K172315 · Straub Medical AG · MCW · Apr 12, 2018 · Cardiovascular
Device Facts
Record ID
K172315
Device Name
Straub Endovascular System
Applicant
Straub Medical AG
Product Code
MCW · Cardiovascular
Decision Date
Apr 12, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4875
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K172315 · Apr 12, 2018
Straub Endovascular System
Straub Medical AG
Meta-analysis of 8 clinical studies (2002-2015)
A meta-analysis of 2,107 patients was used to provide a quantitative review and synthesis of clinical outcomes, including technical/clinical success, target lesion revascularization, restenosis rates, ankle-brachial index, Rutherford score, and adverse events.
Meta-analysis; Clinical performance; Retrospective data
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Meta-analysis of 8 clinical studies; Meta-analysis; Study Period: 2002-2015
When operated with a Rotarex S single use catheter, the Straub Endovascular System is intended for use as an atherectomy device and to break up and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, pulmonary, iliac or renal vasculature.
Device Story
Straub Endovascular System is a percutaneous, over-the-wire atherectomy device used to remove thrombus and occluding material from peripheral arteries. System components include a Rotarex S catheter, guidewire, collecting bag, and a drive system (control unit, motor, foot switch). The catheter features a rotating head with two metal cylinders and a helix; the head rotates at 40,000–60,000 rpm. The rotating tip abrades occluding material; a vortex created by the head assists in erosion. Detached material is aspirated through side openings into the catheter lumen and transported to an external collecting bag. Used in clinical settings by physicians; the device provides mechanical removal of arterial obstructions, potentially improving blood flow and patient outcomes. The system is provided sterile for single use.
Clinical Evidence
Clinical performance was established via a meta-analysis of 8 clinical studies (2002–2015) involving 2,107 patients. Endpoints included technical and clinical success, target lesion revascularization, restenosis rates, ankle-brachial index, Rutherford score, and procedure-related adverse events. Additionally, a GLP porcine animal study assessed device handling, clinical pathology, gross necropsy, histopathology, and vascular injury scores.
Technological Characteristics
System includes a Rotarex S catheter (flexible braided shaft, rotating helix, metal cylinder head), drive system (control unit, motor, foot switch), and guidewire. Materials are biocompatible per ISO 10993-1. Sterilization via Ethylene Oxide (ISO 11135-1). Electrical safety/EMC compliant with IEC 60601-1 and IEC 60601-1-2. Rotational speed: 40,000–60,000 rpm. Mechanical aspiration mechanism.
Indications for Use
Indicated for atherectomy and thrombus removal in upper and lower extremity peripheral arteries. Contraindicated for use in coronary, carotid, pulmonary, iliac, or renal vasculature.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that.
April 12, 2018
Straub Medical AG % Ms. Barbara Atzenhoefer Principal Medical Research Scientist, Regulatory NAMSA 400 Hwy 169 South, Suite 500 Minneapolis, MN 55426
Re: K172315
Trade/Device Name: Straub Endovascular System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW, DQX Dated: March 9, 2018 Received: March 12, 2018
Dear Ms. Atzenhoefer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
1000-1050.
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172315
Device Name Straub Endovascular System
Indications for Use (Describe)
When operated with a Rotarex S single use catheter, the Straub Endovascular System is intended for use as an atherectomy device and to break up and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, pulmonary, iliac or renal vasculature.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Straub Medical. The word "straub" is written in a dark blue, sans-serif font, with a registered trademark symbol next to it. Below the word "straub" is a teal rectangle with the word "MEDICAL" written in a white, sans-serif font. There are also three horizontal teal lines above the teal rectangle.
# 510(k) Summary
## Applicant Information
| Date Prepared: | April 10, 2018 |
|-----------------|---------------------------------------------------------------------|
| Applicant: | Straub Medical AG<br>Straubstrasse 12<br>CH 7323 Wangs, Switzerland |
| Contact Person: | Amy Ehlenfeldt, Manager Quality and Regulatory Compliance |
| Telephone: | +41 (0) 81 720 04 60 |
| Fax: | +41 (0) 81 720 04 61 |
#### Subject Device
| Name of Device: | Straub Endovascular System |
|----------------------|---------------------------------|
| Common/Usual Name: | Peripheral Atherectomy Catheter |
| Classification Name: | Intraluminal Artery Stripper |
| Regulation: | 21 CFR §870.4875 |
| Regulatory Class: | Class II |
| Product Codes: | MCW and DQX |
#### Predicate
MEDRAD Jetstream Atherectomy System K133023
Asahi Intecc Co. Ltd. Asahi Asato 30 guidewire K071721
## Device Description
The Straub Endovascular System consists of:
- Rotarex®S catheter set which includes a Rotarex®S catheter, guidewire, collecting bag and drape. ● All are provided sterile and are for single use
- . Drive System, consisting of the Control Unit, Motor and Foot switch. The Drive System serves to drive and control the rotational catheters.
Rotarex S is an over-the-wire, percutaneous catheter. The catheter consists of a flexible braided shaft, a rotating head, and a rotating helix which runs the length of the catheter. A lumen for the passage of the supplied guidewire runs the entire length of the helix and through the head of the catheter. The catheter head is made up of two overlying metal cylinders, with two side openings.
The outer cylinder is connected to the rotating helix, and the inner cylinder to the catheter shaft. The helix and the catheter head rotate at approximately 40.000-60.000 rpm depending on the model, by means of a gear box in the catheter housing and a motor contained within the catheter handle driven by the Drive System. The rotating outer cylinder is fitted with facets at its foremost tip, which when rotating, serve to abrade occluding material lying in front of it. Concomitantly, the rotation of the catheter head creates a vortex in the blood which assists to further erode occluding material from the vessel lumen. The detached material is aspirated into the side openings of the overlying cylinders where it is further broken down within the head and then carried through the inner lumen into a collecting bag outside of the body.
{4}------------------------------------------------
## Indications for Use
When operated with a Rotarex S single use catheter, the Straub Endovascular System is intended for use as an atherectomy device and to break up and remove thrombus from upper and lower extremity peripheral arteries. It is not intended for use in coronary, carotid, pulmonary, iliac or renal vasculature.
#### Comparison of Technological Characteristics
As compared to the predicate devices, the Straub Endovascular System has similar components, indications for use, and identical classification codes as the selected predicate devices. The Straub Endovascular System is comprised of similar patient contact materials and has undergone appropriate biocompatibility testing for its intended use. As with the predicate devices, the Rotarex®S Catheter Set components are provided sterile and are intended for single-use.
With respect to general device design and technological characteristics, the Straub Endovascular System is substantially equivalent to the selected predicate device, since it similarly uses a cutting tip controlled by an external control unit and allows for the removal of tissue, thrombus, and fluid via aspiration. A summary of the similarities and differences are shown below.
| Catheter Predicate - Similarities and Differences | | |
|---------------------------------------------------|---------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Characteristic | Straub Endovascular System | Predicate |
| Intended Use | Atherectomy and thrombus removal | Atherectomy and thrombus removal |
| System Components | Catheter, console/control unit, guidewire,<br>Collecting bag, sterile drape for motor | Guidewire, Collecting bag, console /<br>control unit |
| Mechanism of Action | Front cutting / rotational | Front cutting / rotational |
| Catheter Diameter | 6F: 2.00 mm<br>8F: 2.67 mm | XC 2.1/3.0: 2.1mm (3.0mm blades out)<br>XC 2.4/3.4: 2.4mm (3.4mm blades out) |
| Catheter Length | 6F: 110 or 135 cm<br>8F: 85 or 110 cm | XC 2.1/3.0: 135 cm<br>XC 2.4/3.4: 120 cm |
| Max. Guidewire Dia. | 0.018" | 0.014" |
| Min. Introducer Size | 6F Rotarex: 6-7 F<br>8F Rotarex: 8-9 F | XC 2.1/3.0: 7 F<br>XC 2.4/3.4: 7 F |
| Enclosed Cutting<br>Blades | Yes | No |
| Target speed | 6F: 60,000 rpm<br>8F: 40,000 rpm | XC 2.1/3.0: 70,000 rpm<br>XC 2.4/3.4: 70,000 rpm |
| Aspiration | Yes | Yes |
| Infusion | No | Yes |
| Sterilization Method | EtO | EtO |
There are no significant technological differences among the subject device and the predicate device. Therefore, the Straub Endovascular System is substantially equivalent to the Jetstream Atherectomy System.
## Performance Testing Summary
The following testing was performed to demonstrate that the system and its components perform as intended:
Design verification and bench validation testing:
Rotarex®S system: Visual and Dimensional Inspections, Dynamic (Durability) Test Under Simulated Use Conditions, Tensile Test, Torsional Strength, Kink Test, Corrosion Resistance, Temperature Rise (heat generation) Testing, Aspiration Capacity, FAS (Helix) Integrity, Stator Coating Integrity, In vitro Calcified Plaque and Thrombus Removal Studies, Rotational Speed stability). Guidewire: Visual and Dimensional Inspections, Corrosion resistance, Fracture Testing, Bending, Tensile Strength, Tip Flexibility, Torque Response, Torque Strength, Radiopacity
K172315 510(k) Summary
{5}------------------------------------------------
Shelf-life was established at 3-years based on accelerated aging studies.
Biocompatibility testing was conducted on the catheter and guidewire in accordance with ISO 10993-1 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process (and related sub-parts). Testing and included Cytotoxicity, Acute Systemic Toxicity, Sensitization, Intracutaneous Reactivity, Complement Activation, Hemolysis and Material-mediated Pyrogenicity.
Sterilization validations of the catheter and guidewire were conducted in accordance with ISO 11135-1 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process Sterilization of health-care products - Ethylene oxide - Requirements for the development. validation and routine control of a sterilization process for medical devices.
Electrical safety and Electromagnetic compatibility testing was conducted and the system determined to be compliant with the requirements of IEC 60601-1: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance and IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
A GLP animal study was conducted to assess the safety of the Straub Endovascular System in a porcine model. The following were assessed: procedure and device handling data, clinical pathology, adverse events, gross necropsy, histopathology and vascular injury score.
## Clinical Testing
A meta-analysis of 8 clinical studies comprising data obtained from 2,107 patients studied from 2002 to 2015 was performed to establish Rotarex® S clinical performance. The statistical summary of the literature data was to provide a quantitative review and synthesis of related, but independent, studies of the Rotarex device. The analyses allowed for the estimation of various clinical outcomes including, but not limited to technical and clinical success, target lesion revascularization, restenosis rates, ankle-brachial index, Rutherford score and procedure-related adverse events.
## Conclusions
The data and information presented within this submission and the similarities between the subject and predicate device support a determination of substantial equivalence, and therefore market clearance of the subject Straub Endovascular System through this 510(k) Premarket Notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.