CHENA-C Spacer System

K213115 · Kahtnu Surgical, Inc. · ODP · Dec 6, 2021 · Orthopedic

Device Facts

Record IDK213115
Device NameCHENA-C Spacer System
ApplicantKahtnu Surgical, Inc.
Product CodeODP · Orthopedic
Decision DateDec 6, 2021
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The CHENA-C Spacer System is an anterior cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) with accompanying radicular symptoms at one level or two contiguous levels from C2-TI. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The CHENA-C Spacer interior should be packed with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and implanted via an anterior approach. The CHENA-C Spacer is intended to be used with supplemental fixation.

Device Story

The CHENA-C Spacer System is an intervertebral body fusion device used in spinal surgery. It consists of cervical interbody spacer implants available in various footprints and heights. The device is implanted via an anterior approach into the cervical spine to assist in interbody fusion. The spacer interior is packed with autogenous or allogenic bone graft to promote fusion. The system is intended for use in conjunction with supplemental fixation. It is used by surgeons in a clinical/OR setting to treat degenerative disc disease. The device provides structural support to the intervertebral space, facilitating bone growth and stabilization of the spinal segment, which helps alleviate radicular symptoms and discogenic pain.

Clinical Evidence

No clinical data. Bench testing only; mechanical evaluation confirmed the subject device does not present a new worst-case scenario compared to the predicate devices.

Technological Characteristics

Intervertebral body fusion device. Materials: Polyetheretherketone (ASTM F2026-17), tantalum (ASTM F560-17), and titanium alloy (ASTM F136). Form factor: Cervical cages with varying footprints and heights. Mechanical stabilization via anterior approach.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) and radicular symptoms at one or two contiguous levels (C2-T1) who failed six weeks of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION". December 6, 2021 Kahtnu Surgical. Inc % Ann Dunahoo Principal Regulatory And Quality Consultant MRC Global, LLC 9085 E. Mineral Circle, Suite 110 Centennial, Colorado 80112 Re: K213115 Trade/Device Name: CHENA-C Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP Dated: November 10, 2021 Received: November 12, 2021 Dear Ms. Dunahoo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ ## K213115 - Ms. Ann Dunahoo requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K213115 Device Name CHENA-C Spacer System #### Indications for Use (Describe) The CHENA-C Spacer System is an anterior cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) with accompanying radicular symptoms at one level or two contiguous levels from C2-TI. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The CHENA-C Spacer interior should be packed with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and implanted via an anterior approach. The CHENA-C Spacer is intended to be used with supplemental fixation. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## K213115 #### 510(k) Summary CHENA-C Spacer System September 23, 2021 | Company: | Kahtnu Surgical, Inc.<br>170 E. Corral Ave. Suite 1<br>Soldotna, AK 99669 | |-------------------------------|------------------------------------------------------------------------------------------------------------| | Primary Contact: | Ann Dunahoo<br>Principal Quality and Regulatory Consultant<br>901-299-9390<br>ann.dunahoo@askmrcglobal.com | | Company/Secondary<br>Contact: | Craig Wilcox<br>President and CEO<br>907-202-3111<br>craig.wilcox@kahtnu.com | | Trade Name: | CHENA-C Spacer System | | Common Name: | Intervertebral Fusion Device With Bone Graft, Cervical | | Classification: | Class II | | Regulation Number: | 21 CFR 888.3080 (Intervertebral body fusion device) | | Panel: | 87-Orthopedic | | Product Code: | ODP | #### Device Description: The Kahtnu (previously Axis Orthopaedics) Chena-C Spacer System is consists of a variety of footprints and heights of cervical interbody spacer implants to assist in interbody fusion. The previously cleared Axis Chena Cervical PEEK Spacer System (K181140) implant components are fabricated from medical implant grade Polyetheretherketone and tantalum per ASTM F2026-17 and ASTM F560-17. The purpose of this Special 510(k) submission is to add titanium alloy (ASTM F136) cervical cages to the Chena-C Spacer System. ### Indications for Use: The CHENA-C Spacer System is an anterior cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) with accompanying radicular symptoms at one level or two contiguous levels from C2-Tl. DDD is defined as discogenic pain > K213115 510k Summary Page 1 of 2 {4}------------------------------------------------ with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The CHENA-C Spacer interior should be packed with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft and implanted via an anterior approach. The CHENA-C Spacer is intended to be used with supplemental fixation. ## Substantial Equivalence: The subject Chena-C Spacer System is substantially equivalent to the following predicate devices: - . Primary Predicate: Axis Orthopedics – Chena Cervical PEEK Spacer System (K181140; S.E. 11/16/2018) - . Secondary Predicate: Alliance Spine, Alamo® C - K173128 (S.E. 3/22/2018) #### . Reference Devices: Axis 5.5 Lumbar Pedicle Screw System (Now Axis Talkeetna) – K180301 (S.E. 4/10/2018) DePuy Synthes ACIS - K120275 (S.E. 5/25/2012) Zimmer Vista®-S Device – K133784 (S.E. 7/7/2014) Stryker AVS® AS PEEK Spacer – K142251 (S.E. 11/19/2014) The Indications for Use are identical between the subject and primary predicate devices. The Materials are identical to the secondary predicate device as well as the reference devices. While the subject device does contain a slight change in geometry to incorporate a larger graft window, it has been shown to be mechanically equivalent to the primary predicate device. Thus, it can be concluded that the subject does not raise new questions about safety and effectiveness. # Performance Testing: Mechanical testing was performed on the predicate Chena Cervical PEEK Spacers. The subject devices have been evaluated and the results of evaluation have shown that no new worst case is presented compared to the predicate devices. Therefore, no testing is required # Conclusion Based on the test results and the comparison to the predicate devices, the subject device is determined to be substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...