical hole in cranium. Benefits include precise access to surgical sites and facilitation of device placement. Modifications from predicate include compatibility with MR8 drill systems, extended depth setting to 20mm, minor tool head geometry changes, and updated user workflow.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included product life verification, summative validation, and biocompatibility testing to demonstrate safety and effectiveness of modifications.
Indicated for incision, cutting, removing, and drilling of soft and hard tissue during cranial surgical procedures to create cranial access holes for surgical access or device placement.
Regulatory Classification
Identification
Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.
Predicate Devices
Midas Rex Legend Depth Stop Attachment and Tools (K191597)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 23, 2021
Medtronic Powered Surgical Solutions Carrie Eddings Principal Regulatory Affairs Specialist 4620 North Beach Street Fort Worth, Texas 76137
Re: K213454
Trade/Device Name: Midas Rex MR8 Depth Stop Attachment, Midas Rex MR8 Depth Stop Tool 2.4mm, Midas Rex MR8 Depth Stop Tool 3.2mm, Midas Rex MR8 Depth Stop Tool 3.2mm Long, Midas Rex MR8 Depth Stop Tool 7.5mm Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, And Their Accessories Regulatory Class: Class II Product Code: HBE, HBB, HBC Dated: October 22, 2021 Received: October 26, 2021
Dear Carrie Eddings:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213454
#### Device Name
Midas Rex MR8 Depth Stop Attachment (MR8-ASDS01 ); Midas Rex MR8 Depth Stop Tool 2. 4mm (MR8-DS1TD24); Midas Rex MR8 Depth Stop Tool 3.2mm (MR8-DS1TD32);Midas Rex MR8 Depth Stop Tool 3.2mm Long (MR8-DS1TD32L); Midas Rex MR8 Depth Stop Tool 7.5mm (MR8-DS1TD75)
#### Indications for Use (Describe)
The Midas Rex MR8 Depth Stop attachment and tools are indicated for the incision, cutting, removing, and drilling of soft and hard tissue during cranial surgical procedures with the intent to create a hole through the cranium to allow surgeons access to desired surgical locations and/or to facilitate insertion, placement of other surgical devices during such procedures.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">✘</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
# November 23, 2021
- Company: Medtronic Powered Surgical Solutions 4620 North Beach Street Fort Worth, Texas 76137 Telephone Number: (817) 788-6400
- Carrie Eddings (Primary) Contact: Principal Regulatory Affairs Specialist Telephone Number: (214) 384-0606 Email: carrie eddings(@medtronic.com
Kyle Hoefling (Alternate) Senior Regulatory Affairs Manager Telephone: (760) 207-2432 Email: kyle.d.hoefling(@medtronic.com
Proprietary Trade Name: Midas Rex MR8 Depth Stop Attachment and Tools Common Name: MR8 Depth Stop System Classification Name: Powered cranial drills, burrs, trephines, and their accessories (21 CFR 882.4310) Classification: Class II Product Code: HBE Primary Predicate: Midas Rex Legend Depth Stop Attachment and Tools, K191597
### Product Description:
The Midas Rex MR8 Depth Stop attachment and tools are intended to create a cranial access hole of a known diameter and depth.
#### Indications for Use:
The Midas Rex MR8 Depth Stop attachment and tools are indicated for the incision, cutting. removing, and drilling of soft and hard tissue during cranial surgical procedures with the intent to create a hole through the cranium to allow surgeons access to desired surgical locations and/or to facilitate insertion, placement of other surgical devices during such procedures.
#### Comparison of the Technological Characteristics:
There have been no significant changes to the Midas Rex™ Legend Depth Stop Attachment and Tools since prior clearance in K191597.
The subject device, Midas Rex™ MR8™ Depth Stop Attachment and Tools, are to be used with the MR8TM High-Speed Drill Systems and the Stealth Autoguide™ System to create
{4}------------------------------------------------
cranial access holes of specific diameter and depth as previously cleared through K191597 for the Midas Rex™ Legend Depth Stop Attachment and Tools.
Changes from the predicate include:
- Compatibility with the predicate MR8 Drill Systems included within K181535 .
- . Extended depth setting to 20mm
- Minor modifications to tool head geometry ●
- Adding a new task in the user workflow .
{5}------------------------------------------------
| Feature | Subject Device (Midas Rex™<br>MR8™ Depth Stop System) | Predicate Device (Midas Rex™ Legend™<br>Depth Stop System, K191597) | Equivalence Discussion |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor | Medtronic Powered Surgical<br>Solutions | Medtronic Powered Surgical Solutions | Identical |
| Product Code | HBE | HBE, HBC, HBB | Identical |
| Drill System | • Pneumatic<br>• Electric | • Pneumatic<br>• Electric | Identical |
| Operating Principle | | | Identical |
| Intended Use | The Midas Rex MR8 Depth Stop<br>system is intended to create a<br>cranial access hole of a known<br>diameter and depth. | The Midas Rex Legend Depth Stop system is<br>intended to create a cranial access hole of a<br>known diameter and depth. | Identical |
| Indications for Use | The Medtronic Midas Rex MR8<br>Drill System is indicated for the<br>incision, cutting, removing, and<br>drilling of soft and hard tissue<br>during cranial surgical procedures<br>with the intent to create a hole<br>through the cranium to allow<br>surgeons access to desired<br>surgical locations and/or to<br>facilitate insertion, placement of<br>other surgical devices during such<br>procedures. | The Medtronic Midas Rex Legend Drill System<br>is indicated for the incision, cutting, removing,<br>and drilling of soft and hard tissue during cranial<br>surgical procedures with the intent to create a<br>hole through the cranium to allow surgeons<br>access to desired surgical locations and/or to<br>facilitate insertion, placement of other surgical<br>devices during such procedures. | Identical |
| System<br>components | • Electric Handpieces<br>• Pneumatic Handpieces<br>• Attachments<br>• Surgical Dissecting Tools | • Electric Handpieces<br>• Pneumatic Handpieces<br>• Attachments<br>• Surgical Dissecting Tools | Identical |
| Patient Contacting<br>Components | • Attachments<br>• Surgical Dissecting Tools | • Attachments<br>• Surgical Dissecting Tools | Identical |
| Attachment<br>Materials | Stainless Steel, Aluminum,<br>Ceramic, Phenolic, Epoxy,<br>Chrome Coated Brass, Polymeric | Stainless Steel, Aluminum, Ceramic, Phenolic,<br>Epoxy, Chrome Coated Brass, Polymeric | Identical |
| Attachment<br>Configuration | Straight, Variable | Straight, Variable | Identical |
| Feature | Subject Device (Midas Rex™<br>MR8™ Depth Stop System) | Predicate Device (Midas Rex™ Legend™<br>Depth Stop System, K191597) | Equivalence Discussion |
| Attachment Tube<br>Length | 12.8cm (tube retracted)<br>14.5cm (tube exposed) | 13.7cm (tube retracted)<br>14.8cm (tube exposed) | Equivalent<br>The attachment tube length is<br>inherent due to the overall design of<br>MR8 attachments and does not<br>impact the performance or safety of<br>the device. |
| Sterilization<br>Method – Reusable<br>Attachment | Steam | Steam | Identical |
| Surgical Dissecting<br>Tool Materials | Tool Steel | Tool Steel | Identical |
| Dissecting Tool<br>Tip Style | Twist Drill | Twist Drill | Equivalent<br>Minor tool tip modifications which<br>do not impact safety or performance. |
| Dissecting Tool<br>Overall Length | 17-22cm | 17-22cm | Identical |
| Dissecting Tool<br>Head Diameter | 2.1 – 7.5 mm | 2.1 – 7.5 mm | Identical |
| Packaging – Sterile<br>Dissecting Tools | Individually packaged in a clear<br>plastic capped tube placed within<br>a polypoly pouch | Individually packaged in a clear plastic capped<br>tube placed within a polypoly pouch | Equivalent<br>The packaging is identical except for<br>the tube cap. On the predicate device<br>it is yellow, and on the subject<br>device, the tube cap is blue. |
| Sterilization<br>Method – Single<br>Use Tools | Gamma with minimum radiation<br>dose of 25 kGy | Gamma with minimum radiation dose of 25 kGy | Identical |
| Shelf Life –<br>Dissecting Tools | 5 Years | 5 Years | Identical |
{6}------------------------------------------------
{7}------------------------------------------------
#### Discussion of the Performance Testing:
The intended use of the subject devices, Midas Rex™ MR8™ Depth Stop Attachments and Tools and predicate are identical, and their technological characteristics are similar. The subject devices are a line extension to the MR8 Drill System (K183515) in the same way that the predicate device was a line extension to the Legend Drill Systems (K163182 and K170312). The changes from the predicate, do not raise any new issues of safety and effectiveness.
Risk management was conducted in accordance with ISO 14971:2019 and did not identify any new risks when compared to the predicate. The following testing was conducted to demonstrate that the modifications to the subject devices are as safe and effective as the predicate.
- . Product Life Verification
- Summative Validation .
- . Biocompatibility Testing
Results of these tests demonstrate that the functionality, integrity, and safety and effectiveness of the subject devices are sufficient for their intended use, indications for use and support a determination of substantial equivalence.
#### Conclusions
The Midas Rex™ MR8™ Depth Stop Attachment and Tools have shown through comparison to be substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.