OnyxCeph3™ is a software only device and is intended to be used for the medical purpose of manaqing and evaluating two-dimensional and threedimensional images in the framework of digital orthodontics by qualified staff only who have received dedicated training for use of the software. Diagnostic and therapeutic decisions cannot be motivated exclusively or even mainly on evaluation results provided by the software.
Device Story
OnyxCeph3™ is a Windows-based software for orthodontic and maxillofacial surgical planning. It ingests 2D images and 3D CT/DVT volume data. The software enables clinicians to perform metric and angular evaluations, cephalometric analysis, treatment simulation, and virtual model planning. Used in dental facilities by trained orthodontic staff, the software provides visual outputs for clinical review. It does not provide automated diagnostic or therapeutic decisions; clinicians use the output to support treatment planning. The device operates locally without requiring internet access, utilizing a key-based activation system. It benefits patients by facilitating precise, image-based orthodontic treatment planning and progress monitoring.
Clinical Evidence
No clinical data provided. Substantial equivalence is demonstrated through non-clinical software validation conducted in accordance with IEC 62304 standards.
Technological Characteristics
Windows-based software; local data storage; key-based activation; supports 2D/3D image import, cephalometric evaluation, treatment simulation, and volume segmentation. Complies with IEC 62304 for software life cycle processes.
Indications for Use
Indicated for qualified orthodontic staff to manage and evaluate 2D and 3D images for digital orthodontic workflows, including metric and angular evaluation of image data.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
January 4, 2023
Image Instruments GmbH % Greg Holland Sr. Partner Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, California 92606
Re: K213263
Trade/Device Name: OnyxCeph3TM Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ, PNN Dated: November 30, 2022 Received: December 5, 2022
Dear Greg Holland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bobak Shirmohammadi -S
For Michael E. Adjodha, M. ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213263
Device Name OnyxCeph3TM
#### Indications for Use (Describe)
OnyxCeph*™ software is intended to be used for the medical purpose of managing and evaluating two-dimensional and three-dimensional images in the framework of digital orthodontics by qualified staff only. The use of OnyxCeph7™ requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
The software can be used to digitally perform certain image-based orthodontic workflows, such as metric and angular evaluation of image data.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> <b> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 5. 510(k) Summary - K213263
| 510(k) Owner | Image Instruments GmbH<br>Niederwaldstr. 3<br>09123 Chemnitz<br>Germany<br>Phone: +49 371 9093140<br>Facsimile: +49 371 9093149 | | OnyxCeph3 TM<br>K213263<br>Subject Device | VISTADENTTM AT<br>COMPLETE K053010<br>Predicate Device | | OnyxCeph3™<br>K213263<br>Subject Device | VISTADENT™ AT<br>COMPLETE K053010<br>Predicate Device |
|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|-------------------------------------------|--------------------------------------------------------|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person | Greg Holland<br>Regulatory Specialists, Inc.<br>3722 Ave. Sausalito<br>Irvine, CA 92606<br>Phone: 949.262.0411<br>Fax: 949.552.2821<br>Email: greg@regulatoryspecialists.com | Intended Use | Orthodontic Software | Orthodontic Software | Indications for<br>Use | OnyxCeph3™ software<br>is intended to be used<br>for the medical<br>purpose of managing<br>and evaluating two-<br>dimensional and three-<br>dimensional images in<br>the framework of<br>digital orthodontics by<br>qualified staff only.<br>The use of<br>OnyxCeph3™ requires<br>the user to have the<br>necessary training and<br>domain knowledge in<br>the practice of<br>orthodontics, as well<br>as to have received a<br>dedicated training in<br>the use of the<br>software.<br>The software can be<br>used to digitally perform<br>certain image- based<br>orthodontic workflows,<br>such as metric and<br>angular evaluation of<br>image data. | VISTADENT™ AT<br>COMPLETE software is<br>a digital database for<br>storing, retrieving and<br>printing images that also<br>has the ability to perform<br>image manipulation and<br>cephalometric analysis. |
| Submission Date | January 3, 2023 | Product Code | LLZ | LLZ | | | |
| Common Name<br>Trade Name<br>Classification Name<br>Regulation<br>Class<br>Panel<br>Primary Product Code<br>Secondary Product Code | Orthodontic Software<br>OnyxCeph3TM<br>Radiological Image Processing System<br>892.2050<br>Class II<br>Dental<br>LLZ<br>PNN | 2 Dimensional<br>Image Import | Yes | Yes | | | |
| Primary Predicate Device | K053010<br>VISTADENTTM AT<br>COMPLETE<br>From: Dentsply INTL.,<br>Inc. | 2 Dimensional<br>Image Adjust | Yes | Yes | | | |
| Reference Device | K171634<br>Ortho SystemTM<br>From: 3Shape A/S<br>K192475<br>NemoFAB<br>From: Software<br>Nemotec S.L. | | | | | | |
## Description
OnyxCeph3™ dental aligner software device is designed for the simulation and planning of orthodontic and maxillofacial surgical combination treatments, based on virtual models and CT / DVT volume data. OnyxCeph3™ is a software device that includes both 3D and 2D versions both suitable for dental facilities to use for orthodontic treatments.
{4}------------------------------------------------
#### Intended Use
OnyxCeph3™ is a software only device and is intended to be used for the medical purpose of manaqing and evaluating two-dimensional and threedimensional images in the framework of digital orthodontics by qualified staff only who have received dedicated training for use of the software. Diagnostic and therapeutic decisions cannot be motivated exclusively or even mainly on evaluation results provided by the software.
### Indications for Use
OnyxCeph3™ software is intended to be used for the medical purpose of manaqing and evaluating two-dimensional and three-dimensional images in the framework of digital orthodontics by qualified staff only. The use of OnyxCeph3™ requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
The software can be used to digitally perform certain image-based orthodontic workflows, such as metric and angular evaluation of image data.
## Technological Characteristics
The predicates and the Image Instruments GmbH OnyxCeph3™ were compared in the following areas and found to have similar technological characteristics and to be equivalent. Both predicates are shown as a comparison in separate tables.
{5}------------------------------------------------
{6}------------------------------------------------
| | OnyxCeph3™<br>K213263<br>Subject Device | VISTADENT™ AT<br>COMPLETE K053010<br>Predicate Device |
|----------------------------------------------|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| 2 Dimensional<br>Cephalometric<br>Evaluation | Yes | Yes |
| 2 Dimensional<br>Image<br>Combination | Yes | Yes |
| 2 Dimensional<br>Image<br>Comparison | Yes | Yes |
| 2 Dimensional<br>Mirroring | Yes | Yes |
| CO/CR-<br>Conversion | Yes | Yes |
| Ricketts VTO | Yes | Yes |
| 2 Dimensional<br>Treatment<br>Simulation | Yes | Yes |
| Windows Based | Yes | Yes |
| Activation by<br>Key | Yes | Yes |
| Client<br>Requirements | Hardware Type, CPU,<br>RAM, Disk, Network,<br>Graphic Board, Desktop<br>Resolution | Hardware Type CPU, RAM,<br>Disk, Network, Graphic<br>Board,<br>Desktop Resolution |
| Client<br>Requirements | Not Required | Internet Access,<br>Internet Speed |
| Server<br>Requirements | Hardware Type, CPU,<br>RAM, Disk, Network,<br>Port Sharing | Hardware Type, CPU,<br>RAM, Disk, Network,<br>Port Sharing |
{7}------------------------------------------------
| | OnyxCeph3™<br>K213263<br>Subject Device | Ortho System<br>3Shape A/S<br>K171634<br>Reference<br>Device |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use<br>Indications | Orthodontic Software<br>OnyxCeph3™ software is<br>intended to be used for the<br>medical purpose of<br>managing and evaluating<br>two-dimensional and<br>three-dimensional images<br>in the framework of digital<br>orthodontics by qualified<br>staff only. The use of<br>OnyxCeph3™ requires the<br>user to have the necessary<br>training and domain<br>knowledge in the practice<br>of orthodontics, as well as<br>to have received a<br>dedicated training in the<br>use of the software.<br>The software can be used<br>to digitally perform certain<br>image- based orthodontic<br>workflows, such as metric<br>and angular evaluation of<br>image data. | Orthodontic Software<br>The 3Shape Ortho<br>System™ is intended for<br>use as a medical front-<br>end device providing<br>tools for management of<br>orthodontic models,<br>systematic inspection,<br>detailed analysis,<br>treatment simulation and<br>virtual appliance design<br>options (Custom Metal<br>Bands, Export of<br>Models, Indirect<br>Bonding Transfer<br>Media) based on 3D<br>models of the patient's<br>dentition before the start<br>of an orthodontic<br>treatment. It can also be<br>applied during the<br>treatment to inspect and<br>analyze the progress of<br>the treatment. It can be<br>used at the end of the<br>treatment to evaluate if<br>the outcome is<br>consistent with the<br>planned/desired<br>treatment objectives.<br>The use of the Ortho<br>System™ requires the<br>user to have the<br>necessary training and<br>domain knowledge in<br>the practice of<br>orthodontics, as well as<br>to have received a<br>dedicated training in the<br>use of the software. |
| Product Code | LLZ | PNN |
| | OnyxCeph3TM<br>K213263<br>Subject Device | Ortho System<br>3Shape A/S<br>K171634<br>Reference<br>Device |
| 2 Dimensional<br>Evaluation | Yes | Yes |
| 2 Dimensional<br>Cephalometric<br>Evaluation | Yes | No |
| 3 Dimensional<br>Evaluation | Yes | Yes |
| Supported<br>anatomical areas | Maxilla and Mandible | Maxilla and Mandible |
| Determine arch<br>shape | Yes | Yes |
| Determine wire<br>length | Yes | Yes |
| Determine tooth<br>width | Yes | Yes |
| Determine Bolton | Yes | Yes |
| Determine space<br>analysis | Yes | Yes |
| Determine overbite | Yes | Yes |
| Determine<br>occlusion | Yes | Yes |
| Extract/Segment<br>Volume Scans | Yes | Yes |
| Combine Bony and<br>Dental Scan Data | Yes | Yes |
| Windows Based | Yes | Yes |
| Data Storage | Local | Local / Cloud |
| Activation by<br>Dongle | No | Yes |
| | OnyxCeph3™<br>K213263<br>Subject Device | Ortho System<br>3Shape A/S<br>K171634<br>Reference<br>Device |
| Activation by Key | Yes | No |
| Client<br>Requirements | Hardware Type, CPU,<br>RAM, Disk, Network,<br>Graphic Board, Desktop<br>Resolution | Hardware Type, CPU,<br>RAM, Disk, Network,<br>Graphic Board,<br>Desktop Resolution |
| Client<br>Requirements | Not required | Internet Access,<br>Internet Speed |
| Server<br>Requirements | Hardware Type, CPU,<br>RAM, Disk, Network,<br>Port Sharing | Hardware Type, CPU,<br>RAM, Disk, Network,<br>Port Sharing |
{8}------------------------------------------------
{9}------------------------------------------------
Although the Indications for Use Statement for the subject device is different from the predicate device, K053010, the intended use of the software is the same, to allow for two-dimensional and three-dimensional evaluation of images. The predicate device includes Internet Access which is not required for the subject device. The subject device includes 3-dimensional evaluation capabilities; whereas the predicate device does not. However, the reference devices support this function.
The reference device, K171634, was included to support the 3-dimensional evaluation capabilities of the subject device. This reference device does not include 2-Dimensional Cephalometric function, and activation by Key. However, the 2-Dimensional Cephalometric function is supported by the predicate device. The reference device offers activation by Dongle. In addition, the reference device supports the use of Data Storage via the Cloud and requires Internet Access whereas the subject device does not.
Additional reference device, K192475 was included to support the devices intended use for supporting the treatment planning process of orthognathic procedures.
In summary, the similarities between the subject device, OnyxCeph3™, and the predicate devices are functionally very similar. The differences identified do not raise different questions of safety and effectiveness as discussed in the table above.
Nonclinical Testing
Software Validation conducted per FDA Guidance for Software Contained in Medical Devices IEC 62304 - Software Life Cycle Processes
#### Clinical Testing
Clinical testing has not been included and is not required to demonstrate substantial equivalence.
#### Conclusion
Based on the similarities between the intended use, indications, principle of operation, features and technical data, the subject device, OnyxCeph3™, is found to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.