BD Vacutainer UltraTouch Push Button Blood Collection Set
K212724 · Becton, Dickinson and Company · JKA · Feb 4, 2022 · Clinical Chemistry
Device Facts
Record ID
K212724
Device Name
BD Vacutainer UltraTouch Push Button Blood Collection Set
Applicant
Becton, Dickinson and Company
Product Code
JKA · Clinical Chemistry
Decision Date
Feb 4, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Indications for Use
The BD Vacutainer® UltraTouch™ Push Button Blood Collection Set is a sterile, multi-sample, single-use fixed winged blood collection set intended to be used by trained healthcare professionals for venipuncture to obtain blood specimens from patients, including those patients with difficult vein access who may have small, fragile, and/or non-palpable veins, into evacuated blood collection tubes and/or blood culture bottles. When used without the male adapter, the device allows the clinician to obtain a blood specimen from the female hub with a syringe, if necessary. The device can be used for short-term, single infusions with consideration given to patient size and appropriateness for the solution being infused. The device is not to be left in place and is to remain under the direct supervision of a clinician. The recommended use of the device is to activate the needle safety feature prior to removal from the venipuncture site. The retraction of the intravenous (IV) end of the needle aids in the prevention of accidental needlestick injury.
Device Story
Winged blood collection set; includes flexible tubing, female luer connector, male luer adapter, and needle safety mechanism. Used by trained healthcare professionals for venipuncture; collects blood into evacuated tubes or culture bottles; allows syringe collection via female hub; supports short-term, single infusions. Safety feature: push-button needle retraction prior to removal from venipuncture site to prevent needlestick injury. Available in various needle gauges (21G, 23G, 25G) and tubing lengths; some models include pre-attached holder. Device is single-use, sterile, and non-pyrogenic.
Clinical Evidence
No new clinical testing performed. Substantial equivalence supported by clinical literature, BD-sponsored studies, and investigator-sponsored studies confirming DVA patients are a subset of the general population and that the device risk profile remains unchanged.
Technological Characteristics
Winged set; materials: polyolefin wings, polypropylene hub/barrel, stainless steel 302 spring, stainless steel 304 cannula, PVC tubing. Sterilization: Gamma (SAL 10^-6). Standards: ISO 594-1/2 (luer), EN ISO 23908 (sharps protection), EN ISO 9626 (tubing), ISO 10993 (biocompatibility), ISO 11137 (sterilization), ISO 11607 (packaging).
Indications for Use
Indicated for use by trained healthcare professionals for venipuncture to obtain blood specimens from patients, including those with difficult vein access (small, fragile, or non-palpable veins), into evacuated tubes or blood culture bottles. May be used for short-term, single infusions.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Predicate Devices
BD Vacutainer® UltraTouch™ Push Button Blood Collection Set (K153309)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the U.S. Food and Drug Administration. The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white. Underneath the square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
August 31, 2022
Becton, Dickinson and Company Katherine Lemus Senior Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417
Re: K212724
Trade/Device Name: BD Vacutainer UltraTouch Push Button Blood Collection Set Regulation Number: 21 CFR 862.1675 Regulation Name: Blood specimen collection device Regulatory Class: Class II Product Code: JKA, FPA
Dear Katherine Lemus:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination for your device cleared on February 4, 2022. Specifically, FDA is updating this SE letter due to the clearance date not appearing on the original letter.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Payal Patel, OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices, 240-402-6029, payal.patel(@fda.hhs.gov.
Sincerely,
Pyd for
Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Becton, Dickinson and Company Katherine Lemus Senior Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417
# Re: K212724
Trade/Device Name: BD Vacutainer UltraTouch Push Button Blood Collection Set Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA, FPA Dated: December 29, 2021 Received: January 5, 2022
# Dear Katherine Lemus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cuffley
Gang Peng for Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
K212724
Device Name
BD Vacutainer® UltraTouch™ Push Button Blood Collection Set
Indications for Use (Describe)
The BD Vacutainer® UltraTouch™ Push Button Blood Collection Set is a sterile, multi-sample, single-use fixed winged blood collection set intended to be used by trained healthcare professionals for venipuncture to obtain blood specimens from patients, including those patients with difficult vein access who may have small, fragile, and/or non-palpable veins, into evacuated blood collection tubes and/or blood culture bottles. When used without the male adapter, the device allows the clinician to obtain a blood specimen from the female hub with a syringe, if necessary. The device can be used for short-term, single infusions with consideration given to patient size and appropriateness for the solution being infused. The device is not to be left in place and is to remain under the direct supervision of a clinician.
The recommended use of the device is to activate the needle safety feature prior to removal from the venipuncture site. The retraction of the intravenous (IV) end of the needle aids in the prevention of accidental needlestick injury.
Type of Use (Select one or both, as applicable) 区 Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
#### 510(K) SUMMARY K212724
## Summary Preparation Date:
2/4/2022
# Submitted by:
Becton, Dickinson and Company 1 Becton Drive Franklin Lakes, NJ 07417-1885
Phone: (201) 847-6800
# Contact:
Katherine Kenner Lemus, MS, RAC-US Senior Regulatory Affairs Specialist email: Katherine.lemus@bd.com
Phone: (801) 541-9274 Work: (201) 847 6800:
# Proprietary Names:
BD Vacutainer® UltraTouch™ Push Button Blood Collection Set:
## Common or Usual Names:
Blood Collection Set, IV Administration Set
## Regulatory Information
Classification Name: Blood Collection Set, IV Administration Set Classification Regulation: 21 CFR §862.6175 and 880.5440 Regulatory Class: Class II Product Code: JKA, FPA
## Predicate Device(s)
BD Vacutainer® UltraTouch™ Push Button Blood Collection Set (K153309)
## Device Description
The BD Vacutainer® UltraTouch™ Push Button Blood Collection Set is a winged blood collection set with flexible tubing, a female luer connector and a male luer adapter. The BD Vacutainer® UltraTouch™ Push Button Blood Collection Set also contains a needle protector.
{5}------------------------------------------------
The BD Vacutainer® UltraTouch™ Push Button Blood Collection Set models available consist of a combination of different needle gauges (21G, 23G, 25G), tubing length (12-inch, 7 inch), and are available with or without a pre-attached holder.
# Indications For Use
The proposed indications for use, are similar to the predicate indications for use, except for the inclusion of both healthy patients with vascular anatomy and/or co-morbidities which may cause difficulty in locating a vein. Healthy patients and patients with these conditions often exhibit small, fragile, non-palpable veins, commonly referred to as difficult vein access (DVA). The addition of DVA in the revised indications for use statement does not significantly change the intended use as butterfly needles have historically been used on patients with small veins and similar statements were included for previous device iterations and were inadvertently removed in subsequent submissions. BD believes that the addition of specific reference to patients with DVA does not raise new questions of safety/effectiveness.
{6}------------------------------------------------
# Substantial Equivalence1/Technical Comparison
The subject and predicate device are substantially equivalent as described in Table 2.
| Characteristic | Subject Device<br>BD Vacutainer® UltraTouch™<br>Push Button Blood Collection<br>Set | Predicate Device<br>BD Vacutainer® UltraTouch™<br>Push Button Blood Collection<br>Set<br>K153309 | Comparison |
|----------------------------------------|-------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Population | General use including patients<br>with difficult vein access (DVA) | General use | Different<br>Comment #1<br>Added specific<br>reference to patients<br>with DVA which is a<br>subset of 'General<br>Use' - no effect on<br>device safety or<br>effectiveness |
| Needle Diameter<br>OD | 21G, 23G and 25G | 21G, 23G and 25G | Same |
| Needle Diameter<br>ID | Ultra-Thin wall | Ultra-Thin wall | Same |
| Needle Point | 5 Bevel | 5 Bevel | Same |
| Needle Length | 3/4 inch | 3/4 inch | Same |
| Wing | Polyolefin | Polyolefin | Same |
| Hub | Polypropylene | Polypropylene | Same |
| Button Ink | UV Curable Ink | UV Curable Ink | Same |
| Front Barrel | Polypropylene | Polypropylene | Same |
| Rear Barrel | Acrylic | Acrylic | Same |
| Rear Barrel<br>Lubricant | Silicone | Silicone | Same |
| Rear Barrel<br>Lubricant Diluent | Isopropyl Alcohol | Isopropyl Alcohol | Same |
| Spring | Stainless Steel 302 | Stainless Steel 302 | Same |
| IV Protector<br>(Cannula<br>Protector) | Polyethylene | Polyethylene | Same |
| IV Cannula/NP<br>Cannula | Stainless Steel 304 | Stainless Steel 304 | Same |
| Tubing | Polyvinyl Chloride | Polyvinyl Chloride | Same |
Table 1: Substantial Equivalence Comparison
{7}------------------------------------------------
| Characteristic | Subject Device | Predicate Device | Comparison |
|--------------------------------------------|-------------------------------------------------------------------|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | BD Vacutainer® UltraTouch™<br>Push Button Blood Collection<br>Set | BD Vacutainer® UltraTouch™<br>Push Button Blood Collection<br>Set<br>K153309 | |
| Cannula<br>Lubricant | Silicone | Silicone | Same |
| Cannula Adhesive | UV cured adhesive | UV cured adhesive | Same |
| Hub-Tubing<br>Adhesive | UV cured adhesive | UV cured adhesive | Same |
| Female Luer<br>Connectors | Acrylonitrile Butadiene Styrene | Acrylonitrile Butadiene Styrene | Same |
| Luer Adapter Hub | Polypropylene | Polypropylene | Same |
| NP Sleeve | Synthetic Isoprene Rubber | Synthetic Isoprene Rubber | Same |
| Luer Adhesive | Heat Curing Epoxy | Heat Curing Epoxy | Same |
| Luer Cannula<br>Lubricant | Medical Grade Silicone | Medical Grade Silicone | Same |
| Luer Cap | Polypropylene | Polypropylene | Same |
| Top Web | Paper | Paper | Same |
| Blister | Polyethylene terephthalate –<br>glycol modified | Polyethylene terephthalate –<br>glycol modified | Same |
| Pre-attached<br>holder | Polypropylene | N/A | Different<br>Comment #2<br>6 model numbers<br>were added to<br>include devices<br>assembled with the<br>legally marketed<br>holder pre-attached<br>for user<br>convenience. This<br>non-significant<br>change does not<br>affect safety or<br>effectiveness |
| Materials | Compliant with ISO 10993<br>series | Compliant with ISO 10993<br>series | Same |
| Non-pyrogenic | Yes | Yes | Same |
| Non-toxic | Yes | Yes | Same |
| Sterile | Yes | Yes | Same |
| Sterility<br>Assurance Level<br>(SAL) 10-6 | Yes | Yes | Same |
{8}------------------------------------------------
| Characteristic | Subject Device<br>BD Vacutainer® UltraTouch™<br>Push Button Blood Collection<br>Set | Predicate Device<br>BD Vacutainer® UltraTouch™<br>Push Button Blood Collection<br>Set<br>K153309 | Comparison |
|-------------------------|----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization<br>Method | Gamma | Gamma | Same |
| Shelf Life | 2 years | 2 years | Same |
| Models | 367363<br>367364<br>367365<br>367391<br>367392<br>367393<br>368684<br>368685<br>368686<br>368687<br>368688<br>368689 | 367363<br>367364<br>367365<br>367391<br>367392<br>367393 | 6 model numbers<br>were added to<br>include devices<br>assembled with the<br>legally marketed<br>holder pre-attached<br>for user<br>convenience. This<br>non-significant<br>change does not<br>affect safety or<br>effectiveness |
# Substantial Equivalence Discussion:
# Comment #1
The subject device indicates that the intended population is for general use, including a subset of patients who may have difficult veins, whereas the predicate device population is general use. Both devices are used on all patient populations regardless of their vein status. Adding a reference to patients with difficult veins has no effect on clinical safety or effectiveness.
# Comment #2
The subject device is available in models with and without a pre-attached holder. The predicate device is only available in models without the pre-attached holder and requires the user to add a holder in order to use the device for blood collection. Adding models with the pre-attached holder to the subject device is a user convenience and does not affect clinical safety or effectiveness.
## Applicable Standards:
#### I. Performance Standards
ISO 594-1-1986 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements
ISO 594-2-1998 Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment -Part 2: Lock fittings
{9}------------------------------------------------
EN ISO 23908:2013 Sharps injury protection - Requirements and test methods - Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
EN ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices -Requirements and test methods
EN ISO 14971:2012 Medical Devices - Application of risk management to medical devices
#### II. Sterilization, Package Integrity, Shipping and Shelf Life
EN ISO 11137-1:2015/A2:2019 Sterilization 15of health care products - Radiation Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices
EN ISO 11137-2:2015 Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose
EN ISO 11137-3:2017 Sterilization of health care products - Radiation - Part 3: Guidance on dosimetric aspects of development, validation and routine control
EN ISO 11737-1:2018 Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on products
EN ISO 11737-2:2019 Sterilization of medical devices - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
ISO 11607-1:2019 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
ISO 11607-2:2019 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes
ASTM F2096-11 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
ASTM F88 Standard Test Method for Seal Strength of Flexible Barrier Materials
ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems
#### III. Biocompatibility Standards
ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
{10}------------------------------------------------
EN ISO 10993-2:2006 Biological evaluation of medical devices - Part 2: Animal welfare requirements
ISO 10993-4:2017 Biological evaluation of medical devices – Part 4: Selection of tests for interactions with blood
ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
ISO 10993-10:2010 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization
ISO 10993-11:2017 Biological evaluation of medical devices – Part 11: Tests for systemic toxicity
ISO 10993-12:2012 Biological evaluation of medical devices - Part 12: Sample preparation and reference materials
ISO 10993-17:2002 Biological evaluation of medical devices - Part 17: Establishment of allowable limits for leachable substances
ISO 10993-18:2020 Biological evaluation of medical devices – Part 18: Chemical characterization of materials
# Technical Performance Summarv
The sponsor has determined that since the subject and predicate blood collection sets are the same, and the BD Vacutainer® Brand Holder is unchanged from clearance under K181730, new performance testing is not required. Specifically, because there has been no significant change to design, product materials, packaging materials, processing/assembly, packaging, or manufacturing that would affect device performance no new performance testing is provided in support of this submission.
## Conclusion
The technical performance characteristics of the subject device are unchanged. The clinical literature, BD's sponsored studies, and Investigator sponsored studies sufficiently support the determination that patients with Difficult Vein Access (DVA) is a subset of the current patient population for the BD Vacutainer® UltraTouch™ Push Button Blood Collection Set. The risk profile of patients with DVA is not substantially different from that of the general patient population and is supported by the extensive clinical literature evidence presented under this submission. Based on information provided in this submission the proposed device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.