The 3Shape Ortho System™ is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual applians (Custom Metal Bands, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
Device Story
Software system for orthodontic case management; inputs include 3D dental scans (STL), DICOM images, and 2D images (JPG, BMP, PNG). Operates by visualizing 3D models, measuring dentition, simulating tooth movements, and designing virtual orthodontic appliances (e.g., indirect bonding trays, metal bands). Used in dental clinics by trained orthodontic professionals. Output consists of digital files for fabrication of appliances by third parties. Assists clinicians in diagnosis, treatment planning, and progress monitoring. Benefits include improved visualization of treatment objectives and customized appliance design.
Clinical Evidence
No clinical data. Bench testing only; software, hardware, and integration verification and validation performed per FDA guidance.
Technological Characteristics
Software-only device programmed in Delphi. Runs on Windows 7/8/10 (64-bit) with minimum 8GB RAM, Intel Core i5, and 250GB HDD. Supports DICOM, STL, JPG, BMP, PNG formats. No patient contact. Features include 2D/3D measurement, arch shape analysis, occlusion mapping, and virtual appliance design.
Indications for Use
Indicated for orthodontic patients requiring management of dental models, inspection, analysis, treatment simulation, and virtual appliance design (custom metal bands, indirect bonding transfer media) prior to, during, and at the end of orthodontic treatment.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Janmuary 17, 2018
Image /page/0/Picture/1 description: The image shows the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are simple and professional in appearance.
3Shape A/S Antje Marquardsen Director Regulatory Affairs Holmens Kanal 7 Copenhagen, 1060 Dk
Re: K171634
Trade/Device Name: Ortho System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: PNN Dated: November 21, 2017 Received: November 24, 2017
Dear Antje Marquardsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
# Page 2 - Antje Marquardsen
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mary S. Runner -S
ForTina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171634
Device Name Ortho System™
#### Indications for Use (Describe)
The 3Shape Ortho System™ is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual applians (Custom Metal Bands, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
The use of the Ortho System™ requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY – Special 510(K) – K171634
# Submitter Information
| Company Name: | 3Shape A/S |
|--------------------------------|---------------------------------------------------|
| Company Address: | Holmens Kanal 7<br>DK-1060 Copenhagen K |
| Company Phone:<br>Company Fax: | +45 7027 2620<br>+45 7027 2621 |
| Contact Person: | Antje Marquardsen<br>Director, Regulatory Affairs |
| Date Summary Prepared: | January 16, 2018 |
| Device Identification | |
| 510(k) number: | K171634 |
| Trade/proprietary Name: | Ortho System™ |
| Regulation Number: | 872.5470 |
| Regulation Name: | Orthodontic Plastic Bracket |
| Classification: | Class II |
# Primary Predicate Device
The Ortho System™ intended for dental retainers has the same intended use, scientific concept, performance specification and technical characteristics as the primary predicate device 3Shape Ortho System™ (K152086) from 3Shape A/S.
Based on the information and supporting documentation provided, the Ortho System™ and the primary predicate device (K152086) have same intended use.
Both software devices are used by Dental Professionals in orthodontic treatment planning (before, during, after treatment) covering management of patients and models, inspection, 2D and 3D measurement and orthodontic analysis of models, 2D & 3D treatment simulation, as well as virtual appliance preparation, handling and export, and they are both providing device output.
Additionally, the indirect bonding functionality of both systems is intended for use with commercially available brackets and wires.
The only modifications compared to the primary predicate device, besides look and feel of the user interface, are:
- 1. A workflow improvement to view present and ideal position of teeth at
{4}------------------------------------------------
the same time (dual view) with the introduction of arch wire in Indirect Bonding.
sshar
- 2. The ScanIt Manager is used for all scanners instead of ScanIt Orthodontics and ScanIt Ortho Impression that both are retracted.
Therefore, the Ortho System™ and the predicate (K152086) are found to be similar in their intended use, supported anatomic areas and the majority of the available features and functionalities.
# Reference Predicate devices
The Ortho System have two reference predicate devices K161884 and K163677, which uses the functions and tools of Ortho System™ to produce different orthodontic dental treatments than the primary predicate device. Both of these devices have the same fundamental scientific technology as the primary predicate device, as they are consolidated in the single medical software.
Based on the on the information and supporting documentation provided, the reference predicate devices do not alter the performance specification nor technical characteristics of the primary predicate device.
# Indications for Use
Ortho System™ is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Custom Metal Bands, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
The use of the Ortho System™ requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
# Device Description
3Shape's Ortho System™ is a software system used for the management of 3D scanned orthodontic models of the patients, orthodontic diagnosis by measuring, analysing, inspecting and visualize 3D scanned orthodontic models, virtual planning of orthodontic treatments by simulating tooth movements, virtual placement of orthodontic brackets on the 3D models and design of orthodontic appliances based on 3D scanned orthodontic models, including transfer methods for indirect bonding of brackets. Output includes only Export Model (also called dental casts), Custom Metal Bands (also called metal bands), and Indirect Bonding Transfer Trays (also called orthodontic bracket placement trays). All devices are to be fabricated from FDA cleared materials by third party under the responsibility of the user.
The device has no patient contact.
{5}------------------------------------------------
## Scientific Concept
The underlying scientific concept of the Ortho System™ is to apply digital imaging tools for in orthodontic case archiving, diagnosis, treatment planning and CAD design of customized appliances.
Virtual positioning of brackets is possible with the use of encrypted libraries of the bracket geometry provided by the manufacturers and available through a dedicated download center in the software.
The system supports the following types of digital data: DICOM, STL, JPG, BMP, PNG.
### Summary of the technological characteristics
Ortho System™ is a software only device programmed in Delphi and has the following PC/laptop hardware requirements:
| Item | Minimum Requirements<br>Ortho System<br>(K171634) | Minimum Requirements<br>3Shape Ortho System<br>(K152086) |
|-------------------------|---------------------------------------------------|----------------------------------------------------------|
| OS: | Windows 7, 8 or 10 64-bit | Windows 7, 8 or 10, 64-bit |
| RAM: | 8 GB | 8 GB |
| Monitor<br>Resolution: | 1280x800 or similar | 1280x800 or similar |
| Video Card<br>Memory: | 1 GB | 1 GB |
| Available<br>HDD Space: | 250 GB | 250 GB |
| CPU: | Intel Core i5 or equivalent | Intel Core i5 or equivalent |
| Network: | Network Internet connection | Network Internet connection |
| Mouse: | With wheel button | With wheel button |
The Ortho System™ Software has the same intended uses and technical characteristics as the 3Shape Ortho System™ (K152086) from 3Shape A/S:
| Feature name | Ortho<br>System™<br>(K171634) | 3Shape<br>Ortho<br>System™<br>(K152086) |
|------------------------------------------------------|-------------------------------|-----------------------------------------|
| Supported anatomic areas | Maxilla | Maxilla |
| | Mandible | Mandible |
| Intended use | | |
| Managing patient and case base data | Yes | Yes |
| Collection of study material | Yes | Yes |
| Alignment of study material | Yes | Yes |
| Measuring study material | Yes | Yes |
| Analyzing study material | Yes | Yes |
| Treatment simulation | Yes | Yes |
| Virtual appliance design | Yes | Yes |
| Supported PC formats | Windows | Windows |
| Managing patient and case base data | | |
| Creating, editing, deleting and copying patient data | Yes | Yes |
| Creating, editing, deleting and copying case data | Yes | Yes |
| Feature name | Ortho<br>System™<br>(K171634) | 3Shape<br>Ortho<br>System™<br>(K152086) |
| Collection of study material | | |
| Surface scan for intra-oral scanner | Yes | Yes |
| Surface scan from STL file | Yes | Yes |
| CT image data | DICOM | DICOM |
| 2D overlay | PNG, JPG,<br>BMP | PNG, JPG, BMP |
| Alignment of study material | | |
| Aligning surface scan and CT image | Yes | Yes |
| Aligning cephalometric images | Yes | Yes |
| Alignment of 2D overlays (e.g. ideal arch) | Yes | Yes |
| Ability to check/adjust DICOM visibility | Yes | Yes |
| DICOM scan segmentation | No | No |
| Measuring study material | | |
| 2D measurement toolbox | Yes | Yes |
| 3D measurement toolbox | Yes | Yes |
| Analyzing study material | | |
| Arch shape | Yes | Yes |
| Wire length | Yes | Yes |
| Tooth width | Yes | Yes |
| Bolton | Yes | Yes |
| Space analysis | Yes | Yes |
| Overjet/overbite | Yes | Yes |
| Occlusion map | Yes | Yes |
| Treatment simulation | | |
| 2D & 3D simulation | Yes | Yes |
{6}------------------------------------------------
# ORTHO SYSTEM™ SOFTWARE - SPECIAL 510(K) SUBMISSION
зshap
{7}------------------------------------------------
## ORTHO SYSTEM™ SOFTWARE - SPECIAL 510(K) SUBMISSION
| Feature name | Ortho<br>System™<br>(K171634) | 3Shape<br>Ortho<br>System™<br>(K152086) |
|-------------------------------------------|-------------------------------|-----------------------------------------|
| Virtual appliance design | | |
| Orthodontic appliance search | Yes | Yes |
| Orthodontic appliance virtual preparation | Yes | Yes |
| Orthodontic appliance design | Yes | Yes |
| Orthodontic appliance export | Yes | Yes |
# Nonclinical Testing
Software, hardware, and integration verification and validation testing was performed in accordance with the FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Issued on May 11, 2005).
The validation suite includes validation of implemented mitigations related to device hazards identified in the risk management procedures.
All test results have been reviewed and approved, showing the Ortho System™ to be substantially equivalent to the primary predicate device.
# Clinical Testing
Clinical testing is not a requirement and has not been performed.
# Conclusion
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for 3Shape. The text "3shape" is written in a dark blue sans-serif font. To the right of the text is a red triangle made up of three smaller triangles.
Based on a comparison of intended use, indications, principle of operations, features and technical data, and the test results, the Ortho System™ is found to be as safe and as effective as the primary predicate device.
Intended use and performance is found to be substantially equivalent to the primary predicate Device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.