The Halo One Thin-Walled Guiding Sheath is indicated for use in peripheral arterial and venous procedures requiring percutaneous introduction of intravascular devices. The Halo One Thin-Walled Guiding Sheath is NOT indicated for use in the neurovasculature nor the coronary vasculature.
Device Story
Halo One Thin-Walled Guiding Sheath functions as both guiding sheath and introducer sheath; facilitates intravascular access via radial, femoral, popliteal, tibial, or pedal sites. Device consists of thin-walled single-lumen tubing with female luer hub, atraumatic distal tip, and radiopaque platinum-iridium marker. Includes detachable hemostasis valve with crosscut silicone membrane, side arm, and 3-way stopcock; supplied with compatible vessel dilator. Available in 4F, 5F, 6F sizes; lengths 10cm to 90cm. Some configurations include access kits with needles and guidewires. Hydrophilic coating optional on distal portion. Used by physicians in clinical settings to provide access for intravascular devices; thin-wall design reduces outer diameter by up to 1F compared to standard sheaths, potentially easing insertion and minimizing access site trauma.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including visual inspection, simulated use, dimensional testing, compatibility, penetration force, trackability, bend/kink resistance, leak testing, needle ultrasound visibility, and packaging integrity. Biocompatibility evaluated per ISO 10993-1 for externally communicating devices with limited blood contact (<24 hours).
Technological Characteristics
Thin-walled single-lumen tubing; platinum-iridium radiopaque marker; silicone membrane hemostasis valve. Materials evaluated per ISO 10993-1. Hydrophilic coating on distal portion. Sizes 4F-6F; lengths 10-90cm. Mechanical device; no energy source or software. Sterilization method not specified.
Indications for Use
Indicated for patients undergoing peripheral arterial and venous procedures requiring percutaneous introduction of intravascular devices. Not for use in neurovasculature or coronary vasculature.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font below.
September 23, 2021
ClearStream Technologies Ltd. Annmarie Fitzgerald Senior Regulatory Affairs Specialist Moyne Upper Enniscorthy, Co. Wexford Ireland
Re: K212708
Trade/Device Name: Halo One Thin-Walled Guiding Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: August 25, 2021 Received: August 26, 2021
Dear Annmarie Fitzgerald:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or
{1}------------------------------------------------
safety reporting (21) CFR CFR 4, Subpart B) for combination products postmarketing (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Finn Donaldson Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212708
Device Name Halo One™ Thin-Walled Guiding Sheath
Indications for Use (Describe)
The Halo One Thin-Walled Guiding Sheath is indicated for use in peripheral and venous procedures requiring percutaneous introduction of intravascular devices. The Halo One Thin-Walled Guiding Sheath is NOT indicated for use in the neurovasculature nor the coronary vasculature.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">❍</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
#### Halo One Thin-Walled Guiding Sheath
21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
#### Submitter Information:
ClearStream Technologies Ltd.
Moyne Upper,
Enniscorthy,
Co. Wexford, Ireland.
Phone: +353 53 9237111
Fax: +353 53 9237100
Contact Person: Annmarie Fitzgerald, Senior RA Specialist, Regulatory Affairs
# Date of Submission: August 25, 2021
# Subject Device:
| Name of Device: | Halo One Thin-Walled Guiding Sheath |
|-----------------------|-------------------------------------|
| Common or Usual Name: | Catheter, Introducer |
| Classification: | Introducer, Catheter |
| Regulatory Class: | II (Product Code: DYB) |
| Regulation Number: | 21 CFR 870.1340 |
| Classification Panel: | Cardiovascular |
| Predicate Device: | |
| 510(k) Number: | K192313 |
| Name of Device: | Halo One Thin-Walled Guiding Sheath |
| Common or Usual Name: | Catheter, Introducer |
| Classification: | Introducer, Catheter |
| Regulatory Class: | II (Product Code: DYB) |
| Regulation Number: | 21 CFR 870.1340 |
TRADE SECRET / CONFIDENTIAL INFORMATION Notify CR Bard Before Releasing this Document.
clearStream
Technologies Ltd
{4}------------------------------------------------
#### Classification Panel: Cardiovascular
# Device Description:
The Halo One Thin-Walled Guiding Sheath is designed to perform as both a guiding sheath and an introducer sheath.
The Halo One Thin-Walled Guiding Sheath consists of a thin-walled (Up to 1F reduction in outer diameter compared to standard sheaths of equivalent French size) sheath made from single-lumen tubing, fitted with a female luer hub at the proximal end and a formed atraumatic distal tip. The thin-wall design reduces the thickness of the sheath wall to help facilitate intravascular access from access sites including but not limited to radial, femoral, popliteal, tibial and pedal.
A detachable hemostasis valve, employing a crosscut silicone membrane and incorporating a side arm terminating in a 3-way stopcock, is connected to the sheath luer hub. The sheath is supplied with a compatible vessel dilator that snaps securely into the hemostasis valve hub. The sheath has a strain relief feature located at the luer hub and a radiopaque platinum-iridium marker located close to the distal tip.
The Halo One Thin-Walled Guiding sheath is supplied in 4F, 5F and 6F compatible sizes and lengths of 90cm. 70cm. 45cm. 25cm and 10 cm. The Halo One Thin-Walled Guiding Sheath 4F, 5F and 6F 25cm and 10cm sheaths will be offered with a 0.018" and 0.035" guide wire compatible dilator option.
The Halo One Thin-Walled Guiding Sheath is also offered as an access kit in 4F,5F and 6F 10cm and 25cm lengths incorporating access needle (21G x 4cm or 19G x 7cm option available) and access guidewire in both 0.018" (0.018" x 80cm or 0.018" x 50cm option available) and 0.035" (0.035" x 80cm or 0.035" x 50cm option available) configurations to the existing predicate device product range.
All sheath configurations (lengths) are provided with a hydrophilic coating over the distal portion of the sheath to provide a lubricious surface to ease insertion. The shorter sheath configurations (25cm and 10cm) are also provided without the coating.
clearStream
Technologies Ltd
{5}------------------------------------------------
# Intended Use / Indications for Use:
# Device Intended Use
The Halo One Thin-Walled Guiding Sheath is indicated for use in arterial and venous procedures requiring percutaneous introduction of intravascular devices.
# Indications for Use
The Halo One Thin-Walled Guiding Sheath is indicated for use in peripheral arterial and venous procedures requiring percutaneous introduction of intravascular devices. The Halo One Thin-Walled Guiding Sheath is NOT indicated for use in the neurovasculature nor the coronary vasculature.
# Technological Comparison to Predicate Device:
The Halo One Thin-Walled Guiding Sheath has the following similarities to the Predicate Device Halo One Thin-Walled Guiding Sheath (K192313, cleared December 19, 2019):
- · Same intended use.
- . Same indications for use
- Same target population/conditions of use .
- . Same operating principle/mechanism of action
- . Same fundamental scientific technology
- . Similar packaging materials and configurations
- Same sterility assurance level and method of sterilization .
The Halo One Thin-Walled Guiding Sheath (subject device) includes the following additions to the product range in comparison to the predicate device Halo One Thin-Walled Guiding Sheath (K192313, cleared December 2019):
- 1. The addition of 6F compatible size and lengths of 90cm, 70cm and 45cm to the range. A vessel dilator which is 0.035" guidewire compatible is provided with each sheath.
- 2. The 4F and 5F 25cm sheaths and 6F 10cm and 25cm sheaths will also be offered with a 0.018" quidewire compatible dilator.
- 3. The Halo One Thin-Walled Guiding Sheath will be further enhanced by the addition of access kits in 4F,5F and 6F 10cm and 25cm lengths incorporating access needle (21G x 4cm or 19G x 7cm option available) and access guidewire in both 0.018" (0.018" x 80cm or 0.018" x 50cm option available) and 0.035" (0.035" x 80cm or 0.035" x 50cm option available) configurations to the existing predicate device product range.
clearStream
Technologies Ltd
{6}------------------------------------------------
# Performance Data:
To demonstrate substantial equivalence of the subject device to the predicate device, both technical characteristics and performance criteria were evaluated. Using FDA Guidance documents on non-clinical testing of medical devices and internal Risk Assessment procedures, testing for the following characteristics and performance criteria were evaluated for the subject device:
- · Visual Inspection of sheath, access quidewire and access needle
- . Simulated use of sheath, access guidewire and access needle
- . Dimensional Testing of Dilator / Sheath
- Compatibility Testing of sheath, access quidewire and access needle .
- . Penetration Force of Dilator / Sheath
- . Trackability of Dilator and Sheath
- . Trackability of device in sheath
- Visual Inspection (Tip-Rollback)
- . Bend Radius / Kink
- . Leak Testing
- . Needle Ultrasound visibility
- . Packaging Testing
- o Visual Inspection
- Bubble Emission of Pouches O
- Visual Inspection of Sterile Barrier Packaging Heat Seal O
- Seal Strength Tensile Method o
The biocompatibility of the Halo One Thin-Walled Guiding Sheath was evaluated based on ISO 10993-1. The device is classified as an externally communication device, circulating blood, limited contact (<24hrs). The results from these tests performed in accordance with standards and FDA guidance, demonstrate that the technical characteristics and performance criteria of the Halo One Thin-Walled Guiding Sheath is substantially equivalent to the predicate and that it can perform in a manner equivalent to devices currently on the market for the same intended use.
# Conclusion:
The subject device, the Halo One Thin-Walled Guiding Sheath, met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocol and / or customer inputs. The clinical and non-clinical tests demonstrate that the Halo One Thin-Walled Guiding Sheath is substantially equivalent to the predicate device, Halo One Thin-Walled Guiding Sheath.
clearStream
Technologies Ltd
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.