K213010 · Feeltech Co., Ltd. · QNQ · May 31, 2023 · General Hospital
Device Facts
Record ID
K213010
Device Name
FEELject LDV (Low dead volume) syringe
Applicant
Feeltech Co., Ltd.
Product Code
QNQ · General Hospital
Decision Date
May 31, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
FEELject LDV(Low dead volume) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.
Device Story
FEELject LDV is a single-use, manually operated disposable syringe with a permanently attached hypodermic needle. Device consists of a calibrated plastic barrel, plunger, rubber piston, and needle. Operates by manual movement of the plunger within the barrel to draw in or expel fluids through the needle. Designed as a low dead volume (LDV) syringe (≤ 0.03 mL) to minimize fluid loss. Used by healthcare professionals or patients in clinical or home settings for subcutaneous, intramuscular, or intravenous fluid administration/withdrawal. Output is the delivery or extraction of fluid. Benefits include reduced medication waste due to low dead space design.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993 series: cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogenicity, hemolysis), sterility (ISO 11737-2), EO residuals (ISO 10993-7), and mechanical performance (ISO 7886-1, ISO 7864, ISO 9626, USP 788). All tests passed.
Technological Characteristics
Materials: Polypropylene (barrel, plunger, hub, cap), Rubber (piston), SUS304 (needle). Manual operation. Single-use. E.O. gas sterilization (SAL 10^-6). Low dead volume (≤ 0.03 mL). 1 mL capacity. Complies with ISO 7886-1, ISO 7864, ISO 9626, and ISO 10993 standards.
Indications for Use
Indicated for injection of fluids into or withdrawal of fluids from parts of the body below the skin surface. No specific age or gender restrictions provided.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 31, 2023
Feeltech Co., Ltd. % Peter Chung President Plus Global 300. Atwood Pittsburgh, Pennsylvania 15213
Re: K213010
Trade/Device Name: FEELject LDV (Low dead volume) syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: QNQ, FMI Dated: April 28, 2023 Received: May 1, 2023
Dear Peter Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alan Guth
CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K213010
Device Name FEELject LDV(Low dead volume) Syringe
Indications for Use (Describe)
FEELject LDV(Low dead volume) Syringe is intended for use to inject fluid into or withdraw fluids from the parts of body below the surface of the skin.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K213010 - 510(k) Summarv
#### [as required by 21 CFR 807.92(c)]
#### 1. Applicant
- Company: Feeltech Co., Ltd. 1)
- Address: 3, 4 Floor, Standard Factory 2-dong, 15, Jayumuyeok2-gil, Gunsan-si, Jeollabuk-do, Korea 2)
- Tel : +82-63-468-6626~8 3)
- Fax : +82-63-468-6623 4)
- Prepared date : May 18, 2023 ર)
- 6) Contact person : Peter Chung, 412-512-8802
- 7) Contact person address : 300 Atwood Street, Pittsburgh, PA, 15213, USA
#### 2. Subject Device Information
- 1) Trade name: FEELject LDV(Low dead volume) Syringe
- Common name : Disposal syringe with needle 2)
- 3) Classification name : Piston syringe
- 4) Product code : QNQ, FMI
- Regulation number : 21 CFR 880.5860 ર)
- Class of device : Class II 6)
- Panel : General hospital 7)
# 3. The legally marketed device to which we are claiming equivalence
Syringe, K192551, Jiangsu Caina Medical Co., Ltd.
#### 4. Device description
A Disposable Syringe with Needle is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a hypodermic single lumen needle which is permanently attached. The device is used to inject fluids into, or withdraw fluids from, the body. It is made of plastic and silicone materials that allows for smooth plunger movement, and manually operated. This is a single-use device. This product is packed by sterile paper and sterilized by E.O gas.
#### 5. Intended Use:
This product is intended for use to inject fluids into or withdraw fluids from parts of the body below the surface of the skin.
#### 6. Comparison Table
| Table 1. Comparison table of proposed to predicate device | | | | |
|-----------------------------------------------------------|--------------|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-----------------|
| | Manufacturer | FeelTech Co., Ltd. | Jiangsu Caina Medical Co., Ltd. | Remark |
| | Item | Proposed device | Predicate device | |
| | 510(K) No. | K213010 | K192551 | N/A |
| | Product | FEELject LDV Syringe | Syringe with permanently<br>attached needle | |
| | Intended use | This product is intended for use to inject<br>fluids into or withdraw from parts of the<br>body below the surface of the skin | This device is intended for use to inject<br>fluids into or withdraw from parts of the<br>body below the surface of the skin | Same |
| | Components | Barrel, Plunger, Plunger rod<br>Needle, needle cap | Barrel, Plunger, Plunger rod<br>Needle, needle cap | Same |
| | Nozzle type | Permanently attached | Permanently attached<br>Luer-slip/Luer-lock | Different<br>#1 |
| Material | Barrel | Polypropylene (PP) | Polypropylene (PP) | |
| | Plunger | Polypropylene (PP) | Polypropylene (PP) | |
| | Piston | Rubber | Rubber | Same |
| | Needle hub | Polypropylene (PP) | Polypropylene (PP) | |
| | Needle | SUS304 | SUS304 | |
| | Needle cap | Polypropylene (PP) | Polypropylene (PP) | |
| | Capacity | 1ml | 0.3, 0.5, 1ml | Different<br>#2 |
| | Dead space | Low dead volume (≤ 0.03 mL) | Unknown | Different<br>#3 |
## Table 1. Comparison table of proposed to predicate device
{4}------------------------------------------------
| Principle of<br>operation | Manual | Manual | Same |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Syringe<br>Performances¹ | Complies with ISO 7886-1 : 2017 Sterile<br>hypodermic syringes for single use - Part<br>1 : Syringes for manual use | Complies with ISO 7886-1 : 2017 Sterile<br>hypodermic syringes for single use - Part<br>1 : Syringes for manual use | Same |
| Needle Length | 25.0, 25.4, 38.1mm | 8, 10, 13 16, 20, 25mm | |
| Gauge | 23, 25G | 21, 23, 25, 26, 27, 28, 29, 30, 31G | Different |
| Needle Tip<br>configuration | Bevel (11°±2°) | Bevel (11°±2°, 15°±2°) | #4 |
| Needle wall type | TW | RW, TW | |
| Needle<br>performance¹ | ISO 7864 : 2016 Sterile hypodermic needles<br>for single use - Requirements and test<br>methods<br>ISO 9626 : 2016 Stainless steel needle<br>tubing for the manufacture of medical<br>devices - Requirements and test methods | ISO 7864 : 2016 Sterile hypodermic needles<br>for single use - Requirements and test<br>methods<br>ISO 9626 : 2016 Stainless steel needle<br>tubing for the manufacture of medical<br>devices - Requirements and test methods | Same |
| | Conforms to the | Conforms to the | |
| Bio-<br>Compatibility¹ | requirements of ISO 10993<br>series standards.<br>Cytotoxicity<br>Acute systemic toxicity<br>Pyrogenicity<br>Sensitization<br>Irritation<br>Hemolysis<br>Intracutaneous reactivity<br>Bacterial Endotoxins | requirements of ISO 10993<br>series standards.<br>Cytotoxicity<br>Acute systemic toxicity<br>Pyrogenicity<br>Sensitization<br>Irritation<br>Hemolysis<br>Intracutaneous reactivity<br>Bacterial Endotoxins | Same |
| Principle of<br>operation | The plunger of syringe can be pulled and<br>pushed along inside the barrel, allowing the<br>syringe to take in and expel the fluids<br>through the connector to the patient. | The plunger of syringe can be pulled and<br>pushed along inside the barrel, allowing the<br>syringe to take in and expel the fluids<br>through the connector to the patient. | Same |
| Sterilization<br>method | E.O Gas sterilization | E.O Gas Sterilization | Same |
| SAL | 10-6 | 10-6 | Same |
| Endotoxin limit¹ | 20 EU / Device | 20 EU / Device | Same |
| Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Same |
* Note 1: Refer to the following section (8. Performance data) for the detail descriptions of conducted testing result.
#### 7. Equivalence Discussion
#### 1) Different #1 – Syringe nozzle type
Subject device only has permanently attached nozzle type. However, predicate device has not only permanently attached nozzle, but also luer-lock and luer-slip. While this configuration is different from each other, the subject device's nozzle configuration is included to the predicate device. Therefore, the differences on nozzle type does not raise new questions of safety and effectiveness.
- Different #2 Capacity 2)
Subject device has only one capacity configuration. (1 ml) Although predicate device has 3 configuration for the capacity (0.3, 0.5, 1 ml) this difference does not affect substantial equivalence because it is covered within the predicate submission.
#### 3) Different #3 - Dead space
Subject device is intended to provide low dead volume (LDV), minimizing the loss of fluid left inside the syringe. Although predicate device does not have this aspect, subject device is better than its requirements for the dead space. In accordance with product code 'QNQ', this device has proved with related performance testing that it conforms to the FDA's requirements. Therefore, the differences between predicate and subject device does not affect the device's effectiveness and safety.
- Different #4 Needle configurations (Needle length, gauge, tip configuration, wall type) The needle configurations (Needle length, gauge, tip configuration, wall type) for proposed devices is different from the predicate devices. However, the predicate device covers the range of needle dimensions.
{5}------------------------------------------------
Therefore, the differences on needle length and gauge does not affect safety and effectiveness.
#### 8. Performance data evaluation
1) Performance testing - Bench
Bench tests were performed. Bench testing included biocompatibility, mechanical testing, sterility testing including EO residues. The tests demonstrated that the device performs in a substantially equivalent manner to the predicate device. The following bench testing is performed to demonstrate the functionality is substantially equivalent.
| Requirement - Test (ISO 7886-1) | Result |
|----------------------------------------------------------|--------|
| Limits for extractable metals | Pass |
| General | Pass |
| Limits for acidity or alkalinity | Pass |
| Conical fitting | Pass |
| Position of nozzle on end of barrel | Pass |
| Tolerance on graduated capacity | Pass |
| Scale | Pass |
| Numbering of scales | Pass |
| Overall length of scale to nominal capacity line | Pass |
| Position of scale | Pass |
| Barrel flanges | Pass |
| Plunger stopper/plunger assembly | Pass |
| Dead space | Pass |
| Freedom from air and liquid leakage past plunger stopper | Pass |
| Force to operate the piston | Pass |
| Fit of plunger stopper/plunger in barrel | Pass |
| Unit packaging and self-contained syringe units | Pass |
| Quantity of silicone oil | Pass |
| Requirement - Test (ISO 7864) | Result |
|-------------------------------------------|--------|
| Cleanliness | Pass |
| Tolerances on length | Pass |
| Needle Point, needles with sharp tip only | Pass |
| Bond between hub and needle tube | Pass |
| Patency lumen | Pass |
| Requirement – Test (ISO 9626) | Result |
|-------------------------------|--------|
| Stiffness | Pass |
| Resistance to breakage | Pass |
| Resistance to corrosion | Pass |
| Requirement - Test (USP 788) | Result |
|-------------------------------|--------|
| Particulate matter injections | Pass |
#### 2) Biocompatibility
| # | Test item | Test method / Test criteria | Result |
|---|-----------------------------------|------------------------------------------------------------------------------------------------------------|--------|
| 1 | Cytotoxicity | ISO 10993-5 Tests for in vitro cytotoxicity | Pass |
| 2 | Skin<br>Sensitization<br>Test | ISO 10993-10 irritation and skin sensitization | Pass |
| 3 | Intracutaneous<br>Reactivity Test | ISO 10993-10 Test for irritation and skin sensitization, maximization test<br>for delayed hypersensitivity | Pass |
| 4 | Acute Systemic<br>Toxicity Test | ISO 10993-11 Test for systemic toxicity - Acute Systemic Toxicity | Pass |
| 5 | Pyrogen Test | ISO 10993-11 Tests for systemic toxicity, Annex(F) Information on<br>material-mediated pyrogens. | Pass |
| 6 | Hemolysis Test | ISO 10993-4 Selection of tests for interactions with blood | Pass |
3) Sterility and LAL test
| # | Test item | Test standard | Result |
|---|----------------|------------------------------------------------------------------------|--------|
| 1 | LAL test | USP39 <85>, Bacterial Endotoxins Test (Unit : EU/Device) | Pass |
| 2 | Sterility test | According to ISO 11737-2 | Pass |
| 3 | E.O Residual | Under the conditions of ISO 10993-7:2008, Ethylene oxide sterilization | Pass |
{6}------------------------------------------------
| | test | residuals, the test articles should meet the test requirements. | | |
|--|------|-----------------------------------------------------------------|--|--|
|--|------|-----------------------------------------------------------------|--|--|
Above conducted performance tests demonstrated that this device is performs in a substantially equivalent manner to the predicate device.
#### 9. Conclusion
The device has completed testing to show that the device meets its intended use and demonstrates substantial equivalence to the predicate device, K192551. Therefore, it is concluded that the subject device, FEELject LDV (Low dead volume) Syringe is substantially equivalent to the legally marketed predicate device, K192551.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.