SYNTHES ACIS/VERTEBRAL SPACER CR

K120275 · Synthes (Usa), LLC · ODP · May 25, 2012 · Orthopedic

Device Facts

Record IDK120275
Device NameSYNTHES ACIS/VERTEBRAL SPACER CR
ApplicantSynthes (Usa), LLC
Product CodeODP · Orthopedic
Decision DateMay 25, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Synthes ACIS is an anterior cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) with accompanying radicular symptoms at one level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The interior of the Synthes ACIS should be packed with autogenous bone graft and implanted via an anterior approach. The Synthes ACIS is intended to be used with supplemental fixation.

Device Story

Synthes ACIS and Vertebral Spacer CR are radiolucent cervical intervertebral body fusion devices; used for spinal stabilization in patients with degenerative disc disease. Devices are implanted via anterior approach; interior packed with autogenous bone graft to facilitate fusion. Implants feature PEEK-Optima LT-1 body with titanium alloy radiopaque markers for intra-operative positioning assessment. Vertebral Spacer CR includes pyramidal teeth on superior/inferior surfaces for construct stabilization. Devices are intended for use with supplemental fixation. Operated by surgeons in clinical/OR settings. Output is physical structural support for vertebral bodies; clinical benefit is stabilization of spinal segment to promote fusion and alleviate radicular symptoms.

Clinical Evidence

No clinical data provided; substantial equivalence established via bench testing only.

Technological Characteristics

Materials: Invibio PEEK-Optima LT-1 (ASTM F2026-10) and titanium alloy (Ti-6Al-4V ELI ASTM F136-2a or Ti-6Al-7Nb ASTM F1295-05). Features: Radiolucent body with radiopaque markers, open architecture for bone graft, pyramidal teeth (Spacer CR). Energy source: None (mechanical). Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) and radicular symptoms at one level from C2-T1, following six weeks of failed non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAY 2 5 2012 #### 6 510(k) Summary | 510(k) Summary –Synthes ACIS | | |-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Name of Firm: | Synthes Spine<br>1302 Wrights Lane East<br>West Chester, PA 19380 | | 510(k) Contact: | Heather Guerin, Ph.D., P.E.<br>Senior Regulatory Affairs Specialist<br>Telephone: 610-719-5432 Facsimile: 610-719-5102<br>Email: guerin.heather@synthes.com | | Date Prepared: | January 27, 2012 | | Trade Name(s): | Synthes ACIS | | Classification: | 21 CFR 888.3080 – Intervertebral fusion device<br>Class II (special controls)<br>Orthopaedic and Rehabilitation Devices Panel (87)<br>Product Code ODP (Intervertebral Fusion Device with Bone Graft,<br>Cervical) | | Predicates: | Synthes Zero-P device (K072981)<br>NuVasive CoRoent System (K081611)<br>BAK Cervical Interbody Fusion System (P980048)<br>Medtronic Cornerstone PSR (K100214)<br>Medtronic Prevail (K073285)<br>Globus PATRIOTTM Spacers (ColonialTM ACDF) (K072991) | | Device<br>Description(s): | The Synthes ACIS is a radiolucent cervical intervertebral body fusion<br>device. The Synthes ACIS is fabricated from Invibio® PEEK-Optima®<br>LT-1 (ASTM F2026-10) with two anterior and one posterior titanium<br>alloy (Ti-6Al-4V ELI; TAV; ASTM F136-2a) radiopaque markers. The<br>markers allow intra-operative radiographic assessment of the position of<br>the implant. | | Intended Use/<br>Indications for<br>Use: | The Synthes ACIS is an anterior cervical interbody fusion device<br>indicated for use in skeletally mature patients with degenerative disc<br>disease (DDD) with accompanying radicular symptoms at one level from<br>C2-T1. DDD is defined as discogenic pain with degeneration of the disc<br>confirmed by history and radiographic studies. These patients should<br>have had six weeks of non-operative treatment. The interior of the<br>Synthes ACIS should be packed with autogenous bone graft and<br>implanted via an anterior approach. The Synthes ACIS is intended to be<br>used with supplemental fixation. | | Comparison of | Synthes ACIS is substantially equivalent to the predicates in design, | | the device to<br>predicate<br>device(s): | function, performance, material, and intended use. | | Performance<br>Date<br>(Non-Clinical<br>and/or Clinical): | Non-Clinical Performance and Conclusions:<br>Synthes conducted the following bench testing (as recommended within<br>FDA Guidance and in accordance with ASTM F2077-03 and ASTM<br>F2267-04): Static Axial Compression; Dynamic Axial Compression;<br>Static Compression Shear; Dynamic Compression Shear; Static Torsion;<br>Dynamic Torsion; and Subsidence.<br>The conclusions drawn from testing demonstrate that the Synthes ACIS<br>device is substantially equivalent in performance to predicate devices.<br><br>Clinical Performance and Conclusions:<br>Clinical data and conclusions were not needed for this device. | {1}------------------------------------------------ the submit the state of the submit of the subject of and the comments of the comments of the comments of the comments of and the control control control control controlled in the controlled - - - - and the comments of the comments of the comments of the comments of and the control control control control controllers and {2}------------------------------------------------ . | 510(k) Summary –Synthes Vertebral Spacer CR | | |-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Name of Firm: | Synthes Spine<br>1302 Wrights Lane East<br>West Chester, PA 19380 | | 510(k) Contact: | Heather Guerin, Ph.D., P.E.<br>Senior Regulatory Affairs Specialist<br>Telephone: 610-719-5432 Facsimile: 610-719-5102<br>Email: guerin.heather@synthes.com | | Date Prepared: | January 27, 2012 | | Trade Name(s): | Synthes Vertebral Spacer CR | | Classification: | 21 CFR 888.3080 – Intervertebral fusion device<br>Class II (special controls)<br>Orthopaedic and Rehabilitation Devices Panel (87)<br>Product Code ODP (Intervertebral Fusion Device with Bone Graft, Cervical) | | Predicates: | Synthes Zero-P device (K072981);<br>NuVasive CoRoent System (K081611);<br>BAK Cervical Interbody Fusion System (P980048);<br>Medtronic Cornerstone PSR (K100214);<br>Medtronic Prevail (K073285);<br>Globus PATRIOTTM Spacers (ColonialTM ACDF) (K072991) | | Device<br>Description(s): | The Synthes Vertebral Spacer CR is a radiolucent cervical intervertebral<br>body fusion device. The Vertebral Spacer CR is fabricated from Invibio®<br>PEEK-Optima® LT-1 (ASTM F2026-10) with two anterior and one<br>posterior titanium alloy (Titanium-6Aluminum-7Niobium ASTM F1295-<br>05) radiopaque markers. The markers allow intra-operative radiographic<br>assessment of the position of the implant.<br><br>The Synthes Vertebral Spacer CR is available in a range of heights, and is<br>available in three different sagittal profiles. Pyramidal teeth that assist in<br>stabilization of the construct are located on the inferior and superior<br>surfaces of the spacers. The open architecture of the device allows it to be<br>packed with autogenous bone graft material, i.e., autograft. | | Intended Use/<br>Indications for<br>Use: | The Synthes Vertebral Spacer CR is an anterior cervical interbody fusion<br>device indicated for use in skeletally mature patients with degenerative<br>disc disease (DDD) with accompanying radicular symptoms at one level<br>from C2-T1. DDD is defined as discogenic pain with degeneration of the<br>disc confirmed by history and radiographic studies. These patients should<br>have had six weeks of non-operative treatment. The interior of the<br>Synthes Vertebral Spacer CR should be packed with autogenous bone<br>graft and implanted via an anterior approach. The Synthes Vertebral<br>Spacer CR is intended to be used with supplemental fixation. | | Comparison of<br>the device to<br>predicate | Synthes Vertebral Spacer CR is substantially equivalent to the predicates<br>in design, function, performance, material, and intended use. | | 510(k) Summary -Synthes Vertebral Spacer CR | | | device(s): | | | Performance<br>Date<br>(Non-Clinical<br>and/or Clinical): | Non-Clinical Performance and Conclusions:<br>Synthes conducted the following bench testing (as recommended within<br>FDA Guidance and in accordance with ASTM F2077-03 and ASTM<br>F2267-04): Static Axial Compression; Dynamic Axial Compression;<br>Static Compression Shear; Dynamic Compression Shear; Static Torsion;<br>Dynamic Torsion; and Subsidence.<br><br>The conclusions drawn from testing demonstrate that the Synthes<br>Vertebral Spacer CR device is substantially equivalent in performance to<br>predicate devices. | | | Clinical Performance and Conclusions:<br>Clinical data and conclusions were not needed for this device. | · {3}------------------------------------------------ . . . . . . . . . . . . . . . : . . . . . . . . . バ {4}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the caduceus symbol. ### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring. MD 20993-0002 Synthes Spine % Heather Guerin, Ph.D., P.E. Senior Regulatory Affairs Specialist 1302 Wrights Lane East West Chester, Pennsylvania 19380 MAY 2 5 2012 Re: K120275 Trade/Device Name: ACIS Spacer and Vertebral Spacer CR Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP Dated: April 24, 2012 Received: April 26, 2012 Dear Dr. Guerin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {5}------------------------------------------------ Page 2 - Heather Guerin, Ph.D., P.E. forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor_You/Industry/default.htm. Sincerely yours, Mark M Mellerson Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ #### Indications for Use Statement 5 510(k) Number: (if known) K 120275 Device Name: Synthes ACIS Indications for Use: The Synthes ACIS is an anterior cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) with accompanying radicular symptoms at one level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of nonoperative treatment. The interior of the spacer of the Synthes ACIS should be packed with autogenous bone graft and implanted via an anterior approach. The Synthes ACIS is intended to be used with supplemental fixation. Prescription Use X (21 CFR 801 Subpart D) AND / OR Over-the-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Vivision of Surgical, Orthopedic, and Restorative Devices KI20275 510(k) Number_ Page 1/2 {7}------------------------------------------------ 510(k) Number: (if known) ## K 120275 Device Name: Synthes Vertebral Spacer CR Indications for Use: The Synthes Vertebral Spacer CR is an anterior cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) with accompanying radicular symptoms at one level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six weeks of non-operative treatment. The interior of the Synthes Vertebral Spacer CR should be packed with autogenous bone graft and implanted via an anterior approach. The Synthes Vertebral Spacer CR is intended to be used with supplemental fixation. Prescription Use X (21 CFR 801 Subpart D) AND / OR Over-the-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) *(Division Sign-Off)* Digision of Surgical, Orthopedic, and Restorative Devices K120275 510(k) Number_ Page 2/2 ರ್ಥಿ
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...