The Lockado™ Repositionable Hemostasis Clip is indicated for Endoscopic clip placement within the Gastrointestinal tract in adult populations only via a straight or side viewing flexible endoscope for the purpose of : (1) Endoscopic marking; (2) Hemostasis for (a) Mucosal / sub-mucosal defects < 3cm, (b) Bleeding ulcers, (c) polyps < 1.5cm in diameter, (d) diverticula in the colon, (e) Arteries < 2 mm, (f) Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection. (3) as a supplementary method, closure of Gl tract luminal perforations 20mm that can be treated conservatively. (4) Anchoring to affix jejunal feeding tubes to the wall of the small bowel;
Device Story
Sterile, single-use endoscopic clipping device; delivery system and clip assembly; used via straight or side-viewing flexible endoscope. Physician-operated in clinical settings. Device enables mechanical closure of tissue defects, hemostasis, marking, or tube anchoring. Repositionable design allows adjustment before final deployment. Output is physical tissue approximation; aids clinical decision-making by providing immediate mechanical intervention for bleeding or perforation. Benefits include reduced risk of delayed bleeding and minimally invasive closure of defects.
Clinical Evidence
Bench testing only. No clinical or animal studies were performed. Performance testing included dimension, release force, clamping strength, tensile strength, rotation, scope compatibility, and mechanical integrity. Biocompatibility testing followed ISO 10993-1; sterilization validated per ISO 11135:2014+A1:2018; MRI safety evaluated per ASTM F2503, F2052, F2119, F2182, and F2213.
Technological Characteristics
Single-use endoscopic clip assembly and delivery system. Compatible with endoscopes having a minimum 2.8mm working channel. Materials conform to ISO 10993-1. Sterilization via Ethylene Oxide (SAL 10^-6). MRI conditional per ASTM F2503, F2052, F2119, F2182, F2213. Available in multiple open widths (8mm, 11mm, 16mm, 22mm) and working lengths (1650mm, 1950mm, 2350mm, 2700mm).
Indications for Use
Indicated for adult patients requiring endoscopic clip placement in the GI tract for marking, hemostasis (mucosal/submucosal defects <3cm, bleeding ulcers, polyps <1.5cm, colonic diverticula, arteries <2mm, prophylactic post-resection bleeding), closure of GI luminal perforations <20mm, or anchoring jejunal feeding tubes.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 14, 2021
Micro-Tech (Nanjing) Co., Ltd. Sally He RA Engineer No.10 Gaoke Third Road Nanjing, Jiangsu 210032 CHINA
Re: K202333
> Trade/Device Name: Lockado™ Repositionable Hemostasis Clip Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal ligator Regulatory Class: Class II Product Code: PKL Dated: November 26, 2020 Received: November 30, 2020
Dear Sally He:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls' provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
#### K202333
Device Name
LockadoTM Repositionable Hemostasis Clip
#### Indications for Use (Describe)
The Lockado™ Repositionable Hemostasis Clip is indicated for Endoscopic clip placement within the Gastrointestinal tract in adult populations only via a straight or side viewing flexible endoscope for the purpose of :
- (1) Endoscopic marking;
- (2) Hemostasis for
- (a) Mucosal / sub-mucosal defects < 3cm,
- (b) Bleeding ulcers,
- (c) polyps < 1.5cm in diameter,
- (d) diverticula in the colon,
- (e) Arteries < 2 mm, `
- (f) Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection.
- (3) as a supplementary method, closure of Gl tract luminal perforations 20mm that can be treated conservatively.
- (4) Anchoring to affix jejunal feeding tubes to the wall of the small bowel;
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
{3}------------------------------------------------
K202333 Page 1 of 6
Image /page/3/Picture/1 description: The image shows a blue and white logo with the letters "MT" in a stylized, three-dimensional design. The "M" is on the left and the "T" is on the right, with the "M" appearing slightly in front of the "T". A registered trademark symbol is located in the upper right corner of the logo. The logo has a modern and corporate feel.
#### Section 5 510K Summary
# 510K Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number: K202333
# 1. Date of Preparation: 2020-09-17
# 2. Sponsor Identification
# Micro-Tech (Nanjing) Co., Ltd.
No.10 Gaoke Third Road, Nanjing National Hi-Tech, Industrial Development Zone, Nanjing,
Jiangsu Province, PRC
## Establishment Registration Number: 3004837686
Contact Person: Sally He
Position: RA Engineer
Tel: +86-25-58646395
Fax: +86-25-58350006
Email: RA.Micro-Tech@outlook.com
## 3. Identification of Proposed Device
Trade Name: Lockado™ Repositionable Hemostasis Clip
Common Name: Hemostasis Clip
## Regulatory Information
Classification Name: Hemorrhoidal ligator
Classification: II
Product Code: PKL
Regulation Number: 876.4400
Review Panel: Gastroenterology/Urology
## 4. Identification of Reference device
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional design. The letters are blocky and appear to be connected, with the "M" slightly overlapping the "T". A registered trademark symbol (®) is visible in the upper right corner of the logo. The logo has a clean and modern look.
## Reference Device
510(k) Number: K182556 Product Name: SureClip™ Repositionable Hemostasis Clip Manufacturer: Micro-Tech (Nanjing) Co., Ltd.
## 5. Identification of Predicate Device
#### Predicate Device
510(k) Number: K151802
Product Name: Resolution 360™ Clip
Manufacturer: Boston Scientific Corporation
## 6. Indications for Use
The Lockado™ Repositionable Hemostasis Clip is indicated for Endoscopic clip placement within the Gastrointestinal tract in adult populations only via a straight or side viewing flexible endoscope for the purpose of :
- (1) Endoscopic marking;
- (2) Hemostasis for
- (a) Mucosal / sub-mucosal defects < 3cm,
- (b) Bleeding ulcers,
- (c) Polyps < 1.5cm in diameter,
- (d) Diverticula in the colon,
- (e) Arteries < 2 mm,
- (f) Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection;
- (3) As a supplementary method, closure of GI tract luminal perforations <20mm that can be treated conservatively.
- (4) Anchoring to affix jejunal feeding tubes to the wall of the small bowel.
## 7. Device Description
The proposed device Lockado™ Repositionable Hemostasis Clip is a sterile, single-use
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are connected and appear to be part of a single, solid shape. A registered trademark symbol (®) is located in the upper right corner of the logo.
Section 5 510K Summary
endoscopic clipping device in adult patients only via a straight or side viewing flexible endoscope, intended to be used for endoscopic marking, hemostasis for mucosal/submucosal defects in digestive tract.
lt consists of two main components, delivery system and clip assembly. And it is offered in different dimensions.
# 8. Comparison of Technological Characteristics
The Lockado™ Repositionable Hemostasis Clip incorporates substantially equivalent device materials, design, configuration, packaging fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the predicate device Resolution 360™ Clip under K151802.
| Item | Proposed Device<br>Lockado™ Repositionable<br>Hemostasis Clip | Predicate Device<br>Resolution 360™ Clip (K151802) | Remark |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product Code | PKL | PKL | Same |
| Regulation No. | 876.4400 | 876.4400 | Same |
| Class | II | II | Same |
| Supplied in Sterile | Yes | Yes | Same |
| Configuration | Delivery system and clip assembly | Delivery system and clip assembly | Same |
| Open width | 8mm, 11mm and 16mm, 22mm | 11mm | Similar |
| Minimal working<br>channel of<br>endoscopy | 2.8mm | 2.8mm | Same |
| Working Length | 1650mm, 1950mm, 2350mm,<br>2700mm | 2350mm | Similar |
| Indications for Use | The Lockado™ Repositionable<br>Hemostasis Clip is indicated for<br>Endoscopic clip placement within the<br>Gastrointestinal tract in adult patients<br>only via a straight or side viewing<br>flexible endoscope for the purpose<br>of :<br>(1) Endoscopic marking: | The Resolution 360™ Clip is<br>indicated for clip placement within<br>the gastrointestinal (GI) tract for the<br>purpose of:<br>1. Endoscopic marking<br>2. Hemostasis for<br>• Mucosal/sub-mucosal<br>defects < 3 cm | Similar |
| Item | Proposed Device<br>Lockado™ Repositionable<br>Hemostasis Clip | Predicate Device<br>Resolution 360™ Clip (K151802) | Remark |
| | (2) Hemostasis for<br>(a) mucosal / sub-mucosal defects < 3cm,<br>(b) bleeding ulcers,<br>(c) polyps < 1.5cm in diameter,<br>(d) diverticula in the colon,<br>(e) Arteries < 2 mm,<br>(f) Prophylactic clipping to reduce the<br>risk of delayed bleeding post lesion<br>resection;<br>(3) As a supplementary method,<br>closure of GI tract luminal<br>perforations <20mm that can be<br>treated conservatively.<br>(4) Anchoring to affix jejunal feeding<br>tubes to the wall of the small bowel. | Bleeding ulcers Arteries < 2 mm` Polys < 1.5 cm in diameter Diverticula in the colon Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection 3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel; and Anchoring to affix fully covered esophageal self-expanding metal stents to the wall of the esophagus<br>4. As a supplemental closure method of luminal perforations < 20 mm that can be treated conservatively | |
| Single Use | Yes | Yes | Same |
| Packaging | Single-use EO sterilized pouch with<br>one device per pouch | Single-use EO sterilized pouch<br>with one device per pouch | Same |
| Shelf Life | Two years | Three years | Different |
| Biocompatibility | Conform to ISO 10993-1 | Conform to ISO 10993-1 | Same |
| Sterilization | EO Sterilized, SAL:10-6 | EO Sterilized, SAL:10-6 | Same |
| Labeling | Conform to 21 CFR part 801 | Conform to 21 CFR part 801 | Same |
| MRI information | Comply with ASTM F 2503, ASTM<br>F 2052, ASTM F2119, ASTM<br>F2182, ASTM F2213 | Comply with ASTM F 2503, ASTM<br>F 2052, ASTM F2119, ASTM<br>F2182, ASTM F2213 | Same |
# Comparison to predicate Devices:
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are interconnected, with the "M" appearing to be in front of the "T". A registered trademark symbol (®) is located in the upper right corner of the logo. The logo has a clean and modern design.
Section 5 510K Summary
# 9. Performance Data
Performance testing was conducted to demonstrate the essential performance of the
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional design. The letters are bold and connected, with the "M" appearing to be in front of the "T". A registered trademark symbol (®) is positioned in the upper right corner of the logo. The logo has a clean and modern appearance.
#### proposed device Lockado™ Repositionable Hemostasis Clip and confirmed that the
proposed device works as intended with the compatible devices.
The bench tests below were tested and evaluated as substantially equivalent to the predicate device.
- > Dimension;
- > Release Force;
- > Clamping Strength;
- > Tensile Strength;
- Clip Assembly Repeated Open/Close; >
- > Clip Open And Close Force
- > Rotation:
- > Scope Compatibility/Usability;
- > Endoscope Damage;
- > Torque;
- > Biopsy Valve Compatibility;
- > Clip Approach;
- Coil to Handle Tensile >
- > Mechanical Integrity of the Clip Assembly
- > MR Conditional Safety
Shelf-life testing was conducted based on an accelerated aging test in accordance with ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. Two-year aging test will be performed to demonstrate longer stability and support the results of the accelerated aging test.
Sterilization validation was carried out in accordance with ISO 11135:2014+A1:2018 "Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices".
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are connected and appear to be made of solid blocks. A registered trademark symbol is located in the upper right corner of the logo.
Biocompatibility testing was performed in accordance with the FDA Guidance, Use of International Standard ISO-10993-1, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process" issued on June 16,2016.
#### 10. Animal Study
No animal study is included in this submission.
#### 11. Clinical Study
No clinical study is included in this submission.
#### 12. Substantially Equivalent (SE) Conclusion
The nonclinical tests demonstrate that the device Lockado™ Repositionable Hemostasis Clip is as safe, as effective, and performs as well as or better than the legally marketed device.
Based on the indications for use, technological characteristics, and safety and performance testing, the Lockado™ Repositionable Hemostasis Clip has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the currently cleared predicate device Resolution 360™ Clip (K151802).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.