CONSTRUX Mini PEEK Spacer System/CONSTRUX Mini PEEK VBR System, CONSTRUX Mini PTC Spacer System, CONSTRUX Mini Ti Spacer System, FORZA PEEK Spacer System, FORZA PTC Spacer System, FORZA Ti Spacer System, FORZA XP Expandable Spacer System, Lonestar Cervical Standalone System, PILLAR PEEK Spacer System, PILLAR SA PEEK Spacer System, PILLAR SA PTC Spacer System, SKYHAWK Lateral Interbody Fusion System
K211704 · Orthofix, Inc. · ODP · Aug 31, 2021 · Orthopedic
Device Facts
Record ID
K211704
Device Name
CONSTRUX Mini PEEK Spacer System/CONSTRUX Mini PEEK VBR System, CONSTRUX Mini PTC Spacer System, CONSTRUX Mini Ti Spacer System, FORZA PEEK Spacer System, FORZA PTC Spacer System, FORZA Ti Spacer System, FORZA XP Expandable Spacer System, Lonestar Cervical Standalone System, PILLAR PEEK Spacer System, PILLAR SA PEEK Spacer System, PILLAR SA PTC Spacer System, SKYHAWK Lateral Interbody Fusion System
Applicant
Orthofix, Inc.
Product Code
ODP · Orthopedic
Decision Date
Aug 31, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When Used as a Cervical Intervertebral Body Fusion System: The CONSTRUX Mini PEEK Spacer System is indicated for spinal fusion procedures at one or two contiguous levels within the cervical spine (C2-T1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. The CONSTRUX Mini PEEK Spacer System is intended for use with autograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation system; the hyperlordotic implants (≥10°) are required to be used with an anterior cervical plate. Patients must have undergone a regimen of at least six weeks of non-operative treatment prior to being treated with the CONSTRUX Mini PEEK Spacer System in the cervical spine. When Used as a Partial Vertebral Body Replacement (VBR) System: The CONSTRUX Mini PEEK Spacer System is indicated for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and to restore the height of a collapsed vertebral body. The CONSTRUX Mini PEEK Spacer System is also indicated for treating fractures of the thoracic and lumbar spine. Lordotic implants greater than a 5° profile are not to be used for partial vertebral body replacement. The CONSTRUX Mini PEEK Spacer System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period of time. The Partial VBR Device is intended to be used with autograft or allograft and supplemental fixation system.
Device Story
System comprises various interbody spacers (PEEK, porous titanium, hybrid) for spinal fusion or partial vertebral body replacement. Implants inserted into disc space or vertebral body to restore height/stability; used with autograft/allograft and supplemental fixation (e.g., anterior cervical plates, pedicle screw systems). Operated by surgeons via anterior, posterior, transforaminal, or lateral approaches. Provides biomechanical support to facilitate fusion or stabilize spine post-resection. Submission adds MR Conditional labeling; no changes to design, materials, or clinical function.
Clinical Evidence
No clinical data. Bench testing only: ASTM F2052-15 (displacement force), ASTM F2213-17 (torque), ASTM F2119-07 (image artifacts), and ASTM F2182-19E2 (RF heating) conducted to support MR Conditional labeling.
Technological Characteristics
Materials: PEEK (ASTM F2026), Titanium alloy (ASTM F136), Tantalum markers (ASTM F560). 3D printed porous titanium end plates. Designs include parallel, lordotic, and hyperlordotic profiles. Non-sterile or sterile configurations. MR Conditional.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the cervical (C2-T1) or lumbar (L2-S1) spine, or for partial vertebral body replacement (T1-L5) due to tumors or fractures. DDD defined as discogenic pain with radiographic confirmation. Requires 6 weeks (cervical) or 6 months (lumbar) of non-operative treatment. Contraindications include patients not meeting skeletal maturity or those without required supplemental fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
CONSTRUX Mini PEEK Spacer System/CONSTRUX Mini PEEK VBR System (K202949)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "HUMAN SERVICES-USA" around the edge. To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
August 31, 2021
Orthofix Inc. Jacki Koch Principle Regulatory Affairs 3451 Plano Parkway Lewisville, Texas 75056
## Re: K211704
Trade/Device Name: CONSTRUX Mini PEEK Spacer System/CONSTRUX Mini PEEK VBR System, CONSTRUX Mini PTC Spacer System, CONSTRUX Mini Ti Spacer System, FORZA PEEK Spacer System, FORZA PTC Spacer System, FORZA Ti Spacer System, FORZA XP Expandable Spacer System, Lonestar Cervical Standalone System, PILLAR PEEK Spacer System, PILLAR SA PEEK Spacer System, PILLAR SA PTC Spacer System, SKYHAWK Lateral Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP, MQP, MAX, OVD, OVE Dated: June 2, 2021 Received: June 3, 2021
Dear Jacki Koch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Brent L. Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
## K211704
Device Name CONSTRUX Mini PEEK Spacer System/CONSTRUX Mini PEEK VBR System
Indications for Use (Describe)
When Used as a Cervical Intervertebral Body Fusion System:
The CONSTRUX Mini PEEK Spacer System is indicated for spinal fusion procedures at one or two contiguous levels within the cervical spine (C2-T1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.
The CONSTRUX Mini PEEK Spacer System is intended for use with autograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation system; the hyperlordotic implants (≥10°) are required to be used with an anterior cervical plate.
Patients must have undergone a regimen of at least six weeks of non-operative treatment prior to being treated with the CONSTRUX Mini PEEK Spacer System in the cervical spine.
When Used as a Partial Vertebral Body Replacement (VBR) System:
The CONSTRUX Mini PEEK Spacer System is indicated for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial vertebrectory) of a diseased vertebral body resected or excised for the treatment of turnors, to achieve anterior decompression of the spinal cord and to restore the height of a collapsed vertebral body. The CONSTRUX Mini PEEK Spacer System is also indicated for treating fractures of the thoracic and lumbar spine. Lordotic implants greater than a 5° profile are not to be used for partial vertebral body replacement.
The CONSTRUX Mini PEEK Spacer System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period of time. The Partial VBR Device is intended to be used with autograft or allograft and supplemental fixation system.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{3}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K211704
Device Name CONSTRUX Mini PTC Spacer System
#### Indications for Use (Describe)
The CONSTRUX Mini PTC Spacer System is indicated for spinal fusion procedures at one or two contiguous levels within the cervical spine (C2-T1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.
The CONSTRUX Mini PTC Spacer System is intended for use with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation system; the hyperlordotic implants (≥10°) are required to be used with an anterior cervical plate.
Patients must have undergone a regimen of at least six weeks of non-operative treatment prior to being treated with the CONSTRUX Mini PTC Spacer System in the cervical spine.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{4}------------------------------------------------
See PRA Statement below.
Form Approved: OMB No. 0910-0120
## Indications for Use
510(k) Number (if known)
#### K211704
Device Name CONSTRUX Mini Ti Spacer System
Indications for Use (Describe)
The CONSTRUX Mini Ti Spacer System is indicated for spinal fusion procedures at one or two contiguous levels within the cervical spine (C2-T1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radios.
The CONSTRUX Mini Ti Spacer System is intended for use with autograft comprised of cancellous and/ or corticocancellous bone graft and supplemental fixation system; the hyperlordotic implants (≥10°) are required to be used with an anterior cervical plate.
Patients must have undergone a regimen of at least six weeks of non-operative treatment prior to being treated with the CONSTRUX Mini Ti Spacer System in the cervical spine.
Type of Use (Select one or both, as applicable)
| <div> <span> <b> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-----------------------------------------------------------------------------------------------------------------------|
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{5}------------------------------------------------
See PRA Statement below.
Form Approved: OMB No. 0910-0120
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K211704
Device Name FORZA PEEK Spacer System
#### Indications for Use (Describe)
The FORZA Spacer System is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated.
The FORZA Spacer System is intended to be used with supplemental fixation systems. As an example, the supplemental fixation that may be used is the Orthofix Firebird Spinal Fixation System.
The FORZA Spacer System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
See PRA Statement below.
Form Approved: OMB No. 0910-0120
## Indications for Use
510(k) Number (if known)
### K211704
Device Name FORZA PTC Spacer System
#### Indications for Use (Describe)
The FORZA PTC Spacer System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade
1 spondylolisthesis at the involved levels. These patients may have had a previous non-fusion surgery at the involved levels.
The FORZA PTC Spacer System is intended for use with autograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation, e.g., the Firebird Spinal Fixation System.
Patients must have undergone a regimen of at least six months of non-operative treatment prior to being treated with the FORZA PTC Spacer System.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211704
Device Name FORZA Ti Spacer System
Indications for Use (Describe)
The FORZA Ti Spacer System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade
1 spondylolisthesis at the involved levels. These patients may have had a previous non-fusion surgery at the involved levels.
The FORZA Ti Spacer System is intended for use with autograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation.
Patients must have undergone a regimen of at least six months of non-operative treatment prior to being treated with the FORZA Ti Spacer System.
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{8}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K211704
Device Name FORZA XP Expandable Spacer System
### Indications for Use (Describe)
The FORZA XP Expandable Spacer System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. These patients may have had a previous non-fusion surgery at the involved level(s).
The FORZA XP Expandable Spacer System is intended for use with autograft comprised of cancellous and/or corticocancellous bone graft; and supplemental fixation system.
Patients must have undergone a regimen of at least six months of non-operative treatment prior to being treated with the FORZA XP Expandable Spacer System.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{9}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K211704
Device Name Lonestar Cervical Standalone System
#### Indications for Use (Describe)
The LONESTAR Cervical Stand Alone System is indicated for spinal fusion procedures at one level in the cervical spine (C2-T1) in skeletally mature patients with degenerative disc disease (DDD), DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.
The LONESTAR Cervical Stand Alone System is used with autograft comprised of cancellous and/or corticocancellous bone graft and the two titanium alloy screws which accompany the implant.
Patients must have undergone a regimen of at least six weeks of non-operative treatment prior to being treated with the LONESTAR Cervical Stand Alone System in the cervical spine.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{10}------------------------------------------------
See PRA Statement below.
Form Approved: OMB No. 0910-0120
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K211704
Device Name PILLAR PEEK Spacer System
#### Indications for Use (Describe)
The PILLAR PEEK Spacer System is indicated for use with bone graft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated.
The PILLAR PEEK Spacer System is intended to be used with supplemental fixation. As an example, the supplemental fixation that may be used is the Orthofix Spinal Fixation System (SFS).
The PILLAR PL PEEK Spacer is used singly or in pairs and is implanted using a posterior approach.
The PILLAR TL PEEK Spacer is used singly or in pairs and is implanted using a transforaminal approach.
The PILLAR AL PEEK Spacer is used singly and is implanted using an anterior approach.
The PILLAR XL PEEK Spacer is used singly and is implanted using a lateral approach.
The PILLAR PEEK Spacer System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
When used as a Partial Vertebral Body Replacement (VBR) Device:
The PILLAR PEEK Spacer System is indicated for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The PILLAR Spacer System is also indicated for treating fractures of the thoracic and lumbar spine.
The PILLAR PEEK Spacer System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of a fusion for a prolonged period of time. The Partial VBR device is intended to be used with autograft or allograft.
The PILLAR PEEK Spacer System is intended for use with supplemental fixation. As an example, the supplemental fixation that may be used is the Orthofix Spinal Fixation System (SFS).
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span></span> </div> |
|----------------------------------------------|----------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span></span> </div> |
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{11}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{12}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
#### K211704
Device Name PILLAR SA PEEK Spacer System
Indications for Use (Describe)
When used as an Intervertebral Body Fusion Device:
The PILLAR SA PEEK Spacer System is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated.
The PILLAR SA PEEK Spacer System is intended for use with the four titanium alloy screws provided with the device. If the physician chooses to use fewer than four of the provided screws, then supplemental fixation must be used to augment stability. As an example, the supplemental fixation system that may be used is the Orthofix Firebird Spinal Fixation System.
The PILLAR SA PEEK Spacer System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
When used as a Partial Vertebral Body Replacement (VBR) Device:
The PILLAR SA PEEK Spacer System is indicated for use in the thoracolumbar spine (TI-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The PILLAR SA PEEK Spacer System is also indicated for treating fractures of the thoracic and lumbar spine.
The PILLAR SA PEEK Spacer System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column, even in the absence of fusion for a prolonged period of time. The PILLAR SA PEEK Spacer System is intended to be used with autograft or allograft.
The PILLAR SA PEEK Spacer System is intended for use with the four titanium alloy screws provided with the device. If the physician chooses to use fewer than four of the supplemental fixation must be used to augment stability. As an example, the supplemental fixation that may be used is the Orthofix Firebird System.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="text-decoration: underline;"><b></b></span> <span style="text-decoration: overline;"><b></b></span> <span style="text-decoration: line-through;"><b></b></span> <span style="text-decoration: underline;"><b></b></span> <span style="font-weight: bold;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{13}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{14}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
### K211704
Device Name PILLAR SA PTC Spacer System
### Indications for Use (Describe)
The PILLAR SA PTC Spacer System is indicated for spinal fusion procedures in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. These patients may have had a previous non-fusion surgery at the involved level(s).
The PILLAR SA PTC Spacer System is intended for use with autograft comprised of cancellous and/or corticocancellous bone graft.
The PILLAR SA PTC Spacer System is intended for use with four of the titanium alloy screws provided with the system. If the physician chooses to use fewer than four of the provided screws then supplemental fixation must be used to augment stability. As an example, a supplemental fixation system that may be used is the Orthofix Firebird® Spinal Fixation System.
Patients must have undergone a regimen of at least six months of non-operative treatment prior to being treated with the PILLAR SA PTC Spacer System.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED
## TINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{15}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K211704
Device Name SKYHAWK Lateral Interbody Fusion System
#### Indications for Use (Describe)
When used as an intervertebral body fusion device, the SKYHAWK Lateral Interbody Fusion System is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six mon-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated.
The SKYHAWK Lateral Interbody Fusion System is intended to be used with supplemental fixation systems that are cleared by the FDA for use in the lumbar spine.
The SK YHA WK Lateral Interbody Fusion System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{16}------------------------------------------------
# 510(k) SUMMARY
# CONSTRUX Mini PEEK Spacer System/CONSTRUX Mini PEEK VBR System CONSTRUX Mini PTC Spacer System CONSTRUX Mini Ti Spacer System FORZA PEEK Spacer System FORZA PTC Spacer System FORZA Ti Spacer System FORZA XP Expandable Spacer System Lonestar Cervical Standalone System PILLAR PEEK Spacer System PILLAR SA PEEK Spacer System PILLAR SA PTC Spacer System SKYHAWK Lateral Interbody Fusion System
### 510(k) Owner Information
| Name:<br>Address: | Orthofix US LLC<br>3451 Plano Parkway<br>Lewisville, TX 75056 |
|--------------------------------------------|---------------------------------------------------------------|
| Telephone Number:<br>Fax Number:<br>Email: | 214-937-2100<br>214-937-3322<br>jackikoch@orthofix.com |
| Registration Number: | 2183449 |
| Contact Person: | Jacki Koch, Principle Regulatory Affairs |
| Date Prepared: | August 30, 2021 |
## Name of Device
Trade Name / Proprietary Name:
> CONSTRUX Mini PEEK Spacer System/ CONSTRUX Mini PEEK VBR System (K202949) CONSTRUX Mini PTC Spacer System (K202666) CONSTRUX Mini Ti Spacer System (K203342) FORZA PEEK Spacer System (K162446) FORZA PTC Spacer System (K200052) FORZA Ti Spacer System (K203576) FORZA XP Expandable Spacer System (K172696) Lonestar Cervical Standalone System (K161280) PILLAR PEEK Spacer System (K162446) PILLAR SA PEEK Spacer System (K162446) PILLAR SA PTC Spacer System (K200052) SKYHAWK Lateral Interbody Fusion System (K162446)
## Product Code(s):
| System Name | Product Codes |
|------------------------------------------------------------------------------------------|---------------|
| CONSTRUX Mini PEEK Spacer System/CONSTRUX Mini PEEK Vertebral<br>Body Replacement System | ODP, MQP |
| CONSTRUX Mini PTC Spacer System | ODP |
| CONSTRUX Mini Ti Spacer System | ODP |
{17}------------------------------------------------
K211704
| | Page 2 of 11 |
|-----------------------------------------|---------------|
| FORZA PEEK Spacer System | MAX, MQP |
| FORZA PTC Spacer System | OVD, MAX, ODP |
| FORZA Ti Spacer System | MAX |
| FORZA XP Expandable Spacer System | MAX |
| Lonestar Cervical Standalone System | OVE |
| PILLAR PEEK Spacer System | MAX, MQP |
| PILLAR SA PEEK Spacer System | MAX, MQP |
| PILLAR SA PTC Spacer System | OVD, MAX, ODP |
| SKYHAWK Lateral Interbody Fusion System | MAX, MQP |
## Classification Name(s)
| System Name | Classification Name |
|------------------------------------------------|-----------------------------------|
| CONSTRUX Mini PEEK Spacer System/CONSTRUX Mini | Intervertebral Body Fusion Device |
| PEEK Vertebral Body Replacement System | Intervertebral Body Fusion Device |
| CONSTRUX Mini PTC Spacer System | Intervertebral Body Fusion Device |
| CONSTRUX Mini Ti Spacer System | Intervertebral Body Fusion Device |
| FORZA PEEK Spacer System | Intervertebral Body Fusion Device |
| FORZA PTC Spacer System | Intervertebral Body Fusion Device |
| FORZA Ti Spacer System | Intervertebral Body Fusion Device |
| FORZA XP Expandable Spacer System | Intervertebral Body Fusion Device |
| Lonestar Cervical Standalone System | Intervertebral Body Fusion Device |
| PILLAR PEEK Spacer System | Intervertebral Body Fusion Device |
| PILLAR SA PEEK Spacer System | Intervertebral Body Fusion Device |
| PILLAR SA PTC Spacer System | Intervertebral Body Fusion Device |
| SKYHAWK Lateral Interbody Fusion System | Intervertebral Body Fusion Device |
## Device Classification
| System Name | Product Codes |
|------------------------------------------------------------------------------------------|--------------------------------|
| CONSTRUX Mini PEEK Spacer System/CONSTRUX Mini<br>PEEK Vertebral Body Replacement System | Class II per 21 CFR § 888.3080 |
| CONSTRUX Mini PTC Spacer System | Class II per 21 CFR § 888.3080 |
| CONSTRUX Mini Ti Spacer System | Class II per 21 CFR § 888.3080 |
| FORZA PEEK Spacer System | Class II per 21 CFR § 888.3080 |
| FORZA PTC Spacer System | Class II per 21 CFR § 888.3080 |
| FORZA Ti Spacer System | Class II per 21 CFR § 888.3080 |
| FORZA XP Expandable Spacer System | Class II per 21 CFR § 888.3080 |
| Lonestar Cervical Standalone System | Class II per 21 CFR § 888.3080 |
| PILLAR PEEK Spacer System | Class II per 21 CFR § 888.3080 |
| PILLAR SA PEEK Spacer System | Class II per 21 CFR § 888.3080 |
| PILLAR SA PTC Spacer System | Class II per 21 CFR § 888.3080 |
| SKYHAWK Lateral Interbody Fusion System | Class II per 21 CFR § 888.3080 |
#### Review Panel: Orthopedic Device Panel
FORZA Ti Spacer System (K203576) Predicate Device:
Additional Predicate Devices: CONSTRUX Mini PEEK Spacer System/ CONSTRUX Mini PEEK VBR System (K202949) CONSTRUX Mini PTC Spacer System (K202666) CONSTRUX Mini Ti Spacer System (K203342) FORZA PEEK Spacer System (K162446) FORZA PTC Spacer System (K200052) FORZA XP Expandable Spacer System (K172696) Lonestar Cervical Standalone System (K161280) PILLAR PEEK Spacer System (K162446)
{18}------------------------------------------------
PILLAR SA PEEK Spacer System (K162446) PILLAR SA PTC Spacer System (K200052) SKYHAWK Lateral Interbody Fusion System (K162446)
Reason for 510(k) Submission:
Orthofix is submitting this Special 510(k) premarket notification for the addition of MR Conditional labeling to the subject medical devices.
The subject addition of MR Conditional labeling does not change the design, intended use, materials, performance specifications or the indications for use as previously cleared.
## Device Description
CONSTRUX Mini PEEK Spacer System/CONSTRUX Mini PEEK VBR System – The CONSTRUX Mini PEEK Spacer System/CONSTRUX Mini PEEK VBR System is comprised of a variety of implants manufactured from PEEK (Polyetheretherketone), as described by ASTM F2026, with titanium markets as described by ASTM F67. The implants are available in multiple sizes to accommodate various patient anatomies. The superior and inferior surfaces of the implant have a pattern of ripples to provide increased stability and help prevent anterior/posterior movement of the device.
The CONSTRUX Mini PEEK VBR System is intended for vertebral body replacement to aid in the surqical correction and stabilization of the spine.
The CONSTRUX Mini PEEK Spacer System/CONSTRUX Mini PEEK VBR System is not intended to be used as a stand-alone device and must be used with supplemental fixation. The implants are used singly and are implanted using an anterior approach.
The CONSTRUX Mini PEEK Spacer System/CONSTRUX Mini PEEK VBR System implants are provided either in a sterile packaging configuration or non-sterile and requires sterilization prior to use.
CONSTRUX Mini PTC Spacer System - The CONSTRUX Mini PTC Spacer System is comprised of a variety of implants that have a PEEK core with integrated porous titanium end plates. CONSTRUX Mini PTC spacers are implanted in the cervical intervertebral disc space and are intended to facilitate vertebral fusion by stabilizing adjacent vertebrae, maintaining disc height and preventing the collapse of one vertebra onto another.
The CONSTRUX Mini PTC Spacer System is not intended to be used as a stand-alone device and must be used with supplemental fixation. The implants are used singly and are implanted using an anterior approach.
The CONSTRUX Mini PTC implants are provided sterile.
CONSTRUX Mini PTC implants are design to be used with CONSTRUX Mini PEEK Spacer System instrumentation. The CONSTRUX Mini PTC implants are not compatible with components or metal from any other manufacturer's system.
CONSTRUX Mini Ti Spacer System - The CONSTRUX Mini Ti Spacer System is comprised of a variety of 3D printed implants that have porous titanium end plates and a functional gradient porous structure. CONSTRUX Mini Ti spacers are implanted in the cervical intervertebral disc space and are intended to facilitate vertebral fusion by stabilizing adjacent vertebrae, maintaining disc height and preventing the collapse of one vertebra onto another.
The CONSTRUX Mini Ti Spacer System is not intended to be used as a stand-alone device and must be used with supplemental fixation. The implants are used singly and are implanted using an anterior approach.
The CONSTRUX Mini Ti implants are provided sterile.
{19}------------------------------------------------
CONSTRUX Mini Ti implants are design to be used with CONSTRUX Mini Spacer System instrumentation. The CONSTRUX Mini Ti implants are not compatible with components or metal from any other manufacturer's system.
FORZA PEEK Spacer System - The FORZA Spacer System consists of implants, trials and instruments. The system is comprised of a variety of implants fabricated and manufactured from polvetheretherketone (PEEK) as described by ASTM F2026 with tantalum markers as described by ASTM F560. PEEK is utilized due to its radiolucent properties, which aid the surgeon in determining if fusion in the operative site has occurred. Since PEEK is transparent in x-rays, tantalum marker pins are inserted into the implants in order to give surgeons a visual aid in determining the location of the implants both intraoperatively and postoperatively.
FORZA Spacer System implants are offered in two geometric shapes – straight and curved, and offered in parallel and lordotic profiles to restore the natural curvature of the implants can be used in single placement or in pairs. Both the curved and straight implants feature a bulleted nose for ease of insertion and anti-migration ripples on both the inferior and superior surfaces to provide increased stability and help prevent anterior/posterior movement of the device.
The FORZA Spacer System is intended for intervertebral body fusion to aid in the surgical correction and stabilization of the spine and are implanted using a posterior approach.
The FORZA Spacer System is not intended to be used as a stand-alone device. The FORZA Spacer System must be used with a supplemental fixation system. The implants are provided sterile but the instruments are provided non-sterilization prior to use.
FORZA PTC Spacer System - The FORZA PTC Spacer System is comprised of a variety of implants that have a PEEK (OPTIMA LT1) core with integrated porous titanium (Ti-6Al-4V) end plates as well as a tantalum marker that acts as a visual aid for the surgeon in determining the location of the implant both intraoperatively and postoperatively.
FORZA PTC Spacer System implants are offered in two geometric shapes - straight and curved, and offered in parallel and lordotic profiles to restore the natural curvature of the spine. The implants can be used in single placement or in pairs. Both the curved and straight implants feature a bulleted nose for ease of insertion and a roughened surface on both the inferior and superior faces of the implant to provide increased stability and help prevent anterior/posterior movement of the device.
The FORZA PTC Spacer System is intended for intervertebral body fusion to aid in the surgical correction of the spine and are implanted using a posterior approach.
The FORZA PTC Spacer System is not intended to be used as a standalone device. The FORZA PTC Spacer System must be used with a supplemental fixation system.
The FORZA PTC Spacer System implants are provided sterile.
FORZA PTC implants are designed for use with FORZA PEEK Spacer System instrumentation. The FORZA PTC spacers are not compatible with components or metal from any other manufacturer's system.
FORZA Ti Spacer System - The FORZA Ti Spacer System is comprised of a variety of 3D printed implants that have porous titanium end plates and a functional gradient porous structure.
FORZA Ti Spacer System implants are offered in two geometric shapes - straight and curved, and offered in parallel and lordotic profiles to restore the natural curvature of the spine. The implants can be used in single placement or in pairs. Both the curved and straight implants feature a bulleted nose for ease of insertion and a roughened surface on both the inferior and
{20}------------------------------------------------
Page 5 of 11 superior faces of the implant to provide increased stability and help prevent anterior/posterior movement of the device.
K211704
The FORZA Ti Spacer System is intended for intervertebral body fusion to aid in the surgical correction of the spine and are implanted using a posterior approach.
The FORZA Ti Spacer System is not intended to be used as a standalone device. The FORZA Ti Spacer System must be used with a supplemental fixation system.
The FORZA Ti Spacer System implants are provided sterile.
FORZA Ti implants are designed for use with FORZA PEEK and FORZA PTC Spacer System instrumentation. The FORZA Ti spacers are not compatible with components or metal from any other manufacturer's system.
FORZA XP Expandable Spacer System - The FORZA XP Expandable Spacer System is comprised of an assortment of non-sterile, single use, titanium allov (Ti-6AI-4V ELI per ASTM F136) and Polyetheretherketone (PEEK) Polymer (PEEK OPTIMA® LT1 per ASTM F2026) spacers with height expansion capability. The expandable interbody spacer is inserted into the lumbar disc space and expanded to fit the patient anatomy.
The implants are offered in parallel, lordotic, and hyperlordotic configurations to help restore the natural curvature of the spine. The implants can be used in single placement or pairs with typical approaches being transforaminal lumbar interbody fusion (TLIF) and posterior lumbar interbody fusion (PLIF).
The implants feature a bulleted nose for ease of insertion ripples on both the inferior and superior surfaces to provide increased stability and help prevent anterior/posterior movement of the device.
The FORZA XP Expandable Spacer System is not intended to be used as a stand-alone device. The system must be used with a supplemental fixation system, is provided non-sterile and requires sterilization prior to use.
Lonestar Cervical Standalone System - The LONESTAR Cervical Stand Alone System is a stand-alone spacer system designed to provide biomechanical strength to a traditional or minimally invasive ACDF procedure with less disruption of patient anatomy and preservation of the anatomical profile. The system helps to preserve the natural sagittal anatomic profile of the cervical spine while providing anterior column support and stability.
The LONESTAR implant consists of a hybrid PEEK and titanium spacer along with titanium bone screws and a titanium cover plate. The spacers are designed with a zero degree anterior profile and are implanted using an anterior approach.
The LONESTAR Cervical Stand Alone implants and instruments are provided non-sterile and will require thorough cleaning and sterilization prior to each use. The implants are not compatible with components or metal from any other manufacturer's system.
PILLAR PEEK Spacer System - The PILLAR PEEK Spacer System is comprised of a variety of implants manufactured from PEEK (Polyetheretherketone) as described by ASTM F2026 with tantalum markers as described by ASTM F560. The implants are available in a variety of sizes and are offered in parallel and lordotic profiles in order to restore the natural curvature of the spine. The implants are available in various heights in either one or two millimeter increments. The superior and inferior surfaces of the implant have a pattern of ripples to provide increased stability and help prevent anterior/posterior movement of the device.
The PILLAR PEEK Spacer System is intended for intervertebral body fusion or partial vertebral body replacement to aid in the surgical correction and stabilization of the spine.
{21}------------------------------------------------
The PILLAR PEEK Spacer System is not intended to be used as a stand-alone device and must be used with supplemental fixation. The system is provided non-sterile and requires sterilization prior to use.
PILLAR SA PEEK Spacer System - The PILLAR SA PEEK Spacer System is comprised of a variety of implants manufactured from PEEK (Polyetheretherketone) as described by ASTM F2026 with tantalum markers as described by ASTM F560. The implants are available in multiple footprints, a variety of heights, and two angles of lordosis: 7° and 12°. The implants incorporate integrated anterior screw holes to allow for medial placement of screws, as well as a titanium plate for securing the screws once in place. The superior and inferior surfaces of the implant have a pattern of ripples that provide increased stability and help prevent of the device.
The PILLAR SA PEEK Spacer System is provided non-sterile.
PILLAR SA PTC Spacer System - The PILLAR SA PTC Spacer System is comprised of a variety of implants that have a PEEK core with integrated porous titanium end plates. The implants incorporate integrated anterior screw holes to allow for medial placement of bone screws as well as a titanium plate for securing the bone screws once in place. The implants are designed with a roughened surface on the inferior and superior faces of the implant to provide increased stability and help prevent anterior/posterior movement of the device.
The PILLAR SA PTC Spacer System is intended for intervertebral body fusion to aid in the surgical correction of the spine and is implanted using an anterior approach.
The PILLAR SA PTC spacers are provided sterile. The cover plate, screws and instruments are provided non-sterile and require sterilization prior to use.
The PILLAR SA PTC implants are designed to be used with PILLAR SA PEEK Spacer System instrumentation. The implants are not compatible with components from any other manufacturer's system.
SKYHAWK Lateral Interbody Fusion System - The SKYHAWK Lateral Interbody Fusion System consists of implants, trials, and instruments and is comprised of a variety of implants fabricated and manufactured from polyetheretherketone (PEEK) as described by ASTM F2026 with tantalum markers as described by ASTM F560. PEEK is utilized due to its radiolucent property, which aids the surgeon in determining if fusion in the operative site has occurred. Since PEEK is transparent in x-rays, tantalum marker pins are inserted into the implants in order to give surgeons a visual aid in determining the implants, both intraoperatively and postoperatively.
The SKYHAWK Lateral Interbody Fusion System implants are offered in parallel and lordotic profiles to restore the natural curvature of the spine; the device may be implanted using a lateral approach.
The SKYHAWK Lateral Interbody Fusion System implants, trials and instruments are provided non-sterile. They require sterilization prior to use.
#### Intended Use / Indications for Use
CONSTRUX Mini PEEK Spacer System/CONSTRUX Mini PEEK VBR System
When Used as a Cervical Intervertebral Body Fusion System:
The CONSTRUX Mini PEEK Spacer System is indicated for spinal fusion procedures at one or two contiguous levels within the cervical spine (C2-T1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.
The CONSTRUX Mini PEEK Spacer System is intended for use with autograft and/or allograff comprised of cancellous and/or corticocancellous bone graft and supplemental fixation system; the hyperlordotic implants (≥10°) are required to be used with an anterior cervical plate.
{22}------------------------------------------------
Patients must have undergone a regimen of at least six weeks of non-operative treatment prior to being treated with the CONSTRUX Mini PEEK Spacer System in the cervical spine.
## When Used as a Partial Vertebral Body Replacement (VBR) System:
The CONSTRUX Mini PEEK Spacer System is indicated for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The CONSTRUX Mini PEEK Spacer System is also indicated for treating fractures of the thoracic and lumbar spine. Lordotic implants greater than a 5° profile are not to be used for partial vertebral body replacement.
The CONSTRUX Mini PEEK Spacer System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of fusion for a prolonged period of time. The Partial VBR Device is intended to be used with autograft or allograft and supplemental fixation system.
### CONSTRUX Mini PTC Spacer System
The CONSTRUX Mini PTC Spacer System is indicated for spinal fusion procedures at one or two contiquous levels within the cervical spine (C2-T1) in skeletally mature patients with deqenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.
The CONSTRUX Mini PTC Spacer System is intended for use with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation system; the hyperlordotic implants (≥10°) are required to be used with an anterior cervical plate.
Patients must have undergone a regimen of at least six weeks of non-operative treatment prior to being treated with the CONSTRUX Mini PTC Spacer System in the cervical spine.
#### CONSTRUX Mini Ti Spacer System
The CONSTRUX Mini Ti Spacer System is indicated for spinal fusion procedures at one or two contiguous levels within the cervical spine (C2-T1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.
The CONSTRUX Mini Ti Spacer System is intended for use with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation system; the hyperlordotic implants (≥10°) are required to be used with an anterior cervical plate.
Patients must have undergone a regimen of at least six weeks of non-operative treatment prior to being treated with the CONSTRUX Mini Ti Spacer System in the cervical spine.
#### FORZA PEEK Spacer System
The FORZA Spacer System is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated.
The FORZA Spacer System is intended to be used with supplemental fixation systems. As an example, the supplemental fixation that may be used is the Orthofix Firebird Spinal Fixation System.
{23}------------------------------------------------
Page 8 of 11 The FORZA Spacer System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
K211704
### FORZA PTC Spacer System
The FORZA PTC Spacer System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. These patients may have had a previous non-fusion surgery at the involved levels.
The FORZA PTC Spacer System is intended for use with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation, e.g., the Firebird Spinal Fixation System.
Patients must have undergone a regimen of at least six months of non-operative treatment prior to being treated with the FORZA PTC Spacer System.
### FORZA Ti Spacer System
The FORZA Ti Spacer System is indicated for spinal fusion procedures in skeletally mature patients with deqenerative disc disease (DDD) at one or two contiquous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. These patients may have had a previous non-fusion surgery at the involved levels.
The FORZA Ti Spacer System is intended for use with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation.
Patients must have undergone a regimen of at least six months of non-operative treatment prior to being treated with the FORZA Ti Spacer System.
## FORZA XP Expandable Spacer System
The FORZA XP Expandable Spacer System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. These patients may have had a previous non-fusion surgery at the involved level(s).
The FORZA XP Expandable Spacer System is intended for use with autograft and/or allograff comprised of cancellous and/or corticocancellous bone graft; and supplemental fixation system.
Patients must have undergone a regimen of at least six months of non-operative treatment prior to being treated with the FORZA XP Expandable Spacer System.
#### Lonestar Cervical Standalone System
The LONESTAR Cervical Stand Alone System is indicated for spinal fusion procedures at one level in the cervical spine (C2-T1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.
The LONESTAR Cervical Stand Alone System is used with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft and titanium alloy screws which accompany the implant.
Patients must have undergone a regimen of at least six weeks of non-operative treatment prior to being treated with the LONESTAR Cervical Stand Alone System in the cervical spine.
{24}------------------------------------------------
### PILLAR PEEK Spacer System
The PILLAR PEEK Spacer System is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated.
The PILLAR PEEK Spacer System is intended to be used with supplemental fixation. As an example, the supplemental fixation that may be used is the Orthofix Spinal Fixation System (SFS).
The PILLAR PL PEEK Spacer is used singly or in pairs and is implanted using a posterior approach.
The PILLAR TL PEEK Spacer is used singly or in pairs and is implanted using a transforaminal approach.
The PILLAR AL PEEK Spacer is used sindly and is implanted using an anterior approach.
The PILLAR XL PEEK Spacer is used singly and is implanted using a lateral approach.
The PILLAR PEEK Spacer System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
When used as a Partial Vertebral Body Replacement (VBR) Device:
The PILLAR PEEK Spacer System is indicated for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The PILLAR Spacer System is also indicated for treating fractures of the thoracic and lumbar spine.
The PILLAR PEEK Spacer System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of a fusion for a prolonged period of time. The Partial VBR device is intended to be used with autograft or allograft.
The PILLAR PEEK Spacer System is intended for use with supplemental fixation. As an example, the supplemental fixation that may be used is the Orthofix Spinal Fixation System (SFS).
## PILLAR SA PEEK Spacer System
When used as an Intervertebral Body Fusion Device:
The PILLAR SA PEEK Spacer System is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated.
The PILLAR SA PEEK Spacer System is intended for use with the four titanium alloy screws provided with the device. If the physician chooses to use fewer than four of the provided screws, then supplemental fixation must be used to augment stability. As an example, the supplemental fixation system that may be used is the Orthofix Firebird Spinal Fixation System.
{25}------------------------------------------------
The PILLAR SA PEEK Spacer System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
When used as a Partial Vertebral Body Replacement (VBR) Device:
The PILLAR SA PEEK Spacer System is indicated for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The PILLAR SA PEEK Spacer System is also indicated for treating fractures of the thoracic and lumbar spine.
The PILLAR SA PEEK Spacer System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column, even in the absence of fusion for a prolonged period of time. The PILLAR SA PEEK Spacer System is intended to be used with autograft or allograft.
The PILLAR SA PEEK Spacer System is intended for use with the four titanium alloy screws provided with the device. If the physician chooses to use fewer than four of the provided screws. then supplemental fixation must be used to augment stability. As an example, the supplemental fixation that may be used is the Orthofix Firebird Spinal Fixation System.
### PILLAR SA PTC Spacer System
The PILLAR SA PTC Spacer System is indicated for spinal fusion procedures in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. These patients may have had a previous nonfusion surgery at the involved level(s).
The PILLAR SA PTC Spacer System is intended for use with autograft and/or allograft comprised of cancellous and/or corticocancellous bone graft.
The PILLAR SA PTC Spacer System is intended for use with four of the titanium alloy screws provided with the system. If the physician chooses to use fewer than four of the provided screws then supplemental fixation must be used to augment stability. As an example, a supplemental fixation system that may be used is the Orthofix Firebird® Spinal Fixation System.
Patients must have undergone a regimen of at least six months of non-operative treatment prior to being treated with the PILLAR SA PTC Spacer System.
#### SKYHAWK Lateral Interbody Fusion System
When used as an intervertebral body fusion device, the SKYHAWK Lateral Interbody Fusion System is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated.
The SKYHAWK Lateral Interbody Fusion System is intended to be used with supplemental fixation systems that are cleared by the FDA for use in the lumbar spine.
The SKYHAWK Lateral Interbody Fusion System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
Summary of the Technological Characteristics of the Device Compared to the Selected Predicate Devices
{26}------------------------------------------------
The Technological Characteristics, design, dimensions, intended use, materials and performance characteristics of the subject devices are unchanged from their previous clearance. The purpose of this 510(k) submission is for the addition of MR Conditional to the device labeling.
## PERFORMANCE DATA – Summary of Non-Clinical Test Conducted for Determination of Substantial Equivalence
In accordance to the FDA Guidance "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment" the following testing was conducted:
- ASTM F2052-15 Standard test method for measurement of magnetically induced ● displacement force on passive implants in the magnetic resonance environment
- ASTM F2213-17 Standard test method for measurement of magnetically induced ● torque on medical devices in the magnetic resonance environment
- . ASTM F2119-07 – Standard test method for evaluation of MR image artifacts from passive implants
- . ASTM F2182-19E2 - Standard test method for measurement of radio frequency induced heating on or near passive implant during magnetic resonance imaging
## Basis of Substantial Equivalence
As stated throughout this 510(k) notification, there are no modifications to the design, intended use, or indications for use. The purpose of this 510(k) notification is for the addition of MR Conditional labeling for the subject devices.
The subject devices are temporary, multiple component systems comprised of a variety of single use components. There have been no changes to the material, intended use or indications for use.
Therefore, the subject devices are identical to themselves as previously cleared.
In accordance with FDA Guidance "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment," the following testing was conducted to determine that the subject devices met requirements necessary for MRI Conditional labeling:
- ASTM F2052-15 "Standard test method for measurement of magnetically induced ● displacement force on passive implants in the magnetic resonance environment"
- . ASTM F2213-17 - "Standard test method for measurement of magnetically induced torque on medical devices in the magnetic resonance environment"
- ASTM F2119-07 "Standard test method for evaluation of MR image artifacts from . passive implants"
- . ASTM F2182-19E2 - "Standard test method for measurement of radio frequency induced heating on or near passive implant during magnetic resonance imaging"
- ASTM F2503 - "Standard practice for marking medical devices and other items for safety in the magnetic resonance environment"
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.