BD Vacutainer UltraTouch Push Button Blood Collection Set
K153309 · Becton, Dickinson and Company · JKA · Feb 12, 2016 · Clinical Chemistry
Device Facts
Record ID
K153309
Device Name
BD Vacutainer UltraTouch Push Button Blood Collection Set
Applicant
Becton, Dickinson and Company
Product Code
JKA · Clinical Chemistry
Decision Date
Feb 12, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1675
Device Class
Class 2
Indications for Use
The BD Vacutainer® UltraTouch™ Push Button Blood Collection Set is a sterile, multiple sample, single-use fixed winged blood collection set intended for venipuncture to obtain blood specimens from patients. When used without the male adapter, the device allows the clinician to obtain blood sampling to the female hub with a syringe, if necessary, or can be used for short-term, single infusions with consideration given to patient size and appropriateness for the solution being infused. The device is not to be left in place and remain under the direct supervision of a clinician. The recommended use of the device is to activate the needle safety feature prior to removal from the venipuncture site. The retraction of the intravenous (IV) end of the needle aids in the prevention of accidental needlestick injury.
Device Story
Sterile, single-use, fixed-winged blood collection set; used for venipuncture to obtain blood specimens or short-term, single infusions. Operated by clinicians. Features active needle retraction mechanism; button depression triggers spring-loaded retraction of IV needle into barrel, locking it to prevent needlestick injuries. Input: venipuncture site; Output: blood sample or fluid infusion. Device includes stainless steel cannula, spring, hub, wings, tubing, and optional male luer adapter. Enhances safety via automated needle containment.
Clinical Evidence
Bench testing only. Verification testing included tubing dimensions (ISO 9626), retraction/lockout, flow rate, IV cannula removal force, leak testing, hemolysis, resistance to breakage (ISO 9626), and penetration/sharpness testing. Results demonstrate equivalence to predicate.
Indicated for venipuncture to obtain blood specimens from patients; may be used for short-term, single infusions with consideration of patient size and solution appropriateness. Not for indwelling use. Requires direct clinician supervision.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Predicate Devices
BD Vacutainer® Push Button Blood Collection Set (K030573)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 12, 2016
Becton, Dickinson and Company Mr. Pasquale Amato Staff Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417
Re: K153309
Trade/Device Name: BD Vacutainer® Ultratouch™ Push Button Blood Collection Set Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: II Product Code: JKA, FPA Dated: December 16, 2015 Received: January 11, 2016
Dear Mr. Amato:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| 510(k) Number (if known)<br>K153309 | <b>Indications for Use</b> |
| Device Name<br>BD Vacutainer® UltraTouch™ Push Button Blood Collection Set | |
| Indications for Use (Describe) | |
| The BD Vacutainer® UltraTouch™ Push Button Blood Collection Set is a sterile, multiple sampie, single-use fixed winged blood collection set intended for venipuncture to obtain blood specimens from patients. When used without the male adapter, the device allows the clinician to obtain blood sampling to the female hub with a syringe, if necessary, or can be used for short-term, single infusions with consideration given to patient size and appropriateness for the solution being infused. The device is not to be left in place and remain under the direct supervision of a clinician. | |
| The recommended use of the device is to activate the needle safety feature prior to removal from the venipuncture site. The retraction of the intravenous (IV) end of the needle aids in the prevention of accidental needlestick injury. | |
| Type of Use (Select one or both, as applicable) | |
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| <b>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*</b> | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | |
| Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br><a href="mailto:PRAStaff@fda.hhs.gov">PRAStaff@fda.hhs.gov</a> | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
| FORM FDA 3881 (8/14) | Page 1 of 1 |
| PSC Publishing Services (301) 443-6740 EF | |
.
{3}------------------------------------------------
# 510(K) SUMMARY K153309
#### Submitted Bv: 1.
Pasquale Amato Staff Regulatory Affairs Specialist
Becton, Dickinson and Company 1 Becton Drive Franklin Lakes, NJ 07417
Phone: 201-847- 4513 Email: Pasquale Amato@BD.com
Date Prepared: February 05, 2016
- 2. Device:
T
| Trade Name: | BD Vacutainer® UltraTouch™ Push Button Blood<br>Collection Set | |
|----------------------|----------------------------------------------------------------|--|
| Common Names: | Blood Collection Set | |
| Classification Name: | Blood Specimen Collection Device | |
| Classification: | Class II, 21 CFR 862.1675 | |
| Product Code: | IKA and FPA | |
### Predicate Device: 3.
BD Vacutainer® Push Button Blood Collection Set - K030573 Manufactured by: Becton, Dickinson and Company
### 4. Device Description:
The BD Vacutainer® Push Button Blood Collection Set has two models. One is a steel winged blood collection set with flexible tubing, female luer connector, with an integrated male luer adapter which connects to a Vacutainer® Brand Needle Holder and is intended to be used with blood collection tubes. The other model is sold identical except without the male luer adapter. This model allows the clinician to obtain blood sampling with a syringe, if necessary, or can be used for short-term, single infusions. The device is not to be left in place and remain under the supervision of a clinician.
{4}------------------------------------------------
The BD Vacutainer® Push Button Blood Collection Set also contains a sharps injury prevention feature. The wing set is designed with an active retraction method. When the button is actively depressed, the needle fully retracts and is enclosed and locked within the barrel of the device. The retraction of the intravenous (IV) end of the needle is designed to protect users from an accidental needle sticks.
The BD Vacutainer® Push Button Blood Collection Set consists of:
- * Stainless steel cannula (Intravenous end and Non-patient end of cannula)
- * Stainless steel spring
- * Hub. front and rear barrel
- * Wings (color coded according to needle gauge)
- や Tubing
- * Female luer connector and an optional male luer adapter
- * Intravenous (IV) needle protector (covers the needle before use)
- & Luer Cap (provided if there is no luer adapter attached)
The intravenous needle of the blood collection set is bonded to one end of the hub. The tubing of the blood collection set is bonded to the other end of the hub and the female luer connector. A spring is assembled over the needle protector and onto the front of the hub. Once the sample is collected the user will depress the button that projects through the barrel. As soon as the button is depressed, the needle assembly moves using spring energy. In this retracted (locked) position, the IV point of the needle is fully contained within the body of the device. This will prevent the needle from coming out of the front barrel once it has retracted as well as preventing accidental overriding of the safety feature.
The optional male luer adapter contains threads for attachment to a Vacutainer® Brand Needle Holder, and a non-patient cannula for puncture of evacuated blood collection tube stoppers. The non-patient end (NP) of the cannula of the luer adapter has a sleeve that recovers over the cannula to stop blood flow during collection of multiple tubes.
The purpose of this Special 510(k) is to expand the needle range to include the BD Vacutainer® UltraTouch™ Push Button Blood Collection Set design with ultra thin wall cannula and a five bevel point. The intended use for the modified device remains the same as the predicate device.
### ર. Indications for Use:
The BD Vacutainer® UltraTouch™ Push Button Blood Collection Set is a sterile, multiple sample, single-use fixed winged blood collection set intended for venipuncture to obtain blood specimens from patients. When used without
{5}------------------------------------------------
the male adapter, the device allows the clinician to obtain blood sampling to the female hub with a syringe, if necessary, or can be used for short-term, single infusions with consideration given to patient size and appropriateness for the solution being infused. The device is not to be left in place and remain under the direct supervision of a clinician.
The recommended use of the device is to activate the needle safety feature prior to removal from the venipuncture site. The retraction of the intravenous (IV) end of the needle aids in the prevention of accidental needlestick injury.
## Technological Characteristics: 6.
The subject BD Vacutainer® UltraTouch™ Push Button Blood Collection Set is equivalent to that of the predicate BD Vacutainer® Push Button Blood Collection Set in intended use, materials and performance characteristics. The Indications for Use are not identical; however, the intended use remains the same for blood collection/sampling.
| Characteristic | Subject Device | Predicate Device |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | BD Vacutainer® UltraTouch™ Push<br>Button Blood Collection Set, 21G, 23G<br>and 25G | BD Vacutainer® Push Button<br>Blood Collection Set 21G, 23G<br>and 25G (k030573) |
| Indications for Use | The BD Vacutainer® UltraTouch™ Push Button Blood Collection Set is a sterile, multiple sample, single-use fixed winged blood collection set intended for venipuncture to obtain blood specimens from patients. When used without the male adapter, the device allows the clinician to obtain blood sampling to the female hub with a syringe, if necessary, or can be used for short-term, single infusions with consideration given to patient size and appropriateness for the solution being infused. The device is not to be left in place and remain under the direct supervision of a clinician.<br><br>The recommended use of the device is to activate the needle safety feature prior to removal from the venipuncture site. The retraction of the intravenous (IV) end of the needle aids in the prevention of accidental needlestick injury. | The BD Vacutainer® Push Button Blood Collection Set is a sterile, multiple-sample, single-use winged blood collection set intended for venipuncture to obtain blood specimens from patients. The BD Vacutainer® UtraTouch™ Push Button Blood Collection Set is also indicated for the intravenous administration of fluids. It may be used for any patient population with consideration given to patient size and appropriateness for the solution being infused and duration of therapy.<br><br>The recommended use of the device is to activate the needle prior to removal from the venipuncture site. The retraction of the intravenous (IV) end of the needle aids in the prevention of accidental needlestick injury. |
| Needle Diameter OD | 21G, 23G and 25G | same |
| Needle Diameter ID | Ultra Thin Wall | Thin Wall |
| Needle Point | 5 bevel | 3 bevel |
| Needle Length | 3/4in | same |
| Wing | Polyolefin | same |
| Hub | Polypropylene | same |
| Button Ink | UV curable Ink | same |
| Front Barrel | Polypropylene | same |
| Rear Barrel | Acrylic | same |
| Rear Barrel Lubricant | Silicone | same |
| Rear Barrel Lubricant | Isopropyl Alcohol | same |
| Diluent | | |
| Spring | Stainless Steel 302 | same |
| IV Protector (Cannula Protector) | Polyethylene | same |
| IV Cannula/NP Cannula | Stainless Steel 304 | same |
| Tubing | PVC | same |
| Cannula Lubricant | Silicone | same |
| Cannula Adhesive | UV cured adhesive | same |
| Hub-Tubing Adhesive | UV cured adhesive | same |
| Female Luer Connector | ABS | same |
| Luer Adapter Hub | Polypropylene | same |
| NP Sleeve | Synthetic Isoprene Rubber | same |
| Luer Adhesive | Heat Curing Epoxy | same |
| Luer Cannula Lubricant | Medical Grade Silicones | same |
| Luer Cap | Polypropylene | same |
| Top Web | Paper | same |
| Blister | PETG Copolyester | same |
| Materials - comply with ISO 10993-1 | Yes | same |
| Non-pyrogenic | Yes | same |
| Non-toxic | Yes | same |
| Sterile | Yes | same |
| SAL 10-6 | Yes | same |
| Sterilization | Gamma | same |
| Shelf Life | 2 years…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.