K211741 · Lake Region Medical · DQX · Nov 22, 2021 · Cardiovascular
Device Facts
Record ID
K211741
Device Name
Pre-Formed Extra Support Guidewire
Applicant
Lake Region Medical
Product Code
DQX · Cardiovascular
Decision Date
Nov 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
Pre-Formed Extra Support guidewire is intended to facilitate the introduction and placement of interventional devices within the chambers of the heart, including those used during transcatheter aortic valve procedures.
Device Story
Pre-Formed Extra Support guidewire facilitates introduction/placement of interventional devices in heart chambers during TAVR. Device features 0.035" diameter, 275cm length, and unique pre-shaped spiral distal tip to ensure placement. Composed of stainless-steel core and coil, with PTFE coating. Used by physicians in clinical settings. Device provides mechanical support and guidance for interventional tools; benefits patient by ensuring accurate placement of valves or other devices during cardiac procedures.
Clinical Evidence
Bench testing and animal study (porcine model) performed. Bench testing included thermal conditioning, package integrity (ASTM F88-15, F1929-15, F2096-11), dimensional inspection, lubricity, pull testing, 3-point bend, coating adhesion, radiopacity, corrosion, particulate residue, fracture, flex, torque fatigue, kink resistance, and biocompatibility (ISO 10993-1). Animal study demonstrated performance equivalent to predicate and met all customer requirement specifications.
Technological Characteristics
Stainless steel core and coil; PTFE coating. 0.035" diameter, 275cm length. Pre-shaped spiral distal tip available in three sizes. Sterilized via ethylene oxide. Testing standards: ASTM D4169, ISTA 2A, ASTM F88-15, ASTM F1929-15, ASTM F2096-11, EN ISO 11070, ISO 10993-1.
Indications for Use
Indicated for patients undergoing interventional procedures within the heart chambers, including transcatheter aortic valve replacement (TAVR).
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 22, 2021
Lake Region Medical Deko Hussein Sr. Regulatory Specialist 340 Lake Hazeltine Drive Chaska, Minnesota 55318
Re: K211741
Trade/Device Name: Pre-Formed Extra Support Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guide wire Regulatory Class: Class II Product Code: DQX Dated: October 18, 2021 Received: October 21, 2021
Dear Deko Hussein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Jaime Raben Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211741
Device Name Pre-Formed Extra Support Guidewire
Indications for Use (Describe)
Pre-Formed Extra Support guidewire is intended to facilitate the introduction and placement of interventional devices within the chambers of the heart, including those used during transcatheter aortic valve procedures.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
|--|----------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
|--|---------------------------------------------|
|--|---------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K211741 510(k) Summary
### Submitter Address:
Establishment Registration Number: Primary Contact Person: Email: Phone: Prepared:
#### Subiect Device
Name of Device (Trade): Common or Usual Name: Classification: Regulatory Class: Product Code:
#### Predicate Device
Device Name: Manufacturer: 510(k) Number: Regulatory Class: Product Code:
#### Reference Device
- Device Name: Manufacturer: 510(k) Number: Regulatory Class: Product Code:
Lake Region Medical 340 Lake Hazeltine Drive Chaska, MN 55318 2126666 Deko Hussein, Sr. Regulatory Specialist deko.hussein@integer.net (952) 641-8302 June 4, 2021
Pre-Formed Extra Support Guidewire Catheter Guidewire Catheter Guidewire (21 CFR 870.1330) Class II DOX
Pre-Formed Guidewire Lake Region Medical K151244 (Cleared June 11, 2015) Class II per 21 CFR 870.1330 DQX
Lunderquist Extra Stiff William Cook Europe ApS K171513 Class II per 21 CFR 870.1330 DQX
#### Device Description
The Pre-Formed Guidewire Extra Support guidewire is designed to facilitate device placement during TAVR procedures and has a unique pre-shaped spiral distal tip to ensure placement within the heart. The guidewire is intended for single use.
The guidewire has a 0.035" diameter and is 275cm in length. The guidewire composed of two primary components: a core, and a coil. Both components are made of stainless steel. The core
{4}------------------------------------------------
wire component is a piece of stainless-steel wire which is ground on the distal end to fit into the coil and provide flexibility. The proximal end of the core wire is coated with grev polytetrafluoroethylene (PTFE) coating. The coil component is fitted over the distal end of the core and is a stainless-steel coiled with green PTFE coating. The coil and core components are secured together using a weld on the distal and a glue joint at the proximal end of the coil, forming the guidewire. The pre-shaped distal end of the guidewire is available in three sizes (extra-small, small and large).
The Pre-Formed Extra Support guidewire is sterilized using ethylene oxide. There are no accessories packaged with the Pre-Formed Extra Support guidewire.
# Indications for Use
Pre-Formed Extra Support guidewire is intended to facilitate the introduction and placement of interventional devices within the chambers of the heart, including those used during transcatheter aortic valve procedures.
### Comparison of Technological Characteristics with the Predicate Devices
The Pre-Formed Extra Support has the same indication for use/intended use and principles of operation as the legally marketed Pre-Formed guidewire (K151244). In addition, the functional characteristics of the Pre-Formed Extra Support are substantially equivalent to the Pre-Formed guidewire, with similar materials, dimensions, and method of construction. The design modifications to the new device include differences in material used on the coating for the core, the proximal joint and the partial coil design on the distal end. Performance and safety testing have shown that these modifications have not raised any new questions of safety or efficacy and the device continues to meet its intended use.
### Summarv of Testing
The determination of substantial equivalence includes an assessment of non-clinical (bench) performance testing. This testing was performed in order to demonstrate that the Pre-Formed Extra Support guidewire met applicable design and performance requirements and to support a determination of substantial equivalence. The following testing was conducted according to established procedures and samples were analyzed according to predetermined acceptance criteria. FDA guidance Coronary, Peripheral and Neurovascular Guidewires Performance Tests and Recommended Labeling (October 10, 2019) was utilized.
- . Thermal Conditional and Packaging Distribution - Tested in accordance with ASTM D4169 (Distribution simulation) and ISTA 2A (environmental conditioning). The predetermined acceptance criteria were met.
- Package Integrity evaluation, including:
- . Visual Inspection - Tested in accordance with EN ISO 11070 & FDA Guidance. The predetermined acceptance criteria were met.
- . Pouch Peel Tear - Tested in accordance with ASTM F88-15. The predetermined acceptance criteria were met.
{5}------------------------------------------------
- . Seal Integrity – Tested in accordance with ASTM F1929-15. The predetermined acceptance criteria were met.
- . Bubble Leak Testing - Tested in accordance with ASTM F2096-11(2019). The predetermined acceptance criteria were met.
- Dimensional Inspection, including:
- Length, Outer Diameter Tested in accordance with EN ISO 11070 & ● FDA Guidance. The predetermined acceptance criteria were met.
- . Curve – Tested in accordance with internal procedures. The predetermined acceptance criteria were met.
- Lubricity Tested in accordance with FDA Guidance. The predetermined acceptance o criteria were met.
- Guidewire Distal Joint Pull Test Tested in accordance with EN ISO 11070 & FDA ● Guidance. The predetermined acceptance criteria were met.
- 3 Point Bend Tested in accordance with internal procedures. The predetermined . acceptance criteria were met.
- Coating Adhesion Tested in accordance with FDA Guidance. The predetermined . acceptance criteria were met.
- . Radiopacity - Tested in accordance with EN ISO 11070 & FDA Guidance. The predetermined acceptance criteria were met.
- ISO Corrosion Test Tested in accordance with EN ISO 11070 & FDA . Guidance. The predetermined acceptance criteria were met.
- . Particulate Residue Test - Tested in accordance with FDA Guidance. The predetermined acceptance criteria were met.
- ISO Fracture Test Tested in accordance with EN ISO 11070. The predetermined o acceptance criteria were met.
- ISO Flex Test Tested in accordance with EN ISO 11070. The predetermined . acceptance criteria were met.
- Torque Fatigue Tested in accordance with FDA Guidance. The predetermined . acceptance criteria were met.
- . Kink Resistance - Tested in accordance with FDA guidance. The predetermined acceptance criteria were met.
- Biocompatibility Tested in accordance with ISO 10993-1 and all applicable parts of this . standard series. The battery of tests included cytotoxicity, sensitization, irritation, systemic toxicity, material mediated pyrogenicity, hemocompatibility, and thrombogenicity. The predetermined acceptance criteria were met.
# Performance Data (Animal)
The Pre-Formed Extra Support guidewire was evaluated in a porcine model in accordance with 21 CFR Part 58 for Good Laboratory Practice (GLP) for Non- Clinical Laboratory studies. The subject device Pre-Formed Extra Support guidewire shares the same intended use and fundamental scientific technology as the predicate and demonstrated all safety and performance objectives as the device performed equivalent to the predicate, Pre-Formed guidewire, and
{6}------------------------------------------------
demonstrated all Customer Requirement Specifications as measured by post procedure feedback forms.
Conclusion: The Pre-Formed Extra Support guidewire met all design input requirements based on the intended use and supports the conclusion that the Pre-Formed Extra Support does not raise new questions of safety and effectiveness. The results of these tests support a determination of substantial equivalence to the predicate device, the Pre-Formed Guidewire (K151244).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.