Medtronic Legend Pneumatic; MR7 Pneumatic; Triton Pneumatic Drill System incorporating various Pneumatic Handpieces; Attachments; Surgical Dissecting Tools; and System Accessories
Medtronic Legend Pneumatic; MR7 Pneumatic; Triton Pneumatic Drill System incorporating various Pneumatic Handpieces; Attachments; Surgical Dissecting Tools; and System Accessories
Applicant
Medtronic Powered Surgical Solutions
Product Code
HBB · Neurology
Decision Date
Mar 23, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4370
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Pneumatic Drill System is a pneumatically operated surgical instrument system. The pneumatic motor provides power to operate removable rotating surgical cutting tools and their accessories intended for use in neurosurgery, including craniotomy and spinal surgery; as well as Ear Nose and Throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial and sternotomy surgeries. Additionally, the Pneumatic Drill System is incision/cutting, removal, drilling, and sawing of soft and hard tissue and bone, and biomaterials during open and minimally invasive spine procedures, which may incorporate application of various surgical techniques during the following lumbar spinal procedures: Lumbar Microdiscectomy Lumbar Stenosis Decompression Posterior Lumbar Interbody Fusion (PLIF) Transforaminal Lumbar Interbody Fusion (TLIF) Anterior Lumbar Interbody Fusion (ALIF) Direct Lateral Interbody Fusion (DLIF)
Device Story
Pneumatic drill system comprises high-speed handpieces, foot/finger controllers, surgical dissecting tools, and attachments. System uses pneumatic energy to power rotating cutting tools for tissue/bone removal. Operated by surgeons in OR settings for neurosurgical, spinal, ENT, and orthopedic procedures. Device provides mechanical power for cutting/sawing; output is controlled by surgeon via foot/finger input. Benefits include precise tissue/bone removal during open or minimally invasive procedures. No software or AI/ML components.
Clinical Evidence
No clinical trials. Evidence consists of a clinical literature review and a cadaveric validation study. Cadaveric testing confirmed the system's acceptability and performance for the expanded spinal indications (LM, PLIF, TLIF, LSD, ALIF, DLIF).
Technological Characteristics
Pneumatically powered high-speed drill system. Components include handpieces, high-pressure/exhaust hoses, foot/finger controllers, and surgical dissecting tools. Energy source: pneumatic. Materials are consistent with previously cleared predicate devices. No software, electronics, or connectivity features.
Indications for Use
Indicated for patients undergoing neurosurgery (craniotomy, spinal surgery), ENT, orthopedic, and general surgical procedures (maxillofacial, craniofacial, sternotomy). Specifically indicated for incision, cutting, removal, drilling, and sawing of soft/hard tissue, bone, and biomaterials during open and minimally invasive spine procedures, including lumbar microdiscectomy, stenosis decompression, and various interbody fusions (PLIF, TLIF, ALIF, DLIF).
Regulatory Classification
Identification
A pneumatic cranial drill motor is a pneumatically operated power source used with removable rotating surgical cutting tools or drill bits on a patient's skull.
Predicate Devices
Oscillating and Reciprocating Saw Drill-Reamer (K870157)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of a human figure with outstretched arms, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 23, 2017
Medtronic Powered Surgical Solutions Deep Pal Regulatory Affairs Manager 4620 North Beach Street Fox Worth, TX 76177
Re: K163182
Trade/Device Name: Medtronic Legend Pneumatic: MR7 Pneumatic: Triton Pneumatic Drill System incorporating various Pneumatic Handpieces; Attachments; Surgical Dissecting Tools; and System Accessories Regulation Number: 21 CFR 882.4370 Regulation Name: Pneumatic Cranial Drill Motor Regulatory Class: Class II Product Code: HBB, KFK, ERL, EQJ, HSZ, GET, HBE, DWH Dated: December 22, 2016 Received: December 23, 2016
Dear Mr. Pal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Michael J. Hoffmann -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163182
### Device Name
Medtronic Legend Pneumatic; Triton Pneumatic Drill System incorporating various Pneumatic Handpieces; Attachments; Surgical Dissecting Tools; and System Accessories
### Indications for Use (Describe)
The Pneumatic Drill System is a pneumatically operated surgical instrument system. The pneumatic motor provides power to operate removable rotating surgical cutting tools and their accessories intended for use in neurosurgery, including craniotomy and spinal surgery; as well as Ear Nose and Throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial and sternotomy surgeries.
Additionally, the Pneumatic Drill System is incision/cutting, removal, drilling, and sawing of soft and hard tissue and bone, and biomaterials during open and minimally invasive spine procedures, which may incorporate application of various surgical techniques during the following lumbar spinal procedures:
- Lumbar Microdiscectomy
- Lumbar Stenosis Decompression
- Posterior Lumbar Interbody Fusion (PLIF)
- Transforaminal Lumbar Interbody Fusion (TLIF)
- Anterior Lumbar Interbody Fusion (ALIF)
- Direct Lateral Interbody Fusion (DLIF)
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">✓</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SECTION 9: 510(k) SUMMARY
This summary is submitted in accordance with the requirements of 21CFR807.92.
#### 9.1 DATE PREPARED
November 10, 2016
#### 9.2 NAME AND ADDRESS OF MANUFACTURER
## Table 4: Manufacturer information
| Establishment Name | Establishment Registration Number |
|---------------------------------------------------------------------------------------------|-----------------------------------|
| Medtronic Powered Surgical Solutions<br>4620 North Beach Street<br>Fort Worth, TX 76137 USA | 1625507 |
| Medtronic Xomed, Inc.<br>6743 Southpoint Drive North<br>Jacksonville, FL 33216 USA | 1045254 |
#### 9.3 CONTACT PERSON
Deep Pal Regulatory Affairs Manager Telephone: 817.788.6685 Facsimile: 817.788.6222 E-Mail: deep.pal@medtronic.com
#### 9.4 PROPRIETARY NAME
Medtronic Legend Pneumatic, MR7 Pneumatic, Triton Pneumatic Drill System incorporating various Pneumatic Handpieces, Attachments, Surgical Dissecting Tools and System Accessories
#### 9.5 COMMON/USUAL NAME
Surgical Drill Motors and Accessories
#### 9.6 DEVICE CLASSIFICATION NAME
### Table 5: FDA device classification information
| Description | FDA Code | Regulation Number |
|----------------------------------------------------------------------------|----------|-------------------|
| Saw, pneumatically powered | KFK | 21CFR878.4820 |
| Motor, drill, pneumatic | HBB | 21CFR882.4370 |
| Drill, surgical, ent (electric or pneumatic) including handpiece | ERL | 21CFR874.4250 |
| Bur, ear, nose and throat | EQJ | 21CFR874.4140 |
| Instrument, surgical, orthopedic, pneumatic powered & accessory/attachment | HSZ | Pre-Amendment |
| Pneumatic motor for general surgery | GET | 21CFR878.4820 |
| Drills, burrs, trephines and accessories (simple, powered) | HBE | 21CFR882.4310 |
| Blades, saw, surgical cardiovascular | DWH | 21CFR878.4820 |
{4}------------------------------------------------
#### 9.7 PREDICATE DEVICE IDENTIFICATION
Table 6: Subject Devices that are in the scope of this Submission
| SUBJECT SYSTEM DESCRIPTION | PREDICATE | |
|-------------------------------------------|------------------------------------------------|------------------|
| | SYSTEM DESCRIPTION | 510(k) NUMBER |
| Triton Pneumatic Drill System | Oscillating and Reciprocating Saw Drill-Reamer | K870157 |
| - Pneumatic Handpiece | - Pneumatic Handpiece | K870157 |
| - Pneumatic Foot Control Unit | - Pneumatic Control Unit | K870157 |
| - Surgical Dissecting Tools | - Surgical Dissecting Tools | K870157 |
| - Attachments | - Attachments | K870157 |
| - System Accessories | - System Accessories | K870157 |
| Legend Pneumatic High Speed System | Legend Pneumatic High Speed System | K020069 |
| - Pneumatic Handpiece | - Pneumatic Handpiece | K020069 |
| - Pneumatic Foot Control Unit | - Pneumatic Control Unit | K020069 |
| - Surgical Dissecting Tools | - Surgical Dissecting Tools | K020069, K072315 |
| - Attachments | - Attachments | K020069 |
| - System Accessories | - System Accessories | K020069 |
| Midas Rex MR7 Pneumatic High Speed System | Midas Rex MR7 Pneumatic High Speed System | K090112 |
| - Pneumatic Handpiece | - Pneumatic Handpiece | K090112 |
| - Pneumatic Foot Control Unit | - Pneumatic Control Unit | K090112 |
| - Surgical Dissecting Tools | - Surgical Dissecting Tools | K020069, K072315 |
| - Attachments | - Attachments | K020069 |
| - System Accessories | - System Accessories | K090112 |
#### 9.8 DEVICE DESCRIPTION
The Pneumatic Drill System is a pneumatically powered high-speed drill system consisting of a choice of various Pneumatic Handpieces (comprising of a High Pressure and Exhaust Hose, and Handpiece) equipped with a foot or finger controller, Pneumatic Foot Control Unit, Surgical Dissecting Tools, System Accessories, and where applicable, Attachments to support various Surgical Dissecting Tools, and System Accessories.
The device design, function, the intended use and the general operating principles, and conditions of use of the overall Medtronic Pneumatic Drill Systems remain similar to those cleared under K870157, K020069, K072315, and K090112.
#### INDICATIONS FOR USE 9.9
The Pneumatic Drill System is a pneumatically operated surgical instrument system. The pneumaic motor provides power to operate removable rotating surgical cutting tools and their accessories intended for use in neurosurgery, including craniotomy and spinal surgery; as well as Ear Nose and Throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial and sternotomy surgeries.
Additionally, the Pneumatic Drill System is incision / cutting, removal, drilling, and sawing of soft and hard tissue, bone, and biomaterials during open and minimally invasive spine procedures, which may incorporate application of various surgical techniques during the following lumbar spinal procedures:
- Lumbar Microdiscectomy o
- o Lumbar Stenosis Decompression
- Posterior Lumbar Interbody Fusion (PLIF) o
- O Transforaminal Lumbar Interbody Fusion (TLIF)
- Anterior Lumbar Interbody Fusion (ALIF) O
- O Direct Lateral Interbody Fusion (DLIF)
{5}------------------------------------------------
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE 9.10
The Pneumatically powered drill system, designed to remove soft and hard tissue, and bone, and biomaterials is the technological principle for both the subject and predicate systems.
The subject and predicate systems are based on the following same technological elements:
- A Pneumatic Drill System Application: Designed to remove soft and hard tissue, bone, and biomaterials during various surgical applications.
- A Operating Principle: The pneumatic energy is supplied to the Handpiece to provide power to operate interchangeable Surgical Dissecting Tools supported by Attachments and intended for use in various surgical procedures to remove soft and hard tissue, bone, and biomaterials.
In terms of the materials used in manufacturing of the patient contacting components of the subject Pneumatic Drill System, the subject Attachments and Surgical Dissecting Tools are similar to the predicate Medtronic Attachments and Surgical Dissecting Tools.
#### 9.11 DISCUSSION OF THE PERFORMANCE TESTING
Testing was completed to ensure the functionality of the Medtronic Pneumatic Drill System for the expanded indications. The following table summarizes the testing completed:
| TEST | DESCRIPTION | RESULTS |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clinical<br>Literature<br>Review | A comprehensive clinical literature<br>search was conducted and reviewed<br>regarding the safe and effective use<br>of the Drill System for the expanded<br>indications for use, which is the<br>subject of this submission. | The cadaver validation study, and<br>review of supporting literature,<br>supports the use of the Medtronic<br>Drill System in LM, PLIF, TLIF, LSD,<br>ALIF, and DLIF spinal surgical<br>procedures, as well as any<br>variations of these procedures. |
| Cadaveric<br>Testing | Users evaluated the acceptability of<br>the subject drill system to its<br>intended use on a variety of<br>procedures using cadavers. | The Medtronic Pneumatic Drill<br>System is acceptable for its<br>intended use in various surgical<br>procedures. |
#### 9.12 CONCLUSIONS
The subject Medtronic Penumatic Drill Systems when compared to the predicate Drill Systems have same intended use, where the subject and the predicate drill systems are intended for use in various surgical procedures to remove soft and hard tissue, and biomaterials. Results of cadaveric testing have demonstrated that the addition of the proposed indications do not present any new issues of safety or effectiveness, and the systems perform as intended during surgical use, similar to the use during various other surgical procedures on currently cleared indications.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.