K211878 · Advanced Medical Solutions Limited · OMD · May 23, 2022 · General, Plastic Surgery
Device Facts
Record ID
K211878
Device Name
LiquiBand XL
Applicant
Advanced Medical Solutions Limited
Product Code
OMD · General, Plastic Surgery
Decision Date
May 23, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4011
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
LiquiBand® XL is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including incisions from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. LiquiBand® XL should be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of the wound during application of the liquid adhesive.
Device Story
LiquiBand® XL is a sterile, single-use topical skin adhesive system. It consists of a monomeric 2-octyl cyanoacrylate adhesive (with D&C Violet No. 2 colorant) and a self-adhering polyester mesh. The mesh provides temporary skin edge alignment; the liquid adhesive is then applied over the mesh to achieve final closure. The device is used by clinicians in a clinical setting. It acts as a microbial barrier while the adhesive film remains intact. The system includes a glass ampoule of adhesive within a plastic applicator body containing a porous block with a quaternary ammonium salt accelerant and a foam tip. It is intended to hold skin edges together to facilitate healing.
Clinical Evidence
No clinical data was submitted or relied upon. Substantial equivalence was demonstrated through bench testing, including peel/shear adhesion (ASTM D3330, D3654), tensile properties (ASTM D882), lap-shear/T-peel/tension/wound closure strength (ASTM F2255, F2256, F2258, F2458), polymerization kinetics, microbial barrier testing, and animal wound healing studies. Biocompatibility was evaluated per ISO 10993-1, including cytotoxicity, sensitization, irritation, pyrogenicity, and systemic toxicity.
Technological Characteristics
Sterile, single-use applicator containing 2-octyl cyanoacrylate monomer and D&C Violet No. 2. Includes polyester mesh with pressure-sensitive adhesive. Accelerant: quaternary ammonium salt. Bench testing per ASTM D3330, D3654, D882, F2255, F2256, F2258, F2458. Biocompatibility per ISO 10993-1. No electronic components or software.
Indications for Use
Indicated for topical closure of easily approximated surgical incisions (including minimally invasive) and simple, cleansed, trauma-induced lacerations. Must be used with deep dermal stitches. No specific age or gender restrictions.
Regulatory Classification
Identification
A tissue adhesive with adjunct wound closure device intended for the topical approximation of skin is a device indicated for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. It may be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of wound during application of the liquid adhesive.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Guidance for Industry and FDA Staff; Class II Special Controls Guidance Document: Tissue Adhesive with Adjunct Wound Closure Device Intended for the Topical Approximation of Skin.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Build Correspondence
Image /page/0/Picture/2 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 23, 2022
Advanced Medical Solutions Limited Martin Mitchell Senior Regulatory Affairs Associate Western Wood Way , Langage Science Park Plymouth, Devon PL7 5BG United Kingdom
Re: K211878
Trade/Device Name: LiquiBand® XL Regulation Number: 21 CFR 878.4011 Regulation Name: Tissue Adhesive With Adjunct Wound Closure Device For Topical Approximation Of Skin Regulatory Class: Class II Product Code: OMD Dated: May 4, 2022 Received: May 4, 2022
Dear Martin Mitchell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211878
Device Name LiquiBand® XL
LiquiBand® XL is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including incisions from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. LiquiBand® XL should be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of the wound during application of the liquid adhesive.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|----------------|
| <span> <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) </span> | <span> </span> |
| <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | <span> </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Advanced Medical Solutions Ltd. The logo features a stylized blue letter "A" with a swoosh extending from the top right. Below the logo, the company name "Advanced Medical Solutions Ltd" is written in blue, sans-serif font. The overall design is clean and professional.
Advanced Medical Solutions Limited Western Wood Way, Langage Science Park, Plymouth, Devon, PL7 5BG, United Kingdom Tel: +44 (0) 1752 209955 Fax: +44 (0) 1752 209956 Web: www.admedsol.com Registered in England 2666957 VAT No. GB 636 5551 27
## 510(k) Summary
| Submitted by: | Advanced Medical Solutions Limited<br>Western Wood Way<br>Langage Science Park<br>Plymouth, Devon<br>PL7 5BG<br>United Kingdom<br>Tel: +44 (0)1752 209955<br>Fax: +44 (0) 1752 209956 |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Martin Mitchell<br>Regulatory Affairs Manager |
| Date of Summary: | 19th May 2022 |
| Device Name: | LiquiBand® XL |
| Common Name: | Cutaneous Tissue Adhesive With Mesh |
| Classification Name: | Tissue adhesive with adjunct wound closure device for topical approximation of skin |
| Regulatory Number: | 21 CFR 878.4011 |
| Device Class: | Class II |
| Product Code: | OMD |
| Predicate Device<br>Device Name:<br>510(k) Clearance: | Exofin Fusion Skin Closure System<br>K191461 |
| Reference Device<br>Device Name:<br>510(k) Clearance: | LiquiBand® Plus<br>K183570 |
Image /page/3/Picture/4 description: The image shows two logos. The first logo is for the UK Authorized Economic Operator, and it features the letters "AEO" in bold black font, with a map of the UK inside of a circle. The second logo is for BSI, and it features a heart-shaped design with the letters "bsi" above it. The certificate number MD 695790 is printed below the BSI logo.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Advanced Medical Solutions Ltd. The logo features a stylized blue letter A with a swoosh above the text. The text "Advanced Medical Solutions Ltd" is written in blue font below the logo.
#### Device Description
LiquiBand® XL is a sterile, liquid topical skin adhesive containing a monomeric (2-octyl cyanoacrylate) formulation and the colorant D & C Violet No. 2. It is provided in a single-use applicator. As applied to skin, the liquid topical skin adhesive is slightly more viscous than water and polymerizes within minutes.
In vitro studies have shown that LiguiBand® XL acts as a barrier to microbial penetration as long as the adhesive film remains intact. Clinical studies were not conducted to demonstrate microbial barrier properties.
LiquiBand® XL also incorporates a self-adhering mesh that is approximated skin edges to provide temporary skin edge alignment to an incision until the liquid adhesive is applied to achieve skin closure.
## Indications for Use
LiquiBand® XL is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including incisions from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. LiquiBand® XL should be used in conjunction with, but not in place of, deep dermal stitches. Additionally, the adjunct wound closure device component maintains temporary skin edge alignment along the length of the wound during application of the liquid adhesive.
# Comparison of Technological Characteristics with the Predicate Device
The technological characteristics of the LiquiBand® XL and the predicate device are similar, in that they both contain the same technological characteristics:
| Device | Subject device: LiquiBand® XL | Predicate device: Exofin Fusion Skin<br>Closure System | Comparison |
|------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Adhesive | 2-octyl cyanoacrylate adhesive with<br>D&C Violet #2 colorant | 2-octyl cyanoacrylate adhesive with<br>D&C Violet #2 colorant | Same |
| Accelerant | Quaternary ammonium salt | Quaternary ammonium salt | Same |
| Applicator | Sterile single use applicator;<br>comprising of a glass ampoule<br>(containing the adhesive) within a<br>plastic applicator body with a<br>porous block (containing the<br>accelerant) with a foam applicator<br>tip | Sterile single use applicator;<br>comprised of an aluminum tube<br>(containing the adhesive) with a<br>porous disk (containing the<br>accelerant) with an applicator tip | Equivalent |
| Mesh | Polyester mesh with pressure<br>sensitive adhesive | Polyester mesh with pressure<br>sensitive adhesive | Same |
Image /page/4/Picture/11 description: The image shows two logos side by side. The first logo on the left has the letters "AEO" in large, bold font, with an image of the United Kingdom inside a circle to the right of the letters. Below the letters, the text "UK Authorised Economic Operator" is printed. The second logo on the right has the letters "bsi" above a heart-shaped design. To the right of this logo is a square containing a crown above a check mark, with the text "UKAS MANAGEMENT SYSTEMS" below. The number "0003" is printed below this text, and the text "Certificate No. MD 695790" is printed below the entire logo.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Advanced Medical Solutions Ltd. The logo features a stylized blue "A" with a swoosh extending from the top right. Below the logo, the text "Advanced Medical Solutions Ltd" is written in blue, with each word on the same line.
Advanced Medical Solutions Limited Western Wood Way. Langage Science Park, Plymouth. Devon, PL7 5BG, United Kingdom Tel: +44 (0) 1752 209955 Fax: +44 (0) 1752 209956 Web: www.admedsol.com Registered in England 2666957 VAT No. GB 636 5551 27
## Performance Data
Testing was performed in accordance with the FDA special controls guidance document for "Tissue Adhesive with Adjunct Wound Closure Device Intended for the Topical Approximation of Skin"
## Performance Testing
The following tests were performed on the LiquiBand® XL to demonstrate substantial equivalence to the predicate device:
- Peel Adhesion of Pressure-Sensitive Tape (ASTM D3330/D3330M-04) ●
- Shear Adhesion of Pressure-Sensitive Tapes (ASTM D3654/D3654M-06) ●
- Tensile Properties of Thin Plastic Sheeting (ASTM D882-12)
- Lap-shear strength (ASTM F2255-05)
- T-peel adhesion strength (ASTM F2256-05)
- Adhesive strength in tension (ASTM F2258-05)
- Wound closure strength (ASTM F2458-05)
- Adhesive degradation study
- Heat of polymerization
- Force to actuate and express
- Viscosity
- Polymerization set time
- Microbial barrier testing
- Quality of adhesive film on mesh
- Accelerant degradation study ●
- . Animal wound healing study
# Biocompatibility
The biological evaluation of the LiguiBand® XL was performed in accordance with FDA guidance on the use of ISO 10993-1. The following test reports were provided in this submission:
- Cytotoxicity
- Sensitization
- . Irritation
- Material mediated pyrogenicity
- Acute systemic toxicity
- Chemical characterization and toxicological risk assessment ●
#### Clinical Testing
- No clinical testing has been submitted, referenced, or relied upon for determining substantial ● equivalence.
#### Conclusions
Based on the intended use, technological characteristics, safety and performance testing, LiquiBand® XL has been demonstrated to be substantially equivalent to the predicate device, Exofin Fusion Skin Closure System.
Image /page/5/Picture/34 description: The image contains three logos. The first logo on the left has the letters AEO in bold black font, with a globe containing the UK to the right of the letters. Below the letters, the text reads "UK Authorised Economic Operator". The second logo is a circle with the letters "bsi" above a heart-shaped figure with a line through it. The third logo is a square with a crown above a circle containing a checkmark. Below the circle, the text reads "UKAS MANAGEMENT SYSTEMS 0003".
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.