K141313 · Reverse Medical Corporation · KRD · Jun 18, 2014 · Cardiovascular
Device Facts
Record ID
K141313
Device Name
REVERSE MEDICAL MICRO VASCULAR PLUG SYSTEM
Applicant
Reverse Medical Corporation
Product Code
KRD · Cardiovascular
Decision Date
Jun 18, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Reverse Medical® Corporation MVP® Micro Vascular Plug System is indicated to obstruct or reduce the rate of blood flow in the peripheral vasculature.
Device Story
The MVP Micro Vascular Plug System is a vascular embolization device used to obstruct or reduce blood flow in peripheral vessels. The device consists of a self-expandable, ovoid-shaped nitinol frame with a PTFE cover, secured by platinum marker bands. It is delivered to the treatment site via a microcatheter using a composite delivery wire. The device is operated by a physician. The system features a mechanical detachment mechanism; the plug is detached from the delivery wire by rotating the wire counter-clockwise. This mechanical method replaces a previous electrolytic detachment system, eliminating the need for external power boxes or cables. The device is intended for use in clinical settings where peripheral vascular embolization is required to manage blood flow.
Clinical Evidence
Bench testing only. No clinical data was provided. Verification testing included dimensional and visual inspection, microcatheter compatibility (flexibility, kinking, deployment/retraction forces), detachment evaluations (torque strength, number of turns), galvanic corrosion (ASTM G71), MRI compatibility (ASTM F-2503), and package integrity/shelf-life testing. All tests met acceptance criteria.
Indicated for patients requiring obstruction or reduction of blood flow in the peripheral vasculature.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
{0}------------------------------------------------
| Sponsor/Applicant: | Reverse Medical Corporation<br>13700 Alton Parkway, Ste. 167<br>Irvine, CA 92618 |
|-------------------------|------------------------------------------------------------------------------------------------------------|
| Date Prepared: | May 17, 2014 |
| Contact Person: | Jane Metcalf<br>Vice President, Quality, Regulatory and Clinical Affairs<br>Email: jmetcalf@reversemed.com |
| Trade Name: | Reverse Medical® MVP® Micro Vascular Plug System |
| Common Name: | Vascular Embolization Device |
| Classification Name: | Vascular Embolization Device |
| Device Classification : | Class 2 |
| Regulation Number: | 870.3300 (product code: KRD) |
| Predicate Devices: | Reverse Medical MVP-3 (K123803) and Reverse Medical<br>MVP-5 (K133282) |
#### Purpose of Submission
The purpose of this special 510(k) submission is to obtain market clearance for two (2) modifications to the MVP System. The first is to replace the electrolytic detachment method with a mechanical detachment method. The second is to reduce the length of the delivery wire from 180 cm to 160 cm.
#### Indication for Use and Intended Use
The Reverse Medical® Corporation MVP® Micro Vascular Plug System is indicated to obstruct or reduce the rate of blood flow in the peripheral vasculature.
#### Device Description
Reverse Medical® MVP® Micro Vascular Plug System consists of a micro vascular occlusion plug that is attached to a composite delivery wire and is intended to be delivered to the treatment site through a catheter. The MVP occlusion plug is a selfexpandable, ovoid-shaped frame made from nitinol and incorporates a PTFE cover over the proximal portion of the ovoid. The plug device is secured at both ends with platinum marker bands. The proximal marker band is attached to a delivery wire that is used to push the plug device through a commercially available catheter to the intended treatment site. After satisfactory deployment of the plug device at the treatment site, the implant is detached from the delivery wire by rotating the wire counter clockwise.
#### Technical Characteristics
The modified devices have the same technological characteristics as the predicate devices except for the detachment method and length of delivery wire. The mechanical
{1}------------------------------------------------
detachment method eliminates the need for a software controlled detachment box and cables. Table 5-1 compares characteristics of the modified devices to the predicates.
| | K123803 | K133282 | Subject of this Submission | |
|---------------------------------|----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Feature | MVP-3<br>Predicate | MVP-5<br>Predicate | MVP-3 Modified | MVP-5 Modified |
| Indications for Use | To obstruct or reduce<br>the rate of blood flow<br>in the peripheral<br>vasculature. | To obstruct or reduce<br>the rate of blood flow<br>in the peripheral<br>vasculature. | To obstruct or reduce<br>the rate of blood flow<br>in the peripheral<br>vasculature. | To obstruct or reduce<br>the rate of blood flow<br>in the peripheral<br>vasculature. |
| Materials of<br>Construction | Nitinol, PTFE,<br>Platinum, SS 301,<br>Solder, Polypropylene<br>sheath, Urethane,<br>Cyanoacrylate | Nitinol, PTFE,<br>Platinum, SS 301,<br>Solder, Polypropylene<br>sheath, Urethane,<br>Cyanoacrylate | Nitinol, PTFE,<br>Platinum, Solder,<br>Polypropylene sheath,<br>Urethane<br>Cyanoacrylate | Nitinol, PTFE,<br>Platinum, Solder,<br>Polypropylene sheath,<br>Urethane<br>Cyanoacrylate |
| Plug (Implant)<br>descripton | Self-expandabable,<br>ovoid shaped frame<br>with a PTFE cover<br>over the proximal<br>portion | Self-expandabable,<br>ovoid shaped frame<br>with a PTFE cover<br>over the proximal<br>portion | Self-expandabable,<br>ovoid shaped frame<br>with a PTFE cover<br>over the proximal<br>portion | Self-expandabable,<br>ovoid shaped frame<br>with a PTFE cover<br>over the proximal<br>portion |
| Plug Diameter,<br>Unconstrained | 5.3 mm | 6.5 mm | 5.3 mm | 6.5 mm |
| Plug Length,<br>Unconstrained | 12 mm | 12 mm | 12 mm | 12 mm |
| Target Vessel<br>Diameter | 1.5-3.0 mm | 3.0-5.0 mm | 1.5-3.0 mm | 3.0-5.0 mm |
| Method of<br>Placement | Delivery wire through<br>a 0.021" ID<br>Microcatheter | Delivery wire through<br>a 0.027" ID<br>Microcatheter | Delivery wire<br>through a 0.021" to<br>0.027" ID<br>Microcatheter | Delivery wire<br>through a 0.027" ID<br>Microcatheter |
| Radiopaque<br>Markers | Platinum marker bands<br>at each end of the plug | Platinum marker<br>bands at each end of<br>the plug | Platinum marker<br>bands at each end of<br>the plug | Platinum marker<br>bands at each end of<br>the plug |
| Proximal End of<br>Plug Config. | Proximal marker band<br>attached to delivery<br>wire | Proximal marker band<br>attached to delivery<br>wire | Proximal marker<br>band attached to<br>delivery wire | Proximal marker<br>band attached to<br>delivery wire |
| Delivery Wire<br>Length | 180 cm | 180 cm | 160 cm | 160 cm |
| Detachment System | Electrolytic | Electrolytic | Mechanical | Mechanical |
| Sterilization<br>Process | EO | EO | EO | EO |
| Accessories | Electrolytic Box and<br>Cables | Electrolytic Box and<br>Cables | Torquer | Torquer |
# Table 5-1 Comparison to the Predicates
{2}------------------------------------------------
## Performance Tests - Non-clinical
Due to the change in detachment method and delivery wire length the following design verification tests were conducted in accordance with Reverse Medical Design Control procedures. All testing was performed on units that were sterilized and met all inspection criteria. Tests on the Reverse Medical MVP System included:
- . Dimensional Inspection
- Visual Inspection
- Microcatheter Compatibility within "Simulated Use Vascular Model" .
- Flexibility within microcatheter o
- Delivery wire kinking assessment o
- Multiple deployments and withdrawals through the microcatheter O
- Force required to deploy and retract device within the microcatheter O
- Detachment Evaluations
- o Number of turns required to detach
- Torque strength of detachment junction o
- Galvanic Corrosion per ASTM G71 .
- MRI Compatibility per ASTM F-2503
- Package Integrity and Shelf-life .
All tests successfully passed acceptance criteria. This demonstrates that the modified devices meet the product specification.
## Basis for Determination of Substantial Equivalence
Upon reviewing the performance data and comparing intended use, design, materials, principle of operation and overall technological characteristics, the modified Reverse Medical MVP System is determined to be substantially equivalent to the currently marketed Reverse Medical MVP System. Differences between the systems do not raise any issues of safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three abstract shapes that resemble a human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 18, 2014
Reverse Medical Corporation % Jane Metcalf VP, Quality Assurance, Regulatory and Clinical Affairs 13700 Alton Parkway, Suite 167 Irvine, California 92618
Re: K141313
> Trade/Device Name: MVP Micro Vascular Plug System Regulation Number: 21 CFR 870.3300 Regulation Name: Device Embolization, Vascular Regulatory Class: 11 Product Code: KRD Dated: May 16, 2014 Received: May 20, 2014
Dear Ms. Metcalf,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, howeyer, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{4}------------------------------------------------
Page 2 - Jane Metcalf
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K141313
#### Device Name
Reverse Medical(R) Corporation MVP(R) Micro Vascular Plug System
#### Indications for Use (Describe)
The Reverse Medical(R) Corporation MVP(R) Micro Vascular Plug System is indicated to obstruct or reduce the rate of blood flow in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY "
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.