K151881 · Vilex IN Tennessee, Inc. · KTT · Mar 24, 2016 · Orthopedic
Device Facts
Record ID
K151881
Device Name
X-Fix Line Additions
Applicant
Vilex IN Tennessee, Inc.
Product Code
KTT · Orthopedic
Decision Date
Mar 24, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Vilex X-Fix is intended for external fixation with the following indications: 1. Stabilization of Fractures & Osteotomy 2. Adult and Pediatric Leg Lengthening 3. Correction of Bone Deformity in Upper & Lower Extremities The P&C Software is intended to be used as a component of multilateral external fixation system for the indications listed above.
Device Story
X-Fix Line Additions comprise external fixation rings, footplates, struts, assembly accessories, and P&C software; used as components of a multilateral external fixation system. System stabilizes fractures/osteotomies, performs leg lengthening, and corrects bone deformities. Operated by orthopedic surgeons in clinical settings. Software component assists in the management of the multilateral fixation system. Device provides mechanical support for bone healing and alignment; benefits include structural stabilization and deformity correction. No clinical data required; safety established via mechanical bench testing and software verification/validation.
Clinical Evidence
No clinical data. Safety and effectiveness supported by mechanical bench testing per ASTM F1541-02 (Sections A6 and A7) and software verification/validation per FDA guidance.
Technological Characteristics
Metallic bone fixation components (rings, footplates, struts) manufactured from stainless steel and aluminum. Mechanical testing conducted per ASTM F1541-02. Includes P&C software component for multilateral fixation system management.
Indications for Use
Indicated for adult and pediatric patients requiring external fixation for fracture stabilization, osteotomy, leg lengthening, and correction of bone deformities in upper and lower extremities.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three faces in profile, stacked on top of each other. The faces are oriented to the right. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the image of the faces.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 24, 2016
Vilex in Tennessee, Incorporated % Abraham Lavi, Ph.D. Consultant Vilex, Incorporated 8374 Market Street, #167 Lakewood Ranch, Florida 34202
Re: K151881
Trade/Device Name: X-Fix Line Additions Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT. OSN Dated: February 19, 2016 Received: February 22, 2016
Dear Dr. Lavi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
K151881 - Page 1 of 1
Image /page/2/Picture/1 description: The image shows the word "VILEX" in red font. The "V" is stylized with a diamond shape above it. The letters are rounded and connected, giving the word a modern and sleek appearance. The background is plain white.
ADVANCE WITH US >>>
#### INDICATIONS FOR USE
# 510(k) NUMBER: K151881
DEVICE NAME: X-Fix Line Additions
INDICATIONS FOR USE:
The Vilex X-Fix is intended for external fixation with the following indications:
- 1. Stabilization of Fractures & Osteotomy
- 2. Adult and Pediatric Leg Lengthening
- 3. Correction of Bone Deformity in Upper & Lower Extremities
The P&C Software is intended to be used as a component of multilateral external fixation system for the indications listed above.
× Prescription Use __ (Per 21 CFR 801.109)
Concurrence of CDRH, Office of Device Evaluation (DOE)
Page 1 of 1
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "VILEX" in red font. The "V" has a diamond shape above it. The letters are all capitalized and are in a sans-serif font. The word is centered in the image.
ADVANCE WITH US >>>
| 510(k) Summary | K151881 |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Vilex in Tennessee, Inc.<br>111 Moffitt Street<br>McMinnville, TN 37110<br>Phone: (931) 474-7550<br>Fax: (931) 474-7551 |
| Contact: | Abraham Lavi |
| Date Prepared: | March 17, 2016 |
| Trade Name: | X-Fix Line Additions |
| Common Name: | External Fixation System |
| Classification: | 888.3030 Single/multiple component metallic bone fixation<br>appliances and accessories |
| Product Code: | KTT/OSN/Orthopedics, Class II |
| Predicate Devices: | Smith & Nephew's Taylor Spatial Frame (K110069, K093047, and K970748)<br>Smith & Nephew's Ilizarov External Fixator (K870961)<br>Orthofix's TL-HEX Truelok Hexapod System (K141078)<br>Vilex's X-Fix (K052196) |
# Description of Device:
The X-Fix Line Additions, subject of this 510(k) submission, include External Fixation Rings and Footplates, External Fixation Struts, External Fixation Assembly Accessories, and P&C Software. The line additions are manufactured from stainless steel and aluminum materials in various sizes to accommodate patient needs. The X-Fix Line Additions are used with the X-Fix system.
#### Indications for Use:
The Vilex X-Fix is intended for external fixation with the following indications:
- 1. Stabilization of Fractures & Osteotomy
- 2. Adult and Pediatric Leg Lengthening
- 3. Correction of Bone Deformity in Upper & Lower Extremities
The P&C Software is intended to be used as a component of multilateral external fixation system for the indications listed above.
{4}------------------------------------------------
## Technological Characteristics:
The technological characteristics for the X-Fix Line Additions are the same as the characteristics of the predicate devices. The line additions are manufactured in similar sizes and designs as the predicate devices. In addition, the line additions are manufactured from the same materials as the predicate devices.
## Performance Data:
The X-Fix Line Additions were evaluated according to FDA's "Reviewers Guidance Checklist for Orthopedic External Fixation Devices" dated February 21, 1997. Applicable mechanical testing was conducted per ASTM F1541-02 (Sections A6 and A7). All testing confirmed that the device is capable of withstanding expected in vivo loading.
Software verification and validation documentation and testing was completed in conformance with FDA guidance "General Principles of Software Validation; Final Guidance for Industry and FDA Staff" dated January 11, 2002. The results of software V&V testing indicate that the software performed as intended.
A review of this testing has demonstrated that there are no new issues related to the safety and effectiveness of the X-Fix Line Additions.
Clinical data was not needed to support the safety and effectiveness of the X-Fix Line Additions.
# Substantial Equivalence
The design features of the X-Fix Line Additions are substantially equivalent to the design features of other predicate devices previously cleared for market. The methods used to establish equivalence are based on the comparison of the intended use, product technical characteristics, performance characteristics, and mechanical testing. The safety and effectiveness of the X-Fix Line Additions are adequately supported by the substantial equivalence information, material information, analysis data, and testing provided within this Premarket Notification. Therefore, it is concluded that the X-Fix Line Additions are substantially equivalent to the noted predicate devices.
#### Conclusions
While the X-Fix Line Additions are not identical to the predicate devices, any differences that may exist do not significantly affect device safety and effectiveness. In addition, the differences do not add new or increased risks and complications. Therefore, it is concluded that the X-Fix Line Additions are substantially equivalent to the predicate devices as outlined previously and should not render the subject device NSE.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.