K151244 · Lake Region Medical · DQX · Jun 11, 2015 · Cardiovascular
Device Facts
Record ID
K151244
Device Name
Pre-Formed Guidewire
Applicant
Lake Region Medical
Product Code
DQX · Cardiovascular
Decision Date
Jun 11, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
Pre-Formed Guidewires are intended to facilitate the introduction and placement of interventional devices within the chambers of the heart, including those used within transcatheter aortic valve procedures.
Device Story
Pre-Formed Guidewire is a 0.035" diameter, 260-300cm long stainless steel wire used to facilitate placement of interventional devices in heart chambers. Device consists of inner stainless steel core and outer stainless steel coil coated in green PTFE; distal end features double-curve shape. Used by clinicians in interventional cardiology settings. Guidewire acts as a track for delivery of interventional devices, such as those for transcatheter aortic valve procedures. Benefits include improved navigation and placement of interventional tools within cardiac chambers. Device is not for use in coronary or cerebrovascular arteries.
Clinical Evidence
Bench testing and a GLP animal study were performed. Bench testing included dimensional, tensile strength, tip flexibility, catheter compatibility, packaging, tip shape retention, particulate, visual, fracture, flex, corrosion resistance, and linear stiffness tests. A GLP animal study evaluated safety and performance compared to the predicate. No new biocompatibility testing was required as materials remained unchanged. Results demonstrate the device is as safe and effective as the predicate.
Technological Characteristics
Materials: Stainless steel (ASTM A313) core and coil; coil coated in green PTFE. Dimensions: 0.035" diameter, 260-300cm length. Tip: Double-curve shape. Standards: ISO 10993, ISO 11070, ISO 11135, ISO 11138. Non-powered, mechanical device.
Indications for Use
Indicated for facilitating introduction and placement of interventional devices within heart chambers, including transcatheter aortic valve procedures. Contraindicated for use in cerebrovasculature or coronary arteries.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Pre-Formed Guidewires (Lake Region Medical) (K130798)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, which is a traditional symbol of medicine, with three human profiles incorporated into the design.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 11, 2015
Lake Region Medical Mathew Pexa Sr. Regulatory Specialist 340 Lake Hazeltine Dr. Chaska, Minnesota 55318
Re: K151244
Trade/Device Name: Pre-Formed Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DOX Dated: May 8, 2015 Received: May 11, 2015
Dear Mr. Pexa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrenner
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Lake Region Medical. The logo consists of a stylized heart shape in red and gray, followed by the words "Lake Region" in a simple, sans-serif font. Below "Lake Region" is the word "MEDICAL" in a smaller, lighter font.
# INDICATIONS FOR USE STATEMENT
## 510(k) NUMBER (IF KNOWN): K151244
### DEVICE NAME: Pre-Formed Guidewire
### INDICATIONS FOR USE:
The Pre-Formed guidewires are intended to facilitate the introduction and placement of interventional devices within the chambers of the heart including those used within transcatheter aortic valve procedures.
PRESCRIPTION
USE
AND/OR
OVER-THE-COUNTER USE
(Part 21 CFR 801 Subpart D)
×
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
> PAGE 1 of ।
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Lake Region Medical. The logo consists of a geometric shape on the left, which is a combination of red and gray triangles forming a heart-like shape. To the right of the geometric shape is the text "Lake Region" in a clean, sans-serif font, with the "i" in "Region" having a red square dot. Below "Lake Region" and slightly offset to the right is the word "MEDICAL" in a smaller, lighter font.
# 510(k) SUMMARY
May 8, 2015
This summary is being included in the submission in lieu of a statement of availability.
| 2.1 | MANUFACTURER / REGISTRATION INFORMATION | | | |
|----------------------------------|-----------------------------------------|-----------------|---------------------------|--|
| Lake Region Medical | | Telephone: | 952-448-5111 | |
| 340 Lake Hazeltine Dr. | | Fax: | 952-448-3441 | |
| Chaska, Mn 55318 | | Contact Person: | Mathew Pexa | |
| FDA REGISTRATION NUMBER: 2126666 | | Title: | Sr. Regulatory Specialist | |
| 2.2 | DEVICE TRADE NAME / PROPRIETARY NAME | | | |
Pre-Formed Guidewires
2.3 DEVICE COMMON NAMES / USUAL NAMES / CLASSIFICATION NAMES CATHETER GUIDEWIRE (DQX)
2.4 CLASS OF DEVICE
These devices are commonly known as guides, guidewires, or spring guidewires. The current classification name and product code is Catheter Guidewire (DQX) and is considered a Class II device per 21 CFR Part 870.1330.
| 2.5 | IDENTIFICATION OF PREDICATE DEVICE(s) |
|-----|---------------------------------------|
|-----|---------------------------------------|
Pre-Formed Guidewires (Lake Region Medical) K130798
#### DEVICE DESCRIPTION 2.6
The 0.035" diameter, 260-300cm length guidewire is composed of two primary wire components: a core and a coil. Both components are manufactured from Stainless Steel per ASTM A313. The core wire is a stainless steel wire which forms the inner body of the guidewire. The coil component is the guidewire's outer layer and is a stainless steel wire coated in Green Polytetrafluoroethylene (PTFE), The coil and the core components are welded together on the distal and proximal ends, forming the guidewire. The distal ends are shaped into a double-curve. The following table shows the characteristics of the product family:
| Outside Diameter: | 0.035" |
|-------------------|--------------------------------------|
| Overall Length: | 260cm to 300cm |
| Tip Shape: | Double-Curve |
| Tip Sizes: | Extra Small - Large |
| 2.7 | COMPLIANCE WITH APPLICABLE STANDARDS |
LRM has determined that no mandatory standards, performance standards, or special controls have been established for these devices under Section 514 of the Medical Device Amendments to Federal Food, Drug, and Cosmetic or by any subsequent regulatory action. However, the following standards are referenced within this filing: ISO 10993, ISO 11070, ISO 11135, ISO 11138.
2.8 INTENDED USE STATEMENT
Pre-Formed Guidewires are intended to facilitate the introduction and placement of interventional devices within the chambers of the heart, including those used within transcatheter aortic valve procedures.
NOTE: This modification does not alter its intended use.
2.9 CONTRAINDICATIONS
This wire is not intended for use in the cerebrovasculature or coronary arteries.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Lake Region Medical. The logo consists of a geometric shape on the left, which is a combination of red and gray squares forming a heart-like shape. To the right of the shape, the words "Lake Region" are written in a simple, sans-serif font, with the word "MEDICAL" in smaller letters below.
#### 2.10 TECHNOLOGICAL CHARACTERISTICS
The design specifications are substantially equivalent to the existing Pre-Formed Guidewires
#### QUALITY SYSTEM CONTROL 2.11
### DESIGN CONTROLS
LRM is in conformance with the design control procedure requirements as specified in 21 CFR Part 820.30. Risk analysis was completed by means of a Failure Mode and Effects Analysis and all verification and validation activities resulted in the ability to demonstrate that the predetermined acceptance criteria were met.
### MATERIALS / SUPPLIER / PRODUCT / PROCESS CONTROLS
LRM has formal quality systems in placed to assure that each product manufactured remains equivalent to the predicate products, and that the changes will not have an adverse effect on safety or effective use of the product. The quality systems include Engineering Change Order Review, Material Qualification, Supplier Qualification, Product Qualification, and Process Qualification. These controls are applied to each product size / group.
#### 2.12 QUALIFICATION TESTING
The conclusions drawn from bench testing and biocompatibility testing demonstrate compliance with the design input summary which shows the device is at least as safe and effective as the current legally marketed device.
### BENCH TESTING
In order to demonstrate equivalence of the guidewire, Lake Region Medical performed bench testing to establish requirements. Test devices were manufactured and inspected according to established requirements for visual/tactile, dimensional and mechanical attributes. The devices were then subjected to the following test methods to show the devices comply with the design input summary:
- Dimensional -
- -FDA Tensile Strength
- -FDA Tip Flexibility
- -FDA Catheter Compatibility
- -Packaging Study
- -Tip Shape Retention
- -Particulate
- ISO Visual "
- -ISO Fracture
- -ISO Flex
- -ISO Corrosion Resistance
- -Linear Stiffness
## BIOCOMPATIBILITY TESTING
There were no changes to materials used in the design or manufacture of the Pre-Formed guidewire family; therefore, no additional biocompatibility tests were completed.
### ANIMAL STUDY
A GLP Animal study was completed to evaluate safety and performance of the guidewire compared to the currently marketed device. The study shows the guidewires are substantially equivalent to the legally marketed device.
#### 2.13 CHANGES FROM PREDICATE DEVICE
- Additional tip shape, Extra Small -
- -Distal core grind diameter minimum made larger
#### 2.14 SUBSTANTIAL EQUIVALENCE DATA
The changes included in this 510(k) for the Pre-Formed guidewire family do not change the indications for use of the Mandrel guidewires and is not a change to the fundamental scientific technology. The information summarized above shows the device will perform as well as the previously marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.