Correx Software

K212552 · Vilex, LLC · KTT · Oct 12, 2021 · Orthopedic

Device Facts

Record IDK212552
Device NameCorrex Software
ApplicantVilex, LLC
Product CodeKTT · Orthopedic
Decision DateOct 12, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesSoftware as a Medical Device, Pediatric

Indications for Use

The Correx Software is intended to be used as a component of multilateral external fixation system for the Vilex External Fixation System with the indications listed below: - Stabilization of Fractures & Osteotomy - Rear and Mid-foot Arthrodesis - Adult and Pediatric Leg Lengthening - Correction of Bone Deformity in Upper & Lower Extremities

Device Story

Correx Software is an optional computational tool for the Vilex External Fixation System; used by surgeons to plan bone segment adjustments. Input: surgeon-entered data derived from A/P and M/L radiographs (ring size, strut configuration, deformity parameters, desired alignment). Operation: software visualizes bone segment position and calculates required strut lengths for specific translations or rotations. Output: daily schedule of strut lengths for frame adjustment. The software does not control hardware or the fixation device directly; the surgeon retains responsibility for all clinical decisions regarding bone alignment and correction magnitude. Benefits include precise planning for complex orthopedic corrections.

Clinical Evidence

Bench testing only. Software verification and validation testing were performed to ensure the product meets all requirements and specifications. No clinical data was provided.

Technological Characteristics

Computational software environment for orthopedic planning. Inputs: manual entry of radiographic measurements and frame parameters. Outputs: calculated strut length schedules. Connectivity: standalone. Software: non-controlling, user-interactive tool for visualization and geometric calculation of bone segment translation/rotation.

Indications for Use

Indicated for adult and pediatric patients requiring stabilization of fractures and osteotomies, rear and mid-foot arthrodesis, leg lengthening, or correction of bone deformities in upper and lower extremities.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ October 12, 2021 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Vilex LLC % Roshana Ahmed Sr. Regulatory Affairs Specialist TELOS Partners LLC 571 Christina Lake Drive Lakeland, Florida 33813 Re: K212552 Trade/Device Name: Correx Software Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: KTT, JDW, OSN Dated: August 13, 2021 Received: August 13, 2021 Dear Roshana Ahmed: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212552 Device Name Correx Software Indications for Use (Describe) The Correx Software is intended to be used as a component of multilateral external fixation system for the Vilex External Fixation System with the indications listed below: - · Stabilization of Fractures & Osteotomy - Rear and Mid-foot Arthrodesis - · Adult and Pediatric Leg Lengthening - · Correction of Bone Deformity in Upper & Lower Extremities | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary #### I. Submitter Vilex, LLC 111 Moffitt Street McMinnville, TN 37110 Contact Person: Brock Johnson, Vice President Phone: (435)-659-1880 Date Prepared: October 12, 2021 #### II. Device | Device Proprietary Name: | Correx Software | |--------------------------|--------------------------------------------------------------------------------| | Common or Usual Name: | Software for Diagnosis/Treatment<br>External Fixation System | | Classification Name: | Single/Multiple Component Metallic Bone Fixation Appliances<br>and Accessories | | Regulation Number: | 21 CFR 888.3030 | | Product Code: | KTT/JDW/OSN | | Device Classification | II | #### III. Predicate Device Substantial equivalence is claimed to the following device: - X-Fix Line Additions, K151881, Vilex in Tennessee, Incorporated ● #### IV. Device Description The Correx Software, intended for use with the Dynex Ring Fixation System (K202054), is an optional-use computational environment that can be used for determining the orientation of bone segments and external fixator elements attached to a bone to treat fractures and bone deformities. The software includes user interface, processing, data storage, and display functions. The Correx Software can be used to visualize the position of a moving bone segment and to compute the strut lengths necessary to implement any desired translation or rotation. The Software program does not control the device. The Software program does not control any hardware (such as rings, struts, wires) directly. The goal of the Software program is to provide the Surgeon with a tool for determining the length of the six struts connecting the two circular {4}------------------------------------------------ rings. The Surgeon must decide where a bone segment should line up and the angulation and translation necessary for the treatment. The magnitude and direction of any rotation or translation is ultimately the responsibility of the Surgeon. #### V. Indications for Use The Correx Software is intended to be used as a component of multilateral external fixation system for the Vilex External Fixation System with the indications listed below: - Stabilization of Fractures & Osteotomy - . Rear and Mid-foot Arthrodesis - Adult and Pediatric Leg Lengthening . - Correction of Bone Deformity in Upper & Lower Extremities . #### VI. Comparison of Technological Characteristics The subject and predicate device share the same intended use and technological characteristics. A comparison of products technological characteristics is provided in the table below. | | Correx Software | P&C Software | |--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| | | | K151881 | | Intended User | Healthcare Professional | Same | | Software Use | Optional | Same | | Principles of<br>Operation | The Software can be used to visualize the<br>position of a moving bone segment and to<br>compute the strut lengths necessary to<br>implement any desired translation or<br>rotation. | Same | | Inputs | Surgeon Inputs based on X-Ray<br>positioning | Same | | Image<br>Requirements | A/P and M/L radiographs after External<br>Frame attached to bone segments | Same | | System Fields<br>(Data Inputs) | Ring Size, Struts, Preference Side,<br>Deformity, Correction, Bone Alignment,<br>Separation Rate | Same | | Output | Daily schedule of strut lengths necessary<br>to adjust the frame so that the bone<br>segments are aligned as desired and<br>translated to the desired position within a<br>minimum number of days | Same | {5}------------------------------------------------ #### VII. Performance Data Software verification and validation testing was undertaken to ensure that the product requirements and specifications were met. ## VIII. Conclusion Based on the information provided above and within this submission, the Correx Software is substantially equivalent to the predicate device. --------------------------------This space intentionally left blank-----------------------------------------------------------------------------------------------------------
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...