K213200 · Artio Medical, Inc. · KRD · Mar 21, 2022 · Cardiovascular
Device Facts
Record ID
K213200
Device Name
Solus Gold Embolization Device
Applicant
Artio Medical, Inc.
Product Code
KRD · Cardiovascular
Decision Date
Mar 21, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Solus Gold Embolization Device is indicated to obstruct or reduce the rate of blood flow in the peripheral vasculature.
Device Story
Vascular plug device for peripheral vasculature blood flow obstruction/reduction. Implant assembly (gold body, silicone valves, platinum/iridium markers) and delivery system (microcatheter, delivery catheter, RHV, 3-way stopcock). Physician selects implant size (3-6 mm); device delivered over .014" guidewire through compatible catheter. Expansion facilitated by injecting saline or saline/contrast mixture into implant. Mechanical detachment from delivery system. Used in clinical settings by physicians. Provides immediate vessel occlusion; gold material enhances radiopacity for visualization. Benefits include effective, durable embolization of peripheral vessels.
Clinical Evidence
No human clinical trials. Evidence based on bench performance testing and a 90-day GLP chronic canine study (n=7). Canine study compared Solus Gold to MVP predicate. Primary endpoints: acute embolization time (0.50 min vs 1.94 min) and chronic occlusion durability at 90 days. Results: 14/14 Solus Gold devices achieved complete occlusion at 90 days vs 2/14 for predicate. Histology showed 100% mean lumen occlusion for Solus Gold vs 79.6% for predicate. No complications observed.
Indicated for patients requiring obstruction or reduction of blood flow in the peripheral vasculature.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
Predicate Devices
Reverse Medical Micro Vascular Plug System (K141313)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 21, 2022
Artio Medical, Inc. Erdie Peralta Vice-President, Clinical And Regulatory Affairs 127 Independence Drive Menlo Park, California 94025
Re: K213200
Trade/Device Name: Solus Gold Embolization Device Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: February 14, 2022 Received: February 15, 2022
Dear Erdie Peralta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Misti Malone, PhD Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213200
Device Name Solus Gold Embolization Device
Indications for Use (Describe)
The Solus Gold Embolization Device is indicated to obstruct or reduce the rate of blood flow in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
K213200 510(k) Number:
### I. SUBMITTER INFORMATION
| Submitter: | Artio Medical, Inc.<br>127 Independence Drive<br>Menlo Park, CA 94025 |
|------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Erdie DePeralta<br>Vice President, Clinical and Regulatory Affairs<br>Phone: (978) 408-0949<br>Email: edeperalta@artiomedical.com |
March 18, 2021 Date Summary Prepared:
### II. SUBJECT DEVICE INFORMATION
| Device Trade Name: | Solus Gold™ Embolization Device |
|----------------------|----------------------------------------------|
| Common Name: | Device, Vascular, For Promoting Embolization |
| Classification Name: | Vascular Embolization Device |
| Regulation Number: | 21 CFR §870.3300 |
| Regulatory Class: | II |
| Product Code: | KRD |
| Submission Type: | Traditional 510(k) |
### III. PREDICATE DEVICE INFORMATION:
Predicate Device: Reverse Medical Micro Vascular Plug System (K141313)
### IV. DEVICE DESCRIPTION:
The Solus Gold Embolization Device is a vascular plug device intended to obstruct or reduce the rate of blood flow in the peripheral vasculature. It is comprised of an implant assembly and a delivery system assembly. The delivery system assembly is comprised of a microcatheter and a delivery catheter. A rotating hemostatic valve (RHV) and a 3-way stopcock is provided on the hub of the delivery catheter for injecting fluid to facilitate expansion of the implant. The Solus Gold implant is available in 3 mm – 6 mm diameters. All implant sizes are provided with the same delivery system, and the physician selects the device size based on the treatment plan. The Solus Gold implant assembly consists of a gold implant body, a distal neck subassembly, and an implant mating component. The proximal end of the implant is joined to the delivery catheter through the implant , which is a platinum tubular segment secured to the inside of the gold implant body with adhesive. The distal neck subassembly is comprised of two stacked silicone valves and a platinum (Pt) marker band that are contained within a stainless-steel housing and bonded to the implant with adhesive. The distal neck subassembly of the implant is mounted to the microcatheter over a preparation tool, which is removed by the user after device preparation. The implant is pleated and folded, loaded onto the delivery catheter and microcatheter, and protected with a sheath. The device is provided sterile and is intended for single use only.
{4}------------------------------------------------
### V. INDICATIONS FOR USE
The Solus Gold Embolization Device is indicated to obstruct or rete of blood flow in the peripheral vasculature.
## VI. COMPARISON OF INTENDED USE, INDICATIONS FOR USE, AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Attribute | Subject Device<br>Solus Gold™ Embolization Device | Predicate Device<br>Reverse Medical Micro Vascular Plug System<br>(K141313, K150108) |
|------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | KRD | Same as subject device |
| Regulation No. | 21 CFR §870.3300 | Same as subject device |
| Regulation Name | Vascular embolization device | Same as subject device |
| Intended Use | To obstruct or reduce the rate of blood flow in<br>the peripheral vasculature | Same as subject device |
| Indications for<br>Use | The Solus Gold Embolization Device is indicated to<br>obstruct or reduce the rate of blood flow in the<br>peripheral vasculature. | The Reverse Medical Micro Vascular Plug System<br>is indicated to obstruct or reduce the rate of<br>blood flow in the peripheral vasculature. |
| Anatomical Site | Peripheral vasculature | Same as subject device |
| Target Vessel<br>Diameter | 1.5 - 5.0 mm | 1.5 – 7.0 mm |
| Implant Diameter x<br>Implant Length<br>(Expanded or<br>Unconstrained) | 3 mm x 6.7 mm, 3.5 mm x 7.6 mm,<br>4 mm x 8.4 mm, 4.5 mm x 9.3 mm,<br>5 mm x 10.2 mm, 5.5 mm x 11.1 mm,<br>6 mm x 11.9 mm | MVP-3: 5.3 mm x 12 mm<br>MVP-5: 6.5 mm x 12 mm<br>MVP-7: 9.2 mm x 16 mm |
| Implant Materials | Gold, platinum/iridium, stainless steel,<br>silicone, adhesive | Nitinol, platinum, ePTFE |
| Delivery System<br>Materials of<br>Construction | PTFE, PEBA, polyurethane, nylon, nitinol, stainless-<br>steel, platinum/iridium, liquid crystal polymer<br>(LCP), adhesive, polyolefin, hydrophilic coating | Stainless steel, solder, polypropylene,<br>urethane, cyanoacrylate |
| Implant Description | Ovoid gold implant body with proximal neck<br>implant mating component and distal neck<br>subassembly, expanded by injection of saline or a<br>mixture of saline and contrast | Ovoid nitinol frame with an ePTFE cover over<br>the proximal portion, self-expanding |
| Implant Radiopacity | Entire gold implant body and proximal and distal<br>platinum/iridium marker bands | Proximal and distal marker bands only |
| Delivery method | Implant comes attached to the delivery catheter<br>and microcatheter, is delivered over a .014"<br>guidewire to the intended treatment site,<br>expanded, and detached from the delivery<br>catheter and microcatheter. | Implant comes attached to a delivery wire, is<br>pushed through a commercially available<br>catheter to the intended treatment site,<br>expanded, and detached from delivery wire. |
| Catheter<br>Compatibility | Compatible with commercially available catheters:<br>All implant sizes: > 0.070" ID catheter (6 Fr) | Compatible with commercially available<br>catheters:<br>MVP-3Q: 0.021" - 0.027" ID catheter<br>MVP-5Q: 0.027" ID catheter<br>MVP-7Q: 0.041" ID catheter |
| Microcatheter<br>Maximum Length | 154 cm | 150 cm |
| Delivery System<br>Maximum Length | Delivery catheter length: 136 cm | Delivery wire length: 160 – 180 cm |
| How Transferred to<br>Catheter | Attached to delivery system, inserted into<br>vasculature over a guidewire and through a<br>compatible commercially available catheter | Attached to delivery wire, inserted into<br>vasculature through a compatible,<br>commercially available catheter |
| Attribute | Subject Device<br>Solus GoldTM Embolization Device | Predicate Device<br>Reverse Medical Micro Vascular Plug System<br>(K141313, K150108) |
| Detachment | Mechanical | Same as subject device |
| How Provided | Sterile, single-use | Same as subject device |
| Sterilization<br>Method | Ethylene oxide | Same as subject device |
{5}------------------------------------------------
### VII. PERFORMANCE DATA
### Sterilization Validation:
Sterilization conditions were validated according to ANSI / AAMI /ISO 11135, Sterilization of health care products — Ethylene oxide — Requirements for development, validation and routine control of a sterilization process for medical devices to provide a Sterility Assurance Level (SAL) of 10-6.
## Packaging & Shelf-Life Validation:
Packaging Validation for the Solus Gold Embolization Device included: Pouch Seal Strength, Packaging Integrity, Unique Device Identification Barcode Quality, and Microbial Barrier testing was conducted at both the T=0 timepoint and the accelerated aging timepoint at T=6 months. Product Integrity Following Transit and Device Condition after Transit was done at T=0. Additionally, the performance testing results of the accelerated aged samples demonstrated that the device performs as intended throughout its labeled shelf-life.
### Biocompatibility:
The biological evaluation for the Solus Gold Embolization Device was conducted in accordance with the FDA recognized consensus standard (ANSI/AAMI/ISO 10993-1) "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" and FDA's applicable guidance document, "Use of International Standard ISO 10993" issued September 2020. The implant of the Solus Gold Embolization Device is categorized as "Long Term Contact with Circulating Blood." The non-implantable delivery system, including the microcatheter and delivery catheter, is categorized as "Externally Communicating Device in Limited Contact with Blood." Pursuant to ISO 10993-1 and the 2020 FDA Guidance, biological risks were addressed by: appropriate testing, a chemical characterization assessment, a literature review or a combination of these approaches.
## Bench Performance Testing:
Pursuant to the Class II Special Controls Guidance Document: Vascular Embolization Devices (issued December 29, 2004), bench testing of the Solus Gold Embolization Device was performed. Bench testing of the Solus Gold Embolization implant included: 1) implant tensile; 3) implant tensile; 3) implant expansion; 4) implant maximum pressure; 5) distal neck sub-assembly flow; 6) implant dimensions; 7) corrosion resistance; and 8) MR compatibility. Delivery catheter and microcatheter verification testing included: 1) microcatheter distal flexibility; 2) microcatheter liquid leak under pressure and delivery catheter liquid leak and burst; 3) catheter air leak under vacuum; 4) catheter kink resistance; 5) catheter torque durability; 6) catheter tensile force; 7) inadvertent implant detachment; 8) luer hub compatibility; 10) coating integrity; and 11) catheter corrosion resistance. Additionally, verification testing of the complete Solus Gold Embolization Device included: 1) simulated use testing; 2) microcatheter retraction tensile force; 3) implant detachment tensile force; and 4) particulate. These bench performance studies confirm that the device performed as intended per the
{6}------------------------------------------------
product specifications and also demonstrated that the Solus Gold Embolization Device is substantially equivalent to the MVP predicate device.
### Human Factors Assessment:
The Solus Gold Embolization Device was evaluated by fifteen (15) intended users in a simulated clinical setting. This testing was conducted pursuant to the applicable requirements of FDA's Guidance Document titled "Applying Human Factors and Usability Engineering to Medical Devices (issued February 3, 2016)." For this evaluation, the users were trained by Artio personnel. After a decay period of at least 1 hour post training, users were provided with a Solus Gold Embolization Device and the proposed instructions for use. Critical tasks were assessed through observation during simulated use and through user knowledge assessments about the device's labeling and procedure steps that were not amendable to simulated use testing. Additionally, the test subjects completed a survey. The human factors assessment demonstrated that the device labeling and training provided to the intended users allowed for the proper use of the Solus Gold Embolization Device in its intended use environment. Further, this assessment confirmed that all identified risks were considered acceptable and that no new risks were identified.
## Pre-Clinical Animal Performance Testing
The Solus Gold Embolization Device and the MVP predicate device were evaluated in a chronic 90-day canine animal study model. Specifically, safety, effectiveness, in vivo performance and handling were conducted to assess device performance with the design requirements and product specification. Additionally, in vivo thrombosis testing (thrombogenicity) was conducted as part of the GLP animal study. The study was conducted under Good Laboratory Practice (GLP) controls pursuant to 21 CFR §58. The study complied with the requirements of Vascular and Neurovascular Embolization Devices - Class II Special Controls Guidance Document for Industry and FDA Staff (Dec. 2004), EN ISO 10993-4:2017 Biological Evaluation of Medical Devices Part 4: Selection of Test for Interactions with Blood and FDA General Considerations for Animal Studies for Cardiovascular Devices – Guidance for Industry and FDA Staff (July 2010). A total of seven (7) canine underwent treatment procedures with the Solus Gold Embolization Device and/or the MVP predicate device (Day 0). One Solus Gold Embolization Device (2.5 mm to 6.0 mm with 0.5 mm increments) or the MVP predicate device (MVP-3Q, MVP-5Q, MVP-7Q devices) was placed per treated artery (internal thoracic artery, renal artery, external iliac artery, and splenic artery. Follow-up angiography occurred on Day 28 ± 3 days after the treatment procedure. Terminal procedures took place on Day 90 ± 7 days. Acute embolization effectiveness was assessed by the time to achieve complete (Grade 0 or Grade 1) target vessel segment occlusion by angiography on Day 0. Chronic embolization effectiveness (durability) was assessed by target vessel segment occlusion by angiography (Grade 0 – 4) and morphometric percent vessel lumen occlusion by histology on Day 90.
The mean time to achieve complete (Grade 0 or Grade 1) target vessel segment occlusion by angiography on Day 0 was .50 minute for the Solus Gold Embolization Device and 1.94 minutes for the MVP predicate device. Complete (Grade 0 or Grade 1) target vessel segment occlusion was observed at 90 days were 14/14 for the Solus Gold Embolization Device and 2/14 for the MVP predicate device. Mean (maximal) percent vessel lumen occlusion by histology at 90 days was 100% for the Solus Gold Embolization Device and 79.6 % for the MVP predicate device. No complications were observed for either the Solus Gold Embolization Device or the MVP predicate device.
{7}------------------------------------------------
### VIII. CONCLUSION
Information in this 510(k) submission demonstrate that the Solus Gold Embolization Device is substantially equivalent to the MVP predicate device. Substantial equivalence is based on comparative bench and animal testing, a comparison of the component materials and comparison of the technological characteristics of the Solus Gold Embolization Device and the MVP predicate device. The Solus Gold Embolization Device is substantially equivalent in its intended use, indications for use, and performance to the MVP predicate device. The differences between the Solus Gold Embolization Device and the MVP predicate device do not raise any new or different questions related to the device safety or effectiveness. Therefore, the Solus Gold Embolization Device is substantially equivalent to the MVP predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.