The Longfian Oxygen Concentrator. Model JAY-5AW is intended to be used by patients with respiratory disorders who require supplemental oxygen. A high concentration of supplemental oxygen is supplied and a nasal cannula is used to channel oxygen from the concentrator to the patient. The Longfian Oxygen Concentrator, Model JAY-5AW, can be used at home or healthcare environments. The Longfian Oxygen Concentrator, Model JAY-5AW, does not nor is it intended to sustain or support life. The device is intended for use in adults.
Device Story
Device uses pressure swing adsorption (PSA) to separate oxygen from ambient air; oilless compressor draws in air; air filtered and cooled; control valve directs air into absorption tower containing synthetic zeolite; oxygen separated and collected in tank; output delivered via flow meter and humidifier through nasal cannula. Used in home or healthcare environments; operated by patient or caregiver. Provides up to 5 LPM oxygen flow at 90-96% concentration. Output allows patients to receive supplemental oxygen for respiratory disorders. Device includes alarms for low/high pressure and power failure; LCD displays timing information.
Clinical Evidence
Bench testing only. Evaluated per ISO 80601-2-69 for safety and essential performance. Biocompatibility testing conducted per ISO 18562-1/2/3 and ISO 10993-5/10/17. Electrical safety and EMC testing per IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11. Oxygen concentration and flow performance verified after 8 hours of continuous operation.
Technological Characteristics
PSA oxygen concentrator; synthetic zeolite sieve bed; oilless compressor; PVC nasal cannula; 0-5 LPM flow; 90-96% oxygen concentration; Class II, Type B electrical classification; AC powered; dimensions 280x3485x510mm; weight 16kg; microprocessor-based control; standards: IEC 60601-1, IEC 60601-1-2, ISO 80601-2-69.
Indications for Use
Indicated for adult patients with respiratory disorders requiring supplemental oxygen. Not intended to sustain or support life.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 8, 2022
Longfian Scitech Co., Ltd. % Jun Peng Principal Consultant P&L Scientific, Inc. 1430 S. Dixie Hwy Suite 105 Coral Gables, Florida 33146
Re: K213210
Trade/Device Name: Oxygen Concentrator, Model JAY-5AW Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: Class II Product Code: CAW Dated: November 3, 2022 Received: November 7, 2022
Dear Jun Peng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James J. Lee -S
James J. Lee, Ph.D. Division Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213210
Device Name Oxygen Concentrator, Model JAY-5AW
#### Indications for Use (Describe)
The Longfian Oxygen Concentrator. Model JAY-5AW is intended to be used by patients with respiratory disorders who require supplemental oxygen. A high concentration of supplemental oxygen is supplied and a nasal cannula is used to channel oxygen from the concentrator to the patient. The Longfian Oxygen Concentrator, Model JAY-5AW, can be used at home or healthcare environments. The Longfian Oxygen Concentrator, Model JAY-5AW, does not nor is it intended to sustain or support life. The device is intended for use in adults.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
510(k) summary of Safety and Effectiveness as required by the Safe Medical Devices Act of 1990 and codified in 21 CFR 807.92 upon which Substantial Equivalence is based:
# K213210
# 1. Submitter Information:
●
Sponsor/510(K) Owner: LONGFIAN SCITECH CO., LTD. 2F&3F, East Section, Building 12, Power Valley Pioneer Park, No.369 Huiyang Street, Baoding 071051, China
Contact: Mr. Shi Lei Technical Manager Phone: +86-312-5909505
Date: October 28, 2022
#### Contact Name: ●
Mr. Jun Peng P&L SCIENTIFIC, INC. 1430 S. Dixie Hwy STE 105, Coral Gables, FL 33146 Phone: (305) 345 6018 Fax: (305) 397 0289 Email: jpeng@plscientificinc.com
# 2. Device Name
Oxygen Concentrator, Model JAY-5AW Trade Name: Common Name: Oxygen concentrator; Classification Name: Oxygen concentrator, portable;
# 3. Classification :
| Classification: | 21 CFR 868.5440; Class II |
|-----------------|---------------------------|
| Product Code: | CAW; |
#### 4. Predicate Devices:
| Oxygen<br>(Model: JAY-5) | Concentrator | K131968,<br>September 5, 2014 | Cleared | LONGFIAN<br>CO., LTD. | SCITECH |
|--------------------------|--------------|-------------------------------|---------|-----------------------|---------|
|--------------------------|--------------|-------------------------------|---------|-----------------------|---------|
# 5. Description of Device
The Longfian Oxygen Concentrator, Model JAY-5AW adopts pressure swing adsorption principle, which can separate oxygen, nitrogen and other gas from the air, at constant temperature. As soon as power is connected, the air is taken in and compressed by oilless
{4}------------------------------------------------
air compressor through filtering, then the compressed air goes through the cooler and it is cooled. After that, the air is taken into absorption tower by control valve and the oxygen can be separated in the absorption tower. At the same time, the high purity oxygen is collected into the oxygen tank, and it goes through the flow meter and humidifier, finally the oxygen that meets medical standards can be supplied. Oxygen is generated by pure physical method.
The device is capable of providing oxygen flow up to 5 LPM. The concentration of supplied oxygen is from 90%-96%. The weight of Jay-5AW is around 16kg (~351bs). There is no influence on indoor oxygen percent during the generator operating.
The biocompatibility contact classification of the device is as permanent contact externally communicating with dry and humified gas pathway.
The life-time/use life of the oxygen concentrator is set at 5 years or 20.000 hours after the product is sold, whichever comes first.
The reusable components:
- · Intake air filter (two pieces, part number:GL-01)
- · Secondary filter (one piece, part number:GL-02)
- nasal oxygen tube and
- humidifier.
JAY-5AW medical oxygen concentrator contains the following items:
- · One (1) JAY-5AW Oxygen Concentrator Unit
- One (1) AC Power Supply
- · One (1) Nasal Oxygen Tube
- · One (1) Firesafe Cannula Valve
- One (1) Humidifier
- · One (1) Operation Manual
# 6. Intended Use
The Longfian Oxygen Concentrator, Model JAY-5AW, is intended to be used by patients with respiratory disorders who require supplemental oxygen. A high concentration of supplemental oxygen is supplied and a nasal cannula is used to channel oxygen from the concentrator to the patient. The Longfian Oxygen Concentrator, Model JAY-5AW, can be used at home or healthcare environments. The Longfian Oxygen Concentrator, Model JAY-5AW, does not nor is it intended to sustain or support life. The device is intended for use in adults.
#### Summary of Comparison in to Technological Characteristics Predicate Device 7.
| Name | Predicate device | Subject device |
|----------|------------------|----------------|
| K number | K131968 | K213210 |
{5}------------------------------------------------
| Device Trade Name | JAY-5 Medical Molecular Sieve Oxygen Concentrator | The Longfian Oxygen Concentrator, Model JAY-5AW |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Oxygen concentrator | Oxygen concentrator |
| Classification Name | Oxygen concentrator, Portable | Oxygen concentrator, Portable |
| Indications for use | The Longfian Oxygen Concentrator, Model Jay-5 is intended to be used by patients with respiratory disorders who require supplemental oxygen. A high concentration of supplemental oxygen is supplied and a nasal cannula is used to channel oxygen from the concentrator to the patient. The Longfian Oxygen Concentrator, Model Jay-5 can be used in a home, institution, vehicle, and various mobile environments. The Longfian Oxygen Concentrator, Model Jay-5 does not nor is it intended to sustain or support life. | The Longfian Oxygen Concentrator, Model JAY-5AW, is intended to be used by patients with respiratory disorders who require supplemental oxygen. A high concentration of supplemental oxygen is supplied and a nasal cannula is used to channel oxygen from the concentrator to the patient. The Longfian Oxygen Concentrator, Model JAY-5AW, can be used at home or an institution. The Longfian Oxygen Concentrator, Model JAY-5AW, does not nor is it intended to sustain or support life. The device is intended for use in adults. |
| Classification | 21 CFR 868.5440 | 21 CFR 868.5440 |
| Feature | | |
| Power supply | AC 220±22V, 50±1Hz | AC 230±23V, 50±1Hz |
| Input power | 500W | 460W |
| Oxygen concentration | 93%±3% | 93%±3% |
| Oxygen flow | 0~5L/min | 0~5L/min |
| Outlet pressure | 40~70kPa | 40~70kPa |
| Noise | ≤53dB(A) | ≤53dB(A) |
| Fuse | T6.3AL/250V | F5AL/250V |
| Operating system | Pressure Swing Adsorption | Pressure Swing Adsorption |
| Electrical classification | Class II Type B | Class II Type B |
| Dimensions | 365×375×600mm<br>(14.37"×14.76"×23.62") | 280×3485×510mm<br>(14.37"×14.76"×23.62") |
| Net weight | 26Kg | 16Kg |
| Alarm | Low & high pressure | Low & high pressure;<br>Power failure |
| LCD display | Accumulating timing;<br>present timing;<br>timing | Accumulating timing;<br>present timing;<br>timing |
| Accessories | Nasal oxygen tube<br>Humidifier | Nasal oxygen tube<br>Humidifier |
| Materials | | |
| Sieve Bed | Synthetic Zeolite | Synthetic Zeolite |
| Nasal Oxygen Tube | PVC | PVC |
| Principles of operation | | |
| Operating system | Time cycle/ Pressure Swing Adsorption | Time cycle/ Pressure Swing Adsorption |
| Electronic safety<br>Software/Hardware | Electrical Safety per IEC-60601<br>Analog and digital electronics<br>with microprocessor | Electrical Safety per IEC-60601<br>Analog and digital electronics with<br>microprocessor |
| <b>Operating , transportation and storage environment</b> | | |
| Operating conditions | Ambient temperature : 10~40 ℃<br>Relative humidity:30%~85%<br>Atmospheric pressure:700<br>1060hPa | Ambient temperature : 10~40 ℃<br>Relative humidity:30%~85%<br>Atmospheric pressure:700 1060hPa |
| Transportation and<br>storage Conditions | Ambient temperature :-20~45 ℃<br>Relative humidity: ≤95%<br>Atmospheric pressure 500-<br>1060hPa | Ambient temperature :-40~45 ℃<br>Relative humidity: ≤95%<br>Atmospheric pressure 700-1060hPa |
{6}------------------------------------------------
The Longfian Oxygen Concentrator, Model JAY-5AW is similar to the predicate devices:
- Has the same intended use and indications for use ●
- Utilizes the same operating principle ●
- Incorporates the same basic design
- Incorporates the same technological characteristics ●
- Tested to the same electrical and electromagnetic safety standards for medical ● electrical equipment
- Manufactured under a quality system ●
The differences are:
- JAY-5AW has different enclosure shape comparing to JAY-5; ●
- The two models use different power of compressor, but compressor principle is the . same, the electronical safety and performance testing has validate the safety and effectiveness of JAY-5AW:
- The differences in dimensions and weight do not raise any issues in safety and ● effectiveness.
# 8. Assessment of Non-Clinical Testing:
Non-clinical testing of the Longfian Oxygen Concentrator, Model JAY-5AW has been performed against requirements for performance, physical attributes, environmental conditions, materials and safety, and to provide objective evidence that the device's intended use is met.
# 9. Test Summary:
The device has been evaluated the safety and performance by lab bench testing as following:
# Biocompatibility testing:
The biocompatibility evaluation for the Longfian Oxygen Concentrator, Model JAY-5AW was conducted in accordance with the ISO18562-1, ISO 18562-2, ISO18562-3, ISO10993-5, ISO10993-10, ISO10993-17.
{7}------------------------------------------------
#### Electrical safety and electromagnetic compatibility (EMC):
Electrical safety and EMC testing were conducted on the Longfian Oxygen Concentrator, Model JAY-5AW. The system complies with the IEC 60601-1. IEC 60601-1-2, IEC 60601-1-8, IEC60601-1-11, and ISO 80601-2-69 standards for electrical safety and the IEC60601-1-2 standard for EMC.
#### Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Level of Concern: The software for this device was considered as a moderate level of concern, because the subject device is a supplement to a patient's normal oxygen intake. If a failure of operation or latent design flaw could directly result in minor injury to the patient or operator. The device is not intended to be life supporting or life sustaining.
#### Performance Bench Testing
The basic safety and essential performance Test were evaluated based on ISO 80601-2-69. And Oxygen concentration and maximum flow performance testing was also conducted to verify that oxygen production capacity of the oxygen generator can meet the design requirements after working continuously at the maximum output for 8 hours
### 10. Conclusion
The subject device and the predicate device have the same intended use. The technological differences do not raise different questions of safety and effectiveness
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.