FORZA Spacer System, PILLAR PEEK Spacer System, PILLAR SA PEEK Spacer System, SKYHAWK Lateral Interbody Fusion System
K162446 · Orthofix, Inc. · MAX · Dec 22, 2016 · Orthopedic
Device Facts
Record ID
K162446
Device Name
FORZA Spacer System, PILLAR PEEK Spacer System, PILLAR SA PEEK Spacer System, SKYHAWK Lateral Interbody Fusion System
Applicant
Orthofix, Inc.
Product Code
MAX · Orthopedic
Decision Date
Dec 22, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The FORZA Spacer System is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated. The FORZA Spacer System is intended to be used with supplemental fixation systems. As an example, the supplemental fixation that may be used is the Orthofix Firebird Spinal Fixation System. The FORZA Spacer System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
Device Story
Spinal spacer systems (FORZA, PILLAR, SKYHAWK) provide structural support for intervertebral body fusion or partial vertebral body replacement. Implants are placed in the spine (lumbar or thoracolumbar) to restore disc height and biomechanical integrity. Devices are used with autograft or allogenic bone graft to promote fusion. Systems are non-standalone and require supplemental internal fixation (e.g., pedicle screws/rods). Implants are manufactured from PEEK (ASTM F2026) with tantalum markers (ASTM F560) for radiographic visualization. Surgeons implant devices via various approaches (anterior, posterior, transforaminal, lateral). Output is structural stabilization of the spinal column. Benefits include restoration of spinal height, decompression of neural tissues, and facilitation of fusion in patients with DDD, tumors, or fractures.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on identical technological characteristics, design, and materials to previously cleared predicate devices. Bench testing included pyrogen/bacterial endotoxin testing (BET) per ANSI/AAMI ST-72:2011.
Technological Characteristics
Implants fabricated from PEEK (ASTM F2026) with tantalum markers (ASTM F560). PILLAR SA includes titanium alloy screws. Devices feature ripple patterns on superior/inferior surfaces for stability. Non-sterile; require sterilization prior to use. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1, with or without Grade I Spondylolisthesis/retrolisthesis, who have failed six months of non-operative treatment. Also indicated for partial vertebral body replacement (T1-L5) for tumors or fractures.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads facing right, representing health, human services, and the USA. The eagle is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 22, 2016
ORTHOFIX Inc. Ms. Jacki Koch Regulatory Affairs Specialist, II 3451 Plano Parkway Lewisville, Texas 75056
Re: K162446
Trade/Device Name: FORZA Spacer System, PILLAR PEEK Spacer System, PILLAR SA PEEK Spacer System, SKYHAWK Lateral Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, MQP Dated: September 26, 2016 Received: September 28, 2016
Dear Ms. Koch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Vincent J. Devlin -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
K162446
Page 1 of 6
510(k) Number (if known) K162446
Device Name FORZA Spacer System
#### Indications for Use (Describe)
The FORZA Spacer System is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated.
The FORZA Spacer System is intended to be used with supplemental fixation systems. As an example, the supplemental fixation that may be used is the Orthofix Firebird Spinal Fixation System.
The FORZA Spacer System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{3}------------------------------------------------
K162446
Page 2 of 6
510(k) Number (if known) K162446
Device Name PILLAR PEEK Spacer System
#### Indications for Use (Describe)
When used as an intervertebral body fusion device, the PILLAR PEEK Spacer System is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated.
The PILLAR PEEK Spacer System is intended to be used with supplemental fixation. As an example, the supplemental fixation that may be used is the Orthofix Spinal Fixation System (SFS).
The PILLAR PL PEEK Spacer is used singly or in pairs and is implanted using a posterior approach.
The PILLAR TL PEEK Spacer is used singly or in pairs and is implanted using a transforaminal approach.
The PILLAR AL PEEK Spacer is used singly and is implanted using an anterior approach.
The PILALR XL PEEK Spacer is used singly and is implanted using a lateral approach.
The PILLAR PEEK Spacer System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
When used as a Partial Vertebral Body Replacement (VBR), The PILLAR PEEK Spacer System is indicated for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and to restore the height of a collapsed vertebral body. The PILLAR Spacer System is also indicated for treating fractures of the thoracic and lumbar spine.
The PILLAR PEEK Spacer System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of a fusion for a prolonged period of time. The Partial VBR device is intended to be used with autograft or allograft.
The PILLAR PEEK Spacer System is intended for use with supplemental fixation. As an example, the supplemental fixation that may be used is the Orthofix Spinal Fixation System (SFS).
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (8/14)
{4}------------------------------------------------
K162446
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# the of the occident applies only to toquientients of the Processor Pool 1000.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* Page 3 of 6
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
K162446
Page 4 of 6
510(k) Number (if known) K162446
Device Name PILLAR SA PEEK Spacer System
#### Indications for Use (Describe)
When used as an intervertebral body fusion device, the PILLAR SA PEEK Spacer System is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated.
The PILLAR SA PEEK Spacer System is intended for use with the four titanium alloy screws provided with the device. If the physician chooses to use fewer than four of the provided screws, then supplemental fixation must be used to augmt stability. As an example, the supplemental fixation system that may be used is the Orthofix Spinal Fixation System (SFS).
The PILLAR SA PEEK Spacer System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
When used as a Partial Vertebral Body Replacement (VBR) the PILLAR SA PEEK Spacer System is indicated for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The PILLAR SA PEEK Spacer System is also indicated for treating fractures of the thoracic and lumbar spine.
The PILLAR SA PEEK Spacer System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column, even in the absence of fusion for a prolonged period of time. The PILLAR SA PEEK Spacer System is intended to be used with autograft or allograft.
The PILLAR SA PEEK Spacer System is intended for use with the four titanium alloy screws provided with the device. If the physician chooses to use fewer than four of the provided screws, then supplemental fixation must be used to augment stability. As an example, the supplemental fixation that may be used is the Orthofix Spinal Fixation System (SFS).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{6}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
Page 5 of 6 *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{7}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
K162446
Page 6 of 6
510(k) Number (if known) K162446
Device Name
SKYHAWK Lateral Interbody Fusion System
#### Indications for Use (Describe)
When used as an intervertebral body fusion device, the SKYHAWK Lateral Interbody Fusion System is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated.
The SKYHAWK Lateral Interbody Fusion System is intended to be used with supplemental fixation systems that are cleared by the FDA for use in the lumbar spine.
The SKYHAWK Lateral Interbody Fusion System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10px">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|------------------------------------------------------------------------------------|
| <span style="font-size:10px">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the word "ORTHOFIX" in bold, black letters. To the left of the word is a blue, abstract symbol that resembles a stylized letter "O" or a spiral. The overall design is clean and professional, suggesting a corporate logo or brand identity.
# 510(k) SUMMARY
# FORZA Spacer System PILLAR PEEK Spacer System PILLAR SA PEEK Spacer System SKYHAWK Lateral Interbody Fusion System
| 510(k) Owner Information | |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name:<br>Address: | Orthofix Inc.<br>3451 Plano Parkway<br>Lewisville, TX 75056 |
| Telephone Number:<br>Fax Number:<br>Email: | 214.937.2100<br>214-937-3322<br>jackikoch@orthofix.com |
| Registration Number: | 2183449 |
| Contact Person: | Jacki Koch, Regulatory Affairs Specialist, II |
| Date Prepared: | December 8, 2016 |
| Name of Device<br>Trade Name / Proprietary<br>Name: | FORZA Spacer System<br>PILLAR PEEK Spacer System<br>PILLAR SA PEEK Spacer System<br>SKYHAWK Lateral Interbody Fusion System |
| Common Name: | Spinal Vertebral Body Replacement Device<br>Spinal Partial Vertebral Body Replacement Device<br>Intervertebral Body Fusion Device |
| Product Code: | FORZA Spacer System - MAX<br>PILLAR PEEK Spacer System – MAX, MQP<br>PILLAR SA PEEK Spacer System – MAX, MQP<br>SKYHAWK Lateral Interbody Fusion System - MAX |
| Regulatory Classification: | Class II – 21 CFR § 888.3080 – Intervertebral body fusion device<br>Class II - 21 CFR § 888.3060 – Spinal intervertebral body fixation orthosis |
| Review Panel: | Orthopedic Device Panel |
| Primary Predicate Device: | FORZA Spacer System - Orthofix – K152475 |
Additional Predicate Devices: PILLAR PEEK Spacer System - Orthofix - K081177
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the Orthofix logo. The logo consists of a blue abstract shape on the left and the word "ORTHOFIX" in black capital letters on the right. The "®" symbol is located in the upper right corner of the word "ORTHOFIX".
PILLAR SA PEEK Spacer System - Orthofix - K081849 SKYHAWK Lateral Interbody Fusion System - Orthofix - K140709
Reason for 510(k) Submission:
Expanded Indications for Use to include the use with allograft material composed of cancellous or corticocancellous bone graft material.
# Device Description
## FORZA Spacer System:
The FORZA Spacer System consists of implants, trials, and instruments. The FORZA Spacer System is comprised of a variety of implants fabricated and manufactured from polyetheretherketone (PEEK) as described by ASTM F2026 with tantalum markers as described by ASTM F560. PEEK was utilized due to its radiolucent properties, which aids the surgeon in determining if fusion is the operative site has occurred. Since PEEK is transparent in x-rays, tantalum marker pins are inserted into the implants in order to give surgeons a visual aid in determining the location of the implants, both intra and postoperatively.
# PILLAR PEEK Spacer System:
The PILLAR Spacer System is comprised of a variety of implants manufactured from PEEK (Polyetheretherketone), as described by ASTM F2026, with tantalum markers as described by ASTM F560. The implants are available in a variety of footprint sizes. Additionally, they are offered in parallel and lordotic profiles in order to restore the natural curvature of the spine. The implants are available in various heights, in either one or two millimeter increments. The superior and interior surfaces of the implant have a pattern of ripples to provide increased stability and help prevent anterior/posterior movement of the device.
The PILLAR Spacer System is intended for intervertebral body fusion or partial vertebral body replacement to aid in the surqical correction and stabilization of the spine.
The PILLAR Spacer System is not intended to be used as a stand-alone device. The PILLAR Spacer System must be used with supplemental internal fixation. The PILLAR Spacer System is provided non-sterile.
# PILLAR SA PEEK Spacer System:
The PILLAR SA PEEK Spacer System is comprised of a variety of implants manufactured from PEEK (Polyetheretherketone), as described by ASTM F2026, with tantalum markers as described by ASTM F560. The implants are available in multiple footprint sizes, a variety of heights, and two angles of lordosis: 7° and 12°. The implants incorporate integrated anterior screw holes to allow for medial placement of screws, as well as titanium plate for securing the screws once in place. The superior and inferior surfaces of the implant have a pattern of ripples that provide increased stability and help prevent movement of the device.
The PILLAR SA PEEK Spacer System is provided non-sterile.
SKYHAWK Lateral Interbody Fusion System:
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in black, sans-serif font. The "®" symbol is located to the upper right of the word "ORTHOFIX".
The SKYHAWK Interbody Fusion System consists of implants, trials, and instruments. The system is comprised of a variety of implants fabricated and manufactured from polyetheretherketone (PEEK) as described by ASTM F2026 with tantalum markers as described by ASTM F560. PEEK is utilized due to its radiolucent properties, which aids the surgeon in determining if fusion is the operative site has occurred. Since PEEK is transparent in x-rays, tantalum marker pins are inserted into the implants in order to give surgeons a visual aid in determining the location of the implants, both intra and postoperatively.
The SKYHAWK Interbody Fusion System implants are offered in parallel and lordotic profiles to restore the natural curvature of the spine; the device may be implanted using a later or anterolateral approach.
The SKYHAWK Interbody Fusion System implants, trials and instruments are provided nonsterile. They require sterilization prior to use.
## Indications for Use
### FORZA Spacer System:
The FORZA Spacer System is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated.
The FORZA Spacer System is intended to be used with supplemental fixation systems. As an example, the supplemental fixation that may be used is the Orthofix Firebird Spinal Fixation System.
The FORZA Spacer System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
## PILLAR PEEK Spacer System:
When used as an intervertebral body fusion device, the PILLAR PEEK Spacer System is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated.
The PILLAR PEEK Spacer System is intended to be used with supplemental fixation. As an example, the supplemental fixation that may be used is the Orthofix Spinal Fixation System (SFS).
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the word "ORTHOFIX" in black letters. To the left of the word is a blue symbol that looks like a stylized letter O. The word is in all caps and has a registered trademark symbol after it.
The PILLAR PL PEEK Spacer is used singly or in pairs and is implanted using a posterior approach.
The PILLAR TL PEEK Spacer is used singly or in pairs and is implanted using a transforaminal approach.
The PILLAR AL PEEK Spacer is used singly and is implanted using an anterior approach.
The PILALR XL PEEK Spacer is used singly and is implanted using a lateral approach.
The PILLAR PEEK Spacer System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
When used as a Partial Vertebral Body Replacement (VBR), The PILLAR PEEK Spacer System is indicated for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The PILLAR Spacer System is also indicated for treating fractures of the thoracic and lumbar spine.
The PILLAR PEEK Spacer System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column even in the absence of a fusion for a prolonged period of time. The Partial VBR device is intended to be used with autograft or allograft.
The PILLAR PEEK Spacer System is intended for use with supplemental fixation. As an example, the supplemental fixation that may be used is the Orthofix Spinal Fixation System (SFS).
## PILLAR SA PEEK Spacer System:
When used as an intervertebral body fusion device, the PILLAR SA PEEK Spacer System is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiquous levels from L2 to S1. These DDD patients may have up to Grade I Spondvlolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated.
The PILLAR SA PEEK Spacer System is intended for use with the four titanium alloy screws provided with the device. If the physician chooses to use fewer than four of the provided screws, then supplemental fixation must be used to augment stability. As an example, the supplemental fixation system that may be used is the Orthofix Spinal Fixation System (SFS).
The PILLAR SA PEEK Spacer System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the word "ORTHOFIX" in bold, black letters. To the left of the word is a blue, abstract shape that resembles a stylized leaf or wing. The word "ORTHOFIX" appears to be the logo for a company or organization. The logo is simple and modern, with a clean typeface and a distinctive graphic element.
When used as a Partial Vertebral Body Replacement (VBR) the PILLAR SA PEEK Spacer System is indicated for use in the thoracolumbar spine (T1-L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors, to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The PILLAR SA PEEK Spacer System is also indicated for treating fractures of the thoracic and lumbar spine.
The PILLAR SA PEEK Spacer System is designed to restore the biomechanical integrity of the anterior, middle and posterior spinal column, even in the absence of fusion for a prolonged period of time. The PILLAR SA PEEK Spacer System is intended to be used with autograft or allograft.
The PILLAR SA PEEK Spacer System is intended for use with the four titanium alloy screws provided with the device. If the physician chooses to use fewer than four of the provided screws, then supplemental fixation must be used to augment stability. As an example, the supplemental fixation that may be used is the Orthofix Spinal Fixation System (SFS).
# SKYHAWK Lateral Interbody Fusion System:
When used as an intervertebral body fusion device, the SKYHAWK Lateral Interbody Fusion Svstem is indicated for use with bone graft (autograft bone or allogenic bone graft composed of cancellous or corticocancellous bone graft) in patients with degenerative disk disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may have up to Grade I Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with deqeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. Patients with previous non-fusion spinal surgery at the treated level may be treated.
The SKYHAWK Lateral Interbody Fusion System is intended to be used with supplemental fixation systems that are cleared by the FDA for use in the lumbar spine.
The SKYHAWK Lateral Interbody Fusion System must be used with autograft or allogenic bone graft composed of cancellous or corticocancellous bone graft.
# Summary of the Technological Characteristics of the Device Compared to the Selected Predicate Devices
The technological characteristics of the FORZA Spacer System (K103111), PILLAR PEEK Spacer System (K082235), PILLAR SA PEEK Spacer System (K081849) and the SKYHAWK Lateral Interbody Fusion System (K140709) are all similar to the predicate devices in terms of design, dimensions, intended use, materials, and performance characteristics. There are no significant differences between the FORZA Spacer System (K103111), PILLAR PEEK Spacer System (K082235), PILLAR SA PEEK Spacer System (K081849) and the SKYHAWK Lateral Interbody Fusion System (K140709) and the predicate device which would adversely affect the use of the product.
PERFORMANCE DATA - Summary of Non-Clinical Test Conducted for Determination of Substantial Equivalence
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the word "ORTHOFIX" in bold, black letters. To the left of the word is a blue, abstract shape. The shape is curved and has a pointed end. The word "ORTHOFIX" is written in all capital letters. There is a trademark symbol to the right of the word.
The purpose of this 510(k) submission is to obtain clearance for the modified Indications for Use for Intervertebral Body Fusion Devices (IBFD), to include the use of alloqenic bone graft composed of cancellous or corticocancellous bone graft in the spine. The subject FORZA Spacer System, PILLAR PEEK Spacer System, PILLAR SA PEEK Spacer System and the SKYHAWK Lateral Interbody Fusion System technological characteristics, design, material and principles of operation are all unchanged from when they were previously cleared.
## The FORZA Spacer System Pyrogen Testing:
The FORZA Spacer System STERILE packed implants were tested using Bacterial endotoxin testing (BET) as specified in ANSI/AAMI ST-72:2011 to achieve an Endotoxin limit less than 20EU per device.
# Basis of Substantial Equivalence
The FORZA Spacer System, PILLAR PEEK Spacer System, PILLAR SA PEEK Spacer System and the SKYHAWK Lateral Interbody Fusion System technological characteristics, design, material and principles of operation are all unchanged from when they were previously cleared.
Furthermore the subject devices have similar intended use, indications for use, technological characteristics, design, same materials and the same principles of operation as the to the predicate FORZA PTC Spacer System (K152475).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.