Radical-7 Pulse CO-Oximeter and Accessories, Rad-97 and Accessories
Applicant
Masimo Corporation
Product Code
MWI · Cardiovascular
Decision Date
Mar 6, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Rad-97 and Accessories is a multi-parameter patient monitor that is intended for multi-parameter physiological patient monitoring in hospital, hospital type facilities, mobile and home environments. The Rad-97 and Accessories can communicate with network systems for supplemental remote viewing (e.g., at a central station). The Rad-97 and Accessories are indicated for the non-invasive spot-checking and continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) of adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused. The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of carboxyhemoglobin saturation (SpCO) of adult, pediatric, and infant patients during no motion conditions. The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of methemoglobin saturation (SpMet) of adult, pediatric, and neonatal patients during no motion conditions. The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of total hemoglobin concentration (SpHb) of adult and pediatric patients during no motion conditions. The Rad-97 and Accessories are indicated for the continuous monitoring rate (RRa) for adult, pediatric, and neonatal patients during no motion conditions. In addition, the Rad-97 and Accessories are indicated to provide the non-invasive spot-checking and continuous monitoring data obtained from the Rad-97 and Accessories for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) to multi-parameter devices for the display on those devices. The Rad-97 and Accessories are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms. The optional NomoLine Capnography product family is intended to other medical backboard devices for monitoring of breath rate and CO2. The NomoLine Capnography product family is intended to a patient breathing circuit for monitoring of inspired gases during anesthesia, recovery and respiratory care. The environment is the operating suite, intensive care unit and patient population is adult, pediatric and infant patients. The optional non-invasive blood pressure (NiBP) module is indicated for the noninvasive measurement of arterial blood pressure. The NiBP module is designed to messure for patient population described in the following table: Patient Population Approximate Age Range Newborn (neonate) Birth to 1 month of age Infant 1 month to 2 years of age Child 2 to 12 years of age Adolescent 12-21 years of age Adult 21 years of age and older Devices with Masimo technology are only to be used with Masimo sensors and cables. The Rad-97 and accessories are indicated for the non-invasive continuous monitoring of PVi as a measure of relative variability of the photoplethysmograph (pleth) of adults during no motion conditions. PVi may be used as a noninvasive dynamic indicator of fluid responsiveness in select populations of mechanically ventilated adult patients. Accuracy of PV in predicting fluid responsiveness is variable and influenced by numerous patient, procedure and device related factors. PVi measures the variation in the plethysmography amplitude but does not provide measurements of stroke volume or cardiac output. Fluid management decisions should be based on a complete assessment of the patient's condition and should not be based solely on PVi. The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of respiratory rate from Pleth (RRp) for adult and pediatric patients during no motion conditions.
Device Story
Rad-97 and Radical-7 are multi-parameter patient monitors; input signals include optical sensors for pulse oximetry (SpO2, PR, SpCO, SpMet, SpHb, PVi, RRp), acoustic sensors for respiration (RRa), capnography for CO2, and oscillometric cuffs for NiBP. Devices transform physiological signals via Masimo SET technology; output data displayed on touchscreen LCD or transmitted to central stations. Used in hospitals, mobile, and home environments by clinicians or patients. Output supports clinical decision-making for vital sign monitoring and fluid responsiveness assessment. Benefits include non-invasive, continuous, or spot-check monitoring of critical physiological parameters.
Clinical Evidence
No clinical data required; submission relies on existing clearance and labeling updates for spot-checking functionality. Bench testing only.
Technological Characteristics
Masimo SET pulse oximetry technology; spectrophotometry for blood constituents; photoplethysmography for pulse/respiration; infrared spectrometry for capnography; oscillometric method for NiBP. Touchscreen LCD interface; visual/audible alarms. Network connectivity for remote viewing. No patient-contacting materials.
Indications for Use
Indicated for non-invasive spot-checking and continuous monitoring of SpO2, PR, SpCO, SpMet, SpHb, RRa, RRp, and PVi in adult, pediatric, infant, and neonatal patients (depending on parameter) in hospital, mobile, and home environments. Includes NiBP for arterial blood pressure and NomoLine Capnography for CO2/breath rate. Contraindicated as sole diagnostic basis for CO poisoning.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
March 6, 2023
Masimo Corporation Kertana Shankar Regulatory Specialist II 52 Discovery Irvine, California 92618
Re: K212161
Trade/Device Name: Radical-7 Pulse CO-Oximeter and Accessories, Rad-97 and Accessories Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, DOA, BZO, JKS, CCK, DPZ, DXN, FLL Dated: May 27, 2022 Received: May 31, 2022
Dear Kertana Shankar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# James J. Lee -S
James J. Lee, PhD Division Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthsia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K212161
Device Name Rad-97 and Accessories
Indications for Use (Describe)
The Rad-97 and Accessories is a multi-parameter patient monitor that is intended for multi-parameter physiological patient monitoring in hospital, hospital type facilities, mobile and home environments.
The Rad-97 and Accessories can communicate with network systems for supplemental remote viewing (e.g., at a central station).
The Rad-97 and Accessories are indicated for the non-invasive spot-checking and continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) of adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused.
The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of carboxyhemoglobin saturation (SpCO) of adult, pediatric, and infant patients during no motion conditions.
The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of methemoglobin saturation (SpMet) of adult, pediatric, and neonatal patients during no motion conditions.
The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of total hemoglobin concentration (SpHb) of adult and pediatric patients during no motion conditions.
The Rad-97 and Accessories are indicated for the continuous monitoring rate (RRa) for adult, pediatric, and neonatal patients during no motion conditions.
In addition, the Rad-97 and Accessories are indicated to provide the non-invasive spot-checking and continuous monitoring data obtained from the Rad-97 and Accessories for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) to multi-parameter devices for the display on those devices.
The Rad-97 and Accessories are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.
The optional NomoLine Capnography product family is intended to other medical backboard devices for monitoring of breath rate and CO2. The NomoLine Capnography product family is intended to a patient breathing circuit for monitoring of inspired gases during anesthesia, recovery and respiratory care. The environment is the operating suite, intensive care unit and patient population is adult, pediatric and infant patients.
The optional non-invasive blood pressure (NiBP) module is indicated for the noninvasive measurement of arterial blood pressure. The NiBP module is designed to messure for patient population described in the following table:
Patient Population Approximate Age Range Newborn (neonate) Birth to 1 month of age Infant 1 month to 2 years of age Child 2 to 12 years of age
{3}------------------------------------------------
Adolescent 12-21 years of age Adult 21 years of age and older
Devices with Masimo technology are only to be used with Masimo sensors and cables.
The Rad-97 and accessories are indicated for the non-invasive continuous monitoring of PVi as a measure of relative variability of the photoplethysmograph (pleth) of adults during no motion conditions.
PV i may be used as a noninvasive dynamic indicator of fluid responsiveness in select populations of mechanically ventilated adult patients. Accuracy of PV in predicting fluid responsiveness is variable and influenced by numerous patient, procedure and device related factors. PVi measures the variation in the plethysmography amplitude but does not provide measurements of stroke volume or cardiac output. Fluid management decisions should be based on a complete assessment of the patient's condition and should not be based solely on PVi.
The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of respiratory rate from Pleth (RRp) for adult and pediatric patients during no motion conditions.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known) K212161
Device Name Radical-7 and Accessories
#### Indications for Use (Describe)
The Radical-7 and Accessories are indicated for the non-invasive spot-checking and continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) of adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.
The Radical-7 and Accessories are indicated for the non-invasive continuous monitoring of carboxyhemoglobin saturation (SpCO) of adult, pediatric, and infant patients during no motion conditions in hospitals and hospital-type facilities. The Radical-7 and Accessories are not intended to be used as the sole basis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.
The Radical-7 and Accessories are indicated for the non-invasive continuous monitoring of methemoglobin saturation (SpMet) of adult, pediatric, and neonatal patients during no motion conditions in hospitals and hospital-type facilities.
The Radical-7 and Accessories are indicated for the non-invasive continuous monitoring of total hemoglobin concentration (SpHb) of adult and pediatric patients during no motion conditions in hospital-type facilities.
The Radical-7 and Accessories are indicated for the non-invasive continuous monitoring of respiratory rate (RRa) for adult, pediatric, and neonatal patients during no motion conditions in hospital-type facilities, home environments, and transport within healthcare facilities.
The Radical-7 and Accessories are indicated for the non-invasive continuous monitoring of Respiratory Rate from photoplethysmogram (RRp) for adult and pediatric patients during no motion conditions in hospital-type facilities, home environments, and transport within healthcare facilities.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Masimo, a medical technology company. The logo features a red circle with a white checkmark-like shape inside, positioned to the left of the company name, "Masimo," which is written in a bold, black, sans-serif font. The overall design is clean and modern, reflecting the company's focus on innovation in healthcare.
| Submitter and Address of<br>Manufacturing Facility: | Masimo Corporation<br>52 Discovery<br>Irvine, CA 92618<br>Phone: (949) 297-7000<br>FAX: (949) 297-7592 |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | February 22, 2023 |
| Contact: | Kertana Shankar<br>Senior Regulatory Specialist<br>Masimo Corporation<br>Phone: (949) 297-7260 |
| Trade Name: | Rad-97 and Accessories |
| Common Name: | Patient Monitor |
| Classification Regulation/<br>Product Code: | 21 CFR 870.2300, Class II/MWI |
| Additional Product Code: | 21 CFR 870.2700, Class II/DQA<br>21 CFR 868.2375, Class II/BZQ<br>21 CFR 862.3200, Class II/JKS<br>21 CFR 868.1400, Class II/CCK<br>21 CFR 870.2710, Class II/DPZ<br>21 CFR 870.1130, Class II/DXN<br>21 CFR 880.2910, Class II/FLL |
| Establishment Registration<br>Number: | 3011353843 |
| Reason for Premarket<br>Notification: | Update indications for use to include spot-checking |
| Predicate Device: | K193626 - Rad-97 and Accessories |
| Reference Predicate Device | K201770 - Rad-G and Accessories |
| Performance Standards | No performance standards for the above device have been<br>promulgated pursuant to Section 514. |
| Trade Name: | Radical-7 and Accessories |
|---------------------------------------------|-------------------------------|
| Common Name: | Oximeter |
| Classification Regulation/<br>Product Code: | 21 CFR 870.2700, Class II/DQA |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Masimo, a medical technology company. The logo features a red circle with a white checkmark-like shape inside, followed by the company name "Masimo" in black, sans-serif font. The checkmark shape is stylized and appears to be a stylized "M".
| Additional Product Code: | 21 CFR 868.2375, Class II/BZQ<br>21 CFR 862.3200, Class II/JKS<br>21 CFR 870.2710, Class II/DPZ |
|---------------------------------------|-------------------------------------------------------------------------------------------------|
| Establishment Registration<br>Number: | 3011353843 |
| Reason for Premarket<br>Notification: | Update indications for use to include spot-checking |
| Predicate Device: | K193242 – Radical-7 Pulse CO-Oximeter and Accessories |
| Reference Predicate Device | K201770 – Rad-G and Accessories |
| Performance Standards | No performance standards for the above device have been<br>promulgated pursuant to Section 514. |
### 1. Device Description
This premarket notification [510(k)] is for the authorization to market the previously cleared Rad-97 (K193626) and Radical-7 (K193242) with an additional indication for spot-checking.
The cleared versions of the subject devices already support both continuous monitoring and spotchecking use through the activation and deactivation of alarms. This submission updates the indications to reflect both uses of the subject devices.
### Rad-97
The Rad-97 is a patient monitor capable of providing multiple parameters. The Rad-97 product family provides the integrated ability of noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), Perfusion Index (Pi), Pleth Variability Index (PVi), carboxyhemoglobin (SpCO), methemoglobin (SpMet), total hemoglobin (SpHb), oxygen content (SpOC), acoustic respiration rate (RRa), Pleth Respiration Rate (RRp), capnography parameters, and noninvasive blood pressure (NiBP) parameters.
The subject device is the same as the predicate cleared under K193626 with exception of the updated indications for spot-checking.
| Feature | Rad-97 Specification |
|------------------------------|-----------------------------------------------------|
| Performance | Accuracy (ARMS)* |
| SpO2, no motion, 70-100% | 1.5%, adults/ pediatrics/ infants/ neonates** |
| SpO2, motion, 70-100% | 1.5% Arms, adults/ infants/ pediatrics/ neonates ** |
| SpO2, low perfusion, 70-100% | 2%, adults/ pediatrics/ infants/ neonates |
The specifications for Rad-97 are as follows:
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Masimo, a medical technology company. The logo features a red circle with a white checkmark-like shape inside, followed by the company name "Masimo" in bold, black letters. The checkmark is stylized and appears to be formed by a curved line. The overall design is simple and modern.
| Pulse Rate, no motion, 25-240 bpm | 3 bpm, adults/ pediatrics/ infants/ neonates |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Pulse Rate, motion, 25-240 bpm | 5 bpm, adults/ pediatrics/ infants/ neonates |
| Pulse Rate, low perfusion, 25-240 bpm | 3 bpm, adults/ pediatrics/ infants/ neonates |
| SpCO, 1-40% | 3%, adults/ pediatrics/ infants |
| SpMet, 1-15% | 1%, adults/ pediatrics/ infants/ neonates |
| SpHb, 8-17 g/dL | 1 g/dL adults/ pediatrics |
| RRa, 4-70 rpm,<br>4-120 rpm | 1 rpm, adults/ pediatrics<br>1 rpm, infants/ neonates |
| RRp, 4-70 rpm | 3 rpm ARMS, 1 rpm Mean Error, adults/ pediatrics |
| CO2 | Single dry gasses at 22±5°C and 1013±40 hPa:<br>0-15 volume %: ±(0.2 volume% +2% or reading)<br>All conditions: ±(0.3 kPa + 4% of reading) |
| NiBP | 0-300 mmHg, ISO 81060-2 |
*ARMS accuracy is a statistical calculation of the difference between device measurements and reference measurements. Approximately two-thirds of the device measurements fell within +/- ARMS of the reference measurements in a controlled study.
** Applicable with RD SET Disposable sensors
### Radical-7
The Radical-7 is a noninvasive monitor that measures arterial oxygen saturation (SpO2), pulse rate (PR), perfusion index (Pi), Pleth Variability Index (PVi), totalhemoglobin (SpHb), carboxyhemoglobin (SpCO), total oxygen content (SpOC), methemoglobin (SpMet), acoustic respiration rate (RRa), and Pleth Respiration Rate (RRp).
The subject device is the same as the predicate cleared under K 193242 with exception of the updated indications for spot-checking.
The Specifications for Radical-7 are as follows:
| Feature | Radical-7 Specification |
|---------------------------------------|--------------------------------------------------|
| Performance | Accuracy (ARMS)* |
| SpO2, no motion, 70-100% | 2%, adults/ pediatrics/ infants; 3% neonates |
| SpO2, motion, 70-100% | 3% adults/ pediatrics/ infants/ neonates |
| SpO2, low perfusion, 70-100% | 2%, adults/ pediatrics/ infants/ neonates |
| Pulse rate, no motion, 25-240 bpm | 3 bpm, adults/ pediatrics/ infants/ neonates |
| Pulse rate, motion, 25-240 bpm | 5 bpm, adults/ pediatrics/ infants/ neonates |
| Pulse rate, low perfusion, 25-240 bpm | 3 bpm, adults/ pediatrics/ infants/ neonates |
| RRa, 4-70 rpm | 1 rpm, adults/ pediatrics |
| 4-120 rpm | 1 rpm, infants/ neonates |
| RRp, 4-70rpm | 3 rpm ARMS, 1 rpm Mean Error, adults/ pediatrics |
| SpCO, 1-40% | 3%, adults/ pediatrics/ infants |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Masimo, a medical technology company. The logo features a red circle with a white waveform inside, followed by the company name in bold, black letters. To the right of the name is some smaller text, but it is not legible in the image. The logo is clean and modern, and the red and black color scheme is eye-catching.
| Feature | Radical-7 Specification |
|-----------------|-------------------------------------------|
| SpMet, 1-15% | 1%, adults/ pediatrics/ infants/ neonates |
| SpHb, 8-17 g/dL | 1 g/dL, adults/ pediatrics |
*ARMS accuracy is a statistical calculation of the difference between device measurements and reference measurements. Approximately two-thirds of the device measurements fell within +/- ARMS of the reference measurements in a controlled study.
#### 2. Intended Use/ Indications for Use
#### Rad-97
The Rad-97 and Accessories is a multi-parameter patient monitor that is intended for multi-parameter physiological patient monitoring in hospital type facilities, mobile and home environments.
The Rad-97 and Accessories can communicate with network systems for supplemental remote viewing and alarming (e.g., at a central station).
The Rad-97 and Accessories are indicated for the non-invasive spot-checking and continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) of adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused.
The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of carboxyhemoglobin saturation (SpCO) of adult, pediatric, and infant patients during no motion conditions.
The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of methemoglobin saturation (SpMet) of adult, pediatric, and neonatal patients during no motion conditions.
The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of total hemoglobin concentration (SpHb) of adult and pediatric patients during no motion conditions.
The Rad-97 and Accessories are indicated for the continuous monitoring of respiratory rate (RRa) for adult, pediatric, and neonatal patients during no motion conditions.
In addition, the Rad-97 and Accessories are indicated to provide the non-invasive spot-checking and continuous monitoring data obtained from the Rad-97 and Accessories for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) to multi-parameter devices for the display on those devices.
The Rad-97 and Accessories are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.
The optional NomoLine Capnography product family is intended to be connected to other medical backboard devices for monitoring of breath rate and CO2. The NomoLine Capnography product family is
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark-like symbol inside, followed by the word "Masimo" in a bold, sans-serif font. To the right of the logo is the text "MASIMO CORPORATION" in a smaller, sans-serif font, followed by the address "52 Discovery, Irvine, CA 92618".
intended to be connected to a patient breathing circuit for monitoring of inspired/expired gases during anesthesia, recovery and respiratory care. The environment is the operating suite, intensive care unit and patient room. The intended patient population is adult, pediatric and infant patients.
The optional non-invasive blood pressure (NiBP) module is indicated for the noninvasive measurement of arterial blood pressure. The NiBP module is designed to measure blood pressure for patient population described in the following table:
| Patient Population | Approximate Age Range |
|----------------------|---------------------------|
| Newborn<br>(neonate) | Birth to 1 month of age |
| Infant | 1 month to 2 years of age |
| Child | 2 to 12 years of age |
| Adolescent | 12-21 years of age |
| Adult | 21 years of age and older |
Devices with Masimo technology are only to be used with Masimo sensors and cables.
The Rad-97 and accessories are indicated for the non-invasive continuous monitoring of PVi as a measure of relative variability of the photoplethysmograph (pleth) of adults during no motion conditions.
PVi may be used as a noninvasive dynamic indicator of fluid responsiveness in select populations of mechanically ventilated adult patients. Accuracy of PV in predicting fluid responsiveness is variable and influenced by numerous patient, procedure and device related factors. PVi measures the variation in the plethysmography amplitude but does not provide measurements of stroke volume or cardiac output. Fluid management decisions should be based on a complete assessment of the patient's condition and should not be based solely on PVi.
The Rad-97 and Accessories are indicated for the non-invasive continuous monitoring of respiratory rate from Pleth (RRp) for adult and pediatric patients during no motion conditions.
### Radical-7
The Radical-7 and Accessories are indicated for the non-invasive spot-checking and continuous monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) of adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
The Radical-7 and Accessories are indicated for the non-invasive continuous monitoring of carboxyhemoglobin saturation (SpCO) of adult, pediatric, and infant patients during no motion conditions in hospitals and hospital-type facilities. The Radical-7 and Accessories are not intended to be
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Masimo, a medical technology company. The logo consists of a red circle with a white checkmark-like shape inside, followed by the company name "Masimo" in a bold, sans-serif font. To the right of the logo, there is some additional text, which is likely the company's address.
used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.
The Radical-7 and Accessories are indicated for the non-invasive continuous monitoring of methemoglobin saturation (SpMet) of adult, pediatric, and neonatal patients during no motion conditions in hospitals and hospital-type facilities.
The Radical-7 and Accessories are indicated for the non-invasive continuous monitoring of total hemoglobin concentration (SpHb) of adult and pediatric patients during no motion conditions in hospitals and hospital-type facilities.
The Radical-7 and Accessories are indicated for the non-invasive continuous monitoring of respiratory rate (RRa) for adult, pediatric, and neonatal patients during no motion conditions in hospitaltype facilities, home environments, and transport within healthcare facilities.
The Radical-7 and Accessories are indicated for the non-invasive continuous monitoring of Respiratory Rate from photoplethysmogram (RRp) for adult and pediatric patients during no motion conditions in hospitals, hospital-type facilities, home environments, and transport within healthcare facilities.
### 3. Technological Characteristics
### Principle of Operation
As part of this submission, there were no changes to the principles of operation for the Rad-97 and Radical-7 from their respective predicates, Rad-97 (K193626) and Radical-7 (K193242).
The devices use the same previously cleared Masimo SET pulse oximetry technology, which relies on the following principle:
- Oxyhemoglobin (oxygenated blood) and deoxyhemoglobin (nonoxygenated blood) differ in . their absorption of red and infrared light (spectrophotometry).
- The amount of arterial blood in tissue changes with your pulse (photoplethysmography). ● Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.
### Mechanism of Action for Achieving the Intended Effect
The mechanism of action for the subject devices is the same as that of their predicates (K193626 and K193242). The optical-based parameters still rely on the application of an optical sensor to a capillary application site to detect physiological signals. which are processed to estimate the parameters (e.g., SpO2 and Pulse Rate). Both devices already support the ability to activate and deactivate parameter alarms through the user interface.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the logo is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618".
| Feature | Rad-97<br>Subject Device | Rad-97,<br>Predicate device | Comparison to the Predicate Device |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K212161 | K193626 | |
| General<br>Information | | | |
| Primary<br>Classification<br>Regulation/ Product<br>code | 21 CFR 878.2300, Class II/MWI | 21 CFR 878.2300, Class II/MWI | Same |
| Additional<br>Classification<br>Regulation/ Product<br>Code(s) | 21 CFR 870.2700, Class II/DQA<br>21 CFR 862.3200, Class II/JKS<br>21 CFR 868.1400, Class II/CCK<br>21 CFR 868.2375, Class II/BZQ<br>21 CFR 870.2710, Class II/DPZ<br>21 CFR 870.1130, Class II/DXN<br>21 CFR 880.2910, Class II/FLL | 21 CFR 870.2700, Class II/DQA<br>21 CFR 862.3200, Class II/JKS<br>21 CFR 868.1400, Class II/CCK<br>21 CFR 868.2375, Class II/BZQ<br>21 CFR 870.2710, Class II/DPZ<br>21 CFR 870.1130, Class II/DXN<br>21 CFR 880.2910, Class II/FLL | Same |
| Indications for Use | The Rad-97 and Accessories is a multi-<br>parameter patient monitor that is<br>intended for multi-parameter<br>physiological patient monitoring in<br>hospital, hospital type facilities, mobile<br>and home environments.<br><br>The Rad-97 and Accessories can<br>communicate with network systems for<br>supplemental remote viewing and<br>alarming (e.g., at a central station). | The Masimo Rad-97 and Accessories<br>are indicated for hospitals, hospital-type<br>facilities, mobile, and home<br>environments.<br><br>The Masimo Rad-97 and Accessories<br>can communicate with network systems<br>for supplemental remote viewing and<br>alarming (e.g., at a central station). | Same with addition of indications for<br>spot-checking. The update in the<br>indications for use for spot-checking<br>was not found to result in a new<br>intended use.<br><br>The indications for use for the subject<br>device have also been revised to include<br>the cleared indications for the<br>Respiratory Rate from<br>photoplethysmogram (RRp) parameter |
| The Rad-97 and Accessories are<br>indicated for the non-invasive spot-<br>checking and continuous monitoring of<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate (PR)<br>of adult, pediatric, and neonatal patients<br>during both no motion and motion<br>conditions, and for patients who are well<br>or poorly perfused. | The Masimo Rad-97 and Accessories<br>are indicated for the continuous non-<br>invasive monitoring of functional<br>oxygen saturation of arterial<br>hemoglobin (SpO2) of adult, pediatric,<br>and neonatal patients during both no<br>motion and motion conditions, and for<br>patients who are well or poorly<br>perfused.. | based on the subject device's clearance<br>as part of K193242. | |
| The Rad-97 and Accessories are<br>indicated for the non-invasive continuous<br>monitoring of carboxyhemoglobin<br>saturation (SpCO) of adult, pediatric, and<br>infant patients during no motion<br>conditions. | The Masimo Rad-97 and Accessories<br>are indicated for the continuous non-<br>invasive monitoring of pulse rate (PR)<br>of adult, pediatric, and neonatal patients<br>during both no motion and motion<br>conditions, and for patients who are<br>well or poorly perfused. | | |
| The Rad-97 and Accessories are<br>indicated for the non-invasive continuous<br>monitoring of methemoglobin saturation<br>(SpMet) of adult, pediatric, and neonatal<br>patients during no motion conditions. | The Masimo Rad-97 and Accessories<br>are indicated for the continuous non-<br>invasive monitoring of<br>carboxyhemoglobin saturation (SpCO)<br>of adult, pediatric, and infant patients<br>during no motion conditions. | | |
| The Rad-97 and Accessories are<br>indicated for the non-invasive continuous<br>monitoring of total hemoglobin<br>concentration (SpHb) of adult and | The Masimo Rad-97 and Accessories<br>are indicated for the continuous non-<br>invasive monitoring of methemoglobin | | |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the Masimo Corporation logo. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black letters. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white "V" inside, followed by the name "Masimo" in black, stylized font. To the right of the name is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".
| pediatric patients during no motion<br>conditions. | saturation (SpMet) of adult, pediatric, and neonatal patients during no motion conditions. |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The Rad-97 and Accessories are<br>indicated for the continuous monitoring<br>of respiratory rate (RRa) for adult,<br>pediatric, and neonatal patients during no<br>motion conditions. | The Masimo Rad-97 and Accessories<br>are indicated for the continuous non-<br>invasive monitoring of total<br>hemoglobin concentration (SpHb) of<br>adult and pediatric patients during no<br>motion conditions. |
| In addition, the Rad-97 and Accessories<br>are indicated to provide the non-invasive<br>spot-checking and continuous monitoring<br>data obtained from the Rad-97 and<br>Accessories for functional oxygen<br>saturation of arterial hemoglobin (SpO2)<br>and pulse rate (PR) to multi-parameter<br>devices for the display on those devices. | The Masimo Rad-97 and Accessories<br>are indicated for the continuous non-<br>invasive monitoring of respiratory rate<br>(RRa) for adult, pediatric, and neonatal<br>patients during no motion conditions. |
| The Rad-97 and Accessories are not<br>intended to be used as the sole basis for<br>making diagnosis or treatment decisions<br>related to suspected carbon monoxide<br>poisoning; it is intended to be used in<br>conjunction with additional methods of<br>assessing clinical signs and symptoms. | In addition, the Masimo Rad-97 and<br>Accessories are indicated to provide the<br>continuous non-invasive monitoring<br>data obtained from the Masimo Rad-97<br>and Accessories for functional oxygen<br>saturation of arterial hemoglobin<br>(SpO2) and pulse rate (PR) to multi-<br>parameter devices for the display on<br>those devices. |
| The optional NomoLine Capnography<br>product family is intended to be | |
| connected to other medical backboard<br>devices for monitoring of breath rate and CO2. The NomoLine Capnography<br>product family is intended to be<br>connected to a patient breathing circuit<br>for monitoring of inspired/expired gases<br>during anesthesia, recovery and<br>respiratory care. The environment is the<br>operating suite, intensive care unit and<br>patient room. The intended patient<br>population is adult, pediatric and infant<br>patients. | The Masimo Rad-97 and Accessories<br>are not intended to be used as the sole<br>basis for making diagnosis or treatment<br>decisions related to suspected carbon<br>monoxide poisoning; it is intended to<br>be used in conjunction with additional<br>methods of assessing clinical signs and<br>symptoms. |
| The optional non-invasive blood<br>pressure (NiBP) module is indicated for<br>the noninvasive measurement of arterial<br>blood pressure. The NiBP module is<br>designed to measure blood pressure for<br>patient population described in the<br>following table: | The optional NomoLine Capnography<br>product family is intended to be<br>connected to other medical backboard<br>devices for monitoring of breath rate<br>and CO2. The NomoLine Capnography<br>product family is intended to be<br>connected to a patient breathing circuit<br>for monitoring of inspired/expired gases<br>during anesthesia, recovery and<br>respiratory care. The environment is the<br>operating suite, intensive care unit and<br>patient room. The intended patient<br>population is adult, pediatric and infant<br>patients. |
| The optional non-invasive blood<br>pressure (NiBP) module is indicated for<br>the noninvasive measurement of arterial<br>blood pressure. The NiBP module is | |
| Patient<br>Population | Approximate Age<br>Range |
| Newborn<br>(neonate) | Birth to 1 month of<br>age |
| Infant | 1 month to 2 years<br>of age |
| Child | 2 to 12 years of<br>age |
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white "V" inside, followed by the word "Masimo" in black, bold font. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618" in a smaller, gray font.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a stylized font, with a red circle containing a white checkmark to the left of the word. To the right of the logo is the text "Masimo Corporation" followed by the address "52 Discovery, Irvine, CA 92618".
| Adolescent | 12-21 years of age | designed to measure blood pressure for patient population described in the following table: | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Adult | 21 years of age and older | | |
| | | | |
| | Devices with Masimo technology are only to be used with Masimo sensors and cables. | Patient<br>Population | Approximate Age<br>Range |
| | | Newborn<br>(neonate) | Birth to 1 month of<br>age |
| | | Infant | 1 month to 2 years<br>of age |
| | The Rad-97 and accessories are indicated for the non-invasive continuous monitoring of PVi as a measure of relative variability of the photoplethysmograph (pleth) of adults during no motion conditions. | Child | 2 to 12 years of age |
| | | Adolescent | 12-21 years of age |
| | | Adult | 21 years of age and<br>older |
| | PVi may be used as a noninvasive dynamic indicator of fluid responsiveness in select populations of mechanically ventilated adult patients. Accuracy of PVi in predicting fluid responsiveness is variable and influenced by numerous patient, procedure and device related factors. PVi measures the variation in the plethysmography amplitude but does not provide measurements of stroke volume or cardiac output. Fluid management decisions should be based on a complete assessment of the patient's condition and | Devices with Masimo technology are only to be used with Masimo sensors and cables.<br>The Rad-97 and accessories are indicated for the non-invasive continuous monitoring of PVi as a measure of relative variability of the photoplethysmograph (pleth) of adults during no motion conditions<br>PVi may be used as a noninvasive dynamic indicator of fluid | |
| should not be based solely on PVi.<br>The Masimo Rad-97 and Accessories are<br>indicated for the continuous non-invasive<br>monitoring of Respiratory Rate from<br>photoplethysmogram (RRp) for adult and<br>pediatric patients during no motion<br>conditions in hospitals, hospital-type<br>facilities, home environments, and<br>transport within healthcare facilities. | mechanically ventilated adult patients.<br>Accuracy of PVi in predicting fluid<br>responsiveness is variable and<br>influenced by numerous patient,<br>procedure and device related factors.<br>PVi measures the variation in the<br>plethysmography amplitude but does<br>not provide measurements of stro…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.