The Masimo Rad-97 and Accessories are indicated for hospitals, hospital-type facilities, mobile, and home environments. The Masimo Rad-97 and Accessories can communicate with network systems for supplemental remote viewing and alarming (e.g., at a central station). The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) of adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused. The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of pulse rate (PR) of adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused. The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of carboxyhemoglobin saturation (SpCO) of adult, pediatric, and infant patients during no motion conditions. The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of methemoglobin saturation (SpMet) of adult, pediatric, and neonatal patients during no motion conditions. The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of total hemoglobin concentration (SpHb) of adult and pediatric patients during no motion conditions. The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of respiratory rate (RRa) for adult, pediatric, and neonatal patients during no motion conditions. In addition, the Masimo Rad-97 and Accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Rad-97 and Accessories for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) to multi-parameter devices for the display on those devices. The Masimo Rad-97 and Accessories are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms. The optional Nomoline Capnography product family is intended to be connected to other medical backboard devices for monitoring of breath rate and CO2. The Nomoline Capnography product family is intended to be connected to a patient breathing circuit for monitoring of inspired gases during anesthesia, recovery and respiratory care. The environment is the operating suite, intensive care unit and patient room. The intended patient population is adult, pediatric and infant patients. The optional non-invasive blood pressure (NIBP) module is indicated for the noninvasive measurement of arterial blood pressure. The NIBP module is designed to measure blood pressure for patient population described in the following table: Patient Population Approximate Age Range Newborn (neonate) Birth to 1 month of age Infant 1 month to 2 years of age Child 2 to 12 years of age Adolescent 12-21 years of age Adult 21 years of age and older Devices with Masimo technology are only indicated for use with Masimo sensors and cables. The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of PVi as a measure of relative variability of the photoplethysmograph (pleth) for adults and pediatrics during no motion conditions in hospitals and hospital-type facilities. PVi may be used as a noninvasive dynamic indicator of fluid responsiveness in select populations of mechanically ventilated adult patients. Accuracy of PVi in predicting fluid responsiveness is variable and influenced by numerous patient, procedure and device related factors. PVi measures the variation in the plethysmography amplitude but does not provide measurements of stroke volume or cardiac output. Fluid management decisions should be based on a complete assessment of the patient's condition and should not be based solely on PVi.
Device Story
Masimo Rad-97 is a pulse CO-oximeter; utilizes Masimo rainbow SET technology; inputs include optical signals from Masimo sensors (SpO2, SpCO, SpMet, SpHb, RRa, PVi) and optional NIBP/capnography modules; transforms light absorption data via Beer-Lambert law to calculate physiological parameters; displays numeric values on touchscreen LCD; used in hospitals, mobile, and home settings by clinicians; provides continuous monitoring data; supports remote viewing via network connectivity; PVi output serves as a dynamic indicator of fluid responsiveness in mechanically ventilated adults; aids clinical decision-making regarding fluid management; requires use with Masimo-specific sensors and cables.
Clinical Evidence
Clinical evidence consists of a literature review of 30 published studies. Six highlighted studies (182 total subjects) evaluated PVi performance in predicting fluid responsiveness in mechanically ventilated adults during surgery using fluid bolus infusions. Reported AUC values ranged from 0.74 to 0.97, supporting PVi as an indicator of fluid responsiveness.
Technological Characteristics
Touchscreen LCD display; AC/battery powered; connectivity via USB, Ethernet, Wi-Fi, Bluetooth; parameters: SpO2, PR, SpCO, SpMet, SpHb, RRa, Pi, CO2, NIBP, PVi; principle: Beer-Lambert law for light absorption; continuous operation; dimensions 9x6.5x4 inches; weight 0.92 kg.
Indications for Use
Indicated for adult, pediatric, neonatal, and infant patients in hospital, mobile, and home environments for continuous non-invasive monitoring of SpO2, PR, SpCO, SpMet, SpHb, RRa, NIBP, and PVi. PVi is indicated for mechanically ventilated adult patients as a dynamic indicator of fluid responsiveness.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 10, 2020
Masimo Corporation Linus Park Vice President, Regulatory Affairs 52 Discovery Irvine, California 92618
Re: K193626
Trade/Device Name: Masimo Rad-97 Pulse CO-Oximeter and Accessories Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, DQA, CCK, DPZ, DXN, FLL, BZQ Dated: July 28, 2020 Received: July 29, 2020
Dear Linus Park:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K193626
Device Name Masimo Rad-97 Pulse CO-Oximeter and Accessories
#### Indications for Use (Describe)
The Masimo Rad-97 and Accessories are indicated for hospital-type facilities, mobile, and home environments.
The Masimo Rad-97 and Accessories can communicate with network systems for supplemental remote viewing and alarming (e.g., at a central station).
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) of adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of pulse rate (PR) of adult, pediatic, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of carboxyhemoglobin saturation (SpCO) of adult, pediatric, and infant patients during no motion conditions.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of methemoglobin saturation (SpMet) of adult, pediatric, and neonatal patients during no motion conditions.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of total hemoglobin concentration (SpHb) of adult and pediatric patients during no motion conditions.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of respiratory rate (RRa) for adult, pediatric, and neonatal patients during no motion conditions.
In addition, the Masimo Rad-97 and Accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Rad-97 and Accessories for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) to multi-parameter devices for the display on those devices.
The Masimo Rad-97 and Accessories are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.
The optional Nomoline Capnography product family is intended to be connected to other medical backboard devices for monitoring of breath rate and CO2. The Nomoline Capnography product family is intended to a patient breathing circuit for monitoring of inspired gases during anesthesia, recovery and respiratory care. The environment is the operating suite, intensive care unit and patient population is adult, pediatic and infant patients.
The optional non-invasive blood pressure (NIBP) module is indicated for the noninvasive measurement of arterial blood pressure. The NIBP module is designed to measure blood pressure for patient population described in the following table:
{3}------------------------------------------------
Patient Population Approximate Age Range Newborn (neonate) Birth to 1 month of age Infant 1 month to 2 years of age Child 2 to 12 years of age Adolescent 12-21 years of age Adult 21 years of age and older
Devices with Masimo technology are only indicated for use with Masimo sensors and cables.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of PVi as a measure of relative variability of the photoplethysmograph (pleth) for adults and pediatics during no motion conditions in hospitals and hospital-type facilities.
PVi may be used as a noninvasive dynamic indicator of fluid responsiveness in select populations of mechanically ventilated adult patients. Accuracy of PVi in predicting fluid responsiveness is variable and influenced by numerous patient, procedure and device related factors. PVi measures the variation in the plethysmography amplitude but does not provide measurements of stroke volume or cardiac output. Fluid management decisions should be based on a complete assessment of the patient's condition and should not be based solely on PVi.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the Masimo Corporation logo. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the logo, the text "MASIMO CORPORATION" is written in gray, followed by the address "52 Discovery, Irvine, CA 92618".
| Submitter and Address of<br>Manufacturing Facility: | Masimo Corporation<br>52 Discovery<br>Irvine, CA 92618<br>Phone: (949) 297-7000 |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | July 28, 2020 |
| Contact: | Linus Park<br>Vice President, Regulatory<br>Masimo Corporation<br>Phone: (949) 297-7337 |
| Trade Name: | Masimo Rad-97 Pulse CO-Oximeter and Accessories |
| Common Name: | Oximeter. |
| Classification Regulation/<br>Product Code: | 21 CFR 870.2300, Class II/MWI |
| Additional Product Code: | 21 CFR 870.2700, Class II/DQA<br>21 CFR 862.3200, Class II/JKS<br>21 CFR 880.2910, Class II/CCK<br>21 CFR 870.2710, Class II/DPZ<br>21 CFR 870.1130, Class II/DXN<br>21 CFR 880.2910, Class II/FLL<br>21 CFR 868.2375, Class II/BZQ |
| Establishment Registration<br>Number: | 2031172 |
| Reason for Premarket<br>Notification: | Expanded indications for PVi |
| Predicate Device: | K183697 - Masimo Rad-97 and Accessories |
| Performance Standards | No performance standards for the above device have been<br>promulgated pursuant to Section 514. |
#### 5.1. Device Description
Masimo Rad-97 System and Accessories (Rad-97 product family), features a touchscreen display that continuously displays numeric values for the connected monitoring parameters. The Rad-97 product family can be operated on AC power or internal rechargeable battery.
The subject device (Rad-97 product family) is the same as the predicate (Rad-97 product family) cleared under K183697. The Rad-97 comprises of the same measurement technologies as cleared in the predicate, which includes the Masimo rainbow SET technology, capnography technology, and noninvasive blood pressure (NIBP) technology. These technologies enable the Rad-97 product family to provide noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), Pleth Variability Index (PVi), carboxyhemoglobin (SpCO), methemoglobin (SpMet), total
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a stylized font, with a red circle containing a white checkmark to the left of the word. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".
hemoglobin (SpHb), oxygen content (SpOC), acoustic respiration rate (RRa) and/or optional capnography parameters or optional noninvasive blood pressure (NIBP) parameters.
The subject of this submission is the addition of indications for PVi, an index previously cleared as an informational index.
| General Information | |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Display | |
| Display Type | Touchscreen LCD |
| Alarm | |
| Type of alarm | Visual/Audible alarm |
| Technological Characteristics | |
| Measured Parameters | NIBP, SpO2, PR, SpCO, SpMet, SpHb, RRa, Pi, CO2 |
| Calculated or Derived Parameters | PVi |
| Performance Specification | |
| SpO2, no motion | 70-100%, 1.5%, adults/pediatrics/infants 3% neonates |
| SpO2, motion | 70-100%, 1.5% Arms ,adults/ infants/ pediatrics; 3% neonates |
| SpO2, low perfusion | 70-100%, 2%, adults/pediatrics/infants/neonates |
| Pulse Rate, no motion | 25-240 bpm, 3 bpm, adults/ infants/ pediatrics/ neonates |
| Pulse Rate, motion | 25-240 bpm, 5 bpm, adults/ infants/ pediatrics /neonates |
| Pulse Rate, low perfusion | 25-240 bpm, 3 bpm, adults/ infants/ pediatrics / neonates |
| SpCO | 1-40%, 3% , adults/ infants/ pediatrics |
| SpMet | 1-15%, 1%, adults/ infants/ pediatrics/ neonates |
| SpHb | 8-17 g/dL, 1g/dL adults/ pediatrics |
| RRa | 4-70 rpm, 1 rpm, adults/pediatrics<br>4-120 rpm, 1 rpm, infants/neonates |
| CO2 | Single dry gasses at 22±5°C and 1013±40 hPa:<br>0-15 volume %: ±(0.2 volume% +2% or reading)<br>All conditions:<br>±(0.3 kPa + 4% of reading) |
| NIBP | 0-300 mmHg, ISO 81060-2 |
| Environmental | |
| Operating temperature | 0 to 35 °C (32 to 95 °F) |
| Storage temperature | -20 to 60 °C (-4 to 140 °F) |
| Operational/ storage humidity | 10 to 95%, non-condensing |
| Operating atmospheric pressure | 540 mbar to 1,060 mbar (540 hPa to 1060 hPa) |
| Mechanical | |
| Instrument Dimensions | 9 x 6.5 x 4 inch (22.9 x 16.5 x 10.2 cm) |
| Instrument Weight | 0.92 kg (2.03 lbs) without NiBP and Nomoline capnography |
| Electrical | |
| AC power | Input Voltage: 100-240 VAC, 47-63 Hz |
| Battery power | Internally rechargeable lithium ion battery |
| I/O Interface | |
| USB | USB interface |
| Nurse call | Analog output |
| Ethernet | Provides network connection to the device using RJ-45 cable |
| Wireless | Wi-Fi, Bluetooth |
| Mode of Operation | |
| Mode of operation | Continuous operation |
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the Masimo Corporation logo. On the left is a red circle with a white checkmark inside. To the right of the circle is the word "Masimo" in bold, black letters. Further to the right is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray.
#### 5.2. Intended Use/ Indications for Use
The Masimo Rad-97 and Accessories are indicated for hospitals, hospital-type facilities, mobile, and home environments.
The Masimo Rad-97 and Accessories can communicate with network systems for supplemental remote viewing and alarming (e.g., at a central station).
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) of adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of pulse rate (PR) of adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of carboxyhemoglobin saturation (SpCO) of adult, pediatric, and infant patients during no motion conditions.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of methemoglobin saturation (SpMet) of adult, pediatric, and neonatal patients during no motion conditions.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of total hemoglobin concentration (SpHb) of adult and pediatric patients during no motion conditions.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of respiratory rate (RRa) for adult, pediatric, and neonatal patients during no motion conditions.
In addition, the Masimo Rad-97 and Accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Rad-97 and Accessories for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) to multi-parameter devices for the display on those devices.
The Masimo Rad-97 and Accessories are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.
The optional Nomoline Capnography product family is intended to be connected to other medical backboard devices for monitoring of breath rate and CO2. The Nomoline Capnography product family is intended to be connected to a patient breathing circuit for monitoring of inspired gases during anesthesia, recovery and respiratory care. The environment is the operating suite, intensive care unit and patient room. The intended patient population is adult, pediatric and infant patients.
{7}------------------------------------------------
Image /page/7/Picture/2 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a red circle containing a white checkmark to the left of the word. To the right of the logo is the text "MASIMO CORPORATION" in a smaller, sans-serif font, followed by the address "52 Discovery, Irvine, CA 92618".
The optional non-invasive blood pressure (NIBP) module is indicated for the noninvasive measurement of arterial blood pressure. The NIBP module is designed to measure blood pressure for patient population described in the following table:
| Patient Population | Approximate Age Range |
|--------------------|---------------------------|
| Newborn (neonate) | Birth to 1 month of age |
| Infant | 1 month to 2 years of age |
| Child | 2 to 12 years of age |
| Adolescent | 12-21 years of age |
| Adult | 21 years of age and older |
Devices with Masimo technology are only indicated for use with Masimo sensors and cables.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of PVi as a measure of relative variability of the photoplethysmograph (pleth) for adults and pediatrics during no motion conditions in hospitals and hospital-type facilities.
PVi may be used as a noninvasive dynamic indicator of fluid responsiveness in select populations of mechanically ventilated adult patients. Accuracy of PVi in predicting fluid responsiveness is variable and influenced by numerous patient, procedure and device related factors. PV i measures the variation in the plethysmography amplitude but does not provide measurements of stroke volume or cardiac output. Fluid management decisions should be based on a complete assessment of the patient's condition and should not be based solely on PVi.
## 5.3. Technological Characteristics
#### Principle of Operation
The subject of this submission is the addition of the indication for the continuous non-invasive monitoring of Pleth Variability index (PVi) which is a measure of the relative variability of the photo plethysmograph (PPG or pleth) due to respiration. Clinically, the variations in blood pressure and blood volume is used for the monitoring of cardiac output as an indicator of hemodynamic stability, which can be affected by the balance of dynamic factors such as cardiac function, fluid changes, and vascular resistance.
Through the use of the pleth waveform used in SpO2 monitoring, the variations associated with the changes in the stroke volume can also be detected. The ability to detect these changes in blood saturation and pulsatile blood flow is based upon the Beer-Lambert law, which establishes the relationship between the absorption of light to the concentration of the absorption media. The variations in the absorption (Pi) are then used to calculate PVi.
The PVi feature relies on the relationship defined by the Frank-Starling curve along with principles used for pulse oximetry to allow for the continuous and noninvasive indication of fluid responsiveness by
{8}------------------------------------------------
Image /page/8/Picture/2 description: The image shows the logo for Masimo Corporation. On the left is a red circle with a white checkmark inside, followed by the word "Masimo" in a bold, sans-serif font. To the right of the logo is the text "MASIMO CORPORATION" in a smaller, sans-serif font, followed by the address "52 Discovery, Irvine, CA 92618".
continuously calculating the detected variability in the pleth waveform associated with a respiration cycle. Following the Frank-Starling curve, the relationship of a high, low, increasing, or decreasing PVi can also be indicative of the relative stability of the stroke volume associated with the amount of ventricular end-diastolic volume or restated as the fluid responsiveness. As many factors such as the patient and procedure may affect the accuracy of the indication of fluid responsiveness, the current submission is for PVi as an indicator of fluid responsiveness for select populations of mechanically ventilated adult patients.
## Mechanism of Action for Achieving the Intended Effect
The mechanism of action for the monitoring of PVi does not change as part of this submission. PVi is monitored by application of a Masimo SpO2 sensor to an appropriate application site. The sensor utilizes the same SpO2 wavelengths of light form the sensor to detect differing absorption levels of light. The photodetector provided on the sensor detects the transmitted light and to establish the pleth waveform that is displayed on the monitor. The PVi is calculated continuously to provide an indication of the relative stability of the pleth.
## 5.4. Summary of Technological Characteristics of Subject Device Compared to Predicate Device
## The subject device, Masimo Rad-97 with indications for PVi, and the primary predicate device, Masimo Rad-97 (K183697), have the following key similarities:
- Both devices utilize Masimo SET and Rainbow SET Technology;
- Both devices have the same intended use as a patient monitoring device;
- Both devices have the same performance specifications;
- . Both devices have the same input/ output interfaces that allows connection with external devices and with networked systems;
- . Both devices have the same principle of operation
## The subject device, Masimo Rad-97 with indications for PVi, and the primary predicate, Masimo Rad-97 (K183697), have the following key differences:
- . Subject device includes proposed indication for PVi, where the predicate device did not have an indication associated with this particular calculation.
#### 5.5. Performance Data
#### Biocompatibility Testing:
There were no new patient contacting parts or materials related to the introduction of indications for the PVi feature. Therefore, no additional biocompatibility testing was not included as part of this submission.
#### Electromagnetic Compatibility, Electrical Safety, Environmental, Mechanical and
{9}------------------------------------------------
Image /page/9/Picture/2 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the logo, the text "MASIMO CORPORATION" is displayed, followed by the address "52 Discovery, Irvine, CA 92618".
### Cleaning
There were no hardware changes related to the introduction of indications for the PVi feature. Therefore, no additional EMC, Electrical Safety, Environmental, Mechanical, and Cleaning tests were not included as part of this submission.
## Software Verification and Validation Testing
There were no software changes related to the PVi feature. Therefore, no additional software testing specific to the PVi feature was included as part of this submission.
#### Wireless and Cybersecurity Testing
There were no wireless or interconnectivity changes related to the PV i feature. There, no additional wireless or cybersecurity testing was included as part of this submission.
## Human Factors Usability Testing
There were no change made to the software or hardware user interfaces related to the PVi feature. Thus, no additional human factors or usability testing was included as part of this submission.
#### Non-clinical Testing
There were no changes to the way the PVi feature is calculated as part of this submission. Therefore, no additional performance bench testing was included as part of this submission.
#### Clinical Testing
To support the acceptability of the PVi feature for its updated indications, a comparison of the performance in the determination of fluid responsiveness using a fluid bolus infusion and monitoring the cardiac output response was conducted. To support the indication, 30 published studies were provided. Highlighted were 6 studies, which included a combined 182 subjects undergoing surgeries, where PVi was monitored for fluid responsiveness using volume expansion changes (fluid bolus infusions) on mechanically ventilated adults. The test results supported the ability of PVi to provide an indication of fluid responsiveness in those studied mechanically ventilated adult patients.
| Study Number | Area Under the Curve (AUC) |
|--------------|----------------------------|
| 1 | 0.785 |
| 2 | 0.95 |
| 3 | 0.97 |
| 4 | 0.82 |
| 5 | 0.74 |
| 6 | 0.934 |
Table 5.5-1 Summary of Comparison
{10}------------------------------------------------
Image /page/10/Picture/2 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a red circle containing a white checkmark to the left of the word. To the right of the word "Masimo" is the text "MASIMO CORPORATION" in a smaller, sans-serif font, followed by the address "52 Discovery, Irvine, CA 92618".
#### 5.6. Conclusion
The data provided as part of this submission for the subject device supports the substantially equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.