The Hemoclip is indicated for endoscopic clip placement within the digestive tract for the purpose of: 1. Endoscopic marking, 2. Hemostasis for: Mucosal/sub-mucosal defects <3cm Bleeding ulcers Arteries<2mm Polyps<1.5cm in diameter Diverticula in the colon 3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel. 4. As a supplementary method, closure of digestive tract luminal perforations <20mm that can be treated conservatively
Device Story
Endoscopic hemostatic clip system; consists of release handle and clip component. Operated by physician via endoscope; sliding handle mechanism opens/closes clip for tissue grasping. Once positioned, handle separates clip from release part; clip remains in tract for 1-2 weeks before natural passage. Subject device features reloadable design allowing repeated connection/release of multiple clips during single procedure. Used in clinical settings for hemostasis, marking, tube anchoring, and perforation closure. Benefits include reduced bleeding risk and minimally invasive tissue management.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing, including shelf life validation (ISO 11607-1/2, ASTM F 1980-16), packaging validation (ISO 11607-1/2), biocompatibility (ISO 10993), and sterility (ISO 11135).
Indicated for patients requiring endoscopic clip placement in the digestive tract for marking, hemostasis (mucosal/sub-mucosal defects <3cm, bleeding ulcers, arteries <2mm, polyps <1.5cm, colonic diverticula), anchoring jejunal feeding tubes, or closure of digestive tract luminal perforations <20mm.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
July 12, 2022
Hangzhou AGS MedTech Co., Ltd. Yanping Fu, RA Supervisor Building 5, Building 6, No. 597 Kangxin Road Yuhang District Hangzhou, Zhejiang 311106 CHINA
Re: K213143
Trade/Device Name: Hemoclip Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal ligator Regulatory Class: Class II Product Code: PKL Dated: May 27, 2022 Received: June 6, 2022
Dear Yanping Fu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213143
Device Name Hemoclip
Indications for Use (Describe)
The Hemoclip is indicated for endoscopic clip placement within the digestive tract for the purpose of:
- 1. Endoscopic marking,
- 2. Hemostasis for:
- · Mucosal/sub-mucosal defects <3cm
- · Bleeding ulcers
- · Arteries<2mm
- · Polyps<1.5cm in diameter
- · Diverticula in the colon
- 3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel.
- 4. As a supplementary method, closure of digestive tract luminal perforations <20mm that can be treated conservatively
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo for AGS Medtech. The logo consists of a circular design with three curved shapes in blue and green. To the right of the circular design, the text "AGS MEDTECH" is written in English, followed by Chinese characters below it. The Chinese characters are likely the translation of the company name.
510(k) Summary Hemoclip
# 510(k) Summary (K213143)
We submit this 510(k) Summary as per 21 CFR 807.92, it meets the content and format regulatory requirements.
| 5.1 Submitter | |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Submitted by/Owner: | Hangzhou AGS MedTech Co., Ltd.<br>Building 5, Building 6, No.597 Kangxin Road Yuhang<br>District, 311106 Hangzhou, Zhejiang, China |
| Establishment Registration Number: | 3010288205 |
| Registration Status: | Active |
| Contact Person: | Yanping Fu<br>Phone: 0086-15958493282<br>Fax: 0086-0571-87671225<br>Email: fuyp@bioags.com |
| Date Prepared: | July. 21, 2021 |
# 5.2 Proposed Device
| Trade Name: | Hemoclip |
|-------------------------|----------------------------------------|
| Device Name: | Hemoclip |
| Common Name: | Hemoclip |
| Regulation class: | Class II |
| Regulation Number: | 876.4400 |
| Regulation Description: | Hemorrhoidal ligator |
| Review Panel: | Gastroenterology/Urology |
| Product Code: | PKL |
| Product Code Name: | Hemostatic Metal Clip For The Gi Tract |
# 5.3 Predicate Device and Reference Device
#### 5.3.1Predicate Device
| Trade Name: | Hemoclip |
|-------------------------|----------------------------------------|
| Device Name: | Hemoclip |
| Common Name: | Hemoclip |
| 510(k) Number: | K172727 |
| Regulation class: | Class II |
| Regulation Number: | 876.4400 |
| Regulation Description: | Hemorrhoidal ligator |
| Review Panel: | Gastroenterology/Urology |
| Product Code: | PKL |
| Product Code Name: | Hemostatic Metal Clip For The Gi Tract |
| 5.3.2 Reference Device | |
| Trade Name: | Single Use Reloadable Clip Applicator, |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains a logo for AGS Medtech. The logo consists of a circular design with three blue curved shapes intertwined, forming a triangular space in the center filled with green. To the right of the circular design, the text "AGS MEDTECH" is written in English, followed by Chinese characters below it.
510(k) Summary Hemoclip
| | Clip, Long Clip, Short Clip, Super Short Clip |
|-------------------------|-----------------------------------------------------------------------------------------|
| Device Name: | Single Use Reloadable Clip Applicator,<br>Clip, Long Clip, Short Clip, Super Short Clip |
| Common Name: | Endoscopic Clipping Device |
| 510(k) Number: | K183590 |
| Regulation class: | Class II |
| Regulation Number: | 876.4400 |
| Regulation Description: | Hemorrhoidal ligator |
| Review Panel: | Gastroenterology/Urology |
| Product Code: | PKL |
| Product Code Name: | Hemostatic Metal Clip For The Gi Tract |
| Trade Name: | Lockado™ Repositionable Hemostasis Clip |
|-------------------------|-----------------------------------------|
| Device Name: | Lockado™ Repositionable Hemostasis Clip |
| Common Name: | Endoscopic Clipping Device |
| 510(k) Number: | K202333 |
| Regulation class: | Class II |
| Regulation Number: | 876.4400 |
| Regulation Description: | Hemorrhoidal ligator |
| Review Panel: | Gastroenterology/Urology |
| Product Code: | PKL |
| Product Code Name: | Hemostatic Metal Clip For The Gi Tract |
## 5.4 Device Description
Hemoclip consists of Release part and Clip part. Clip part consists of Clip and Frap Tube. Release part consists of Spring End, Spring hose near clip end, Connect end, Plastic Coated Spring Tube / Spring Tube and Handle part. EO Sterilization and use for single use only. For our proposed Hemoclip, the release part could connect and release the clips repeatedly.
We include 5 series for this submission, 55501series, 55502 series, 55504 series and 55505 series. 55501 series and 55502 series are models consist of release part and clip part, and the release part and clip part are already combined, no additional separate clips include, 55501series and 55502 series are reloadable, the release part could connect and release clips repeatedly. 55503 series are models only consist of separate clip part, and the release part of 55501, 55502, 55503 and 55504 series can connect with the separate clip (55503 series) repeatedly. 55504 and 55505 series are models consist of one preloaded Hemoclip (consist of release part and clip part, and the release part and clip part are already combined) and additional separate clips, 55504 series and 55505 series are reloadable, the release part can connect separate clips and release clips repeatedly.
# 5.5 Indication for use statement
The Hemoclip is indicated for endoscopic clip placement within the digestive tract for the purpose of:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for AGS Medtech. The logo consists of a blue and green circular design on the left, with the text "AGS MEDTECH" and "安杰思医学" to the right of the design. The circular design appears to be made up of three curved shapes, with the blue shapes on the outside and the green shape in the center.
510(k) Summary Hemoclip
- 1. Endoscopic marking,
- 2. Hemostasis for:
- Mucosal/sub-mucosal defects <3cm
- Bleeding ulcers
- Arteries<2mm
- Polyps<1.5cm in diameter
- Diverticula in the colon
- Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection
- 3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel.
4. As a supplementary method, closure of digestive tract luminal perforations<20mm that can be treated conservatively.
## 5.6 Comparison of Technology Characteristics
Our proposed device Hemoclip is substantially equivalent to the predicate device. The differences between the Hemoclip and the predicate device do not raise any questions regarding its safety and effectiveness. The differences are listed in the table below:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains a logo for AGS Medtech. The logo consists of a circular design with three blue curved shapes forming a triangle around a green center. To the right of the logo, the text "AGS MEDTECH" is written in English, followed by Chinese characters below it. The overall design is clean and professional, suggesting a company in the medical technology field.
K213143
Page 4 of 5
Section 5 510(k) Summary Hemoclip
| Table 5.6 Comparison of technical characteristics | | | | |
|---------------------------------------------------|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Item | Proposed device<br>Hangzhou AGS MedTech Co.,<br>Ltd. | Predicate device (K172727)<br>Hangzhou AGS MedTech Co.,<br>Ltd. | Comparion |
| Technical | Principle of operation | The clip is opened or closed by<br>operating the sliding handle.<br>After positioning and clamping<br>the related tissues, the sliding<br>handle is operated to separate<br>the clip components from the<br>release part, and the release<br>part then exits the human<br>digestive tract. The clip<br>remained in the alimentary<br>tract for about 1-2 weeks, fell<br>off naturally and was<br>discharged through the<br>intestine and anus.<br>And the release part could<br>connect with additional<br>separate clip part repeatedly, to<br>be a new whole Hemoclip, and<br>realize the intended use. | The clip is opened or closed by<br>operating the sliding handle.<br>After positioning and clamping<br>the related tissues, the sliding<br>handle is operated to separate<br>the clip components from the<br>release part, and the release<br>part then exits the human<br>digestive tract. The clip<br>remained in the alimentary<br>tract for about 1-2 weeks, fell<br>off naturally and was<br>discharged through the<br>intestine and anus. | Different.<br>Our<br>proposed<br>device 555<br>family could<br>repeatedly<br>connect and<br>release clip. |
| | Configuration | For 55501 and 55502 series,<br>consists of release part and clip<br>part;<br>For 55504 and 55505 series,<br>there's additional clips for<br>change; For 55503 series, it's<br>the separate clip part. | Consists of release part and<br>clip part; | Different.<br>Our<br>proposed<br>device's<br>release part<br>could<br>connect and<br>release the<br>clips<br>repeatedly. |
| | Opening span | 9 mm,11 mm,13 mm,16<br>mm,18mm | 9mm, 11mm, 13mm,16mm | Different.<br>We add<br>18mm<br>opening span<br>for more<br>options for<br>doctors. |
| | Working<br>Length | 1650mm, 1950mm, 2150mm,<br>2300mm | 1650mm, 1950mm, 2300mm | Similar. Our<br>proposed<br>device |
| | Item | Proposed device<br>Hangzhou AGS MedTech Co.,<br>Ltd. | Predicate device (K172727)<br>Hangzhou AGS MedTech Co.,<br>Ltd. | Comparion |
| | Connect and<br>release the clip<br>repeatedly<br>performance | Yes | No | Hemoclip<br>have more<br>working<br>length.<br>Different.<br>Our<br>proposed<br>device have<br>connect and<br>release the<br>clip<br>repeatedly<br>performance. |
#### Тэы teristic fr al ch 560 -ic hnic
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a logo for AGS Medtech. The logo consists of a blue and green circular design on the left, with the text "AGS MEDTECH" in black on the right. Below the English text, there is Chinese text in black. The logo appears to be for a medical technology company.
Section 5 510(k) Summary Hemoclip
# 5.7 Applicable Guidance Document
NA
# 5.8 Performance Data
The Hemoclip meets all design specifications and medical device standards for shelf life validation (ISO 11607-1/2 and ASTM F 1980-16), packaging validation (ISO 11607-1/2), biocompatibility (ISO 10993) and sterility (ISO 11135). The non-clinical performance meets the design specification and shows substantial equivalence to the predicated device.
## 5.9 Clinical Test
No Clinical test is included in this submission.
# 5.10 Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Hangzhou AGS MedTech Co., Ltd has demonstrated that proposed device Hemoclip is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.