K192270 · Renovia, Inc. · HIR · Nov 22, 2019 · Obstetrics/Gynecology
Device Facts
Record ID
K192270
Device Name
LEva Pelvic Digital Health System
Applicant
Renovia, Inc.
Product Code
HIR · Obstetrics/Gynecology
Decision Date
Nov 22, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1425
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The leva Pelvic Digital Health System is intended for: 1) Strengthening of the pelvic floor muscles; 2) Rehabilitation and training of weak pelvic floor muscles for the treatment of stress, mixed and mild to moderate urgency urinary incontinence (including overactive bladder) in women. This device interacts with the user via smart phone technology
Device Story
Leva-02 PDHS is a prescription, reusable, intra-vaginal device for pelvic floor muscle (PFM) training. It consists of a probe containing six accelerometers, a storage case with battery, and a mobile application. The device detects PFM contractions via accelerometers and transmits real-time performance data wirelessly to the patient's smartphone. The app provides graphical and numeric feedback to the user, allowing them to monitor progress during training. Used by patients at home to strengthen pelvic floor muscles; the healthcare provider uses the output to guide rehabilitation. The system facilitates PFM training without electrical stimulation.
Clinical Evidence
No clinical data was required or provided. Substantial equivalence was supported by bench testing, including hardware verification, software validation (IEC 62304), usability/human factors testing, biocompatibility (ISO 10993), and electrical safety/EMC testing (IEC 60601 series).
Technological Characteristics
Intra-vaginal probe constructed of Thermoplastic Elastomer (TPE) and Acrylonitrile Butadiene Styrene (ABS). Sensing principle: six internal accelerometers to measure relative intensity of PFM contraction. Connectivity: wireless transmission to mobile app. Power: battery-operated. Software: minor level of concern. Standards: IEC 62304, ISO 10993, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, IEC 62366-1.
Indications for Use
Indicated for women with weak pelvic floor muscles requiring strengthening, rehabilitation, and training for the treatment of stress, mixed, and mild to moderate urgency urinary incontinence, including overactive bladder.
Regulatory Classification
Identification
A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.
Predicate Devices
Leva Pelvic Digital Health System (leva-01) (K180637)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 22, 2019
Renovia Inc. Gina Prochilo-Cawston Director of Regulatory 263 Summer St., 5th Floor Boston, MA 02210
Re: K192270 Trade/Device Name: Leva Pelvic Digital Health System Regulation Number: 21 CFR 884.1425 Regulation Name: Perineometer Regulatory Class: II Product Code: HIR Dated: October 28, 2019 Received: October 29, 2019
Dear Gina Prochilo-Cawston:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Angel A. Soler-García, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K192270
Device Name leva Pelvic Digital Health System
Indications for Use (Describe) The leva Pelvic Digital Health System is intended for:
1) Strengthening of the pelvic floor muscles;
2) Rehabilitation and training of weak pelvic floor muscles for the treatment of stress, mixed and mild to moderate urgency urinary incontinence (including overactive bladder) in women.
This device interacts with the user via smart phone technology
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
*An agency may not conduct or sponsor, and a person is not requred to respond to, a collection of information unless it displays a currently valid OMB number
FORM FDA 3881 (7/17)
Page 1 of 1
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
# 510(k) Summary
A 510(k) Summary was prepared in accordance with 21 CFR 807.92 and is provided on this page.
#### Submitter
| Name and Address: | Renovia Inc.<br>263 Summer Street<br>Boston, MA 02210 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact: | Gina Prochilo-Cawston<br>Director of Regulatory<br>Renovia Inc.<br>263 Summer St.<br>Boston, MA 02210<br>Phone: (857) 324-3089<br>Email: gcawston@renoviainc.com |
| Date Prepared: | November 18, 2019 |
| Device Information | |
| Device Trade Name: | leva Pelvic Digital Health System |
| Device Model #: | leva-02 |
| Common Name: | Pelvic Muscle Exerciser |
| Classification Name: | Perineometer |
| Product Code: | HIR |
| Classification Number: | 884.1425 |
884.1425 Regulatory Class: Class II GU - Gastroenterology/Urology Review Panel: Medical Specialty: OB - Obstetrics/Gynecology
### Predicate Device Information
| 510(k) Number: | K180637 |
|------------------------|-----------------------------------|
| Device Trade Name: | Leva Pelvic Digital Health System |
| Device Model #: | leva-01 |
| Common Name: | Pelvic Muscle Exerciser |
| Classification Name: | Perineometer |
| Product Code: | HIR |
| Classification Number: | 884.1425 |
| Regulatory Class: | Class II |
| Review Panel: | GU - Gastroenterology/Urology |
| Medical Specialty: | OB - Obstetrics/Gynecology |
| Manufacturer: | Renovia Inc. |
#### l. Device Description
The leva-02 PDHS is a prescription intra-vaginal device designed to allow the user (or woman) to rehabilitate and strengthen their pelvic floor muscles (PFM) as well as allow them to monitor their progress during pelvic floor muscle training (PFMT). The leva-02 is designed to wirelessly facilitate pelvic floor muscle training (PFMT) in women and to transmit real-time performance data through a dedicated mobile application that has been downloaded to the patient's mobile device. The levo-02 is designed to be used vaginally and is intended to be used repeatedly by a single patient.
{4}------------------------------------------------
The leva-02 consists of a probe, storage case, associated batteries and the Renovia Digital Health App (App). Thermoplastic elastomer (TPE) was used as the material overlay for the electronics and six accelerometers are contained within the probe. Additional electronics are contained in the storage case to transmit data wirelessly between the device and the App.
#### II. Indications for Use
The /eva Pelvic Digital Health System is intended for:
- 1) Strengthening of the pelvic floor muscles;
- 2) Rehabilitation and training of weak pelvic floor muscles for the treatment of stress, mixed and mild to moderate urgency urinary incontinence in women.
This device interacts with the user via smart phone technology.
## III. Comparison of Technological Characteristics
The following table compares the leva Pelvic Digital Health System (leva-02) to the predicate device (leva-01) with respect to the indications for use and primary technological characteristics:
| Element | Subject Device<br><i>leva</i> Pelvic Digital Health System<br>Model: <i>leva</i> -02 | Predicate Device<br><i>leva</i> Pelvic Digital Health System<br>Model: <i>leva</i> -01 |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)Number | This submission | K180637 |
| Manufacturer | Renovia Inc. | Renovia Inc. |
| Common/Usual Name | Perineometer | Perineometer |
| Classification Number | 884.1425 | 884.1425 |
| Device Class | Class II | Class II |
| Product Code | HIR | HIR |
| Intended Use | The <i>leva</i> Pelvic Digital Health System is<br>intended for:<br>1) Strengthening of the pelvic floor<br>muscles;<br>2) Rehabilitation and training of weak<br>pelvic floor muscles for the<br>treatment of stress, mixed and mild<br>to moderate urgency urinary<br>incontinence in women.<br>This device interacts with the user via<br>smart phone technology. | The <i>leva</i> Pelvic Digital Health System is<br>intended for:<br>1) Strengthening of the pelvic floor<br>muscles;<br>2) Rehabilitation and training of weak<br>pelvic floor muscles for the<br>treatment of stress, mixed and mild<br>to moderate urgency urinary<br>incontinence in women.<br>This device interacts with the user via<br>smart phone technology. |
| Principle of Operation | Provides indication of relative<br>intensity of pelvic floor muscle<br>contraction using accelerometers | Provides indication of relative<br>movement of pelvic floor muscle<br>contraction using accelerometers |
| Muscle Stimulation | No | No |
| Intended Anatomical Location | Vagina | Vagina |
| Single Patient Device | Yes | Yes |
| Reusable | Yes | Yes |
| Sterile | Clean, Non-sterile | Clean, Non-sterile |
| Information Display | Graphical and numeric based on<br>applied bending, anatomical overlay | Graphical and numeric based on<br>applied bending, anatomical overlay |
| Device Materials | Thermoplastic Elastomer (TPE- probe | Silicone |
| Element | Subject Device<br><i>leva</i> Pelvic Digital Health System<br>Model: <i>leva</i> -02 | Predicate Device<br><i>leva</i> Pelvic Digital Health System<br>Model: <i>leva</i> -01 |
| Direct Contact | covering material)<br>Acrylonitrile Butadiene Styrene (ABS-<br>probe battery pack) | |
{5}------------------------------------------------
#### IV. Summary of Non-Clinical Testing and Risk Analysis
The following non-clinical performance testing and risk analysis was performed to support the leva-02 PDHS:
#### A. Hardware
Hardware verification was performed in accordance with predefined test procedures and acceptance criteria and included the following assessments:
- . Dimensional analysis
- . Accelerometer performance
- . Device integrity (e.g., pull force, bend, material) under repeated motion
- . Wireless communications
- . Component connections (i.e., probe, battery pack)
- . Cleanability
Hardware verification demonstrated that the subject device, leva-02, is as safe, as effective, and performs as well as the predicate device, leva-01.
#### B. Software
Software validation was performed in accordance with IEC 62304 and completed with no outstanding anomalies. Software documentation was provided in accordance with FDA guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (2005) for a minor software level of concern. The leva-02 PDHS was tested and found to conform to the requirements of the following standards:
- IEC 62304:2006 (First Edition) + A1:2015, Medical device software life cycle processes
Software verification demonstrated that the subject device, leva-02, is as safe, as effective, and performs as well as the predicate device, leva-01.
#### C. Usability
Usability and Human Factors testing was performed using layperson volunteers. Testing included:
- Reading Instructions for Use and Quick Start Guide .
- . Setting up the system
- . Using system correctly
- . Cleaning, storage, disposal
All participants successfully completed all testing demonstrated that the subject device, levice, levice, levice, leva-02, is as safe, as effective, and performs as well as the predicate device, leva-01.
### D. Biocompatibility
Patient-contacting material was subjected to biocompatibility testing in compliance with ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, for mucosal surface for less than 24 hours duration including:
- . Cytotoxicity (ISO 10993-5)
- . Sensitization (ISO 10993-10)
- Irritation (ISO 10993-10)
Biocompatibility testing demonstrated that the subject device, leva-02, is as safe, as effective, and performs as well as the predicate device, leva-01.
{6}------------------------------------------------
#### Electrical Safety and Electromagnetic Compatibility E.
The leva-02 PDHS was tested and found to conform to the requirements of the following standards:
- . IEC 60601-1:2005 + A1: 2012, Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
- . IEC 60601-1-6:2010, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- . IEC 60601-1-11, General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- . IEC 60601-1-2 ed 4.0 (2014-02), Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- . IEC 61000, Electromagnetic Compatibility
Electrical Safety and Electromagnetic Compatibility testing demonstrated that the subject device, leva-02, is as safe, as effective, and performs as well as the predicate device, leva-01.
## F. Additional Testing
The leva-02 PDHS was found to conform to the following additional standards and requirements:
- 0 IEC 62366-1:2015, Medical Devices -- Part 1: Application of Usability Engineering to Medical Devices
- . CISPR 11:2016+A2:2019, Industrial, scientific and medical equipment - Radiofrequency disturbance characteristics - Limits and methods of measurement
- . 47CFR15.247 Subpart C: 02/2019, Federal Communications Commission, Radio Frequency Devices, Intentional Radiators
- . 47CFR15.247 Subpart B: 02/2019, Federal Communications Commission, Radio Frequency Devices, Unintentional Radiators
- . RSS-247 Issue 2 February 2017, Canada Radio Equipment Standards Specifications, Digital Transmission Systems (DTSs), Frequency Hopping Systems (FHSs) and License-Exempt Local Area Network (LE-LAN) Devices
- . ICES-003 Issue 6 Published: January 2016, updated April 2017, Canada Interference Causing Equipment Standards, Information Technology Equipment (Including Digital Apparatus) — Limits and Methods of Measurement
- . RSS-Gen Issue 5 April 2018, Canada Radio Equipment Standards, Radio Standards Specifications, General Requirements for Compliance of Radio Apparatus
- . RSS-102 Issue 5 March 2015, Canada Radio Equipment Standards Specifications, Radio Frequency (RF) Exposure Compliance of Radiocommunication Apparatus (All Frequency Bands)
- . CISPR 25 Ed 3:2008. Vehicles, boats and internal combustion engines. Radio disturbance characteristics -limitsandmethods ofmeasurement for the protection for on-board receivers.
- . EN 50121-3-2:2015. Railway Applications - Electromagnetic Capability.
- . RTCA DO-160G; Section 21.4 RF Radiated Emission; Published: December 8, 2010. Environmental Conditions and Test Procedures for Airborne Equipment.
The additional testing demonstrated that the subject device, leva-02, is as safe and as effective and performs as well as the predicate device, leva-01.
### G. Packaging
The leva-02 PDHS packaging system will be subjected to ISTA 2A-2011 "Partial Simulation Performance Test Procedure for Packaged Products 150 lb. (68 kg) or Less" testing, prior to commercial launch. To ensure the packaging adequately protects the device during shipping and storage this testing will include:
- . Atmospheric Preconditioning
- . Compression
- . Initial Random Vibration
{7}------------------------------------------------
- . Impact
- Final Random Vibration
## H. Risk Analysis
In addition to non-clinical performance testing, a risk analysis was performed in accordance with ISO 14971:2012 to identify potential hazards and hazardous situate and evaluate the potential risks, and implement risk control measures to mitigate the potential risks of the device, where possible.
The risk analysis demonstrated that the subject device, leva-02, is as safe, as effective, and performs as well as the predicate device, leva-01.
## V. Clinical Testing
Clinical testing was not required to support a substantial equivalence determination for the leva-02 PDHS.
#### VI. Conclusion
Based on the comparison and analysis above, Renovia has demonstrated that the leva-02 PDHS is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.