K192313 · C.R. Bard, Inc. · DYB · Dec 19, 2019 · Cardiovascular
Device Facts
Record ID
K192313
Device Name
Halo One Thin-Walled Guiding Sheath
Applicant
C.R. Bard, Inc.
Product Code
DYB · Cardiovascular
Decision Date
Dec 19, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The Halo One Thin-Walled Guiding Sheath is indicated for use in peripheral arterial and venous procedures requiring percutaneous introduction of intravascular devices. The Halo One Thin-Walled Guiding Sheath is not indicated for use in the neurovasculature or the coronary vasculature.
Device Story
The Halo One Thin-Walled Guiding Sheath is a catheter introducer used in peripheral arterial and venous procedures. It functions as both a guiding sheath and an introducer sheath to facilitate intravascular access. The device consists of a thin-walled, braided single-lumen tube with a female luer hub, a detachable hemostasis valve with a crosscut silicone membrane, a side arm with a 3-way stopcock, and a vessel dilator. It features a radiopaque platinum-iridium marker at the distal tip and a strain relief at the hub. The device is inserted percutaneously by a clinician to provide a conduit for other intravascular devices. The thin-wall design reduces outer diameter, aiding access at sites including radial, femoral, popliteal, and pedal arteries. The device is supplied in various lengths and French sizes, with some configurations featuring a hydrophilic coating for lubricity. It is a single-use, sterile device.
Clinical Evidence
Bench testing only. No clinical data provided. Performance was validated through in vitro testing including visual inspection, simulated use, dimensional testing, radiopacity, penetration force, trackability, kink resistance, valve/sheath leak testing, tensile strength, hub stress cracking, packaging integrity, particulate characterization, and a full suite of biocompatibility testing (cytotoxicity, sensitization, reactivity, systemic toxicity, hemocompatibility, and pyrogenicity).
Indicated for patients undergoing peripheral arterial and venous procedures requiring percutaneous introduction of intravascular devices. Not for use in neurovasculature or coronary vasculature.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 19, 2019
C.R. Bard, Incorporated Elizabeth Delahunty Regulatory Affairs Manager Moyne Upper Enniscorthy, Ireland
Re: K192313
Trade/Device Name: Halo One Thin-Walled Guiding Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: August 23, 2019 Received: August 26, 2019
#### Dear Elizabeth Delahunty:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Misti Malone Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K192313
Device Name Halo One Thin-Walled Guiding Sheath
Indications for Use (Describe)
The Halo One Thin-Walled Guiding Sheath is indicated for use in peripheral arterial and venous procedures requiring percutaneous introduction of intravascular devices. The Halo One Thin-Walled Guiding Sheath is not indicated for use in the neurovasculature or the coronary vasculature.
| Type of Use (Select one or both, as applicable) | <table><tr><td><span> <svg height="12" width="12"> <rect fill="white" height="12" stroke="black" stroke-width="1" width="12"></rect> <path d="M2,2 L10,10 M2,10 L10,2" stroke="black" stroke-width="1"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </td><td><span> <svg height="12" width="12"> <rect fill="white" height="12" stroke="black" stroke-width="1" width="12"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </td></tr></table> | <span> <svg height="12" width="12"> <rect fill="white" height="12" stroke="black" stroke-width="1" width="12"></rect> <path d="M2,2 L10,10 M2,10 L10,2" stroke="black" stroke-width="1"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span> <svg height="12" width="12"> <rect fill="white" height="12" stroke="black" stroke-width="1" width="12"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <svg height="12" width="12"> <rect fill="white" height="12" stroke="black" stroke-width="1" width="12"></rect> <path d="M2,2 L10,10 M2,10 L10,2" stroke="black" stroke-width="1"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span> <svg height="12" width="12"> <rect fill="white" height="12" stroke="black" stroke-width="1" width="12"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
| FORM FDA 3881 (7/17) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 | EF |
|----------------------|-------------|----------------------------------------|----|
|----------------------|-------------|----------------------------------------|----|
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary
#### K192313
### Halo One Thin-Walled Guiding Sheath
# 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
#### Submitter Information:
ClearStream Technologies Ltd. Moyne Upper, Enniscorthy, Co. Wexford, Ireland. Phone: +353 53 9237111 Fax: + 353 53 9237100 Contact Person: Elizabeth Delahunty, RA Manager
#### Date of Submission: August 23, 2019
#### Subject Device:
Name of Device: Halo One Thin-Walled Guiding Sheath Common or Usual Name: Catheter Introducer Classification Name: Introducer, Catheter Regulatory Class: II (Product Code DYB) Requlation Number: 21 CFR 870.1340
#### Predicate Device:
510(k) Number: K161183 Name of Device: Halo One Thin-Walled Guiding Sheath Common or Usual Name: Catheter Introducer Classification Name: Introducer, Catheter Requlatory Class: II (Product Code DYB) Regulation Number: 21 CFR 870.1340
#### Device Description:
The Halo One Thin-Walled Guiding Sheath is designed to perform as both a guiding sheath and an introducer sheath.
{4}------------------------------------------------
The Halo One Thin-Walled Guiding Sheath consists of a thin-walled (Up to 1F reduction in outer diameter compared to standard sheaths of equivalent French size) sheath made from braided single-lumen tubing, fitted with a female luer hub at the proximal end a formed atraumatic distal tip. The thin-wall design reduces the thickness of the sheath wall to help facilitate intravascular access from access sites including but not limited to radial, femoral, popliteal, and pedal.
A detachable hemostasis valve, employing a crosscut silicone membrane and incorporating a side arm terminating in a 3-way stopcock, is connected to the sheath luer hub. The sheath is supplied with a compatible vessel dilator that snaps securely into the hemostasis valve hub. The sheath has a strain relief feature located at the luer hub and a radiopaque platinum-iridium marker located close to the distal tip.
The sheath is supplied in 4F, 5F and 6F compatible sizes and lengths of 90cm, 70cm, 45cm, 25cm and 10 cm.
A vessel dilator which is 0.035" guide wire compatible is provided with each sheath. The 4F and 5F 10cm sheaths will also be offered with a 0.018" guide wire compatible dilator.
All sheath configurations (lengths) are provided with a hydrophilic coating over the distal portion of the sheath to provide a lubricious surface to ease insertion. The shorter sheath configurations (25cm and 10cm) are also provided without this coating.
#### Indications for Use of Device:
The Halo One Thin-Walled Guiding Sheath is indicated for use in peripheral arterial and venous procedures requiring percutaneous introduction of intravascular devices. The Halo One Thin-Walled Guiding Sheath is NOT indicated for use in the neurovasculature or the coronary vasculature.
# Technological Comparison to Predicate Devices:
The Halo One Thin-Walled Guiding Sheath (subject device) has the following similarities to the predicate device, the Halo One Thin-Walled Guiding Sheath (clearance to market via K161183 on June 03, 2016):
- . Same intended use
- . Same indications for use
- . Same target population
- . Same operating principle
- Same fundamental scientific technology
- . Same sterility assurance level and method of sterilization
{5}------------------------------------------------
The Halo One Thin-Walled Guiding Sheath (subject device), incorporates the following differences:
- . Design Differences:
- 1. Changed sheath shaft the manufacturing process.
- 2. Changed braiding design.
- 3. Changed sheath tip design.
- 4. Changed dilator tip design rollback (buckling) while providing a transition to the guidewire.
- Material Differences: •
- 1. Changed material composition in the introducer shaft construction to address the failure mode which the predicate material caused.
- 2. Changed durometer of material to provide improved flexibility and kink resistance.
# Performance Data:
To demonstrate substantial equivalence of the subject device, the Halo One Thin-Walled Guiding Sheath to the predicate device, the technological characteristics and performance criterion were evaluated. Using FDA Guidance Documents on non-clinical testing of medical devices and internal Risk Assessment procedures, the following in vitro tests were performed on the subject device:
- . Visual Inspection (Outer Surface)
- Dilator Outer Surface o
- Sheath Outer Surface O
- Simulated Use .
- Packaging Removal o
- Haemostasis Valve Connection O
- Sheath Inner Surface/Dilator Compatibility O
- Dilator Flushability O
- Valve Flushability O
- Sheath Flushability o
- Guidewire Compatibility O
- Dilator Disengagement by Hand O
- Tape Adhesion O
- Dimensional Testing
- Sheath ID O
- Sheath Length O
- Marker Band to Tip Position O
{6}------------------------------------------------
- Sheath OD O
- Dilator OD o
- Dilator Extension Length o
- Radiopacity
- Penetration Force of Dilator/Sheath .
- . Trackability of Dilator and Sheath
- . Visual Inspection (Tip Rollback-Dilator & Sheath)
- Bend Radius/ Kink
- . Valve Leak
- . Sheath Leak
- Sheath and Dilator Tensile Forces
- Hub Torque/Stress Cracking
- Hub Stress Cracking (48 Hour Test)
- . Packaging
- o Visual Inspection
- Bubble Emission of Pouches o
- Visual Inspection of Sterile Barrier Packaging Heat Seals o
- Seal Strength Tensile Method о
- . Particulate Characterization
- Cytotoxicity •
- . Sensitization
- . Intracutaneous Reactivity
- . Acute Systemic Toxicity
- Hemocompatibility
- . Material Mediated Pyrogenicity
The biocompatibility of the Halo One Thin-Walled Guiding Sheath was evaluated based on ISO 10993-1. The device is classified as an Externally Communicating Devices, Circulating Blood, Limited Contact (<24 hrs). The results from these tests demonstrate that the technological characteristics and performance criteria of the Halo One Thin-Walled Guiding Sheath are substantially equivalent to the predicate device, and that it can perform in a manner equivalent to devices currently on the market for the same intended use.
#### Conclusions:
The subject device, the Halo One Thin-Walled Guiding Sheath, met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The clinical and non-clinical tests
{7}------------------------------------------------
demonstrate that the Halo One Thin-Walled Guiding Sheath is substantially equivalent to the predicate device, Halo One Thin-Walled Guiding Sheath.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.