The sponsor conducted a literature review of real-world clinical studies to evaluate the safety and efficacy of the Rotarex™ Atherectomy System in treating in-stent restenosis (ISR), iliac artery occlusions, and bypass graft debulking compared to other standard-of-care treatments (PTA, thrombectomy devices, and open surgery).
Literature review; In-stent restenosis; Peripheral artery disease; Retrospective clinical outcomes
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Literature review of Rotarex™ Atherectomy System clinical outcomes; Systematic literature review of clinical studies
Patients with in-stent restenosis, iliac artery occlusions, and bypass graft occlusions
Percutaneous transluminal angioplasty (PTA), Ekos, Indigo, and open surgery
Safety, technical success, and clinical outcome
Indications for Use
The Rotarex™ Atherectomy System is intended for use as an atherectomy device and to break up and remove thrombus from native peripheral arteries or peripheral arteries fitted with stents, stent grafts or natificial bypasses.
Device Story
Rotarex™ Atherectomy System is a percutaneous, over-the-wire device for removing atheromatic plaque and thrombi. System includes a single-use catheter, guidewire, collecting bag, and a drive system (control unit, motor, foot switch). Catheter features a rotating head with abrading facets and an internal helix; driven by a motor at 40,000–60,000 RPM. Mechanical rotation creates aspiration, drawing material into the catheter head for removal. Used in clinical settings by physicians. Output is the physical removal of occlusive material from peripheral arteries, improving vessel patency. Benefits include debulking of occlusions in native arteries, stents, stent grafts, and bypasses, potentially reducing need for open surgery.
Clinical Evidence
No new clinical trials were conducted. Substantial equivalence supported by a review of existing medical literature regarding the Rotarex™ Atherectomy System's use in treating in-stent restenosis (ISR), iliac artery occlusions, and bypass graft debulking. Literature compared outcomes against PTA, other thrombectomy devices (Ekos, Indigo), and open surgery. Results supported the benefit-risk profile for the expanded indications.
Technological Characteristics
Catheter components: flexible outer covering, rotating head (two metal cylinders with side openings), rotating helix. Materials: latex and phthalate-free. Guidewire: Nitinol core, PTFE coating, hydrophilic distal coating. Energy: electromechanical (motor-driven). Sterilization: ethylene oxide. Connectivity: standalone drive system. Sizes: 6F and 8F diameters; 85–135 cm lengths.
Indications for Use
Indicated for patients requiring atherectomy or thrombectomy in native peripheral arteries or peripheral arteries fitted with stents, stent grafts, or native/artificial bypasses. Contraindicated in patients unsuitable for atherectomy/thrombectomy, or for use in cardiopulmonary, coronary, carotid, cerebral, renal, or venous vasculature, or in undersized vessels.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 14, 2021
C.R. Bard, Inc. Aaron Conovaloff Regulatory Affairs Manager 1625 West 3rd St Tempe, Arizona 85281
Re: K211738
Trade/Device Name: Rotarex™ Atherectomy System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW, DQX Dated: August 17, 2021 Received: August 18, 2021
Dear Aaron Conovaloff:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211738
Device Name Rotarex™ Atherectomy System
Indications for Use (Describe)
The Rotarex™ Atherectomy System is intended for use as an atherectomy device and to break up and remove thrombus from native peripheral arteries or peripheral arteries fitted with stents, stent grafts or natificial bypasses.
Contraindications:
- · In patients not suitable for atherectomy/thrombectomy
- · In the cardiopulmonary, coronary, carotid, cerebral and renal vasculature
- · In vessels that are undersized for the particular device used
- In the venous vasculature
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Rotarex™ Atherectomy System
#### 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (1)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
#### Submitter Information:
| Applicant: | Bard Peripheral Vascular, Inc<br>1625 West 3rd Street<br>Tempe, Arizona 85281 |
|------------|-------------------------------------------------------------------------------|
| Phone: | 602-830-5453 |
| Fax: | 321-949-0436 |
| Contact: | Aaron Conovaloff, Regulatory Affairs Manager |
| Date: | June 4, 2021 |
### Subiect Device Name:
| Device Trade Name: | Rotarex™ Atherectomy System |
|-----------------------|------------------------------|
| Common or Usual Name: | Intraluminal artery stripper |
| Product Code: | MCW, DQX |
| Classification: | Class II |
| Review Panel: | Cardiovascular |
| Regulation Number: | 21 CFR 870.4875 |
#### Predicate Device:
- Straub Endovascular System (K172315; cleared April 12, 2018) .
#### Reference Device:
- B-Laser Atherectomy System (K181642; cleared October 5, 2018) .
#### Device Description:
The Rotarex™ Atherectomy System is made up of a single use Rotarex™ Atherectomy Catheter Set and the Drive System, consisting of the control unit, motor and foot switch. For more detail on the Drive System, please see the manual provided with the Drive System. Read and understand all instructions prior to attempting any atherectomy procedure with the Rotarex™ Atherectomy System or any of its components.
The Rotarex™ Atherectomy Catheter Set is composed of multiple components, including the Rotarex™ Atherectomy Catheter, guidewire, collecting bag, and sterile drape. Rotarex™ Atherectomy Catheters are over-the-wire, single use, percutaneous devices for the removal of
{4}------------------------------------------------
atheromatic plaque and thrombi in native arteries fitted with stents, stent grafts or native or artificial bypasses. The catheters are latex and phthalate free, and consist of a flexible outer covering, a rotating head, and a rotating helix which runs the length of the catheter. A lumen for the passage of the supplied guidewire runs the entire length of the helix and through the head of the catheter.
The catheter head is made up of two overlying metal cylinders, with two side openings. The outer cylinder is connected to the rotating helix, and the inner cylinder to the catheter shaft. The helix and the catheter head rotate at approximately 40,000-60,000 rpm depending on the model, by means of a gear box in the catheter housing and a motor contained within the catheter handle driven by the Drive System. The rotating outer cylinder is fitted with abrading facets at its foremost tip.
The Rotarex™ Atherectomy System Guidewires are 0.018" and are included with each Rotarex™ Atherectomy Catheter Set. These guidewires consist of a Nitinol core and a PTFE coating with a 9.5cm hydrophilic distal coating and a 4 cm angled flex tip.
All components within the Rotarex™ Atherectomy Catheter Sets are supplied sterile for single use only. The method of sterilization is ethylene oxide.
| Catheter<br>Size | Minimum<br>Vessel<br>Diameter | Catheter<br>External<br>Diameter | Nominal<br>Rotation<br>(RPM) | Maximum<br>Aspiration<br>(ml/min) | Guidewire<br>Size &<br>Length |
|------------------|-------------------------------|----------------------------------|------------------------------|-----------------------------------|-------------------------------|
| 6F<br>110 cm | 3 mm | 2.0 mm | 60,000 | 45 | 0.018"<br>270 cm |
| 6F<br>135 cm | 3 mm | 2.0 mm | 60,000 | 45 | 0.018"<br>320 cm |
| 8F<br>85 cm | 5 mm | 2.7 mm | 40,000 | 75 | 0.018"<br>220 cm |
| 8F<br>110 cm | 5 mm | 2.7 mm | 40,000 | 75 | 0.018"<br>270 cm |
# Indications for Use of Device:
The Rotarex™ Atherectomy System is intended for use as an atherectomy device and to break up and remove thrombus from native peripheral arteries or peripheral arteries fitted with stents, stent grafts or native or artificial bypasses.
# Contraindications:
- In patients not suitable for atherectomy/thrombectomy
- In the cardiopulmonary, coronary, carotid, cerebral and renal vasculature ●
- In vessels that are undersized for the particular device used
- . In the venous vasculature
# Comparison to Predicate Devices:
The subject Rotarex™ Atherectomy System has the following similarities to the predicate Rotarex® S catheter (a component of the Straub Endovascular System: K172315, cleared April 12, 2018):
- . Intended use
- . Target population/conditions of use (anatomical location of use, how device interacts with
{5}------------------------------------------------
other devices, interaction with patient)
- Principles of operation/mechanism of action ●
- . Fundamental scientific technology
- Sterilization method ●
- . Sterility assurance level
The subject Rotarex™ Atherectomy System has the following similarities to the reference B-Laser Atherectomy System (K181642, cleared October 5, 2018):
- Intended use .
- . Indications for use
- . Target population/conditions of use (anatomical location of use, how device interacts with other devices, interaction with patient)
- . Sterilization method
The subject Rotarex™ Atherectomy System incorporates the design of the Rotarex® S catheter predicate with the in-stent restenosis indication of the reference B-Laser Atherectomy System device. Note that the design, materials, and methods of manufacture of the subject device and the Rotarex® S catheter predicate are identical; the only differences are the expansion of the indications for use to include use in peripheral arteries fitted with stents or stent grafts and in native or artificial bypasses, and to remove several of the contraindications in the IFU.
The removed contraindications are not critical to the intended use of this device. Some of these contraindications are being removed completely, either in light of data to support their removal or because the corresponding risk information is already communicated in other areas of the IFU. Other contraindications are being restated as warnings have been determined to be a more appropriate way to present the corresponding risk information. Because the corresponding risk information is still presented appropriately in the IFU, removal of these contraindications does not affect the safety and effectiveness of the device when used as labeled.
### Performance Data:
To demonstrate substantial equivalence of the subject device to the predicate device, its technological characteristics and performance criteria were evaluated. Using FDA Guidance Documents on non-clinical testing of medical devices and internal Risk Assessment procedures, the following in vitro tests were performed on the subject device:
- Simulated use of atherectomy device in a stent .
- Heat generation .
- . Embolization analysis
#### Clinical Literature
A significant body of medical literature confirms the frequent usage of the Rotarex™ Atherectomy System for the treatment of ISR as well as for the debulking of occlusions in illac arteries and bypass grafts. Literature was reviewed to examine the safety and efficacy of the Rotarex™ Atherectomy System compared to percutaneous transluminal angioplasty (PTA) when used for ISR treatment. In addition, the safety, technical success, and clinical outcome for iliac arteries was addressed by description of the results from all-comer Rotarex™ Atherectomy System studies. The debulking of bypass grafts using the Rotarex™ Atherectomy System was compared with thrombectorny devices, Ekos (Boston Scientific) and Indigo (Penumbra), and with open surgery. The results of the literature review were used to determine the acceptability of the benefit-risk profile of Rotarex™ Atherectorny Systems used forthe treatment of ISR and debulking of occlusions in iliac-arteries and bypass-grafts.
{6}------------------------------------------------
### Conclusions:
The results from both in vitro studies and clinical literature demonstrate that the technological characteristics and performance criteria of the subject Rotarex™ Atherectomy System are substantially equivalent to the predicate device, and that it can perform in a manner equivalent to devices currently on the market for the same intended use.
The subject device, the Rotarex™ Atherectomy System, met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. Rotarex™ Atherectomy System is substantially equivalent to the legally marketed predicate device, the Rotarex® S catheter.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.