The Longfian Oxygen Concentrator, Model Jay-5 is intended to be used by patients with respiratory disorders who require supplemental oxygen. A high concentration of supplemental oxygen is supplied and a nasal cannula is used to channel oxygen from the concentrator to the patient. The Longfian Oxygen Concentrator, Model Jay-5 can be used in a home, institution, vehicle, and various mobile environments. The Longfian Oxygen Concentrator, Model Jay-5 does not nor is it intended to sustain or support life.
Device Story
Oxygen concentrator using pressure swing adsorption (PSA) to separate oxygen from ambient air; utilizes oilless air compressor, cooler, and absorption tower with synthetic zeolite sieve bed. Produces 90%-96% purity oxygen at flow rates up to 5 LPM. Used in homes, institutions, vehicles, and mobile environments; operated by patients or caregivers. Device includes flow meter, humidifier, and nasal cannula for delivery. Provides steady oxygen flow; includes alarms for low/high pressure and power failure. Intended to provide supplemental oxygen for respiratory disorders; does not support life. Healthcare providers use output to manage patient oxygen therapy; benefits include portable, reliable supplemental oxygen supply.
Clinical Evidence
Bench testing only. Performance validated against IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 8359 (oxygen concentrator safety), ISO 10993-1 (biocompatibility), and ASTM F 1464-93. Additional testing included VOCs (EPA TO-15), PM 2.5, ozone (21 CFR 801.415), and software validation.
Technological Characteristics
Pressure swing adsorption (PSA) system; synthetic zeolite sieve bed; oilless air compressor. Materials: PVC (nasal tube), synthetic zeolite. Electrical: Class II Type B, AC 230V. Dimensions: 365x375x600mm. Connectivity: None. Software: Microprocessor-based control for timing and alarms. Standards: IEC 60601-1, IEC 60601-1-2, ISO 8359, ISO 10993-1, ASTM F 1464-93.
Indications for Use
Indicated for patients with respiratory disorders requiring supplemental oxygen. Not intended to sustain or support life.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 5, 2014
Longfian Scitech Co., Ltd. c/o Mr. Jun Peng P&L Scientific, Inc. 6840 SW 45th LN Unit 5 Miami, FL 33155
Re: K131968
Trade/Device Name: OXY.LIFE Oxygen Concentrator, Models Jay-5 Regulation Number: 21 CFR 868.5440 Regulation Name: Oxygen concentrator, portable Regulatory Class: Class II Product Code: CAW Dated: August 1, 2014 Received: August 7, 2014
Dear Mr. Peng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 Mr. Peng
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows the name Tejashri Purohit-Sheth, M.D. twice. Below the second instance of the name, the text "Clinical Deputy Director" is present. The text "DAGRID/ODE/CDRH FOR" is located below the title.
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
K131968 510(k) Number (if known):
Device Name: The Longfian Oxygen Concentrator, Model Jay-5
### Indications for Use:
The Longfian Oxygen Concentrator, Model Jay-5 is intended to be used by patients with respiratory disorders who require supplemental oxygen. A high concentration of supplemental oxygen is supplied and a nasal cannula is used to channel oxygen from the concentrator to the patient. The Longfian Oxygen Concentrator, Model Jay-5 can be used in a home, institution, vehicle, and various mobile environments. The Longfian Oxygen Concentrator, Model Jay-5 does not nor is it intended to sustain or support life.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use: ✓ (Per 21 CFR 801.109) OR
Over-the Counter Use:
Image /page/2/Picture/10 description: The image shows the FDA logo. The logo is in a light gray color. The letters "FDA" are stacked on top of each other.
Anya C. Harry -S 2014.09.04 15:55:34 -04'00'
{3}------------------------------------------------
### 510(k) Summary
510(k) summary as required in 21 CFR 807.92 upon which Substantial Equivalence is based:
#### The Assigned 510(k) Number is: k131968
- 1. Submitter Information:
- Sponsor/510(K) Owner: . LONGFIAN SCITECH CO., LTD. Longxing Building A(413) No.77 Longxing Road Baoding 071051, China Phone: +86 10 312-3169262
Date: November 12, 2013
#### Contact Name: ●
Mr. Jun Peng P&L SCIENTIFIC, INC. 6840 SW 45TH LN UNIT 5, MIAMI, FL 33155 Phone: (305) 609 4701 Fax: (305) 397 0289 Email: jpeng@plscientificinc.com
#### 2. Device Name
Trade Name: OXY.LIFE Oxygen Concentrator, Models Jay-5 Common Name: Oxygen concentrator; Classification Name: Oxygen concentrator, portable;
#### 3. Classification:
| Classification: | 21 CFR 868.5440; Class II |
|-----------------|---------------------------|
| Product Code: | CAW; |
#### 4. Predicate Devices:
| Yuyue OxygenMaxTM 7F-5<br>Oxygen Concentrator | K083559,<br>September 4, 2009 | Cleared | JIANGSU<br>MEDICAL EQUIPMENT<br>& SUPPLY CO., LTD. | YUYUE |
|-----------------------------------------------|-------------------------------|---------|----------------------------------------------------|-------|
|-----------------------------------------------|-------------------------------|---------|----------------------------------------------------|-------|
#### 5. Description of Device
JAY-5 medical oxygen concentrator adopts pressure swing adsorption principle, which can separate oxygen, nitrogen and other gas from the air, at constant temperature. As soon as power is connected, the air is taken in and compressed by oilless air compressor through filtering, then the compressed air goes through the cooler and it is cooled. After that, the air is taken into absorption tower by control valve and the oxygen can be separated in the absorption tower. At the same time, the high purity oxygen is collected into the oxygen tank, and it goes through the flow meter and humidifier, finally the
{4}------------------------------------------------
oxygen that meets medical standards can be supplied. Oxygen is generated by pure physical method.
The concentrator can supply 1-2 patients simultaneously, with steady oxygen flowing out and reliable, adjustable flow. The device is capable of providing oxygen flow up to 5 LPM. The concentration of supplied oxygen is from 90%-96%. The key parts of the generator adopt anti-tiring and anti-aging design, and the planned life of the whole generator reaches up to 20,000 hours. The weight of Jay-5 is around 25kg (~55lbs). There is no influence on indoor oxygen percent during the generator operating.
JAY-5 medical oxygen concentrator contains the following items:
- · One (1) Jay-5 Oxygen Concentrator Unit
- · One (1) AC Power Supply
- · One (1) Nasal Oxygen Tube
- One (1) Humidifier
- One (1) Operation Manual
## 6. Intended Use
The Longfian Oxygen Concentrator, Model Jay-5 is intended to be used by patients with respiratory disorders who require supplemental oxygen. A high concentration of supplemental oxygen is supplied and a nasal cannula is used to channel oxygen from the concentrator to the patient. The Longfian Oxygen Concentrator, Model Jay-5 can be used in a home, institution, vehicle, and various mobile environments. The Longfian Oxygen Concentrator, Model Jay-5 does not nor is it intended to sustain or support life.
| Name | Predicate device | Subject device |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K number | K083559 | K131968 |
| Device Trade Name | Yuyue OxygenMax™ 7F-5<br>Oxygen Concentrator | JAY-5 Medical Molecular Sieve<br>Oxygen Concentrator |
| Common Name | Oxygen concentrator | Oxygen concentrator |
| Classification Name | Oxygen concentrator, Portable | Oxygen concentrator, Portable |
| Indications for use | The YuYue OxygenMax™ 7F-5 Oxygen<br>Concentrator is intended solely for<br>medical use in oxygen therapy programs<br>under the supervision of a physician. It is<br>intended to provide supplemental oxygen<br>to patients with respiratory disorders for<br>use in the home or health care facility.<br>This device is available by prescription<br>only and is not intended to support or<br>sustain life | The Longfian Oxygen Concentrator,<br>Model Jay-5 is intended to be used by<br>patients with respiratory disorders who<br>require supplemental oxygen. A high<br>concentration of supplemental oxygen is<br>supplied and a nasal cannula is used to<br>channel oxygen from the concentrator to<br>the patient. The Longfian Oxygen<br>Concentrator, Model Jay-5 can be used in<br>a home, institution, vehicle, and various<br>mobile environments. The Longfian<br>Oxygen Concentrator, Model Jay-5 does<br>not nor is it intended to sustain or support<br>life. |
| Classification | 21 CFR 868.5440 | 21 CFR 868.5440 |
#### Summary of Comparison in to Technological Characteristics Predicate Device 7.
{5}------------------------------------------------
| Feature | | |
|----------------------------------------------------|----------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Power supply | AC 220±22V, 50±1Hz | AC 230±23V, 50±1Hz |
| Input power | 500W | 500W |
| Oxygen concentration | 90%-94% | 93%±3% |
| Oxygen flow | 0.5~5L/min | 0~5L/min |
| Outlet pressure | 40~70kPa | 40~70kPa |
| Noise | ≤53dB(A) | ≤53dB(A) |
| Fuse | / | T6.3AL/250V |
| Operating system | Pressure Swing Adsorption | Pressure Swing Adsorption |
| Electrical classification | Class II Type B | Class II Type B |
| Dimensions | 445×372×680mm | 365×375×600mm<br>(14.37"×14.76"×23.62") |
| Net weight | 25.8Kg | 26Kg |
| Alarm | Low & high pressure | Low & high pressure;<br>Power failure |
| LCD display | Accumulating timing | Accumulating timing;<br>present timing;<br>timing |
| Accessories | Nasal oxygen cannula<br>Humidifier | Nasal oxygen tube<br>Humidifier |
| Materials | | |
| Sieve Bed | Synthetic Zeolite | Synthetic Zeolite |
| Nasal Oxygen Tube | PVC | PVC |
| Principles of operation | | |
| Operating system | Time cycle/ Pressure Swing Adsorption | Time cycle/ Pressure Swing Adsorption |
| Electronic safety | Electrical Safety per IEC-60601 | Electrical Safety per IEC-60601 |
| Software/Hardware | Analog and digital electronics with microprocessor | Analog and digital electronics with microprocessor |
| Operating , transportation and storage environment | | |
| Operating conditions | / | Ambient temperature : 10~40℃<br>Relative humidity:30%~85%<br>Atmospheric pressure:700 1060hPa |
| Transportation and storage Conditions | / | Ambient temperature :-20~45℃<br>Relative humidity:≤95%<br>Atmospheric pressure:500-1060hPa |
The Longfian Oxygen Concentrator, Model Jay-5 is similar to the predicate devices:
- . Has the same intended use and indications for use
- . Utilizes the same operating principle
- Incorporates the same basic design .
- Incorporates the same technological characteristics
{6}------------------------------------------------
- Tested to the same electrical and electromagnetic safety standards for medical electrical equipment
- Manufactured under a quality system
Differences between the subject device and predicate do not affect the safety and effectiveness of the device.
## 8. Assessment of Non-Clinical Testing:
Non-clinical testing of the Longfian Oxygen Concentrator, Models Jay-5 has been performed against requirements for performance, physical attributes, environmental conditions, materials and to provide objective evidence that the device's intended use is met.
## 9. Summary Performance Data
- . IEC 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Safety and essential performance, 1988; Amendment 1, 1991-11, Amendment 2, 1995. (General)
- IEC 60601-1-2, (Second Edition, 2001), Medical Electrical Equipment Part 1-2: ● General Requirements for Safety - Collateral Standard: Electromagnetic
- ISO 8359:1996 Oxygen concentrators for medical use -- Safety requirements ●
- ISO 10993-1:2009 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing
- ASTM F 1464-93:2005 Oxygen Concentrators for medical use-Safety Requirements
- Volatile organic compounds (VOC) test per EPA TO-15 ●
- PM 2.5 test per US EPA requirement ●
- Ozone test per 21 CFR 801.415 ●
- . Software validation
## 10. Conclusion
After analyzing both bench and external laboratory testing data, the intended use and supporting data can conclude that the device in the submission is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.