K212567 · Chengdu Daxan Innovative Medical Tech. Co., Ltd. · EZD · Dec 7, 2022 · Gastroenterology, Urology
Device Facts
Record ID
K212567
Device Name
PVC Hydrophilic Urethral Catheter
Applicant
Chengdu Daxan Innovative Medical Tech. Co., Ltd.
Product Code
EZD · Gastroenterology, Urology
Decision Date
Dec 7, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PVC Hydrophilic Urethral Catheter is launched for clean intermittent catheterization -CIC treatment and indicated for use by patients with chronic urine retention. The catheter is inserted into the bladder through the urethra for emptying the bladder.
Device Story
Single-use medical-grade PVC urethral catheter; coated with polyvinyl pyrrolidone (PVP) hydrophilic polymer. Activated by immersion in water or squeezing integrated water sachet for 30 seconds to reduce insertion friction. Available in French sizes 8-18 for male and female patients. Used for clean intermittent catheterization (CIC) to drain bladder in patients with chronic urine retention. Operated by patient or clinician. Reduces urethral trauma during insertion via lubricious surface.
Clinical Evidence
Bench testing only. Performance testing included flow rate, catheter strength, connector security, and coefficients of friction. Biocompatibility testing conducted per ISO 10993-1:2009, including cytotoxicity (ISO 10993-5:2009), irritation (ISO 10993-10:2010), sensitization (ISO 10993-10:2010), and subacute toxicity (ISO 10993-11:2017). Results confirmed device meets all pre-determined acceptance criteria.
Technological Characteristics
Medical-grade PVC catheter with polyvinyl pyrrolidone (PVP) hydrophilic coating. Activated by water. French sizes 8-18. Biocompatibility per ISO 10993 standards. Single-use.
Indications for Use
Indicated for patients with chronic urine retention requiring clean intermittent catheterization (CIC) to empty the bladder.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Predicate Devices
Hi-Slip Plus Hydrophilic Urethral Catheter (K062444)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 7, 2022
Chengdu Daxan Innovative Medical Tech. Co., Ltd Kevin Huang RA Manager Unit 6, Building 1, No.18, North Bayi Road, Wenjiang District Chengdu. Sichuan 611135 China
Re: K212567
> Trade/Device Name: PVC Hydrophilic Urethral Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZD Dated: November 17, 2022 Received: November 22, 2022
Dear Kevin Huang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Angel A. Soler-garcia -S
for Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212567
Device Name PVC Hydrophilic Urethral Catheter
Indications for Use (Describe)
The PVC Hydrophilic Urethral Catheter is launched for clean intermittent catheterization -CIC treatment and indicated for use by patients with chronic urine retention. The catheter is inserted into the bladder through the urethra for emptying the bladder.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | 2022/11/16 |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Chengdu Daxan Innovative Medical Tech. Co.,Ltd<br>Unit 6, Building 1, No.18, North Bayi Road, Wenjiang District,<br>611135 Chengdu, Sichuan Province, China |
| Contact Person: | Kevin huang<br>Regulatory Affairs Manager<br>Chengdu Daxan Innovative Medical Tech. Co.,Ltd<br>Email: kevin_huang1021@sina.com<br>Tel: +86-760-22819569<br>Fax: +86-760-22819569 |
| Trade or Proprietary Name: | PVC Hydrophilic Urethral Catheter |
| Common or Usual Name: | Catheter, Straight |
| Regulation Number: | 21 CFR 876.5130 |
| Classification Name: | Urological Catheters and Accessories |
| Product Code: | EZD |
| Device Classification: | Class II |
| Review Panel: | Gastroenterology-Urology |
| Predicate Device(s): | K062444 - Hi-Slip Plus Hydrophilic Urethral Catheter |
#### 1. Intended Use
The PVC Hydrophilic Urethral Catheter is launched for clean intermittent catheterization -CIC treatment and indicated for use by patients with chronic urine retention. The catheter is inserted into the bladder through the urethra for emptying the bladder.
#### 2. Device Description
The PVC Hydrophilic Urethral Catheter is a single use urinary catheter designed for clean intermittent catheterization, drainage of bladder. The device consists of disposable PVC (medical grade) catheter coated with hydrophilic polymer with/without water sachet (pure water). The catheters are coated with a coating containing polyvinyl pyrrolidone (PVP) - a kind of hydrophilic polymer, which binds the water molecules to the catheter to reduce the risk of friction while inserting the catheter into the urethra. Coating is activated by the water by squeezing the water sachet or purified water. When the catheter
{4}------------------------------------------------
is immersed in water for 30 seconds, it becomes slippery and ready to use. The PVC Hydrophilic Urethral Catheter is supplied in French size ranging from 8 to 18. It is available for male and female.
### 3. Substantial Equivalence-Comparison to Predicate Devices
The PVC Hydrophilic Urethral Catheter is similar to the predicate Hi-Slip Plus Hydrophilic Urethral Catheter K06244. The similar indications for use and technological characteristics of the PVC Hydrophilic Urethral Catheter as compared to the predicate device support a determination of substantial equivalency.
### Similarities Between Proposed and Predicate Device
The proposed PVC Hydrophilic Urethral Catheter and the predicate device, K062444 -Hi-Slip Plus Hydrophilic Urethral Catheter, have the similar intended use, principle of operation, performance characteristics and scientific technology.
#### Differences Between Proposed and Predicate Device
There is no difference between the proposed and the predicate devices.
Therefore, the proposed PVC Hydrophilic Urethral Catheter are substantially equivalent to the K062444 - Hi-Slip Plus Hydrophilic Urethral Catheter. The proposed device has the same classification information, similar intended use and technological characteristics as compared to the predicate devices.
## 4. Summary of Non-Clinical Performance Testing
The following performance testing was conducted for the PVC Hydrophilic Urethral Catheter:
## 1) General performance testing including:
- · Analysis of Flow rate
- · Analysis of strength of the catheter
- · Analysis of Connector Security
- · Analysis of Coefficients of friction
The results of these tests provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for use as a urethral catheter.
Testing data and results are included in this submission, and demonstrated that the PVC Hydrophilic Urethral Catheter meets all the pre-determined testing and acceptance criteria.
## 2) Bio-compatibility testing as per ISO 10993-1:2009 including:
· Cytotoxicity as per ISO 10993-5:2009
{5}------------------------------------------------
- Irritation as per ISO 10993-10:2010
- · Sensitization as per ISO 10993-10:2010
- · Subacute toxicity as per ISO 10993-11:2017
Bio-compatibility testing reports are included in this submission, and demonstrated that the device components that are in contact with the patient are bio-compatible.
#### Conclusions Drawn from the Non-Clinical Testing
The results of these tests demonstrate that the device is as safe, as effective, and performs as well as the identified predicates and support a determination of substantial equivalence.
#### 5. Conclusion
The PVC Hydrophilic Urethral Catheter is substantially equivalent to predicate devices K062444 - Hi-Slip Plus Hydrophilic Urethral Catheter. Based on the intended use, principle of operation, performance characteristics, and technological characteristics, the proposed PVC Hydrophilic Urethral Catheter is substantially equivalent to and as safe, as effective and performs as the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.